K254283 · Prismatik Dentalcraft, Inc. · EIH · Jan 30, 2026 · Dental
Device Facts
Record ID
K254283
Device Name
BruxZir® Shaded 16 PLUS
Applicant
Prismatik Dentalcraft, Inc.
Product Code
EIH · Dental
Decision Date
Jan 30, 2026
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.6660
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.
Device Story
BruxZir® Shaded 16 PLUS consists of yttria-stabilized zirconia milling blanks used by dental technicians to produce full-contour crowns and bridges. Input involves digital design via CAD/CAM systems; blanks are milled into desired restoration shapes. Unlike the predicate, which requires post-sintering staining/glazing for shade, this device achieves final VITA shades directly after high-temperature sintering. Restorations are used in clinical dental settings to replace natural dentition, providing strength, color, and translucency. Benefits include improved esthetic workflow efficiency by eliminating additional coloring steps.
Clinical Evidence
No clinical data. Bench testing only. Verification included flexural strength (ISO 6872:2024), visual shade match, color consistency (ΔE < 2.00), CTE, chemical solubility, radioactivity, and biocompatibility (ISO 10993-1:2025, ISO 10993-5:2009). All results met predetermined acceptance criteria.
Technological Characteristics
Material: Yttria-stabilized zirconia with trace colorants. Form: Milling blanks (GL Single Unit, GL Bridge Block). Principle: CAD/CAM milling followed by high-temperature sintering. Flexural strength: >800 MPa (ISO 6872:2024 Type II, Class 5). CTE: 10-11 x 10^-6/°C. Chemical solubility: <100 μg/cm². Radioactivity: U-238 ≤ 1.0 Bq/g. Sterilization: Non-sterile.
Indications for Use
Indicated for dental technicians constructing custom all-ceramic restorations for anterior and posterior locations.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
January 30, 2026
Prismatik Dentalcraft, Inc.
Park So Hyun
Sr. Regulatory Affairs Manager
2144 Michelson Dr.
Irvine, California 92612
Re: K254283
Trade/Device Name: BruxZir® Shaded 16 PLUS
Regulation Number: 21 CFR 872.6660
Regulation Name: Porcelain powder for clinical use
Regulatory Class: Class II
Product Code: EIH
Dated: December 30, 2025
Received: December 31, 2025
Dear Park So Hyun:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K254283 - Park So Hyun
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K254283 - Park So Hyun
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
MICHAEL E. ADJODHA -S
Michael E. Adjodha, MChE, RAC, CQIA
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
| Indications for Use | | |
| --- | --- | --- |
| Please type in the marketing application/submission number, if it is known. This
textbox will be left blank for original applications/submissions. | K254283 | ? |
| Please provide the device trade name(s). | | ? |
| BruxZir® Shaded 16 PLUS | | |
| Please provide your Indications for Use below. | | ? |
| The device is indicated for use by dental technicians in the construction of custom made all ceramic
restorations for anterior and posterior location. | | |
| Please select the types of uses (select one or both, as
applicable). | ☑ Prescription Use (21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ? |
{4}
PRISMATIK DENTALCRAFT, INC.
# 510(k) Summary - K254283
Date Prepared: December 30, 2025
# CONTACT DETAILS (21 CFR 807.92(a)(1))
Applicant Name: Prismatik Dentalcraft, Inc.
Applicant Address: 2144 Michelson Drive, Irvine, CA 92612, USA
Applicant Contact Telephone: 949-535-3038
Applicant Contact: So Hyun Park, Sr. Regulatory Affairs Manager, MS
Applicant Contact Email: so.park@glidewelldental.com
Secondary Contact: Shelly Gallup, Sr.VP – RA/QA
Secondary Contact Email: shelly.gallup@glidewelldental.com
Secondary Contact Telephone: 949-533-9412
# DEVICE NAME (21 CFR 807.92(a)(2))
Device Trade Name: BruxZir® Shaded 16 PLUS
Common Name: Zirconia milling block or dental CAD/CAM block
Classification Name: Porcelain powder for clinical use
Regulation Number: 872.6660
Regulatory Class: Class II
Product Code: EIH
# LEGALLY MARKETED PREDICATE DEVICE (21 CFR 807.92(a)(3))
Predicate 510(k) #: K130924
Predicate Device Trade Name: BruxZir® Shaded
Product Code: EIH
Reference 510(k) #: K220816
Reference Device Trade Name: BruxZir® NOW
Product Code: EIH
# DEVICE DESCRIPTION SUMMARY (21 CFR 807.92(a)(4))
BruxZir Shaded 16 PLUS blanks are used for the production of full-contour zirconia and zirconia-based substructures for crowns and bridges (restorations). Multiple thicknesses and shades (A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4, BL1, BL3, and White) are
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® Shaded 16 PLUS
December 2025
{5}
PRISMATIK DENTALCRAFT, INC.
available for milling into BruxZir restorations. The manufactured restorations are made utilizing the CAD/CAM system for design and manufacturing. The designed and manufactured restorations are then sintered and glazed. BruxZir Shaded 16 PLUS restorations are designed to match the body shade in the glazed state; however, precolor and stain should be applied if polychromatic (gingival to incisal) blending or other esthetic effects are desired. The sintered material exhibits maximum strength, color, and translucency similar to natural dentition.
# INTENDED USE/INDICATIONS FOR USE (21 CFR 807.92(a)(5))
The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.
# INDICATIONS FOR USE COMPARISON (21 CFR 807.92(a)(5))
The subject device, BruxZir® Shaded 16 PLUS, has the same indications for use as the predicate device, BruxZir® Shaded (K130924). Both devices are indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.
# TECHNOLOGICAL COMPARISON (21 CFR 807.92(a)(6))
The subject device, BruxZir® Shaded 16 PLUS, is substantially equivalent in technical characteristics to the predicate device, BruxZir® Shaded (K130924) in terms of overall design principles, material, and performance.
The subject device, BruxZir® Shaded 16 PLUS, is similar in terms of overall design to the predicate device, BruxZir® Shaded (K130924). The predicate device, BruxZir® Shaded (K130924), is only offered in a milling blank whereas the subject device, BruxZir® Shaded 16 PLUS, is offered in different designs other than a milling blank such as GL Single Unit and GL Bridge Block, which are designed to be used for milling single-unit restorations and bridges (restorations of three or more units), respectively. The fundamental principle of operation of the subject device, BruxZir® Shaded 16 PLUS, and the predicate device, BruxZir® Shaded (K130924), is the same. The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. Although no color needs to be added additionally to either device before sintering, basic staining and glazing techniques need to be used after sintering to achieve the desired shade for the predicate device, BruxZir® Shaded (K130924). On the other hand, the final shades are achievable without additional coloring steps for the subject device, BruxZir® Shaded 16 PLUS.
The subject device, BruxZir® Shaded 16 PLUS, is similar in terms of material composition to the predicate device, BruxZir® Shaded (K130924). The main ceramic component is composed of yttria-stabilized zirconia with varying trace components of colorants. As the generic type of material has not changed, it does not alter the performance specifications of the subject device. The slight differences in overall chemical composition do not affect the biocompatibility and performance characteristics. Also, the change in the manufacturing process from colloidal process (slip casting) to press process does not affect the mechanical properties of the final finished device.
The technological comparison table below outlines and provides the similarities between the subject device, BruxZir® Shaded 16 PLUS, and the predicate device, BruxZir® Shaded (K130924).
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® Shaded 16 PLUS
December 2025
Page 2 of 5
{6}
PRISMATIK DENTALCRAFT, INC.
Both the subject device and the predicate device have similar physical/mechanical and biocompatibility properties that met the requirements of ISO 6875:2024 (Type II, Class 5) and ISO 10993-1:2025 and ISO 10993-5:2009. Any differences between the subject device and the predicate device do not raise any new concerns of safety and effectiveness.
| Attributes | | Subject Device (TBD) | Predicate Device (K130924) | Comparison |
| --- | --- | --- | --- | --- |
| Device Name | | BruxZir® Shaded 16 PLUS | BruxZir® Shaded | N/A |
| Manufacturer | | Prismatik Dentalcraft, Inc. | Prismatik Dentalcraft, Inc. | Same |
| Product Code | | EIH | EIH | Same |
| Prescription Device | | Yes | Yes | Same |
| Indications for Use | | The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location. | The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location. | Same |
| Design Characteristics | Chemical Composition | The main ceramic component is composed of yttria-stabilized zirconia. The different shades have varying trace colorants as required to match the desired shades. | The main ceramic component is composed of yttria-stabilized zirconia. The different shades have varying trace colorants as required to match the desired shades. | Similar; Similar yttria proportion of the main ceramic components (~3Y) and different colorants for the subject device. |
| | Design | Various sizes and thicknesses | Various sizes and thicknesses | Similar; Different sizes and thicknesses are offered for the subject device. |
| | Principle of Operation | The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. No color needs to be added before sintering. The final target shades are achievable without additional coloring steps. | The manufactured dental restorations are made utilizing a CAD/CAM system for design and manufacture. The designed and manufactured dental restorations are then sintered at a high temperature. No color needs to be added before sintering. Basic staining and glazing techniques need to be used after sintering to achieve the desired final target shades. | Similar; The target shades are achievable without additional coloring steps for the subject device. |
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® Shaded 16 PLUS
December 2025
{7}
PRISMATIK DENTALCRAFT, INC.
| Attributes | | Subject Device (TBD) | Predicate Device (K130924) | Comparison |
| --- | --- | --- | --- | --- |
| | Shades | A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4, BL1, BL3, White | 100, 200, 300, 400 | Different; Final VITA shades are achievable without additional coloring steps for the subject device. More shades are offered for the subject device. |
| | Final VITA Shades | A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4, BL1, BL3, White | A1, A2, A3, A3.5, A4, B1, B2, B3, B4, C1, C2, C3, C4, D2, D3, D4 | Similar; More final target shades are offered for the subject device. |
| | Flexural Strength | >800 MPa | >800 MPa | Same |
| | Coefficient of Thermal Expansion (CTE/25-500°C) | 10-11 x 10^{-6}/°C | 10-11 x 10^{-6}/°C | Same |
| | Chemical Solubility | <100 μg/cm^{2} | <100 μg/cm^{2} | Same |
| | Radioactivity | The activity concentration of Uranium-238 is no more than 1.0 Bq/g^{-1}. | The activity concentration of Uranium-238 is no more than 1.0 Bq/g^{-1}. | Same |
| | Biocompatibility | Biocompatible and non-toxic | Biocompatible and non-toxic | Same |
| | Sterility | Non-sterile | Non-sterile | Same |
| | Manufacturing Process | Press process | Colloidal process | Different manufacturing process |
# NON-CLINICAL AND/OR CLINICAL TESTS SUMMARY & CONCLUSIONS (21 CFR 807.92(b))
Design verification and validation were conducted by Prismatik Dentalcraft, Inc. The tests conducted are listed below.
- Flexural Strength according to ISO 6872:2024
- Visual Shade Match
- Color Consistency
- Coefficient of Thermal Expansion according to ISO 6872:2024
- Chemical Solubility according to ISO 6872:2024
- Radioactivity according to ISO 6872:2024
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® Shaded 16 PLUS
December 2025
{8}
PRISMATIK DENTALCRAFT, INC.
- Packaging Validation according to ASTM D4169-23e1
- Biocompatibility according to ISO 10993-1:2025 and ISO 10993-5:2009
No clinical data is included in this submission.
The test samples were manufactured using the standard, validated manufacturing and quality control processes for BruxZir® Shaded 16 PLUS to ensure that the final finished devices met the following specified acceptance criteria.
- Flexural Strength must be greater than 800 MPa after sintering, as measured per ISO 6872:2024, Type II Class 5.
- Grade of “Pass” for Visual Shade Match for all shades.
- Color consistency within a block, between blocks of the same batch and between batches – ΔE calculation needs to be less than 2.00 between samples measured.
- The Coefficient of Thermal Expansion must be between 10⁻¹¹ × 10⁻⁶/°C and the standard deviation of CTE must be no greater than 0.5 × 10⁻⁶/°C per ISO 6872:2024.
- The solubility must be lower than 100 µg/cm² per ISO 6872:2024.
- The activity concentration of Uranium-238 must not exceed 1.0 Bq·g⁻¹ per ISO 6872:2024.
- Must be biocompatible, safe, and non-toxic.
The results of the verification and validation activities confirmed that the design of the subject device met all the predetermined criteria according to the design specifications. The testing results were used to support substantial equivalence by addressing differences in technological characteristics between the subject device, BruxZir® Shaded 16 PLUS, and the predicate device, BruxZir® Shaded (K130924). Testing confirmed that the modified device met all the design specifications and that the design changes were implemented successfully without introducing any unexpected issues.
The subject device, BruxZir® Shaded 16 PLUS, is substantially equivalent to the predicate device, BruxZir® Shaded (K130924), with respect to intended use, indications for use, fundamental technological characteristics, including design, materials and principles of operation. Based on this information, these modifications do not raise any new questions of safety and effectiveness when compared to the predicate device.
Prismatik Dentalcraft, Inc. 510(k) Notification
BruxZir® Shaded 16 PLUS
December 2025
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.