K022490 · KARL STORZ Endoscopy-America, Inc. · GCT · Sep 29, 2003 · Gastroenterology, Urology
Device Facts
Record ID
K022490
Device Name
KSEA MEDI PACK
Applicant
KARL STORZ Endoscopy-America, Inc.
Product Code
GCT · Gastroenterology, Urology
Decision Date
Sep 29, 2003
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Indications for Use
The Medi Pack is designed to deliver illumination, provide camera use, and display and store medical images obtained during endoscopic surgical or diagnostic procedures.
Device Story
Medi Pack is a compact video camera system for endoscopic procedures; integrates camera control unit, cold light source, documentation module, 6.4-inch LCD monitor, keyboard, and camera head. Device provides illumination, captures video/images, and enables storage/display of medical images. Used in clinical settings by physicians/surgeons during endoscopic surgery or diagnostics. Output allows real-time visualization and documentation of internal anatomy, aiding clinical decision-making and patient record-keeping.
Clinical Evidence
No clinical data provided; substantial equivalence based on design and intended use comparison.
Technological Characteristics
Compact video camera system; includes camera control unit, cold light source, documentation module, 6.4-inch LCD monitor, keyboard, and camera head. Integrated system for illumination and image processing.
Indications for Use
Indicated for use in endoscopic surgical or diagnostic procedures requiring illumination, camera imaging, and image display/storage.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
Related Devices
K071292 — VIEW MD OFFICE SYSTEM, MODEL 41066.5540 · Richard Wolf Medical Instruments Corp. · Jul 19, 2007
K081585 — ENDOSCOPY VIDEO CAMERA SYSTEM · Vision Systems Group, A Division of Viking Systems · Jun 20, 2008
K182696 — Telepack X LED · KARL STORZ Endoscopy-America, Inc. · Nov 20, 2018
K152513 — 8X-10-XXXX Camera System · Santa Barbara Imaging Systems · Dec 15, 2015
K032430 — ACUEITY MEDICAL LIGHT SOURCE · Acueity, Inc. · Sep 5, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is written around the top half of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Karl Storz Endoscopy-America, Inc. Mr. Paul Lee Senior Regulatory Affairs Specialist 600 Corporate Pointe, 5th Floor Culver City, CA 90230
Re: K022490
Trade/Device Name: KSEA Medipack, Model 20042020 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FET, GCT Dated (Date on orig SE ltr): July 14, 2003 Received (Date on orig SE ltr): July 15, 2003
Dear Mr. Lee.
This letter corrects our substantially equivalent letter of September 29, 2003.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
JUL 2 7 2015
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be
{1}------------------------------------------------
Page 2 -
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known):
K022490
Device Name: Medi Pack
Indications for Use: The Medi Pack is designed to deliver illumination, provide camera use, and display and store medical images obtained during endoscopic surgical or diagnostic procedures.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use: | <div style="display:inline-block; vertical-align:middle;">/</div> |
|--------------------------|-------------------------------------------------------------------|
| (Per 21 CFR 801.109) | |
| OR Over-The-Counter Use: | |
| (Optional Format 1-2-96) | |
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
| 510(k) Number | K022490 |
|---------------|---------|
|---------------|---------|
003
{3}------------------------------------------------
## SEP 2 9 2003
を使えてきました。 この日は、 この日は、 この時代
## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectivencss information is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KSEA's knowledge.
| Applicant: | Karl Storz Endoscopy - America, Inc.<br>600 Corporate Pointe Drive<br>Culver City, CA 90230<br>(310) 338-8100 |
|------------|---------------------------------------------------------------------------------------------------------------|
| Contact: | James A. Lee. Ph.D. |
Schior Regulatory Affairs Specialist
Common Name: Device Identification: Camera System
> Trade Name: (optional) Karl Storz Medi Pack
Indication: The Medi Pack is designed to deliver illumination, provide camera use, and display and store medical images obtained during endoscopic surgical or diagnostic procedures.
Device Description: The KSEA Medi Pack is a compact video camera system consisting of a camera control unit, a cold light source, a documentation module, a 6.4-inch high performance LCD video monitor, a keyboard, and a camera head.
The Karl Storz Medi Pack is substantially equivalent to the predicate Substantial Equivalence: devices since the basic features and general intended uses are similar. The minor differences between the Karl Storz Medi Pack and the predicate devices raise no new issues of safety and effectiveness, as these differences have no effect on the performance, function or general intended use of these devices.
Signed:
Juebbe
James A. Lec, Ph.D. Senior Regulatory Affairs Specialist
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.