K Number
K251279

Validate with FDA (Live)

Date Cleared
2025-06-12

(49 days)

Product Code
Regulation Number
874.4680
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series is intended to be used to collect tissue specimens in the tracheobronchial tree in combination with a bronchoscope.

Device Description

The DISPOSABLE CYTOLOGY BRUSH BC-202D-1210/2010/3010/5010, BC-203D-2006 have been designed to collect specimens or cells endoscopically for cytologic examination in conjunction with bronchoscopes. The Subject Device (SD) is inserted into the bronchoscope through a biopsy valve to reach the target area, where, by rubbing the brush on the target area, specimens or cells can be collected. The Subject Device is then withdrawn from the bronchoscope channel with the collected samples.

The Subject Device consists of:

  • Handle
  • Insertion portion

The handle consists of a ring and a grip. The insertion portion consists of a plastic tube, stainless steel wire, nylon brush (length 6mm or 10mm, diameters 1.2 – 5 mm) and stainless steel distal tip. The grip is attached to the tube and the ring is attached to the wire. The user can move the brush at the tip of the wire back and forth by holding the grip and moving the ring back and forth. The distal tip is designed to prevent damage to the inside of the tube and/or bronchial wall.

AI/ML Overview

This document is a 510(k) clearance letter for a medical device, specifically a disposable cytology brush. It is not an AI/ML device. Therefore, the questions regarding acceptance criteria and studies for AI/ML performance (e.g., sample size for test/training sets, expert consensus, MRMC studies, standalone algorithm performance, effect size of human reader improvement with AI) are not applicable to this document.

The document focuses on demonstrating substantial equivalence to a legally marketed predicate device through:

  • Biocompatibility testing: Ensuring the materials are safe for human contact.
  • Performance testing (Bench/Non-Clinical): Evaluating the physical and mechanical performance of the brush.
  • Sterilization and Shelf Life Testing: Confirming the device remains sterile and functional over its stated shelf life.

Here's how the provided information relates to the performance acceptance criteria and proof for this medical device:


Acceptance Criteria and Device Performance for Disposable Cytology Brush

For this medical device, substantial equivalence is proven by meeting established performance specifications through non-clinical (bench) testing, biocompatibility testing, and sterilization/shelf life validation. There are no AI-specific acceptance criteria or studies mentioned as this is a physical medical device, not an AI/ML software device.

1. Table of Acceptance Criteria and Reported Device Performance

Test CategorySpecific Test/CriterionAcceptance Criteria (Implied/Standard)Reported Device Performance
BiocompatibilityCytotoxicity Study (ISO 10993-5)No significant cytotoxic effects.Passed specifications.
Intracutaneous Irritation Study (ISO 10993-23)No significant irritation.Passed specifications.
Guinea Pig Maximization Sensitization Test (ISO 10993-10)No significant sensitization.Passed specifications.
Material-mediated Pyrogen Testing (USP <151>)No pyrogenic response.Passed specifications.
Acute Systemic Toxicity Study (ISO 10993-11)No significant systemic toxicity.Passed specifications.
Performance (Bench)Expansion and Contraction of the Brush SectionPerform as intended (smooth movement, reaches full extension/retraction).Passed specifications.
Insertion & Withdrawal (from Endoscope)Smooth insertion and withdrawal without damaging the device or bronchoscope.Passed specifications.
Performance after Repeated Insertion & WithdrawalMaintain functionality after multiple cycles.Passed specifications.
Performance after Repeated Brush StrokesMaintain integrity and specified force during brushing (quantitative force measurement implies a range).Passed specifications.
Performance after Repeated Brushing (brush condition)Brush maintains structural integrity and effective cell collection capability.Passed specifications.
Performance after Repeated BendingDevice withstands bending stress without kinking or failure.Passed specifications.
Brush Strength (Grip-Tube junction)Withstand specified force without detachment or failure.Passed specifications.
Brush Strength (Tip-Brush Wire junction)Withstand specified force without detachment or failure.Passed specifications.
Sterilization/Shelf LifeSterilization Validation (ISO 11135:2014)Achieves required Sterility Assurance Level (SAL).Passed specifications.
Ethylene Oxide Residuals (ISO 10993-7:2008)Residuals below specified limits.Passed specifications.
Package Integrity (ISO 11607-1/2:2019, ASTM F1980-21)Package maintains sterile barrier after accelerated aging and simulated distribution.Passed specifications.
Product Performance (accelerated aging/simulated distribution)Device functionality maintained over the stated shelf-life (5 years) after simulated stresses.Passed specifications. (Supports 5-year shelf life)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: The document does not specify exact sample sizes for each bench test. For these types of physical device tests, sample sizes are typically determined by established standards (e.g., ISO, ASTM) and statistical validity for mechanical testing (e.g., n=30, or as per specific test method requirements).
  • Data Provenance: The tests are explicitly described as "Non-Clinical (Bench) Performance Testing" and "Biocompatibility Testing." This means the data comes from laboratory experiments and material testing, not human or observational data. There is no mention of country of origin for the data; it would be generated in the manufacturer's or contracted test labs. All tests are inherently "prospective" in the sense that they are conducted specifically to validate the device's performance.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

  • Not applicable. This is a physical medical device. Ground truth for its performance is established by objective engineering measurements, material science principles, and biological safety standards (e.g., ensuring a certain tensile strength, confirming sterility, testing for cytotoxicity per ISO standards). It does not involve human expert interpretation of images or patient data in the way an AI/ML device would.

4. Adjudication Method for the Test Set

  • Not applicable. See point 3. Testing results are pass/fail based on predetermined numerical or qualitative specifications from recognized standards.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • Not applicable. This is not an AI/ML device used for diagnostic interpretation. Therefore, there is no human reader component to improve or compare against.

6. Standalone (Algorithm Only) Performance

  • Not applicable. There is no algorithm or software for standalone performance evaluation. The device is a physical tool.

7. Type of Ground Truth Used

  • Bench Test Specifications: The "ground truth" for this device's performance lies in its ability to meet predefined engineering specifications (e.g., force measurements, cyclical durability, material integrity) and biological safety standards (e.g., absence of toxicity, successful sterilization). This is determined through standardized laboratory testing methods rather than expert consensus on clinical cases or pathology.

8. Sample Size for the Training Set

  • Not applicable. There is no "training set" as this is not an AI/ML algorithm that learns from data.

9. How the Ground Truth for the Training Set Was Established

  • Not applicable. See point 8.

FDA 510(k) Clearance Letter - Disposable Cytology Brush BC-202D/203D Series

Page 1

U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov

Doc ID # 04017.07.05

June 12, 2025

Olympus Medical Systems Corporation
℅ Eve Smith
Regulatory Affairs Associate II
Olympus Corporation of the America
3500 Corporate Parkway
Center Valley, Pennsylvania 18034

Re: K251279
Trade/Device Name: Disposable Cytology Brush BC-202D/203D Series
Regulation Number: 21 CFR 874.4680
Regulation Name: Bronchoscope (flexible or rigid) and accessories
Regulatory Class: Class II
Product Code: BTG
Dated: April 24, 2025
Received: April 24, 2025

Dear Eve Smith:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

Page 2

K251279 - Eve Smith
Page 2

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Page 3

K251279 - Eve Smith
Page 3

Sincerely,

Ethan L. Nyberg -S

Ethan Nyberg, Ph.D.
Assistant Director
DHT1C: Division of Anesthesia,
Respiratory, and Sleep Devices
OHT1: Office of Ophthalmic, Anesthesia,
Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health

Enclosure

Page 4

Indications for Use

Please type in the marketing application/submission number, if it is known. This textbox will be left blank for original applications/submissions.

Please provide the device trade name(s).

DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series

Please provide your Indications for Use below.

The DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series is intended to be used to collect tissue specimens in the tracheobronchial tree in combination with a bronchoscope.

Please select the types of uses (select one or both, as applicable):

  • ☑ Prescription Use (Part 21 CFR 801 Subpart D)
  • ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series
Page 10 of 41

Page 5

TRADITIONAL 510(K) NOTIFICATION

Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006

510(k) Summary

Page 1 of 5

General Information

Applicant: OLYMPUS MEDICAL SYSTEMS CORP.
2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507
Phone: (+81) 42-642-2111
Fax: (+81) 42-642-2307
Establishment Registration Number: 8010047

Applicant Contact: Ms. Seiko Yunoki
Correspondent: Olympus Corporation of America
3500 Corporate Parkway
Center Valley, PA 18034

Date Prepared: June 10, 2025

Device Description

Device Name: Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006
Generic/Common Name: Single-use Cytology Brush
Regulation Number: 874.4680
Regulatory Class: Class II
Classification Name: Bronchoscope (Flexible or Rigid) and accessories
Product Codes: BTG
Review Panel: Ear, Nose and Throat

Predicate Device

Device Name510(k) Submitter510(k) No.
Single Use Cytology Brush BC-205DOlympus Medical Systems CorporationK190293

Intended Use/Indications for Use

The DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series is intended to be used to collect tissue specimens in the tracheobronchial tree in combination with a bronchoscope.

Device Description

The DISPOSABLE CYTOLOGY BRUSH BC-202D-1210/2010/3010/5010, BC-203D-2006 have been designed to collect specimens or cells endoscopically for cytologic examination in conjunction with bronchoscopes. The Subject Device (SD) is inserted into the bronchoscope through a biopsy valve to reach the target area, where, by rubbing the brush on the target area, specimens or cells can be collected. The Subject Device is then withdrawn from the bronchoscope channel with the collected samples.

Page 6

TRADITIONAL 510(K) NOTIFICATION

Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006

510(k) Summary

Page 2 of 5

The Subject Device consists of:

  • Handle
  • Insertion portion

The handle consists of a ring and a grip. The insertion portion consists of a plastic tube, stainless steel wire, nylon brush (length 6mm or 10mm, diameters 1.2 – 5 mm) and stainless steel distal tip. The grip is attached to the tube and the ring is attached to the wire. The user can move the brush at the tip of the wire back and forth by holding the grip and moving the ring back and forth. The distal tip is designed to prevent damage to the inside of the tube and/or bronchial wall.

Comparison of Technological Characteristics

Table 1 compares the subject device to the predicate device with respect to indications for use and technological characteristics, providing detailed information regarding the basis for the determination of substantial equivalence.

Table 1. Subject and Predicate Device Comparison Table

Feature/CharacteristicSubject Device (SD)Predicate Device (PD)
Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006Olympus Single Use Cytology Brush BC-205D K190293
Intended Use/Indications for UseThe DISPOSABLE CYTOLOGY BRUSH BC-202D/203D Series is intended to be used to collect tissue specimens in the tracheobronchial tree in combination with a bronchoscope.The cytology brush has been specifically designed to collect specimens or cells endoscopically for cytologic examination in conjunction with Olympus bronchoscopes with channel ϕ1.7mm or larger.
Regulation Number874.4680874.4680
Regulation NameBronchoscope (flexible or rigid) and accessoriesBronchoscope (flexible or rigid) and accessories
Regulatory ClassClass IIClass II
Product CodeBTGBTG
Classification PanelEar, Nose, & ThroatEar, Nose, & Throat
Basic principlePulling the ring of the operating unit will retract the brush into the tube and by pushing the ring, the brush can be extended from the tube. Cells can be collected by rubbing the target area with the brush protruding from the tube.Pulling the ring of the operating unit will retract the brush into the tube and by pushing the ring, the brush can be extended from the tube. Cells can be collected by rubbing the target area with the brush protruding from the tube.
Brush Diameter [mm]BC-202D-1210 = Ø1.2BC-202D-2010 = Ø2.0BC-202D-3010 = Ø3.0BC-202D-5010 = Ø5.0BC-203D-2006 = Ø2.0Ø2.0
Working length (mm)11501150
Maximum insertion portion diameter [mm]BC-202D-1210, BC-202D-2010, BC-202D-3010, BC-202D-5010 = Ø1.8Ø1.4

Page 7

TRADITIONAL 510(K) NOTIFICATION

Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006

510(k) Summary

Page 3 of 5

Feature/CharacteristicSubject Device (SD)Predicate Device (PD)
Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006Olympus Single Use Cytology Brush BC-205D K190293
BC-203D-2006 = Ø1.1
Brush length (mm)BC-202D-1210, BC-202D-2010, BC-202D-3010, BC-202D-5010 = 10BC-203D-2006 = 610
Compatible BronchoscopeBC-202D-1210, BC-202D-2010, BC-202D-3010, BC-202D-5010:・Length and Model: Bronchoscopes with working length of 600mm or less・Bronchoscopes with Channel Inner Diameter: [mm] ø2BC-203D-2006:・Length and Model: Bronchoscopes with working length of 600mm or less;・Bronchoscopes with Channel Inner Diameter: [mm] ø1.2・Length and Model: Bronchoscopes with working length of 600mm or less・Channel Inner Diameter [mm] ø1.7 or more
TechniquesTransbronchial biopsyTransbronchial biopsy
Shelf-LifeFive YearsThree Years
ReprocessingSingle UseSingle Use
Sterilization MethodEOEO

The DISPOSABLE CYTOLOGY BRUSH BC-202D-1210/2010/3010/5010, BC-203D-2006 are substantially equivalent to the legally marketed predicate device given the similarities in intended use/indications for use and technological features with the predicate device:

  • similar intended use/indications for use,
  • device characteristics (design, materials, and operations) are similar or identical to the predicate device, and
  • does not introduce any new or novel treatments or standard of care that differs from predicate device in commercial use.

Dimensional changes (brush diameters, brush length) were made to the Subject Devices to maximize compatibility with the paired bronchoscopes. Although these are differences between the Subject Devices and Predicate Device, Olympus conducted Non-Clinical (Bench) Performance Testing to demonstrate substantial equivalence of the subject device to the predicate device.

Page 8

TRADITIONAL 510(K) NOTIFICATION

Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006

510(k) Summary

Page 4 of 5

Summary of Performance Testing

The following performance testing was conducted in support of substantial equivalence determination.

Biocompatibility Testing

Biocompatibility testing was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". The biocompatibility testing included the following tests:

  • Cytotoxicity Study using the ISO Elution method (ISO 10993-5)
  • Intracutaneous Irritation Study (ISO 10993-23)
  • Guinea Pig Maximization Sensitization Test (ISO 10993-10)
  • Material-mediated Pyrogen Testing (USP <151>)
  • Acute Systemic Toxicity Study in Mice (ISO 10993-11)

All Biocompatibility testing passed specifications, supporting Substantial Equivalence of the Subject and Predicate devices.

Performance Testing – Bench (Non-Clinical)

Bench tests as listed below were conducted to ensure that the subject device performs as intended and demonstrates substantial equivalence.

  • Expansion and Contraction of the Brush Section
  • Insertion and Withdrawal of the Subject Device from the Endoscope
  • Performance after Repeated Insertion and Withdrawal
  • Performance after Repeated Brush Strokes (quantitative force measurement)
  • Performance after Repeated Brushing (brush condition)
  • Performance after Repeated Bending
  • Brush Strength (Grip-Tube junction)
  • Brush Strength (Tip-Brush Wire junction)

All non-clinical bench testing passed specifications, supporting Substantial Equivalence of the Subject and Predicate devices.

Sterilization and Shelf Life Testing

The Subject Device is provided sterile to the end user. The following was performed:

  • Sterilization validation per ISO 11135:2014
  • Ethylene oxide residuals per ISO 10993-7:2008

Page 9

TRADITIONAL 510(K) NOTIFICATION

Disposable Cytology Brush BC-202D-1210/2010/3010/5010, BC-203D-2006

510(k) Summary

Page 5 of 5

Package Integrity and Product Performance testing were conducted after simulated distribution and accelerated aging for the terminally sterilized medical device per:

  • ISO 11607-1:2019, ISO 11607-2:2019 and ASTM F1980-21

The shelf life testing passed specifications, supporting Substantial Equivalence of the Subject and Predicate devices.

Conclusion

In summary, DISPOSABLE CYTOLOGY BRUSH BC-202D-1210/2010/3010/5010, BC-203D-2006 are substantially equivalent to the predicate device.

§ 874.4680 Bronchoscope (flexible or rigid) and accessories.

(a)
Identification. A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.