MedicCO2LON (MedicCO2LON)

K240847 · Ide Vision, Ltd. · FCX · Oct 30, 2024 · Gastroenterology, Urology

Device Facts

Record IDK240847
Device NameMedicCO2LON (MedicCO2LON)
ApplicantIde Vision, Ltd.
Product CodeFCX · Gastroenterology, Urology
Decision DateOct 30, 2024
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The MedicCO2LON device is intended to be used for colonic distension with CO2 gas for CT Colonography and conventional Colonoscopy procedures. MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose. The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device. The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure.

Device Story

MedicCO2LON is an automated, table/trolley-mounted insufflation device for colonic distension using CO2 gas. It connects to an external CO2 supply and delivers gas to the colorectal cavity via a standard administration set. The device regulates gas flow and pressure (up to 30mmHg) based on user-defined settings via a graphic touch screen. It displays real-time output pressure, gas flow rate, delivered volume, and supply level. The device is operated by trained clinicians during CT Colonography or conventional colonoscopy. It incorporates redundant mechanical and electronic safety pressure relief mechanisms to prevent over-pressurization. By providing controlled, automated insufflation, it assists in visualizing the colon, potentially improving diagnostic accuracy and patient comfort compared to manual methods.

Clinical Evidence

Bench testing only. No clinical data provided. Performance testing verified pressure accuracy (+/- 1mmHg), flow accuracy (+/- 10% at 3LPM), volume accuracy (+/- 20%), and the functionality of redundant mechanical and electronic overpressure relief systems.

Technological Characteristics

Electro-mechanical pneumatic insufflator. Construction: PC/ABS external shell, metal internal frame. Power: Internal mains-connected power supply (3.3V, 12V, 24V DC). Interface: LED backlit color LCD touch screen. Connectivity: IEC mains inlet, low-pressure CO2 input, controlled gas output. Safety: Independent redundant mechanical pressure relief valve (set 75-78mmHg) and electronic pressure relief (50mmHg for 5 seconds).

Indications for Use

Indicated for colonic distension with CO2 gas during CT Colonography and conventional Colonoscopy procedures. For use by trained clinical personnel.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. October 30, 2024 Nicholas See IDE Vision, Ltd. Lodge Way Portskewett Caldicot. NP265PS United Kingdom Re: K240847 Trade/Device Name: MedicCO2LON (MedicCO2LON) Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FCX Dated: October 3, 2024 Received: October 3, 2024 Dear Nicholas See: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory {2}------------------------------------------------ assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Sivakami Venkatachalam -S for Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ## Indications for Use Submission Number (if known) K240847 Device Name MedicCO2LON (MedicCO2LON) Indications for Use (Describe) The MedicCO2LON device is intended to be used for colonic distension with CO2 gas for CT Colonography and conventional Colonoscopy procedures. MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose. The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device. The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW * The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ Image /page/4/Picture/1 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is in a sans-serif font and is stylized. # 510K Summary Date prepared 9th May 2024, Updated 23rd October 2023 ## General Information: | Contact Person: | Nicholas Adrian See | |-----------------|-------------------------------------------------------------------------------------------------------| | Address: | IDE Vision Ltd<br>Lodge Way<br>Portskewett<br>Caldicot<br>NP26 5PS<br>Monmouthshire<br>United Kingdom | | Telephone: | +(44) 1291 425425 | | Email: | nick@doppler.co.uk | ## Device Identification: | Trade Name: | MedicCO2LON (MedicCO2LON) | |----------------------|---------------------------| | Common Name: | Insufflator, Colonic | | Classification Name: | Endoscope and accessories | | | 21 CFR 876.1500 | | Product Code: | FCX | ## Predicate Device Information: | Predicate Device: | BRACCO PROTO <span style="vertical-align: sub;">2</span> L TOUCH Colon Insufflator | |-------------------|------------------------------------------------------------------------------------| | | 510(k) K132192 | {5}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is sans-serif and the "i" in iDEVision is lowercase. - IDE Vision Ltd Lodge Way, Portskewett, Caldicot NP26 5PS United Kingdom ## Device Description: MedicCO2LON is an automated insufflation device designed for administering and requlating colonic distension by insufflation with carbon dioxide qas, in preparation for and during CT Colonography (CTC or Virtual Colonoscopy) and conventional Colonoscopy. #### Construction: The MedicCO2LON colonic insufflator is constructed using a plastic (PC/ABS) external shell covering a metal internal frame, internal components are low voltage (3.3V, 12V and 24V DC) powered from an internal mains connected power supply. None of the parts of the MedicCO2LON contact the patient directly and the user only has transient contact with the unit, with it being table or trolley mounted. #### Use period: The units are for short term use (normally no more than 30 min) and provides. to the patient, CO2 via an external pressure regulator to the device to insufflate the patient using a standard administration set at up to 30mmHg ### Intended Use: The MedicCO2LON device is intended to be used for colonic distension with CO.gas for CT Colonography and conventional Colonoscopy procedures. MedicCO2LON is designed to function with specific colonic insufflation administration sets that allow passage of gas from the device to the colorectal cavity. MedicCO2LON should only be used with administration sets specifically designed for this purpose. The device should only be used by personnel trained in the practice of undertaking CT Colonography and conventional Colonoscopy examinations and the use of an automatic insufflation device. The device should only be used for CT Colonography or conventional Colonoscopy procedures and should not be used for any other patient examination procedure. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized and modern. # Product Comparison: | Parameter | MedicCO2LON | PROTOC02L TOUCH | |--------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) | | K132192 | | Indication for Use | Administer and regulate CO2 as a<br>distention media for the colon<br>during CT Colonography (CTC or<br>Virtual Colonoscopy) and<br>conventional Colonoscopy. | Administer and regulate CO2 as a<br>distention media for the colon<br>during CT Colonography (CTC or<br>Virtual Colonoscopy) and<br>conventional Colonoscopy. | | Applied Gas | CO2 | CO2 | | Design | Electro-Mechanical Pneumatic<br>System to regulate both Flow and<br>Pressure of CO2 | Electro-Mechanical Pneumatic<br>System to regulate both Flow and<br>Pressure of CO2 | | Anatomical Sites | Rectal administration of CO2 via<br>enema tip or catheter lumen | Rectal administration of CO2 via<br>enema tip or catheter lumen | | Radiation | No ionizing radiation emitted | No ionizing radiation emitted | | Thermal | No thermal energy introduced into<br>patient. | No thermal energy introduced into<br>patient. | | Power supply | Switch Mode Power supply housed<br>within unit | Switch Mode Power supply<br>housed within unit | | User Connections | IEC Mains inlet,<br>Low pressure CO2 input<br>Controlled Gas output | IEC Mains inlet,<br>Low pressure CO2 input<br>Controlled Gas output | | Display type | LED Back lit colour LCD,<br>Display Area 152 x 91mm | Back lit colour LCD, | | Displayed data | Current output pressure<br>Gas Flow Rate<br>Delivered gas volume<br>Gas supply level | Current output pressure<br>Gas Flow Rate<br>Delivered gas volume<br>Gas supply level | {7}------------------------------------------------ Image /page/7/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "D" in "iDEVision" having a unique design. | Controls type | Graphic touch screen controls | Graphic touch screen controls | |--------------------------|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------| | Controls data | Start / Stop / Resume Flow | Start / Stop Flow | | | Reset Delivered Volume | Reset Delivered Volume | | | Set Target pressure | Set Target Pressure | | | Set Maximum Flow | | | Gas Control | | | | Maximum total gas volume | 12.0 L | No maximum limit | | Gas Volume Profile | Flow stops at: 4,6,8,10,12L | Flow stops at 3L to 10L | | Default target flow rate | 3.0 LPM | 3.0LPM | | Gas Flow Start Profile | 0.5lpm per 0.5l delivered volume to<br>target flow rate | 1.0lpm for 0 to 0.5l delivered<br>volume<br>2.0lpm for 0.5 to 1.0l delivered<br>volume<br>3.0lpm for greater than 1.0l<br>delivered volume | | Target pressure | 0 to 30mmhg<br>Default 20 mmHg | 0 to 35mmHg | | Volume Accuracy | +/-20% of current reading display | Unknown | |-------------------------------------------------|-------------------------------------------------|--------------------------------------------| | Flow Adjustment | 0.5LPM | 1.0LPM | | Flow Accuracy | +/-10% at 3LPM | +20% @ 3LPM | | Pressure Accuracy | +/- 1mmHg of current reading | +/- 10% | | Safely overpressure Relief | Independent Redundant<br>Mechanical Relief | Independent Redundant<br>Mechanical Relief | | | Safety Pressure Relief<br>pressure (mechanical) | 1.5psi (78mmHg) nominal | | Safety Pressure Relief<br>Accuracy (mechanical) | +/- 20% of trip pressure | Unknown | {8}------------------------------------------------ Image /page/8/Picture/1 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "i" having a dot above it, the "D" being a blocky shape, and the "Vision" being in a more traditional font. The logo is simple and modern. | Safety Pressure Relief<br>pressure (Electronic) | 50mmHg for 5 seconds | 50mmHg for 5 seconds | |-------------------------------------------------|----------------------|----------------------| |-------------------------------------------------|----------------------|----------------------| In all cases the insufflation gas is CO2 and this is externally connected to the units which then deliver it via an external Administration Kit. The control interfaces and displays vary slightly with 1 unit using touch screen LCD and the ProtoCO₂L using an LED display and mechanical controls. The data displayed is the same in style but the MedicCO2Lon adds a display of the delivery gas flow rate to provide the doctor with an indication as to the rate the gas is entering the patient. Both units deliver a volume of gas of at least 12L with the MedicCO2LON delivering 12L and the ProtoCO2L having no limit. The normal delivered volume for insufflation is between 2 – 4L, allowing for leaks 12L will be enough for all patients. Should more gas be required it implies a serious leak and this should be investigated rather than allowing limitless gas to enter the patient. Pressure on the MedicCO2LON has a target of 30mmHg with the ProtoCO2L having a 25mmHg maximum. In all cases these values can be set to different levels by the doctor and the MedicCO2LON has pre-sets that can be set to any level the doctors require up to 30mmHg and these values are recalled for each patient. The flow rate on both units is the same with a maximum of 3LPM, the MedicCO2LON allows an increase in 0.5LPM steps while the ProtoCO2L uses 1LPM steps. The smaller the steps the less stress applied to the patient. Both units use mechanical safety pressure discharge valves set at 75 – 78mmHg. ## Conclusion: The MedicCO2LON device for colonic distension with CO.gas for CT Colonography and conventional Colonoscopy procedures is considered to be substantially equivalent to the predicate device. It does not incorporate any significant changes in intended use, method of operation, material, or design that could affect the safety or effectiveness of the device. #### Clinical: The units are used for the same clinical condition and purpose. They are used at the same site in the body; {9}------------------------------------------------ Image /page/9/Picture/0 description: The image is a logo for iDEVision. The logo is white text on a black background. The text is stylized and modern. They all use the same Administration Kit to supply the gas into the patient. The units have similar critical performance according to expected clinical effect its intended use. #### Technical: The MedicCO2LON and the other unit are used under similar conditions of use. The MedicCO2LON has similar specifications and properties. The MedicCO2LON is of similar design being a standalone mains connected insufflator supplied externally with CO2. The MedicCO2LON has similar principles of operation, Pressure, Flow, Volume and connection to the Patient. #### Biological: The MedicCO2LON uses the same materials in contact with the same human tissues in that the Administration Set (not manufactured my Ultrasound Technologies Ltd) is the same or similar to all others on the market. The Kits are ALL interchangeable and depend on the hospitals purchasing arrangements. #### Tested Predicate Comparison. All characteristics have been tested and are detailed in the documents listed: A summery is as follows: | Function | MedicCO2LON<br>Performance | Test report document<br>Identification | Predicate device<br>Performance | |------------------------------|----------------------------|----------------------------------------|---------------------------------| | Displayed Data | Current Input Pressure | MCODOC121-102<br>Appendix Image | Current output pressure | | | Gas Flow Rate | MCODOC121-102<br>Appendix Image | Not defined | | | Delivered Gas Volume | MCODOC121-102<br>Appendix Image | Delivered Gas Volume | | | Gas Supply level | MCODOC121-102<br>Appendix Image | Gas Supply level | | Touch Screen Control<br>Data | Start / Stop / Resume flow | MCODOC121-102<br>Appendix Image | Start / Stop flow | | | Reset Delivered Volume | MCODOC121-102<br>Appendix Image | Reset Delivered Volume | {10}------------------------------------------------ Image /page/10/Picture/1 description: The image shows the logo for iDEVision. The logo is white text on a black background. The text is stylized, with the "i" having a dot above it and the "D" being a geometric shape. | | Set Target Pressure | MCODOC121-102<br>Appendix Image | Set Target Pressure | |-------------------------------------------------|-------------------------------------------------|--------------------------------------------|------------------------------------------| | | Set Maximum Flow | MCODOC121-102<br>Appendix Image | Not defined | | | Maximum total gas<br>volume | 12.0L | MCODOC166-101 Page 7<br>section 3.3 | | Gas Volume Profile | | Stops at 4,6,8,10,12L | MCODOC121-102 Volume<br>Accuracy test | | | Default target flow rate | 3.0LPM | Power up default on<br>screen | | Gas Flow Start Profile | | 0.5LPM per 0.5L Volume. | MCODOC121-102 Flow<br>rate Accuracy test | | | Target Pressure | 0 to 30mmHg | MCODOC121-102<br>Pressure Accuracy test | | Volume Accuracy | | +/- 20% at current<br>reading. | MCODOC121-102 Volume<br>Accuracy test | | | Flow Adjustment | 0.5LPM | MCODOC121-102 Flow<br>rate Accuracy test | | Flow Accuracy | | +/- 10% at 3LPM | MCODOC121-102 Flow<br>rate Accuracy test | | | Pressure Accuracy | +/- 1mmHg of current<br>reading | MCODOC121-102<br>Pressure Accuracy test | | Safety overpressure relief | | Independent redundant<br>mechanical relief | MCODOC166-101 Page 6<br>section 3.2 | | | Safety pressure relief<br>(mechanical) | 1.5psi (78mmHg) | MCODOC166-101 Page 6<br>section 3.2 | | Safety pressure relief<br>accuracy (mechanical) | | +/- 20mmHg of trip<br>pressure | MCODOC166-101 Page 6<br>section 3.2 | | | Safety pressure relief<br>accuracy (electronic) | 50mmHg for 5 seconds | MCODOC166-101 Page 6<br>section 3.2 |
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...