PENTAX Medical Video Processor EPK-i5500c, PENTAX Medical Video Upper GI Scope EG29-i10c, PENTAX Medical Video Colonoscope EC34-i10cL, PENTAX Medical Video Colonoscope EC38-i10cL
K190805 · Pentax of America, Inc. · PEA · Dec 20, 2019 · Gastroenterology, Urology
Device Facts
Record ID
K190805
Device Name
PENTAX Medical Video Processor EPK-i5500c, PENTAX Medical Video Upper GI Scope EG29-i10c, PENTAX Medical Video Colonoscope EC34-i10cL, PENTAX Medical Video Colonoscope EC38-i10cL
Applicant
Pentax of America, Inc.
Product Code
PEA · Gastroenterology, Urology
Decision Date
Dec 20, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The PENTAX Medical Video Colonoscope EC34-i10cL is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the lower gastrointestinal tract. This anatomy includes, but is not restricted to, the organs, tissues, and subsystems: Large Bowel to the cecum, terminal ileum of the small bowel. This endoscope is introduced via the rectum, as decided by the physician, when indications consistent with the need for the procedure are observed in patient populations with greater than 5kg of body weight. The PENTAX Medical Video Colonoscope EC38-i10cL is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the lower gastrointestinal tract. This anatomy includes, but is not restricted to, the organs, tissues, and subsystems: Large Bowel to the cecum, terminal ileum of the small bowel. This endoscope is introduced via the rectum, as decided by the physician, when indications consistent with the need for the procedure are observed in patient populations with greater than 20kg of body weight. The PENTAX Medical Video Upper GI Scope EG29-i10c is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the upper gastrointestinal tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, and duodenum. This endoscope is introduced via the mouth when indications consistent with the need for the procedure are observed in patient populations with greater than 20kg of body weight. The PENTAX Medical Video Processor EPK-i5500c is intended to be used with the PENTAX Medical camera heads, endoscopes, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation. It processes signals transmitted from a video endoscope into images that are displayed on an inspection monitor. This processor has air/water pump function. This device includes PENTAX i-scan™, a digital, post-processing imaging enhancement technology. i-scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
Device Story
System comprises video processor EPK-i5500c and compatible endoscopes (EG29-i10c, EC34-i10cL, EC38-i10cL). Endoscopes feature distal CMOS sensors capturing reflected light from body cavities illuminated by LEDs. Processor converts raw image data into video signals for display on external monitors. Includes i-scan digital post-processing for image enhancement, used as an optional adjunct to white light endoscopy. Physician operates endoscope via control body (angulation knobs, suction/irrigation valves) and inserts into GI tract. Processor provides air/water pump functions. Output assists physicians in endoscopic diagnosis and therapeutic procedures. Benefits include improved visualization of GI anatomy and potential for enhanced tissue observation via i-scan. Not intended to replace histopathological sampling.
Clinical Evidence
Bench testing only. Performance validated through optical characterization, backflow prevention, compatibility, and product life testing.
Technological Characteristics
System includes video processor and flexible endoscopes. Features CMOS imaging sensors, LED illumination, and air/water/suction channels. Connectivity via electrical contacts to processor. i-scan technology performs digital post-processing of RGB pixel components. Class II device.
Indications for Use
Indicated for optical visualization and therapeutic access to the upper GI tract (esophagus, stomach, duodenum) in patients >20kg, and the lower GI tract (large bowel to cecum, terminal ileum) in patients >5kg (EC34-i10cL) or >20kg (EC38-i10cL).
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K150618 — PENTAX Medical EPK-i7010 Video Processor with GI Family · Pentax of America, Inc. · Nov 24, 2015
K231249 — PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Upper GI Scope EG29-i20c, PENTAX Medical Video Colonoscope EC38-i20cL · Pentax of America, Inc. · Jul 28, 2023
K191282 — PENTAX Medical EPK-i7010 and PENTAX EPK-i5010 Video Processors with GI Family · Pentax of America, Inc. · Nov 8, 2019
K232860 — PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Bronchoscope EB11-J10, PENTAX Medical Video Bronchoscope EB15-J10, PENTAX Medical Video Bronchoscope EB19-J10U, PENTAX Medical Ultrasound Video Bronchoscope EB19-J10U · Pentax of America, Inc. · Jan 18, 2024
K211882 — HD-550 Video Endoscope System · Sonoscape Medical Corp. · Apr 1, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 20, 2019
PENTAX of America, Inc. % Beryl Jeanne Regulatory Specialist NAMSA Inc. 400 Highway 169. Suite 500 Minneapolis, MN 55426
Re: K190805
Trade/Device Name: PENTAX Medical Video Processor EPK-i5500c PENTAX Medical Video Upper GI Scope EG29-i10c PENTAX Medical Video Colonoscope EC34-i10cL PENTAX Medical Video Colonoscope EC38-i10cL Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: PEA, FDS, FDF Dated: November 20, 2019 Received: November 21, 2019
Dear Beryl Jeanne:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shanil P. Haugen, Ph.D. Acting Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K190805
## Device Name
PENTAX Medical Video Colonoscope EC34-i10cL; PENTAX Medical Video Colonoscope EC38-i10cL; PENTAX Medical Video Upper GI Scope EG29-i10c; PENTAX Medical Video Processor EPK-i5500c
## Indications for Use (Describe)
The PENTAX Medical Video Colonoscope EC34-i10cL is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the lower gastrointestinal tract. This anatomy includes, but is not restricted to, the organs, tissues, and subsystems: Large Bowel to the cecum, terminal ileum of the small bowel. This endoscope is introduced via the rectum, as decided by the physician, when indications consistent with the need for
the procedure are observed in patient populations with greater than 5kg of body weight.
The PENTAX Medical Video Colonoscope EC38-110cL is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the lower gastrointestinal tract. This anatomy includes, but is not restricted to, the organs, tissues, and subsystems: Large Bowel to the cecum, terminal ileum of the small bowel. This endoscope is introduced via the rectum, as decided by the physician, when indications consistent with the need for the procedure are observed in patient populations with greater than 20kg of body weight.
The PENTAX Medical Video Upper GI Scope EG29-i10c is intended to provide optical visualization of (via a video monitor), and therapeutic access to, the upper gastrointestinal tract. This anatomy includes, but is not restricted to, the organs; tissues; and subsystems: esophagus, stomach, and duodenum.
This endoscope is introduced via the mouth when indications consistent with the need for the procedure are observed in patient populations with greater than 20kg of body weight.
The PENTAX Medical Video Processor EPK-i5500c is intended to be used with the PENTAX Medical camera heads, endoscopes, monitors, and other ancillary equipment for endoscopic diagnosis, treatment, and video observation. It processes signals transmitted from a video endoscope into images that are displayed on an inspection monitor. This processor has air/water pump function.
This device includes PENTAX i-scan™, a digital, post-processing imaging enhancement technology. i-scan is intended to be used as an optional adjunct following traditional white light endoscopy and is not intended to replace histopathological sampling.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
-
{3}------------------------------------------------
## 510(k) Summary
| 510(k) Number | K190805 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Preparation Date: | November 20, 2019 |
| Submitter: | PENTAX of America, Inc. |
| Contact Person: | US Agent:<br>William Goeller<br>Vice President, Quality Assurance and Regulatory Affairs<br>PENTAX of America, Inc.<br>3 Paragon Drive<br>Montvale, New Jersey 07645-1782<br><br>Official Correspondent:<br>Beryl St. Jeanne<br>Regulatory Consultant<br>NAMSA<br>400 Highway 169 South, Suite 500<br>Minneapolis, MN 55426 |
| Device Name: | PENTAX Medical Video Processor EPK-i5500c<br>PENTAX Medical Video Upper GI Scope EG29-i10c<br>PENTAX Medical Video Colonoscope EC34-i10cL<br>PENTAX Medical Video Colonoscope EC38-i10cL |
| Device Classification: | Regulation: 21 CFR 876.1500<br>Regulation Description: Endoscope and accessories<br>Regulatory Class: Class II<br>Product Code: PEA, FDS, FDF<br>Classification Panel: Gastroenterology/Urology<br>Devices: Endoscope, Accessories, Image Post-<br>Processing For Color Enhancement; Gastroscopic and Accessories,<br>Flexible/Rigid; Colonoscope and<br>Accessories, Flexible/Rigid |
| | PENTAX Medical Video Processor EPK-i5500c<br>This device is intended to be used with the PENTAX Medical camera<br>heads, endoscopes, monitors, and other ancillary equipment for<br>endoscopic diagnosis, treatment, and video observation.<br>It processes signals transmitted from a video endoscope into images<br>that are displayed on an inspection monitor. This processor has<br>air/water pump function.<br>This device includes PENTAX i-scan™, a digital, post-processing<br>imaging enhancement technology. i-scan is intended to be used as an<br>optional adjunct following traditional white light endoscopy and is not<br>intended to replace histopathological sampling. |
| Indications for Use | PENTAX Medical Video Upper GI Scope EG29-i10c<br>The PENTAX Medical Video Upper GI Scope EG29-i10c is intended<br>to provide optical visualization of (via a video monitor), and<br>therapeutic access to, the upper gastrointestinal tract. This anatomy<br>includes, but is not restricted to, the organs; tissues; and subsystems:<br>esophagus, stomach, and duodenum.<br>This endoscope is introduced via the mouth when indications<br>consistent with the need for the procedure are observed in patient<br>populations with greater than 20kg of body weight. |
| | PENTAX Medical Video Colonoscope EC34-i10cL<br>The PENTAX Medical Video Colonoscope EC34-i10cL is intended<br>to provide optical visualization of (via a video monitor), and<br>therapeutic access to, the lower gastrointestinal tract. This anatomy<br>includes, but is not restricted to, the organs, tissues, and subsystems:<br>Large Bowel to the cecum, terminal ileum of the small bowel.<br>This endoscope is introduced via the rectum, as decided by the<br>physician, when indications consistent with the need for the procedure<br>are observed in patient populations with greater than 5kg of body<br>weight. |
| | PENTAX Medical Video Colonoscope EC38-i10cL<br>The PENTAX Medical Video Colonoscope EC38-i10cL is intended<br>to provide optical visualization of (via a video monitor), and<br>therapeutic access to, the lower gastrointestinal tract. This anatomy<br>includes, but is not restricted to, the organs, tissues, and subsystems:<br>Large Bowel to the cecum, terminal ileum of the small bowel.<br>This endoscope is introduced via the rectum, as decided by the<br>physician, when indications consistent with the need for the procedure<br>are observed in patient populations with greater than 20kg of body<br>weight. |
| Device Description | PENTAX Medical Video Processor EPK-i5500c<br>The PENTAX Medical Video Processor EPK-i5500c consists of a<br>video system, monitor and ancillary equipment. This processor is<br>intended for endoscopic diagnostic, treatment and video observation.<br>The PENTAX Medical Video Processor EPK-i5500c contains a<br>contrast enhancement feature: PENTAX i-scan technology.<br>All visualization is done with the white light mode first. White light<br>(RGB) illuminates the tissue and transfers the captured light through<br>the Complementary MOS (CMOS) of the video scope. Note that the<br>white light visualization mode is always used first by the physician.<br>For i-scan image enhancement, the modification of the combination of<br>RGB components for each pixel occurs when the i-scan function is<br>turned on in the PENTAX Medical Video Processor EPK-i5500c. The<br>resulting i-scan image is then displayed on the observation monitor.<br>The EPK-i5500c is compatible with PENTAX Medical endoscopes<br>EG29-i10c, EC34-i10cL and EC38-i10cL.<br>PENTAX Medical Video Upper GI Scope EG29-i10c |
| The PENTAX Medical Video Upper GI Scope EG29-i10c is used<br>with PENTAX Medical Video Processor EPK-i5500c.The PENTAX<br>Medical Video Upper GI Scope EG29-i10c is composed of the<br>following main parts: an Insertion Portion, Control Body and Scope<br>Connector. The Insertion Portion is inserted into the body cavity of<br>patient. The Insertion Portion includes the Distal End and Bending<br>Section. The Objective Lens, Light Guide, Instrument Channel,<br>Air/Water Nozzle, and Water Jet Nozzle are located on the Distal End<br>of the Insertion Portion. The Control Body is held by the user's hand.<br>The Control Body includes the Angulation Control Knob, Angulation<br>Lock Knob/Lever, Air/Water Cylinder, Suction Cylinder, Remote<br>Control Button, Magnification Control Lever, and Instrument Channel<br>Inlet. The Air/Water Feeding Valve is attached to the Air/Water<br>Cylinder, and the Suction Control Valve is attached to the Suction<br>Cylinder. The Inlet Seal is attached to the Instrument Channel Inlet.<br>The Scope Connector is connected to the Video Processor via an<br>Electrical Contacts. The Bending Section is bent by the Angulation<br>Control Knob to operate the endoscope angulation. The Angulation<br>Lock Knob/Lever is used to adjust the rotation torque of the<br>Angulation Control Knob.<br>The Air/Water Feeding System is used to deliver the air and water to<br>the Objective Lens from the Air/Water Nozzle. When the hole at the | |
| top of Air/Water Feeding Valve is covered, air is delivered. When the<br>Air/Water Feeding Valve is pushed, the water is delivered. The<br>Suction Control System is used to suction the fluid and air in body<br>cavity from the Instrument Channel. When the Suction Control Valve<br>is pushed, the fluid and air are suctioned. The Remote Button is used<br>to operate the function of video processor and external device from<br>the control body, as necessary. | |
| Endotherapy Device such as Biopsy Forceps is inserted from the<br>Instrument Channel Inlet into the body cavity through the instrument<br>channel. The Water Jet System is used to stream forward the sterile<br>water from Water Jet Nozzle. | |
| The Light Guide of the Distal End is used to illuminate the body<br>cavity with LEDs. The CMOS built into the Distal End receives<br>reflected light (image data) from the body cavity, and sends the image<br>data to the Video Processor through the video cable. The image data<br>are converted into the image signal by the Video Processor, and the<br>image inside the body cavity is displayed on the Monitor. | |
| <b>PENTAX Medical Video Colonoscopes EC34-i10cL and EC38-<br/>i10cL</b> | |
| The PENTAX Medical Video Colonoscopes EC34-i10cL and EC38-<br>i10cL are used with PENTAX Medical Video Processor EPK-i5500c. | |
| The endoscopes are composed of the following main parts: an<br>Insertion Portion, Control Body and Scope Connector. | |
| The Insertion Portion is inserted into the body cavity of patient. The<br>Insertion Portion includes the Distal End and Bending Section. The<br>Objective Lens, Light Guide, Instrument Channel, Air/Water Nozzle,<br>and Water Jet Nozzle are located on the Distal End of the Insertion<br>Portion. The Control Body is held by the user's hand. The Control<br>Body includes the Angulation Control Knob, Angulation Lock<br>Knob/Lever, Air/Water Cylinder, Suction Cylinder, Remote Button,<br>Suction Selector Cylinder, and Instrument Channel Inlet. The<br>Air/Water Feeding Valve is attached to the Air/Water Cylinder, and<br>the Suction Control Valve is attached to the Suction Cylinder. The<br>Suction Channel Selector is attached to the Suction Selector Cylinder.<br>The Inlet Seal is attached to the Instrument Channel Inlet. The Scope<br>Connector is connected to the Video Processor via an Electrical<br>Contacts. | |
| The Bending Section is bent by the Angulation Control Knob to<br>operate the endoscope angulation. The Angulation Lock Knob/Lever | |
| | is used to adjust the rotation torque of the Angulation Control Knob.<br>The Air/Water Feeding System is used to deliver the air and water to<br>the Objective Lens from the Air/Water Nozzle. When the hole at the<br>top of Air/Water Feeding Valve is covered, the air is delivered. When<br>the Air/Water Feeding Valve is pushed, the water is delivered. The<br>Suction Control System is used to suction the fluid and air in body<br>cavity from the Instrument Channel. When the Suction Control Valve<br>is pushed, the fluid and air are suctioned. The endoscopes have two<br>channels; the primary channel and secondary one. Suction Channel<br>Selector is used to switch between the primary and secondary<br>channels as necessary. The Remote Button is used to operate the<br>function of video processor and external device from the control body,<br>as necessary. |
| | Endotherapy Device such as Biopsy Forceps is inserted from the<br>Instrument Channel Inlet into the body cavity through the instrument<br>channel. The Water Jet System is used to stream forward the sterile<br>water from Water Jet Nozzle. |
| | The Light Guide of the Distal End is used to illuminate the body<br>cavity with LEDs. The CMOS built into the Distal End receives<br>reflected light (image data) from the body cavity, and sends the image<br>data to the Video Processor through the video cable. The image data<br>are converted into the image signal by the Video Processor, and the<br>image inside the body cavity is, displayed on the Monitor. |
| | K122470<br>PENTAX EPK-i5010 VIDEO PROCESSOR<br>PENTAX Medical Company |
| Predicate Devices: | K131902<br>PENTAX VIDEO UPPER G.I. SCOPES (EG FAMILY)<br>PENTAX Medical Company |
| | K131855<br>PENTAX VIDEO COLONOSCOPES (EC FAMILY)<br>PENTAX Medical Company…
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.