METIC™- Airway Balloon Catheter

K231818 · M/S Meril Life Sciences Pvt. , Ltd. · KTI · Nov 15, 2023 · Ear, Nose, Throat

Device Facts

Record IDK231818
Device NameMETIC™- Airway Balloon Catheter
ApplicantM/S Meril Life Sciences Pvt. , Ltd.
Product CodeKTI · Ear, Nose, Throat
Decision DateNov 15, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.4680
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

METIC™ - Airway Balloon Catheter is intended to dilate strictures of the airway.

Device Story

The METIC™ Airway Balloon Catheter is a single-lumen catheter featuring a high-pressure balloon at the distal tip. It is used by physicians to dilate airway strictures (stenosis) that obstruct breathing. The device is inserted into the airway via the mouth under endoscopic visualization. A stylet facilitates advancement to the target site and is removed prior to inflation. The balloon is inflated hydraulically using sterile water to a specified pressure to dilate the stricture, then deflated for removal. Multiple dilations may be performed as needed. The device is supplied sterile for single-patient use.

Clinical Evidence

No clinical data. Substantial equivalence is supported by bench testing, including visual inspection, dimensional measurement, balloon burst and compliance, inflation/deflation time, bond strength, flexibility, kink resistance, torquability, and fatigue testing.

Technological Characteristics

Single-lumen catheter with Nylon 12 balloon (Vestamid care ML21) and Polyether block amide (Pebax 72D) tubing. Stainless steel (SS304V) stylet. Sterilized via Ethylene Oxide per ISO 11135. Hydraulic inflation principle. Dimensions: 425mm usable length, 450mm overall length. Balloon diameters 5.0-16.0mm. Complies with ISO 10993 for biocompatibility.

Indications for Use

Indicated for the dilation of airway strictures in adult and pediatric populations.

Regulatory Classification

Identification

A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. November 15, 2023 M/s Meril Life Sciences Pvt. Ltd. Bhavin Choradiya Senior Manager- Regulatory Affairs Survey No. 135/139, Bilakhia House, Muktanand Marg, Chala, Vapi Vapi, Gujarat 396191 India Re: K231818 Trade/Device Name: METIC™ Airway Balloon Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: KTI Dated: June 20, 2023 Received: June 21, 2023 Dear Bhavin Choradiya: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" {1}------------------------------------------------ (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. # Joyce C. Lin -S for Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices {2}------------------------------------------------ Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K231818 Device Name METICTM - Airway Balloon Catheter Indications for Use (Describe) METIC™ - Airway Balloon Catheter is intended to dilate strictures of the airway. | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| |X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary # Type of 510(k) Traditional # Submitter M/s. Meril Life Sciences Private Limited Survey No. 135/139, Bilakhia House, Muktanand Marg, Chala, Vapi – 396191 Gujarat, India. Tel. No: +91-260-3052100, Fax: +91-260-3052125 Web site: www.merillife.com # Establishment Registration No. 3009613036 # Contact Person Dr. Bhavin Choradiya E-mail: bhavin.choradiya@merillife.com # Device details | Trade Name/ Proprietary Name | METIC™ – Airway Balloon Catheter | |------------------------------|-------------------------------------------------| | Panel | Airway Dilation Balloon Catheter | | Common/ Usual Name | Balloon Dilation Catheter | | Classification Name | Bronchoscope (Flexible or Rigid)and Accessories | | Regulation Number | 21 CFR 874.4680 | | Classification | Class II | | Product Code | KTI | ## Predicate Device | | Predicate device | | | |-------------------------------------|----------------------------------------------------------------|--------------------|-------------| | Subject device | Trade Name | Manufacturer | 510 (k) No. | | METIC™ – Airway<br>Balloon Catheter | Inspira Air® Balloon<br>Dilation System<br>(Primary Predicate) | Acclarent Inc. | K090660 | | | Aeris® Balloon Dilation<br>Catheter<br>(Second Predicate) | Bryan Medical Inc. | K150951 | {5}------------------------------------------------ # Indications for Use METIC™ – Airway Balloon Catheter is intended to dilate strictures of the airway. Federal law (USA) restricts this device to sale by or on the order of a physician or licensed healthcare practitioner. # Device Description The METIC™ – Airway Balloon Catheter is comprised of a single lumen catheter with a high pressure balloon near the distal tip. A stylet is provided to facilitate advancement of the balloon dilation catheter to the desired location. The stylet must be removed before inflation of the high pressure balloon. A Luer at the proximal end is used for the placement of the stylet and injecting sterile water into the balloon. Image /page/5/Figure/6 description: This image is a technical drawing of a medical device, specifically a balloon catheter. The drawing includes labels pointing to key components such as the "BALLOON", "INFLATION PORT", "STYLET", "HUB", and "LUER CAP". The drawing also includes measurements, such as "425±10" and "450±10", indicating the length of certain parts of the device, and "\$\\\$2.40" which indicates the diameter of a part of the device. # Mechanism of Action Airway structures, also called airway stenosis, are an obstruction of the trachea (airway or windpipe) due to constriction or narrowing of the passageway. Airway strictures affect an individual's ability to breathe and cough up secretions in the lungs. METIC™ - Airway Balloon Catheter goes into the airway via mouth under visualization of endoscope. Balloon is deployed at the stenosis area. Once placed to the right location balloon is than inflated using saline to desired atmospheric pressure to dilate the airway. After certain second it balloon is deflated. If required multiple dilation is performed in order to achieve desired dilation of airway. {6}------------------------------------------------ ## PRINCIPLE OF OPERATION Metic™ Airway Balloon catheter operates on the principle of hydraulic pressurization applied through an inflatable balloon attached to the distal end. It has a single lumen shaft. The catheter features a balloon of appropriate compliance for the clinical application, constructed from a polymer. The balloon is designed to uniformly expand to a specified diameter and length at a specific pressure as labeled, with well characterized rates of inflation and deflation and a defined burst pressure. The device generally features a type of endoscopic marker to indicate catheter position relative to the tip of quiding catheter during use. Since the operating principles of Airway Balloon catheter and PTCA catheter are similar, the Guidance for industry & FDA staff class II special control guidance document for certain Percutaneous Transluminal coronary angioplasty (PTCA) catheters, September 8, 2010 was used for test methodology and sampling. However, the acceptance criteria are based on intended use specifications for an airway balloon and as per the predicate devices. #### Technological Characteristics #### Metic™ Airway Balloon Catheter Specification | Characteristics | Specification | | |------------------------|------------------------------------------------------------------------------------|-----------------| | Balloon Length (mm) | 24 mm and 40 ± 1mm | | | Balloon Diameters (mm) | 5.00, 7.00, 9.00, 10.00, 12.00, 14.00 & 16.00 mm ± 10% at balloon Nominal pressure | | | Balloon profile | Balloon Size | Profile | | | 5.0 X 24mm | 2.60 to 3.00 mm | | | 7.0 X 24mm | 3.10 to 3.70 mm | | | 9.0 X 24mm | | | | 10.0 X 40mm | 3.80 to 4.50 mm | | | 12.0 X 40mm | | | | 14.0 X 40mm | | | 16.0 X 40mm | | | {7}------------------------------------------------ | Characteristics | Specification | | | | | | | | | | | | | | | |-------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--| | Catheter Usable Length | 425 ± 10 mm | | | | | | | | | | | | | | | | Catheter Overall Length | 450 ± 10 mm | | | | | | | | | | | | | | | | Balloon burst test | Actual Burst Pressure: | | | | | | | | | | | | | | | | | Balloon Size Burst Pressure 5.0 X 24mm ≥ 22 ATM 7.0 X 24mm 9.0 X 24mm 10.0 X 40mm 12.0 X 40mm 14.0 X 40mm 16.0 X 40mm ≥ 15 ATM | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Balloon Diameter @<br>Nominal Pressure (mm) | Balloon Size Nominal Pressure 5.0 X 24mm 10 ATM 7.0 X 24mm 9.0 X 24mm 10.0 X 40mm 12.0 X 40mm 14.0 X 40mm 7 ATM 16.0 X 40mm | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Balloon Fatigue Test | | Catheter shall withstand 10 cycles minimum to<br>Rated Burst Pressure | | | | | | | | | | | | | | | | | Balloon Size Rated Burst Pressure 5.0 X 24mm 17 ATM 7.0 X 24mm 9.0 X 24mm 10.0 X 40mm 12.0 X 40mm 10 ATM 14.0 X 40mm 16.0 X 40mm | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | {8}------------------------------------------------ | Characteristics | Specification | |--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------| | Balloon inflation Time<br>(Medium – Water) | $≤$ 25 seconds | | Balloon Deflation | < 10 seconds | | Balloon Refoldability Test | Minimum 3 pleats should appear when looked<br>from the tip after balloon inflation & Deflation<br>up to 10 cycles at Nominal Pressure. | # Patient contacting material | Component name | Material of<br>Construction | Unique material<br>Identifier | Patient<br>Contact | |--------------------------------|-----------------------------|-------------------------------|--------------------| | METIC™ Airway Balloon Catheter | | | | | Balloon | Nylon 12 | Vestamid care ML21 | Yes | | Catheter tubing | Polyether block<br>amide | Pebax 72 D | Yes | | Stylet | Stainless Steel | SS304V | No | | Catheter Hub | Polycarbonate | - | No | # Contact classification as per ISO 10993-1 | Device Name | Category | Contact | Contact Duration | |-----------------------------------|-----------------------|----------------------|--------------------------------| | METIC™ Airway<br>Balloon Catheter | Surface<br>contacting | Mucosal<br>membranes | A – Limited Exposure<br>< 24 H | # Clinical Study Results Not applicable. Based on the substantial equivalence discussion and the testing results from bench tests, it was determined that a clinical study was not required to demonstrate substantial equivalence with the predicate devices. {9}------------------------------------------------ # Substantial Equivalence The Metic™ Airway Balloon Catheter is substantially equivalent with the predicate devices contained in the following table. | Characteristics | Subject Device | Predicate Device | | Safety /<br>Effectiveness | |----------------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------| | Device Name | Metic™ Airway<br>Balloon Catheter | Inspira Air® Balloon<br>Dilation System<br>(510K no. K090660)<br>Primary Predicate | Aeris® Balloon Dilation<br>Catheter<br>(510K no. K150951)<br>Second Predicate | - | | Manufacturer | Meril Life Sciences<br>Pvt. Ltd., India | Acclarent Inc. | Bryan Medical Inc. | - | | Common Name | Airway Balloon<br>Catheter | Airway Balloon<br>Catheter | Balloon Dilation<br>Catheter | Same | | Class | II | II | II | Same | | Product Code | KTI | KTI | KTI | Same | | Regulation Number | 21 CFR 874.4680 | 21 CFR 874.4680 | 21 CFR 874.4680 | Same | | Single Patient Use | Yes | Yes | Yes | Same | | Direct Patient Contact | Yes | Yes | Yes | Same | | Balloon Diameter (mm) | 5.0,7.0, 9.0,10.0,<br>12.0,14.0,16.0 | 5.0,7.0, 8.5,10.0,<br>12.0,14.0,16.0 | 5.0,7.0, 8.0, 9.0,10.0,<br>12.0,14.0,16.0 | Same | | Balloon Length (mm) | 24 ± 1 mm | 24 ± 1 mm | 30 ± 1 mm | Same | | | 40 ± 1 mm | 40 ± 1 mm | 40 ± 1 mm | Same | | Characteristics | Subject Device | Predicate Device | Predicate Device | Safety /<br>Effectiveness | | Maximum Inflation<br>Pressure | 10-17 ATM | 8 – 16 ATM | 10 & 17 ATM | Inflation pressure<br>range is within the<br>range of predicate<br>device. | | Flexible Shaft | Yes | Yes | Yes | Same | | Indications for Use | The MeticTM Airway<br>Balloon Catheter is an<br>instrument intended to<br>dilate strictures of the<br>airway. | The Inspira AIR®<br>Balloon Dilation<br>System is an<br>instrument intended to<br>dilate strictures of the<br>airway tree. | Aeris® Airway Balloon<br>Dilation Catheter is<br>intended for use in<br>adult and pediatric<br>populations to dilate<br>strictures of the airway<br>tree. | Same | | Technological<br>Characteristics | Enables dilation of<br>airway Strictures | Enables dilation of<br>airway Strictures | Enables dilation of<br>airway strictures | Same | | Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide | Same | | Packaging | Thermoformed tray in<br>pouch | Thermoformed tray in<br>pouch | Thermoformed tray in<br>pouch | Same | | Labelled as non-<br>pyrogenic? | No | No | No | Same | | Principles of Operation | Hydraulic pressurization<br>applied through an<br>inflatable balloon | Hydraulic pressurization<br>applied through an<br>inflatable balloon | Hydraulic pressurization<br>applied through an<br>inflatable balloon | Same | | Biocompatibility | Complies with ISO<br>10993 | Complies with ISO<br>10993 | Complies with ISO<br>10993 | Same | {10}------------------------------------------------ {11}------------------------------------------------ #### Labelling The Labels and IFUs of Metic™ Airway Balloon Catheter are in accordance with complies with requirements of 21 CFR 801.109 and ISO 15223 - 1: Medical Devices -Symbols to be used with medical device labels, labelling, and information to be supplied - Part 1: General requirements. #### Sterilization Metic™ Airway Balloon Catheter is sterilized by traditional Ethylene Oxide sterilization methods as described in Sterility Review Guidance K90-1. Method followed for Ethylene Oxide sterilization is as per ANSI /AAMI/ISO 11135 (Sterilization of health care products – Ethylene Oxide - Part 1 (Requirements for development, validation and routine control of a sterilization process for medical devices). The Ethylene Oxide sterilization process is validated as per ANSI /AAMI/ ISO 11135-1 Sterilization of health care products – Ethylene Oxide - Part 1 (Requirements for development, validation and routine control of a sterilization process for medical devices), Method C (Overkill Method: This method is based on demonstrating that the sterilization of a microbial challenge (biological indicator) exceeds the challenge posed by the bioburden of the product. This method is commonly used to demonstrate total inactivation of a 106 Bl at a half-cycle exposure time. When exposure time is doubled, a minimum 12 SLR is delivered during EO exposure). ## SAL = 10-6 Ethylene Oxide residuals remaining on the device is in accordance with ISO 10993-7: Biological evaluation of medical devices – Part 7: Ethylene oxide sterilization residuals" for ethylene oxide and ethylene chlorohydrin The maximum residual levels after EO sterilization are as under. - 1. Ethylene Oxide < 4mg/device - 2. Ethylene Chlorhydrin < 9mg/device {12}------------------------------------------------ # Packaging Metic™ Airway Balloon Catheter is supplied sterile. The components are packed in below packaging configuration. The Sterile barrier packaging is designed to maintain sterility and permit fast and easy opening in a single peel. Metic™ Airway Balloon Catheter is available as an integrated System. ## Packaging Confiquration | S. N. | Component | Sterile barrier | Secondary Packaging | |-------|------------------------------------|--------------------------------------------|------------------------------------------------------| | 1. | MeticTM Airway<br>Balloon Catheter | Thermoformed tray<br>sealed with Tyvek lid | Cardboard carton along<br>with “Instruction for use” | ## Shelf-life The proposed shelf life of Metic™ Airway Balloon Catheter is 5 years. The shelf life study for Metic™ Airway Balloon Catheter was planned as per ASTM 1980 recommended conditions. The study was conducted with elevated temperature condition and with real time conditions. As per ASTM 1980 following conditions were used to establish the shelf life of the Metic™ Airway Balloon Catheter. The real time shelf time study for 2 years has been completed. The result of elevated temperature and real-time study has not shown any significant changes in product characteristics. #### Test Conditions | Study | Storage Conditions | Time period | |--------------------------------|------------------------|-------------| | Metic™ Airway Balloon Catheter | | | | Accelerated | 40 ± 2°C / 75 ± 5 % RH | 457 days | | Long term | 30 ± 2°C / 65 ± 5 % RH | 60 months | {13}------------------------------------------------ #### Biocompatibility study Biological evaluation of Metic™ Airway Balloon Catheter as per Use of International Standard ISO 10993-1 ("Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process". (September 8, 2023). Guidance for Industry and Food and Drug Administration Staff) #### Performance Since the operating principles of Airway Balloon catheter and PTCA catheter are similar, the Guidance for industry & FDA staff class II special control guidance document for certain Percutaneous Transluminal coronary angioplasty (PTCA) catheters, September 8, 2010 was used for test methodology and sampling. However, the acceptance criteria were based on the intended use specifications for an airway balloon and as per the predicate devices. The safety and effectiveness of METIC™ – Airway Balloon Catheter has been evaluated for the following performance and safety requirements: - Visual Inspection - Dimensional Measurement - · Balloon Burst and Balloon Compliance - Inflation and Deflation Time - Catheter Bond Strength - Flexibility and Kink - Torquability - Catheter Fatigue - Balloon Hub Measurement - Proximal Luer fitting - Balloon Preparation, Deployment and Retraction {14}------------------------------------------------ #### Conclusion The performance, biocompatibility, sterilization, packaging and shelf life study conducted on METIC™ – Airway Balloon Catheter demonstrated the device is as safe and as effective as the predicate device. Hence, METIC™ - Airway Balloon Catheter is substantially equivalent to currently marketed devices having the same intended use and similar materials and performance and will perform as intended in the specified use conditions. #### Compliance with Voluntary Standards The Metic™ Airway Balloon Catheter comply with the following standards: - ISO 14971: 2019, Medical devices Application of risk management to medical devices. - ISO 11607-1:2020, Packaging for terminally Sterilized Medical Devices Part 1: . Packaging For Terminally Sterilized Medical Devices - Part 1: Requirements for Materials, Sterile Barrier Systems and Packaging Systems - · ISO 11135: 2014, Sterilization of Health-Care Products Ethylene Oxide -Requirements for the Development, Validation and Routine Control of Sterilization Process for Medical Devices. Second Edition, 2014-07-09 - · ISO 10993-1: 2018, Biological evaluation of medical devices Part 1: Evaluation and testing within a risk management process. 2014-07-09 - . ISO 10993-7 Second Edition 2008-10-15. Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals - AAMI / ANSI / ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical . Devices -- Part 5: Tests For In Vitro Cytotoxicity - ISO 10993-10 Fifth Edition 2021, Biological Evaluation Of Medical Devices Part 10: Tests For Irritation And Skin Sensitization.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...