The intended use of the device is to allow the physician to increase the stiffness of the enteroscope when additional rigidity is required. Do not use this product for any other purposes.
Device Story
Device is a nitinol wire with a tapered distal end and ball tip; proximal end features a metal tag for gripping and setting insertion length. Used during retrograde or antegrade double-balloon enteroscopy (DBE) procedures. Physician inserts wire into the endoscope's working channel to provide structural rigidity; wire remains contained within the channel and does not protrude from the distal tip. By increasing stiffness, the device helps reduce endoscope looping and facilitates advancement through the small intestine. Device is removed prior to endoscope withdrawal. Benefits include improved maneuverability and easier navigation of the small bowel during complex endoscopic procedures.
Clinical Evidence
Bench testing only. Performance evaluated through stiffness testing across compatible endoscope models. Device met all pre-defined acceptance criteria.
Technological Characteristics
Nitinol wire, centerless ground with tapered distal end and ball tip. Metal tag at proximal end secured via set screw. Non-sterile. Dimensions and UDI marked on tag. Compatible with FUJIFILM EN-580T and EN-840T enteroscopes.
Indications for Use
Indicated for use by physicians to increase the stiffness of an enteroscope during endoscopic procedures when additional rigidity is required to reduce looping and assist with telescoping maneuvers.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
K100081 — ENTEROSCOPY OVERTUBE, MODEL 00712140 · United States Endoscopy Group, Inc. · Apr 12, 2010
K964780 — BARD DIRECTOR GUIDEWIRE (000560) · C.R. Bard, Inc. · Feb 26, 1997
K060923 — NAVIAID BGE DEVICE, MODEL 21-001 · Smart Medical Systems , Ltd. · Aug 15, 2006
K231323 — Ancora-SB · Aspero Medical, Inc. · Aug 31, 2023
Submission Summary (Full Text)
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
September 26, 2025
Fujifilm Healthcare Americas Corporation
Kotei Aoki
Manager, Regulatory Affairs
81 Hartwell Ave.
Suite 300
Lexington, Massachusetts 02421
Re: K251204
Trade/Device Name: FUJIFILM Stiffening Wire Device (SW-2000)
Regulation Number: 21 CFR 876.1500
Regulation Name: Endoscope And Accessories
Regulatory Class: Class II
Product Code: FDA
Dated: August 29, 2025
Received: August 29, 2025
Dear Kotei Aoki:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K251204 - Kotei Aoki
Page 2
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
{2}
K251204 - Kotei Aoki
Page 3
Sincerely,
Shanil P. Haugen -S
Shanil P. Haugen, Ph.D.
Assistant Director
DHT3A: Division of Renal, Gastrointestinal,
Obesity and Transplant Devices
OHT3: Office of Gastrorenal, ObGyn,
General Hospital, and Urology Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
FORM FDA 3881 (8/23)
Page 1 of 1
PSC Publishing Services (301) 443-6740
EF
| DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. |
| --- | --- |
| 510(k) Number (if known) K251204 | |
| Device Name FUJIFILM Stiffening Wire Device (SW-2000) | |
| Indications for Use (Describe) The intended use of the device is to allow the physician to increase the stiffness of the enteroscope when additional rigidity is required. Do not use this product for any other purposes. | |
| Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | |
| The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | |
| "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{4}
K251204 Page 1 of 2
| 510(k) #: K251204 | 510(k) Summary | Prepared on: 2025-09-03 |
| --- | --- | --- |
| Contact Details | | 21 CFR 807.92(a)(1) |
| Applicant Name | FUJIFILM Healthcare Americas Corporation | |
| Applicant Address | 81 Hartwell Ave. Suite 100 Lexington MA 02421 United States | |
| Applicant Contact Telephone | 765-246-2931 | |
| Applicant Contact | Mr. Kotei Aoki | |
| Applicant Contact Email | hcusregulatoryaffairs@fujifilm.com | |
| Correspondent Name | FUJIFILM Healthcare Americas Corporation | |
| Correspondent Address | 81 Hartwell Ave. Suite 100 Lexington MA 02421 United States | |
| Correspondent Contact Telephone | 781-323-5306 | |
| Correspondent Contact | Ms. Komal Patel | |
| Correspondent Contact Email | hcusregulatoryaffairs@fujifilm.com | |
| Device Name | | 21 CFR 807.92(a)(2) |
| Device Trade Name | FUJIFILM Stiffening Wire Device (SW-2000) | |
| Common Name | Endoscope and accessories | |
| Classification Name | Enteroscope And Accessories | |
| Regulation Number | 876.1500 | |
| Product Code(s) | FDA | |
| Legally Marketed Predicate Devices | | 21 CFR 807.92(a)(3) |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K051068 | ZUTRON COLONOSCOPE STIFFENING DEVICE | FDF |
| Device Description Summary | | 21 CFR 807.92(a)(4) |
| Predicate device selection: The ZUTRON COLONOSCOPE STIFFENING DEVICE (K051068) was chosen as the predicate device. The predicate device was picked based on similar technological characteristics and performance when compared to the subject device. In accordance to draft guidance "Best Practices for Selecting a Predicate Device to Support a Premarket Notification [510(k)] Submission" the predicate device meets the best practices described in Section V of the guidance. A. Intended Use/indications for Use The intended use of the device is to allow the physician to increase the stiffness of the enteroscope when additional rigidity is required. Do not use this product for any other purposes. | | |
{5}
K251204 Page 2 of 2
B. Technological characteristics
The product consists of a nitinol wire of specific diameter and length that is centerless ground to create a tapered distal end that terminates in a ball tip. The tapered distal-end is meant to provide added flexibility to the product for ease of insertion into the endoscope, as well as reduced resistance to bending the angulation of the scope tip while in use. There is a metal tag at the proximal end of device that is use for gripping the device and setting the insertion length in the working channel of the endoscope. The tag will also contain markings that identify the company name (Fujifilm), model number, dimensions, and Unique Device Identifier (UDI). The tag is assembled to the nitinol wire via mechanical means, such as a set screw that cannot be loosened or removed. The product will be packaged and sold in a non-sterile state. The stiffening wire is designed to be contained within the endoscope instrument channel and not to protrude from the distal tip of the endoscope during use.
C. Principal of Operation
The FUJIFILM Stiffening Wire can be used for Retrograde double-balloon enteroscopy (DBE) and Antegrade DBE procedure.
For the Retrograde procedure once entry into the transverse colon is confirmed, withdraw the endoscope to straighten out the endoscope and instrument channel. Gently advance the stiffening wire device down the entire length of the instrument channel. If the stiffening wire does not advance with reasonable force, straighten the endoscope and retry. Advance endoscope to cecum and into ileum. Remove the stiffening wire device prior to endoscope withdrawal or at the physician's discretion.
For Antegrade procedure gently advance stiffening wire device down the desired length of the instrument channel. Position within the channel in the area corresponding to the segment of the endoscope that requires additional stiffness or straightening. Using DBE technique, advance the endoscope down the small intestine to the ileum. The physician at their discretion, will decide when to best use the stiffening wire device to aid in advancing to the ileum by adding endoscope stiffness to reduce looping and assist with the telescoping maneuvers of small bowel endoscopy. Remove the stiffening wire device prior to endoscope withdrawal or at the physician's discretion.
| Intended Use/Indications for Use | 21 CFR 807.92(a)(5) |
| --- | --- |
The intended use of the device is to allow the physician to increase the stiffness of the enteroscope when additional rigidity is required. Do not use this product for any other purposes.
| Indications for Use Comparison | 21 CFR 807.92(a)(5) |
| --- | --- |
The difference between the intended use of the subject device and predicate device does not raise new questions of safety or efficacy of the subject device. The predicate device is intended to be used with a colonoscope whereas the subject device is intended to be used with an enteroscope. All testing was conducted using an enteroscope and no new concerns for safety and efficacy was seen.
| Technological Comparison | 21 CFR 807.92(a)(6) |
| --- | --- |
Fujifilm Stiffening Wire Device is compatible for FUJIFILM scopes EN-580T and EN-840T. Zutron Colonoscope Stiffening Device is compatible for FUJIFILM, Pentax, and Non-variable stiffness Olympus endoscopes. Although there are differences in the compatible endoscopes, the compatibility was tested during the design verification process and and no new concerns for safety and efficacy was seen. Compatible endoscopes are listed on the Operation Manual. Therefore, we believe the differences are minor and conclude that the subject device is as safe and effective as the predicate device.
The Fujifilm Stiffening Wire Device (SW-2000) demonstrates substantially equivalent performance to the Zutron Colonoscope Stiffening Device predicate device.
| Non-Clinical and/or Clinical Tests Summary & Conclusions | 21 CFR 807.92(b) |
| --- | --- |
The following performance for FUJIFILM Stiffening Wire device was evaluated. Testing was done for stiffness for each compatible endoscope.
(1) Compatible Scopes
(2) Endoscope Stiffness
The device met the pre-defined acceptance criteria listed in the test report.
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.