The Brush Biopsy Sets are intended to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula or calyces under direct vision.
Device Story
Brush Biopsy Set and Deflectable Brush Biopsy Set are endoscopic cytology brushes used to collect tissue samples from the upper urinary tract (ureter, renal pelvis, infundibula, calyces). The device is inserted through a rigid or flexible endoscope; the physician operates the device under direct visualization. The assembly consists of a stainless steel shaft, a catheter (PTFE or vinyl), and a distal nylon brush. The deflectable version features a 3-ring handle at the proximal end, allowing the brush tip to deflect up to 15° to reach target lesions. The device is supplied sterile for single-use. By enabling targeted tissue collection, the device assists in the pathological diagnosis of urinary tract lesions, facilitating clinical decision-making regarding patient treatment.
Clinical Evidence
Bench testing included biocompatibility, dimensional verification, simulated use, bristle integrity, corrosion testing in artificial urine, tensile strength, sterilization validation, packaging/distribution, and shelf-life stability. Acute performance was evaluated in a porcine urinary model.
Technological Characteristics
Materials: stainless steel (shaft), nylon (brush), PTFE or vinyl (catheter). Dimensions: 3.2 Fr (115 cm) or 5 Fr (67 cm). Deflectable model features 3-ring handle for 15° tip deflection. Energy source: none (manual). Connectivity: none. Sterilization: sterile, single-use.
Indications for Use
Indicated for patients requiring pathology specimen collection from lesions in the ureter, renal pelvis, infundibula, or calyces via endoscopic access.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left and features a stylized human figure. The FDA logo is on the right and includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 1, 2019
Cook Incorporated Chelsea Woods Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47402
Re: K182231
> Trade/Device Name: Brush Biopsy Set and Deflectable Brush Biopsy Set Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: FDX Dated: March 25, 2019 Received: March 26, 2019
Dear Chelsea Woods:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K182231
Device Name
Brush Biopsy Set and Deflectable Brush Biopsy Set
#### Indications for Use (Describe)
The Brush Biopsy Sets are intended to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula or calvces under direct vision.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo and the text "510(k) Summary". The Cook Medical logo is a red square with the word "COOK" in white letters on the top half. The bottom half of the logo has the word "MEDICAL" in white letters. The text "510(k) Summary" is below the logo in black letters.
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX BLOOMINGTON, IN 47402-0489 U.S 12.339.2235 TOLLEREE: 800.457.4500 WWW.COOKMEDICAL.COM
K182231 Page 1 of 3
# Brush Biopsy Set and Deflectable Brush Biopsy Set 21 CFR §876.1500 Date Prepared: April 25, 2019
#### Submitted By:
Submission: Applicant: Primary Contact: Secondary Contact: Applicant Address:
Primary Contact Phone: Secondary Contact Phone: Contact Fax:
# Device Information:
Trade Name:
Common Name: Classification Name: Classification Regulation: Device Class/Classification Panel:
Traditional 510(k) Premarket Notification Cook Incorporated Chelsea Woods Andrew Breidenbach Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 339-2235 x104707 (812) 339-2235 x105147 (812) 332-0281
# Brush Biopsv Set and Deflectable Brush Biopsy Set Endoscopic Cytology Brush Endoscope and accessories 21 CFR §876.1500, Product Code FDX Class II, Gastroenterology/Urology
## Predicate Devices:
- Primary predicate device: Vance Deflectable Biopsy Brush Set (K810372)
- Secondary predicate device: Urological Biopsy Brush Set (K770913)
## Device Description:
The Brush Biopsy Set and Deflectable Brush Biopsy Set are designed to be used through rigid or flexible endoscopes in order to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula, or calyces under direct vision. The Brush Biopsy Set consists of a stainless steel shaft, a catheter, and a nylon brush at the distal tip. The Brush Biopsy Set is available in two sizes: 3.2 French with a length of 115 centimeters, or 5 French with a length of 67 centimeters. The 3.2 French version has a stainless steel wire
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Cook Medical. The logo is set against a red background. The word "COOK" is written in white, sans-serif font in the upper portion of the logo. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font.
750 DANIELS WAY, P.O. LOOMINGTON, IN 47402-04 12 339 2235 TOULESEE, 800 4 WWW.COOKMEDIC
shaft, a catheter constructed of radiopaque polytetrafluoroethylene, and a connector cap and pin vise located at the proximal end. The 5 French version has a stainless steel coil shaft, a catheter constructed of vinyl radiopaque tubing, and a side-arm adapter attached to the proximal end.
The Deflectable Brush Biopsy Set consists of a stainless steel shaft, a radiopaque vinyl catheter, and a nylon brush at the distal tip. The Deflectable Brush Biopsy Set is 5 French with a length of 67 centimeters. The Deflectable Brush Biopsy Set has a 3-ring handle at the proximal end that allows for deflection of the brush tip to within 15° from the shaft.
The stainless steel shaft of the brush assembly is straight for the length of the catheter, and protrudes 7 centimeters from the distal tip of the catheter on the 5 French Deflectable Brush Biopsy Set to allow for deflection.
The sets will be supplied sterile and are intended for one-time use. The sets are packaged in a peel-open pouch with a three-year shelf life.
# Indications for Use:
The Brush Biopsy Sets are intended to obtain pathology specimens from lesions in the ureter, renal pelvis, infundibula or calyces under direct vision.
# Comparison to Predicate Devices:
The Brush Biopsy Set and Deflectable Brush Biopsy Set are substantially equivalent to the predicate devices, Vance Deflectable Biopsy Brush Set (K810372), and Urological Biopsy Brush Set (K770913) in that the devices have the same intended uses and fundamental technological characteristics and are similar in design, dimensions, and materials of construction.
The differences from the predicate Vance Deflectable Biopsy Brush Set (K810372) include:
- . Indications for Use
- . Catheter sizes
- Packaging, sterilization method, and shelf life (not stated in predicate 510(k)s) .
The differences from the predicate Urological Biopsy Brush Set (K770913) include:
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is in white, bold, sans-serif font, and is placed at the top of the logo. Below the word "COOK" is the word "MEDICAL" in white, sans-serif font, set against a black trapezoid.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
K182231 Page 3 of 3
- Indications for Use ●
- Catheter material ●
- . Catheter sizes
- Proximal fitting on 5 Fr Brush Biopsy Set ●
- Packaging, sterilization method, and shelf life (not stated in predicate 510(k)s) ●
Characteristics of the subject device sets that differ from the predicate devices are supported by testing and analysis.
## Performance Data:
The following testing was performed in order to demonstrate that the subject Brush Biopsy Set and Deflectable Brush Biopsy Set met applicable design and performance requirements:
- Biocompatibility .
- . Dimensional Verification of the Catheter
- Simulated Use and Bristle Integrity
- Corrosion Following Exposure to Artificial Urine ●
- Tensile Strength ●
- . Catheter Shaft
- Wire Shaft to Distal Tip .
- . Proximal Wire Shaft
- . Sterilization
- Packaging/Distribution ●
- Shelf Life/Stability ●
- Acute Performance in a Porcine Urinary Model
# Conclusion:
The results of these tests support a conclusion that the Brush Biopsy Set and Deflectable Brush Biopsy Set will perform as intended. The subject devices do not raise new questions of safety or effectiveness as compared to the predicate devices. Therefore, the data provided in this submission support a determination of substantial equivalence.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.