K Number
K130924

Validate with FDA (Live)

Device Name
BRUXZIR SHADED
Date Cleared
2013-05-16

(43 days)

Product Code
Regulation Number
872.6660
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

Device Description

BruxZir® Shaded Blanks are used for the production of full-contour zirconia and zirconia-based substructures for crowns and bridges. Four (4) thicknesses (12 mm, 15 mm, 20 mm, 25 mm) and four (4) shades (100, 200, 300, 400) are available for milling into BruxZir full-contour crowns and substructures.

AI/ML Overview

The provided document is a 510(k) summary for a dental device, BruxZir® Shaded. It describes the device, its intended use, and a comparison to a predicate device to establish substantial equivalence. However, this document does not contain a detailed study proving the device meets specific acceptance criteria in the way typically required for AI/software-as-medical-device (SaMD) evaluations.

The "Acceptance Criteria" and "Study" described in this document are focused on the non-clinical performance and biocompatibility of a physical dental material, rather than the performance of an AI algorithm or a SaMD.

Here's an attempt to extract and interpret the information based on the provided text, while also explicitly stating what is not present given the context of AI/SaMD.

1. Table of Acceptance Criteria and Reported Device Performance

Test Description (Acceptance Criteria)Reported Device Performance
Cytotoxicity Study (ISO 10993-5)Pass (no reaction on cells)
Sensitization Study (ISO 10993-10)Pass (no reaction on tested subject)
Irritation Study (ISO 10993-10)Pass (no erythema or edema on test subject)
Mechanical Properties: Flexural StrengthSuperior, claimed (not quantified against a specific benchmark in this section)
Mechanical Properties: Fracture Toughness (KIC)Superior, claimed (not quantified against a specific benchmark in this section)
Mechanical Properties: Resistance to Thermal ShockGreat, claimed (not quantified against a specific benchmark in this section)
Esthetics: Color and TranslucencyMost innovative, higher translucency (not quantified against a specific benchmark in this section)

Note on "Acceptance Criteria": For the mechanical and esthetic properties, while "superior," "great," and "higher translucency" are stated, specific numerical acceptance criteria (e.g., minimum flexural strength in MPa) and corresponding numerical test results are not provided in this summary. The "Pass" for biocompatibility tests implies meeting the ISO standard requirements.

2. Sample size used for the test set and the data provenance

  • Test Set (Biocompatibility): The document states "tested subject" for sensitization and irritation. The exact sample size (number of subjects/samples) for these biocompatibility tests is not specified in the provided text.
  • Data Provenance: The tests are described as "Non-clinical test data" and "laboratory testing." The country of origin is not specified. The nature is retrospective or prospective is not specified but typically, such non-clinical lab tests are conducted prospectively on newly manufactured material samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This question is not applicable to the provided document. The "device" in this context is a physical dental material (zirconia blocks), not an AI system that requires expert-established ground truth for performance evaluation in an AI/SaMD context. The biocompatibility tests are based on standardized laboratory protocols (ISO 10993 series), not expert consensus on interpretations.

4. Adjudication method for the test set

This question is not applicable as it pertains to expert disagreement resolution in diagnostic or interpretive tasks, which is not relevant for the non-clinical material testing described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This question is not applicable. The device is a dental restoration material, not an AI system. Therefore, no MRMC study involving human readers or AI assistance would have been performed or is relevant to this submission.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This question is not applicable. The device is a dental restoration material, not an AI algorithm.

7. The type of ground truth used

  • For Biocompatibility: The "ground truth" for the biocompatibility tests (Cytotoxicity, Sensitization, Irritation) is implicitly defined by the ISO 10993 standards (ISO 10993-5 and ISO 10993-10). A "Pass" result indicates that the material's reaction (or lack thereof) met the established criteria within these international standards, which serve as the reference or "ground truth" for biocompatibility.
  • For Mechanical/Esthetic Properties: The document claims "superior microstructure and homogeneity," "higher translucency value," "superior mechanical properties," "high flexural strength," "high fracture toughness," and "great resistance to thermal shock." While these are performance claims, the specific "ground truth" against which these were measured (e.g., a specific reference material, a numerical threshold from a standard) is not explicitly detailed for each claim in this summary.

8. The sample size for the training set

This question is not applicable. The device is a dental restoration material; there is no AI algorithm being "trained."

9. How the ground truth for the training set was established

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

B.

C.

Trade/Proprietary Name:

Page 24 of 142

K130924

MAY 1 6 2013

006-510 (K) Summary-807.92(c)

This 510 (K) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR 807.92.

SUBMITTER INFORMATION A.

Company Name:Prismatik Dentalcraft, Inc.
Company Address:2212 Dupont Dr., Suite IJK,Irvine, CA 92612
Company Phone:949-225-1269
Company FAX:949-553-0924
Facility Registration Number:3005477956
Primary Contact Person:Armin Zehtabchi, (949) 225-1234Senior RA Specialist
Secondary Contact PersonMarilyn Pourazar, (949) 225-1269Senior Director, RA/QA
Date Summary Prepared:April 12, 2013
DEVICE IDENTIFICATION
Trade/Proprietary Name:BruxZir® Shaded
21 CFR Reference:21 CFR 872.6660
21 CFR Common Name:Porcelain powder for clinical use
Classification:Class II
Panel:Dental
IDENTIFICATION OF PREDICATE DEVICE

Prismatik™ Clinical Zirconia (Prismatik™ CZ)-K060104

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left side and the company name on the right side. The company name is written in a bold, sans-serif font.

DEVICE DESCRIPTION D.

BruxZir® Shaded Blanks are used for the production of full-contour zirconia and zirconia-based substructures for crowns and bridges. Four (4) thicknesses (12 mm, 15 mm, 20 mm, 25 mm) and four (4) shades (100, 200, 300, 400) are available for milling into BruxZir full-contour crowns and substructures.

E. DEVICE MANUFACTURING PROCESS:

BruxZir® Shaded are manufactured using a colloidal processing without mechanical pressing or the use of organic binders which lead to decreased esthetics. Other dental Zirconia materials on the market today use organic binders to hold the powder together during the high pressure pressing to form the milling blanks. Conversely, BruxZir®'s colloidal process utilizes vacuum slip casting to form the blocks. This process creates a block with superior microstructure and homogeneity when compared to pressed blocks.

Vacuum slip casting is a ceramic forming technique which uses ceramic slip and a porous mold made from Plaster of Paris. The ceramic powder consolidates and accumulates on the top of the porous mold while the water is absorbed into it, thus forming a high density green state block. One of the several advantages for the colloidal process employed in the creation of BruxZir® products is that smaller particles dispersed using TMAOH and are compacted homogenously through the colloidal process. This results in a material that has a higher translucency value. This allows the end product to have a closer appearance to a natural tooth.

The BruxZir® Shaded block is available in different shapes such as Standard, D Blocks and Z Blocks. The sizes are: 12mm, 15mm, 20mm, 25mm, and 30mm. The drawings of different blocks are provided in this submission.

F. INDICATIONS FOR USE

The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

SUBSTANTIAL EQUIVALENCE G.

The following comparison chart outlines and provides the similarities and the substantial equivalency of the Prismatik™ Clinical Zirconia (Prismatik™ CZ)-K060104 and the BruxZir® Shaded. Prismatik believes that the comparative data presented in the preceding paragraphs, demonstrate that BruxZir® Shaded is essentially the same as currently marketed devices for the same indication for use, and supports our claim of substantial equivalence to predicate Class II devices under the classification of Porcelain powder for clinical use (21 CFR 872.6660) that have previously been found to be substantially equivalent, and that any differences between the BruxZir® Shaded and the predicate device do not introduce any new issues of safety or effectiveness. Both the modified device and the predicate device consist of general porcelain powder material (Product Code:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left and the company name on the right. The company name is written in a bold, sans-serif font.

EIH). Both the predicate device and the modified device have the same intended use.

Elements ofComparisonPrismatik™ Clinical Zirconia(Prismatik™ CZ)-K060104(predicate device)PrismatikBruxZir® Shaded (modifieddevice)
General MaterialPowder, porcelainSame
Raw MaterialSupplierTosoh CorporationSame
IndicationsThe device is indicated for useby dental technicians in theconstruction of custom madeall ceramic restorations foranterior and posterior locationSame
BiocompatibilityYesSame
SterilityNon-sterileSame
Labeling of theIntended UseSameSameA minor change to the name fromPrismatik™ Clinical Zirconia(Prismatik™ CZ to BruxZir® Shaded
DesignVarious Sizes and Thicknesses12mm, 15mm, 20mm, 25mm,and 30mmSame
PerformancePerformance and technologicalcharacteristics have beenpreviously cleared for market(K060104)Same

Comparison of Predicate Devices

NON-CLINICAL TESTING H.

Applicable testing was performed in accordance with the Design Verification Plan including a Risk Analysis addressing the impact of changes to the device. The test results indicate the Prismatik BruxZir® Shaded Blank is comparable to the predicate device.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is a triangle with a smaller triangle cut out of it. The company name is written in a bold, sans-serif font.

I.

BIOCOMPATIBILITY

Non-clinical test data was used to support the substantial equivalency.

The functionality of the BruxZir® Shaded Blocks as well as their conformance to design input was further determined by laboratory testing. BruxZir® Shaded has been tested for Cytotoxicity (ISO 10993-5), Sensitization (ISO 10993-10), and Irritation (ISO 10993-10) tests to meet the biocompatibility requirement and the reports are as follow:

  • · The Cytotoxicity Report shows that there was no reaction on any of the cells.
  • · The Sensitization Report shows that there was no reaction on the tested subject.
  • · The Irritation Report shows that there was no erythema or edema on the test subject.
Test DescriptionResults
Cytotoxicity Study using the IX MEM extractionmethod at 37°CPass
ISO Intracutaneous Study, Extract 0.9% sodiumchloride USP solution (SC) and sesame oil, NF (SO)Pass
Irritation and Skin Sensitization Study, Extract 0.9%sodium chloride USP and sesame oil, NF (SO)Pass

The BruxZir® product is manufactured from TZ-Series of zirconia powders free of organic binder material which exhibits superior mechanical properties, and it is a great material for dental restorations for the following four (4) physical properties such as high flexural strength, high fracture toughness (KIC), great resistance to thermal shock and most innovative, is the color and translucency.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is centered horizontally in the image. The background of the image is white.

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 16, 2013

Mr. Armin Zehtabchi Senior Regulatory Affairs Specialist Prismatik Dentalcraft, Incorporated 2212 Dupont Drive, Suite IJK IRVINE CA 92612

Re: K130924

Trade/Device Name: BruxZir® Shaded Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: II Product Code: EIH Dated: April 12, 2013 Received: April 16, 2013

Dear Mr. Zehtabchi:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

Page 2 - Mr. Zehtabchi

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kwame Ulmer -
S FDA

for

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Respiratory. Infection Control and

Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a black triangle on the left and the company name on the right. The triangle is partially obscured by a white shape, creating a prism effect. The text is in a bold, sans-serif font.

005-Indications for Use Statement

510 (K) Number (if known): K130924

BruxZir® Shaded Device Name:

Indications for Use: The device is indicated for use by dental technicians in the construction of custom made all ceramic restorations for anterior and posterior location.

Prescription Use: Yes No No (Part 21 CFR 801 Subpart D)

Over-the-Counter Use: Yes | No > (Part 21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Runner, DOSMA
-----------------------
Digitally signed by Mary S. Runner -
S
DN: c=US, o=U.S. Government,
ou=HHS, ou=FDA, ou=People,
cn=Mary S. Runner S,
0.9.2342.19200300.100.1.1=130008
-7950
Date: 2015.03.08 12:36:47-04'00

(Division Sign-Off)

Division of Anesthesiology, General Hospital

Infection Control, Dental Devices

510(k) Number:K130924
-------------------------

§ 872.6660 Porcelain powder for clinical use.

(a)
Identification. Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.(b)
Classification. Class II.