The Schoelly Video Bronchoscope System is intended for use by physicians for diagnostic and therapeutic procedures in nasopharyngeal endoscopy, bronchoscopy, tracheoscopy, and laryngoscopy. The Schoelly Video Bronchoscope is intended to provide visualization via a video monitor.
Device Story
Flexible, steerable endoscope with integral working/suction channel and camera control unit (CCU). Captures visual data via distal CMOS imaging sensor; transmits signal to CCU for display on external monitor/PC. Features integrated LED light sources in handle with dual wide-angle and 'DeepVu' focused spot illumination modes. Used by physicians in clinical settings for visualization of airway/nasopharyngeal anatomy. Enables sampling of secretions, irrigation, and insertion of instruments via working channel. Provides real-time visualization to assist clinical decision-making during diagnostic and therapeutic procedures. Benefits include improved visualization of distal structures and integrated suction/instrument access.
Clinical Evidence
Bench testing only. Includes electrical safety, EMC, biocompatibility, and optical performance testing. Reprocessing validation performed per FDA guidance. No clinical data presented.
Technological Characteristics
Flexible endoscope with plastic outer surfaces. CMOS imaging sensor at distal tip. Integrated LED light sources with fiberoptic bundles. Suction/working channel. Connectivity to CCU, monitor, and PC. Reprocessing via manual cleaning and sterilization. Design verification per recognized consensus standards.
Indications for Use
Indicated for physicians performing diagnostic and therapeutic procedures in nasopharyngeal endoscopy, bronchoscopy, tracheoscopy, and laryngoscopy. No specific age or gender restrictions provided.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 17, 2014
Schoelly Fiberoptic GmbH Dr. Sandra Baumann Senior Manager Regulatory Affairs Robert-Bosch-Str. 1 - 3 79211 Denzlingen Germany
Re: K141366
Trade/Device Name: Video Bronchoscope System Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: EOO Dated: November 11, 2014 Received: November 17, 2014
Dear Dr. Baumann:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina Eydelman Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K141366
Device Name Video Bronchoscope System
#### Indications for Use (Describe)
The Schoelly Video Bronchoscope System is intended for use by physicians for diagnostic and therapeutic procedures in nasopharyngeal endoscopy, bronchoscopy, tracheoscopy, and laryngoscopy. The Schoelly Video Bronchoscope is intended to provide visualization via a video monitor.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary - K141366
#### I. SUBMITTER
| Owner's Name: | Schoelly Fiberoptic GmbH (Registration: 8043903) |
|-------------------|--------------------------------------------------------|
| Address: | Robert-Bosch-Str. 1 – 3<br>79211 Denzlingen<br>Germany |
| Telephone Number: | +49-7666-980-0 |
| Fax Number: | +49-7666-908-380 |
| Contact Person: | Dr. Sandra Baumann |
#### II. DEVICE
| Subject Device Name: | Schoelly Video Bronchoscope System |
|----------------------|---------------------------------------------------------------------|
| Trade Name: | Schoelly Video Bronchoscope System |
| Common/Usual Name: | Video Bronchoscope System |
| Classification Name: | EOQ - Bronchoscope (flexible or rigid)<br>21 CFR 874.4680: Class II |
#### III. PREDICATE/REFERENCE DEVICES
#### Predicate Device Name:
| Trade Name: | Karl Storz Video Bronchoscope System |
|-------------------------|-------------------------------------------------------------------------|
| Common/Usual Name: | Video Bronchoscope System |
| Classification Name: | EOQ - Bronchoscope (flexible or rigid)<br>21 CFR 874.4680; Class II |
| Premarket Notification: | K071530 (Karl Storz Endoscopy-America, Inc.)<br>SE date August 24, 2007 |
In addition to the Karl Storz predicate device, the following reference device was used in this submission to support substantial equivalence and address technical comparison of effectiveness of a CMOS imaging sensor (used in the proposed device) vs. a CCD imaging sensor (used in the predicate device) with regards to image quality and resolution.
#### Reference Device Name:
| Trade Name: | Ambu aScope3 |
|-------------------------|----------------------------------------|
| Common/Usual Name: | Video Bronchoscope System |
| | EOQ – Bronchoscope (flexible or rigid) |
| | 21 CFR 874.4680; Class II |
| Premarket Notification: | K130845 (Ambu A/S), |
| | SE date November 01, 2013 |
#### IV. DEVICE DESCRIPTION
The Schoelly Video Bronchoscope System consists of a flexible and steerable endoscope with an integral working/suction channel and a camera control unit (CCU) for regulation of light intensity and connection to a commercially available monitor, PC, and medical video and image capture unit for image display or image documentation.
The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and a working channel port enabling the insertion of instruments for access to the endoscope tip through the working channel.
{4}------------------------------------------------
For sampling of tracheobronchial secretion or irrigation, the endoscope further incorporates a suction channel with the suction valve port located at the endoscope handle and is accompanied by a suction valve/seal accessory . Suction and working channel converge within the endoscope handle, thus within the endoscope shaft there is only one lumen towards the endoscope tip. The suction power can be adjusted at the suction valve which is rotatable by 360°.
The endoscope further comprises a ventilation system to protect the endoscope shaft during sterilization. The exhaust valve at the endoscope handle can also be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories.
LED light sources are integrated in the endoscope handle to illuminate the anatomy under investigation. Light is transmitted through fiberoptic bundles to provide standard dual wide angle illumination as well as a narrower light beam which acts as a focused spot light thereby allowing for an improved visualization of more distal anatomical structures (DeepVu illumination). The DeepVu function can be activated by pressing a button at the endoscope handle.
The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope and transferred to the CCU. Images can be frozen and stored on a PC or medical video and image capture unit by pressing a button at the endoscope handle.
The Schoelly Video Bronchoscope System is delivered in a non-sterile condition and is already CE marked.
#### V. INDICATIONS FOR USE
The Schoelly Video Bronchoscope System is intended for use by physicians for diagnostic and therapeutic procedures in nasopharyngeal endoscopy, tracheoscopy, and laryngoscopy. The Schoelly Video Bronchoscope is intended to provide visualization via a video monitor.
#### VI. COMPARISON OF TECHNOLOGCIAL CHARACTERISTICS WITH THE PREDICATE DEVICE AND THE REFERENCE DEVICE
The Karl Storz predicate device is comprised of the same system components (flexible endoscope with working/suction channel and CCU) and an additional light source unit. It has the same principle of operations and identical or almost identical optical parameters and geometric dimensions. It is made out of the same primary materials and the same standards were applied for design verification and validation as compared to the proposed Schoelly Video Bronchoscope System.
Endoscopes of the proposed and the Karl Storz predicate device are based on the same technology for tip bending and ventilation for shaft protection during reprocessing. Light transmission in both devices relies on the same technical principle (light source and fiberoptic bundles). Both the proposed Schoelly and the Karl Storz predicate device have a suction and working channel converging within the endoscope handle and allowing for sampling of liquid, irrigation, and insertion of instruments.
The image signal of the Schoelly device is received by a CMOS imaging sensor; the Karl Storz predicate device utilizes a CCD imaging sensor instead. The Ambu aScope 3 that was cleared by FDA for marketing in K130845 is a Bronchoscope that has exactly the same clinical application as the proposed Schoelly device. In K130845, Ambu A/S demonstrated that the Ambu aScope 3 is substantially equivalent to the Ambu aScope 2 (K110962) with respect to the imaging and illumination system whereas latter comprises exactly the same CMOS imaging sensor at the endoscope's distal tip as the proposed Schoelly Video Bronchoscope
The proposed Schoelly endoscope has LED light sources integrated in the handle; the light source of the Karl Storz predicate Video Bronchoscope is not integrated; an external light source must be connected to the endoscope. The Ambu aScope 3 Bronchoscope uses integrated LED light sources as well and has the same indications for use as the proposed Schoelly Video Bronchoscope System. Light
{5}------------------------------------------------
output that can be obtained with the standard dual wide angle illumination respectively the DeepVu illumination of the proposed device is at least as high as values that were obtained for this reference device.
## VII. PERFORMANCE DATA
Performance data demonstrated that the Schoelly Video Bronchoscope System has met pre-determined acceptance criteria and is substantially equivalent to the predicate device. The risks associated with use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits are the same for both the proposed and predicate devices. Testing has been conducted in accordance with recognized consensus standards to demonstrate electrical safety. EMC, biocompatibility, and optical performance characteristics and to validate the software encompassed in the device.
### Reprocessing
The endoscope of the proposed Schoelly Video Bronchoscope System is the subject of completed reprocessing validations including manual cleaning and sterilizations have been conducted in accordance with recognized consensus standards as well as the FDA Draft Guidance Processing/Reprocessing Medical Devices in Healthcare Settings (May 2, 2011).
### VIII. CONCLUSION
The Schoelly Video Bronchoscope System meets all the pre-determined testing and acceptance criteria to effectively demonstrate substantial equivalence to the predicate device
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.