Intended Use

The 8X-10-XXXX Camera System is indicated for use in diagnostic and operative endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening. The 8X-10-XXXX Camera System is indicated for use with a compatible 8X-10-XXXX Camera Head and other accessory devices including an endoscope, optical coupler, and light cable.

Device Description

The 8X-10-XXXX Camera System is a high-definition camera used to capture still and video images of endoscopic or general surgications. The system also incorporates an optional internal light source featuring a Turret Light Guide Adapter which accepts various light guides. The 8X-10-XXXX Camera System consists of a console and a compatible 8X-10-XXXX camera head.

AI/ML Overview

The document provided describes the 8X-10-XXXX Camera System, an endoscopic camera system.

Here's an analysis of the acceptance criteria and the study information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

The document does not explicitly state numerical acceptance criteria for performance; instead, it relies on demonstrating compliance with recognized safety standards and bench testing showing that the device "met all requirements" and is "substantially equivalent" to the predicate device.

Acceptance Criteria CategorySpecific Criteria (Implicit/Explicit)Reported Device Performance
SafetyCompliance with recognized safety standards: - IEC 60601-1: Medical electrical equipment - Part 1: General requirements for basic safety and essential performance - IEC 60601-1-2: Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests - IEC 60601-2-18: Medical Electrical Equipment, Particular Requirements for Endoscopic EquipmentThe 8X-10-XXXX Camera System "demonstrates substantial equivalence in safety by tested compliance with" IEC 60601-1 and IEC 60601-1-2. It "has been tested and found to comply with the relevant international Medical Device Standards for Safety" including IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-18.
PerformanceBench testing specifications for functionality, visualization, illumination, image/video capture, and other primary/secondary device functions. (Implicit: Must perform comparably to the predicate device)."Bench testing of performance specifications were completed and demonstrate that the device met all requirements." "Bench comparison testing between the predicate device and the 8X-10-XXXX Camera System demonstrated that the devices are substantially equivalent."
Intended Use/IndicationsDevice is indicated for use in diagnostic and operative endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening, with compatible camera head and accessories. (Implicit: Performance must support this use.)The Indications for Use for the 8X-10-XXXX Camera System are "Same" as the predicate device. The Intended Use is "Substantially Equivalent" to the predicate device.
Biocompatibility(Implicit: Materials in contact with the body must be biocompatible)"Biocompatibility" is "Same" for the subject device and the predicate device, implying it meets the necessary standards.
Sterility(Implicit: Sterilization methods for reusable components must be effective and validated.)"Sterility" is "Same" for the subject device and the predicate device, implying it meets the necessary standards. The endoscopic camera head "may be sterilized by steam autoclave or other prescribed sterilization methods."

2. Sample Size Used for the Test Set and Data Provenance

  • Test Set Sample Size: Not explicitly stated as this was a non-clinical study. The evaluation appears to involve testing of physical devices and components (bench testing).
  • Data Provenance: The study described is non-clinical bench testing. It is not based on patient data (e.g., country of origin, retrospective/prospective) but on lab-based performance and safety evaluations of the medical device itself.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • This information is not provided. Given that the listed tests are primarily engineering and compliance tests (IEC standards, bench testing of specifications), the "ground truth" would be established by adherence to these standards and the device's measured performance against its own specifications and the predicate device's performance. Experts would likely be engineers, test technicians, and regulatory compliance specialists, but their number and specific qualifications are not detailed.

4. Adjudication Method for the Test Set

  • This information is not provided. As mentioned above, the "ground truth" and "adjudication" for this type of testing typically rely on objective measurements against engineering specifications and international standards, rather than expert consensus on diagnostic outcomes.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs without AI Assistance

  • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No comparison of clinical performance data was used for demonstration of substantial equivalence." This device is an endoscopic camera system, which provides visual output, but the submission focuses on its performance as a device (safety, image quality, functionality) rather than its impact on human reader diagnostic accuracy in a clinical setting. There is no mention of AI assistance in this document.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

  • No, this is not applicable. The device is a camera system, not an AI algorithm. Its performance is measured as a standalone physical device.

7. The Type of Ground Truth Used

  • The ground truth used for this submission is based on:
    • Compliance with recognized international safety and performance standards (e.g., IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18).
    • Engineering specifications and objective measurements from bench testing.
    • Comparison to the performance and characteristics of a legally marketed predicate device (OVS I Video System K123359) to demonstrate "substantial equivalence."

8. The Sample Size for the Training Set

  • Not applicable. This submission describes the evaluation of a physical medical device (an endoscopic camera system) through non-clinical bench testing and comparison to a predicate device. It does not involve a "training set" in the context of machine learning or AI algorithms.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. As stated above, there is no "training set" in this context.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure, composed of three overlapping profiles. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 15, 2015

Santa Barbara Imaging Systems James Candy Director 340 Storke Road, Suite 101 Goleta, California 93117

Re: K152513

Trade/Device Name: 8X-10-XXXX Camera System Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: Class II Product Code: GCJ Dated: October 30, 2015 Received: November 5, 2015

Dear James Candy:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K152513

Device Name 8X-10-XXXX Camera System

Indications for Use (Describe)

The 8X-10-XXXX Camera System is indicated for use in diagnostic and operative endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening. The 8X-10-XXXX Camera System is indicated for use with a compatible 8X-10-XXXX Camera Head and other accessory devices including an endoscope, optical coupler, and light cable.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Santa Barbara Imaging Systems. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of four blue and yellow arrow-like shapes pointing towards the center. The text "Santa Barbara" is in a larger, teal font, and "IMAGING SYSTEMS" is in a smaller, teal font below it.

340 Storke Road, Suite 101 Goleta, CA 93117 USA

August 28. 2015 Traditional 510(k) Summary

510(k) Summary 5

In accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92, Santa Barbara Imaging Systems, Incorporated is hereby submitting this 510(k) summary.

Submitter [510(k) owner]

Santa Barbara Imaging Systems, Incorporated.

340 Storke Road, Suite 101

Goleta, CA 93117

Company Contact

James Candy Director, Research and Development P 805.770.2110 F 805.770.2093 jcandy@sbimagingsystems.com

Submitted Device Information

Trade Name: 8X-10-XXXX Camera System Common Name: Endoscope And Accessories Classification Name: Laparoscope, General & Plastic Surgery

Classification Information

Classification: Class II Classification Regulation: 21 CFR 876.1500 Classification Product Code: GCJ

Legally Marketed Predicate Devices

The 8X-10-XXXX Camera System manufactured by Santa Barbara Imaging Systems (SBIS) is substantially equivalent to the following device currently in commercial use:

Device:OVS I Video System
Manufacturer:Olive Medical Corporation
Address:2302 S Presidents Drive, Suite D, Salt Lake City, Utah, 84120
510(k) number:K123359

Submitted Device Description

The 8X-10-XXXX Camera System is a high-definition camera used to capture still and video images of endoscopic or general surgications. The system also incorporates an optional internal light source featuring a Turret Light Guide Adapter which accepts various light guides. The 8X-10-XXXX Camera System consists of a console and a compatible 8X-10-XXXX camera head.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Santa Barbara Imaging Systems. The logo consists of a blue and yellow geometric shape on the left, followed by the text "Santa Barbara" in blue, and "IMAGING SYSTEMS" in a smaller font size below it. The address "340 Storke Road, Suite 101" is printed below the logo.

Goleta, CA 93117 USA

Intended Use

The 8X-10-XXXX Camera System, with optional integrated LED light source and image/video capture, is to be used when performing a variety of minimally invasive surgical procedures and for general medical visualization and video archive applications. The 8X-10-XXXX Camera System incorporates a remote camera head which displays the image, as presented through an endoscope, microscope, integrated or coupled optics, onto a viewing monitor. Displayed images and videos may be captured and stored internally or transferred or transmitted via a variety of means, controlled through the device's integrated touch panel or an optional secondary remote control mobile device.

The 8X-10-XXXX is intended to be used in a controlled operating room environment with compatible devices by qualified medical personnel. The camera heads are provided non-sterile. The endoscopic camera head may be sterilized by steam autoclave or other prescribed sterilization methods. The system has a 3 year expected service life.

Indications for Use:

The 8X-10-XXXX Camera System is indicated for use in diagnostic and operative endoscopic procedures to provide illumination and visualization of an interior cavity of the body through either a natural or surgical opening. The 8X-10-XXXX Camera System is indicated for use with a compatible 8X-10-XXXX Camera Head and other accessory devices including an endoscope, optical coupler, and light cable.

{5}------------------------------------------------

Goleta, CA 93117 USA

Substantial Equivalence

The 8X-10-XXXX Camera System is substantially equivalent to the predicate device. The 8X-10-XXXX Camera System does not raise any new questions of safety or effectiveness.

The following table summarizes the characteristics of the 8X-10-XXXX Camera System (SBIS Subject Device) and the OVS1 Video System (Predicate Device) that were evaluated.

SBIS 8X-10-XXXX Camera SystemOVS I Video System (K123359)Comparison
Device DescriptionDevice DescriptionSubstantiallyEquivalent
Indications for UseIndications for UseSame
Intended UseIntended UseSubstantiallyEquivalent
Skill Level Required for UserSkill Level Required for UserSame
ConfigurationConfigurationSame
Primary Device FunctionsPrimary Device FunctionsSubstantiallyEquivalent
Primary Device Controls (via integratedFront Panel Display or Camera HeadButtons)Primary Device Controls (via integratedFront Panel Display or Camera HeadButtons)SubstantiallyEquivalent
Secondary Device FunctionsSecondary Device FunctionsSubstantiallyEquivalent
Compatible Equipment/AccessoriesCompatible Equipment/AccessoriesSubstantiallyEquivalent
TechnologyTechnologySubstantiallyEquivalent
PerformancePerformanceSubstantiallyEquivalent
MaterialMaterialSubstantiallyEquivalent
Energy SourceEnergy SourceSame
How SuppliedHow SuppliedSame
Safety StandardSafety StandardSame
BiocompatibilityBiocompatibilitySame
SterilitySterilitySame

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for Santa Barbara Imaging Systems. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of four blue triangles pointing towards the center, with yellow diamonds in the center. The text "Santa Barbara" is in a larger, bolder font, while "IMAGING SYSTEMS" is in a smaller font below it.

August 28. 2015 Traditional 510(k) Summary

340 Storke Road, Suite Goleta, CA 93117 USA Non-Clinical Testing:

The 8X-10-XXXX Camera System demonstrates substantial equivalence in safety by tested compliance with:

  • IEC 60601 -1: Medical electrical equipment -Part 1: General requirements for basic safety . and essential performance; and
  • . IEC 60601-1-2 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility -Requirements and tests.

Bench testing of performance specifications were completed and demonstrate that the device met all requirements. Bench comparison testing between the predicate device and the 8X-10-XXXX Camera System demonstrated that the devices are substantially equivalent.

Clinical Testing:

No comparison of clinical performance data was used for demonstration of substantial equivalence.

Equivalence Standards Compliance

The 8X-10-XXXX Camera System has been tested and found to comply with the relevant international Medical Device Standards for Safety.

  • IEC 60601-1 Medical Electrical Equipment, General Standards ●
  • IEC 60601-1-2 Medical Electrical Equipment, Requirements for Electromagnetic Compatibility
  • IEC 60601-2-18 Medical Electrical Equipment, Particular Requirements for Endoscopic Equipment

James Candy Director, Research and Development 340 Storke Road, Suite 101 Goleta, CA 93117 P 805.770.2110 F 805.770.2093

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.