K Number
K132009

Validate with FDA (Live)

Date Cleared
2014-04-09

(282 days)

Product Code
Regulation Number
874.4760
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhinoflaryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

Device Description

The Schoelly CMOS Video Nasopharyngoscope System consists of a flexible and steerable endoscope and a camera control unit (CCU) for regulation of light intensity and connection to a monitor, PC, medical video recorder or printer for image display or image documentation.

The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and an integrated LED light source. Light is transmitted through fiberoptic bundles illuminating the anatomy under investigation. The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope shaft and transferred to the CCU.

The endoscope further incorporates a ventilation system to protect the shaft. The exhaust valve at the endoscope handle can further be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories.

The Schoelly CMOS Video Nasopharyngoscope System is delivered in a non-sterile condition and is already CE marked.

AI/ML Overview

The provided document K132009 for the Schoelly CMOS Video Nasopharyngoscope System is a 510(k) Premarket Notification. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving the device meets specific acceptance criteria based on studies involving clinical outcomes or diagnostic performance metrics like sensitivity, specificity, or AUC.

Therefore, the document does not contain the information requested regarding acceptance criteria related to diagnostic performance or image interpretation, nor does it describe a study assessing such performance (like a standalone study or MRMC study).

The "acceptance criteria" mentioned in the document refer to general safety and performance standards for a medical device of its type, confirmed through non-clinical testing, rather than criteria for diagnostic accuracy or human reader improvement.

Here's a breakdown of why each requested point cannot be addressed from the provided text:

  1. A table of acceptance criteria and the reported device performance: Not available. The document states, "Performance data demonstrated that the Schoelly CMOS Video Nasopharyngoscope System has met pre-determined acceptance criteria and is substantially equivalent to the predicate devices." However, these acceptance criteria are related to engineering, safety (IEC standards), biocompatibility (ISO 10993), and reprocessing validation, not image interpretation performance or diagnostic accuracy.

  2. Sample size used for the test set and the data provenance: Not applicable. There is no diagnostic "test set" described as the submission is for a device for visualization, not for an AI/CAD system or a diagnostic tool requiring such a test set. The non-clinical tests involved physical device testing, reprocessing validation, and electrical safety, not the analysis of medical images or patient data.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth establishment experts are mentioned because there's no diagnostic test set being evaluated.

  4. Adjudication method: Not applicable. No adjudication is mentioned for diagnostic results.

  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This type of study is for evaluating AI-assisted diagnostic tools. The Schoelly CMOS Video Nasopharyngoscope System is a visualization device, not an AI or CAD system.

  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. The device is not an algorithm that would have standalone performance.

  7. The type of ground truth used: Not applicable. No diagnostic ground truth is established or used for performance evaluation in this 510(k). The evaluation is around the physical and functional aspects of the endoscope itself.

  8. The sample size for the training set: Not applicable. There is no AI component, and thus no training set.

  9. How the ground truth for the training set was established: Not applicable. There is no AI component, and thus no training set or ground truth establishment for it.

In summary, the provided 510(k) document for the Schoelly CMOS Video Nasopharyngoscope System is a regulatory submission for a medical device (endoscope) that emphasizes substantial equivalence to predicate devices based on non-clinical engineering, safety, and reprocessing performance, rather than clinical diagnostic performance metrics and studies involving image interpretation by human experts or AI.

{0}------------------------------------------------

Schoelly Fiberontic GmbH 510(k) Premarket Notification (Traditional)

06/21/2013 CMOS Video Nusopharyngoscope System

K132009

Section 5-510(k) Summary

General Information
Preparation date:06/21/2013
Owner's Name:Schoelly Fiberoptic GmbH (Registration: 8043903)
Address:Robert-Bosch-Str. 1-379211 DenzlingenGermany
Telephone Number:+49-7666-980-0
Fax Number:+49-7666-908-380
Contact Person:Dr. Sandra Baumann
Subject Device Name:
Trade Name:Schoelly CMOS Video Nasopharyngoscope System
Trade Name:Schoelly CMOS Video Nasopharyngoscope System
Common/Usual Name:Video Nasopharyngoscope System
Classification Name:EOB - Nasopharyngoscope (flexible or rigid)21 CFR 874.4760; Class II
Predicate Device Name:
Trade Name:Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM
Trade Name:Karl Storz CMOS Video Rhino-Laryngoscope System Model 11101CM
Common/Usual Name:Video Rhino-Laryngoscope System
Classification Name:EOB - Nasopharyngoscope (flexible or rigid)21 CFR 874.4760; Class II
Premarket Notification:K103467 (Karl Storz Endoscopy-America, Inc.), SE date June 28, 2012
Predicate Device Name:
Trade Name:Schoelly Naso-Laryngo-Pharyngoscope
Trade Name:Schoelly Naso-Laryngo-Pharyngoscope
Common/Usual Name:Nasopharyngoscope
Classification Name:EOB - Nasopharyngoscope (flexible or rigid)21 CFR 874.4760; Class II
Premarket Notification:K083553 (Schoelly Imaging, Inc.), SE date January 16, 2009

Device Description

The Schoelly CMOS Video Nasopharyngoscope System consists of a flexible and steerable endoscope and a camera control unit (CCU) for regulation of light intensity and connection to a monitor, PC, medical video recorder or printer for image display or image documentation.

The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and an integrated LED light source. Light is transmitted through fiberoptic bundles illuminating the anatomy under investigation. The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope shaft and transferred to the CCU.

The endoscope further incorporates a ventilation system to protect the shaft. The exhaust valve at the endoscope handle can further be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories.

The Schoelly CMOS Video Nasopharyngoscope System is delivered in a non-sterile condition and is already CE marked.

{1}------------------------------------------------

06/21/2013 CMOS Video Nasopharyngoscope System

Predicate Devices

The Karl Storz predicate device is based on the same technology has the same essential design and dimension characteristics, the same essential optical parameters and the same principle of operations. It is made out of the same primary materials and the same standards were applied and met for design verification and validation as compared to the proposed Schoelly CMOS Video Nasopharyngoscope System.

The Schoelly predicate device differs with respect to the image transfer technology; however this device is built out of identical components/systems for light transmission, angulation of the endoscope's distal tip or ventilation. Further, the manufacturing process for major parts of the endoscope is identical as compared to the endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System.

Indications for Use

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

The indications for use of the proposed devices are identical to the indications for use of the Karl Storz predicate device.

Non-clinical Performance Testing

Performance data demonstrated that the Schoelly CMOS Video Nasopharyngoscope System has met pre-determined acceptance criteria and is substantially equivalent to the predicate devices. The device is as safe, as effective, and performs as well as or better than the predicate devices. The risks associated with use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits of predicate devices are the same as compared to the proposed one.

Testing has been conducted as per IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62304, ISO 8600, and ISO 10993 to address the safety and performance aspects of the device.

Reprocessing

The endoscope of the proposed Schoelly CMOS Video Nasopharyngoscope System is the subject of completed reprocessing validations including manual cleaning, high-level disinfection and STERRAD® 100S sterilization.

Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilties: A Guide for Device Manufacturers) and AAMI TIR30:2003 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices). Devices that have been used for testing had been exposed to extensive clinical use including tip angulation, soiling, and reprocessing to address the FDA Draft Guidance Processing Medical Devices in Healthcare Settings (dated: May 2, 2011).

Sterilization studies have been performed in accordance with ISO 14937.

Conclusion

The Schoelly CMOS Video Nasopharyngoscope System meets all the pre-determined acceptance criteria of the testing performed to confirm substantial equivalence to the predicate devices.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 9, 2014

Schoelly Fiberoptic GmbH c/o Ms. Pamela Papineau Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, MA 01432

Re: K132009

Trade/Device Name: Schoelly CMOS Video Nasopharyngoscope System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (flexible or rigid) and accessories Regulatory Class: Class II Product Code: EOB Dated: March 3, 2014 Received: March 5, 2014

Dear Ms. Papineau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Ms. Pamela Papineau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Eric A. Mann -S

for Malvina B. Eydelman. M.D. Director Division of Ophthalmic and Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

510(k) Number (if known): K132009

Device Name: Schoelly CMOS Video Nasopharyngoscope System

Indications for Use:

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhinoflaryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

.

Concurrence of Center for Devices and Radiological Health (CDRH)

Sunny Park 2014.04.08 14:30:52 -04'00'

Page 1 of 1

§ 874.4760 Nasopharyngoscope (flexible or rigid) and accessories.

(a)
Identification. A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.