High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors

K220179 · Foshan Coxo Medical Instrument Co., Ltd. · EFB · Sep 22, 2023 · Dental

Device Facts

Record IDK220179
Device NameHigh-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors
ApplicantFoshan Coxo Medical Instrument Co., Ltd.
Product CodeEFB · Dental
Decision DateSep 22, 2023
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.4200
Device ClassClass 1
AttributesTherapeutic

Indications for Use

High-Speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Straight Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Geared Angle Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Air Motors: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth.

Device Story

Hand-held dental instruments; air-motor driven; used to hold rotary attachments (burs/drills) for cutting, grinding, and drilling. Operated by dental professionals in clinical settings. Inputs: compressed air; some models include electric connection for LED illumination. Outputs: mechanical rotation of dental burs. Device facilitates tooth preparation and restoration procedures. Benefits: enables precise removal of dental material and finishing of restorations.

Clinical Evidence

No clinical data. Bench testing only, including biocompatibility (ISO 10993-5, -10, -23, -11), electrical safety (IEC 60601-1, -1-2, 80601-2-60), and performance testing per ISO 14457.

Technological Characteristics

Materials: copper (chromium plated), aluminum, rubber, stainless steel. Sensing/Actuation: air-driven turbine/motor. Connectivity: some models include electric connection for LED. Sterilization: moist heat (ISO 17665-1). Standards: ISO 14457, ISO 9168, ISO 17664, ANSI/AAMI ES60601-1.

Indications for Use

Indicated for use by trained dental professionals for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations, and polishing teeth.

Regulatory Classification

Identification

A dental handpiece and accessories is an AC-powered, water-powered, air-powered, or belt-driven, hand-held device that may include a foot controller for regulation of speed and direction of rotation or a contra-angle attachment for difficult to reach areas intended to prepare dental cavities for restorations, such as fillings, and for cleaning teeth.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 22, 2023 Foshan COXO Medical Instrument Co., Ltd. % Ray Wang Official Correspondent Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District Beijing, Beijing 102401 China Re: K220179 Trade/Device Name: High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors Regulation Number: 21 CFR 872.4200 Regulation Name: Dental Handpiece And Accessories Regulatory Class: Class I, reserved Product Code: EFB, EGS Dated: September 1, 2023 Received: September 1, 2023 Dear Ray Wang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Bobak Shirmohammadi -S For Michael E. Adjodha, M. ChE., CQIA Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use ## 510(k) Number (if known) K220179 ## Device Name High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/Air Motors ## Indications for Use (Describe) High-Speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Straight Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. ## Geared Angle Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Air Motors: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. **Type of Use (Select one or both, as applicable)** | <span style="font-size: 12px;"> <span style="text-decoration: overline;">☑</span> Remediation Use (Part 21 CFR 601.3 Subpart D)</span> | |----------------------------------------------------------------------------------------------------------------------------------------| | <span style="font-size: 12px;">☐ Own-Use Exception (21 CFR 601.3 Subpart C)</span> | > Prescription Use (Part 21 CFR 801 Subpart D) | | Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) Summary This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92. The assigned 510(k) Number: K220179 - Date of Preparation 1. 09/21/2023 - 2. Sponsor # Foshan COXO Medical Instrument Co., Ltd. BLDG 4, District A Guangdong New Light Source Industrial Base, South of Luocun Avenue Nanhai District Foshan 528226 Guangdong China Contact Person: Zheng yongjian Position: Legal Representative Tel: 13702544788 Fax: 0757-81800058 Email: 13702544788@163.com - 3. Submission Correspondent Mr. Ray Wang Beijing Believe-Med Technology Service Co., Ltd. Rm.912, Building #15, XiYueHui, No.5, YiHe North Rd., FangShan District, Beijing City, China, 102401 Tel: +86-18910677558 Fax: +86-10-56335780 Email: information@believe-med.com {4}------------------------------------------------ - 4. Identification of Proposed Device Trade Name: High-Speed Air Turbine Handpieces/Straight Handpieces/Geared Angle Handpieces/ Air Motors Common Name: Dental Handpiece and Accessories Model(s): High-Speed Air Turbine Handpieces CX207, CX207-G, CX207-2, CX207-A, CX207-A-2, CX207-B, CX207-B-2, CX207-C, CX207-C-2, CX207-W, CX207-W-2 Straight Handpieces CX235-2, CX235-2A, CX235-2B, CX235-2F, CX235-2G, CX235-2C, CX235- 2S2 Geared Angle Handpieces CX235-1B, CX235-1C, CX235-1F, CX235-1G, CX235C1, CX235C2, CX235C3, CX235C4, CX235C5, CX235C6, CX235C7, CX235C8, CX235-2S, CX235-2S1 Air Motors: CX235-3B, CX235-3F, CX235-3C Regulatory Information: Regulation Name: Dental Handpiece And Accessories Classification: I; Product Code: EFB; Additional Product Code: EGS; Regulation Number: 21 CFR 872.4200; Review Panel: Dental; Indications for Use Statement: ### High-Speed Air Turbine Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. ## Straight Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. #### Geared Angle Handpieces: It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. Air Motors: {5}------------------------------------------------ It is intended for removing carious material, excess filling material, cavity and crown preparation, finishing tooth preparations and restorations, root canal preparations and polishing teeth. #### Device Description 5. The proposed device is used to hold rotary instruments such as burs and drills for cutting, grinding, and drilling operations. The dental handpieces are hand-held instruments driven by air motor, thus driving the rotation of dental attachments for the purpose of drilling and tooth cutting. Some models include an electric connection to power an LED light #### 6. Identification of Predicate Device Primary Predicate Device: 510(k) Number: K181691 Product Name: High-Speed Air Turbine Handpiece / Low-Speed Air Turbine Handpiece Manufacturer: Foshan Wenjian Medical Instrument Co., Ltd. Reference Device: 510(k) Number: K202786 Product Name: Dental Low-speed Handpieces and Accessories Manufacturer: MicroP Technology (Taiwan), Inc. - Non-Clinical Test Conclusion 7. Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards: - > ANSI/AAMI ES60601-1:2005/(R)2012 And A1:2012, Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance; - > IEC 60601-1-2:2014, Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance- Collateral standard: Electromagnetic compatibility-Requirements and tests: - > IEC 80601-2-60: 2019. Medical electrical equipment - Part 2-60: Particular requirements for the basic safety and essential performance of dental equipment; - ISO 10993-5:2009, Biological Evaluation of Medical Device-Part 5: Tests for Vitro > cytotoxicity; - > ISO 10993-10:2021, Biological Evaluation of Medical Device-Part 10: Test for irritation and delay-type hypersensitivity: - > ISO 10993-23:2021, Biological Evaluation of Medical Device – Part 23: Tests for irritation: - > ISO 10993-11:2017, Biological Evaluation of Medical Device – Part 11: Test for systemic toxicity; - > Pyrogen Study in Rabbits using USP General Chapter <151> Pyrogen Test; - A ISO 14457 Second edition 2017-10 Dentistry - Handpieces and motors; {6}------------------------------------------------ - > ISO 17664 Second edition 2017-10 Processing of health care products - Information to be provided by the medical device manufacturer for the processing of medical devices - > ISO 17665-1 First edition 2006-08-15 Sterilization of health care products - Moist heat - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices - > ISO 9168 Third edition 2009-07-15 Dentistry - Hose connectors for air driven dental handpieces - 8. Clinical Test Conclusion No clinical study is included in this submission. {7}------------------------------------------------ # 9. Substantially Equivalent (SE) Comparison | ITEM | Proposed Device | Primary Predicate Device | Reference Device | Remark | | |---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|-------------| | Manufacturer | Foshan COXO Medical Instrument Co., Ltd. | Foshan Wenjian Medical<br>Instrument Co., Ltd. | MicroP Technology (Taiwan), Inc. | -- | | | Device Name | High-Speed Air Turbine Handpieces/ Straight Handpieces/ Geared Angle<br>Handpieces/ Air Motors | High-speed air turbine handpiece<br>Low-speed air turbine handpiece | Dental Low-speed Handpieces and<br>Accessories | -- | | | Model | High-Speed Air Turbine Handpieces:CX207, CX207-G, CX207-2,<br>CX207-A, CX207-A-2, CX207-B, CX207-B-2, CX207-C, CX207-C-2,<br>CX207-W, CX207-W-2<br>Straight Handpieces: CX235-2, CX235-2A, CX235-2B, CX235-2F,<br>CX235-2G, CX235-2C, CX23 5-2S2<br>Geared Angle Handpieces: CX235-1B, CX235-1C, CX235-1F,<br>CX235-1G, CX235C1, CX235C2, CX235C3, CX235C4, CX235C5,<br>CX235C6, CX235C7, CX235C8, CX235-2S, CX235-2S1<br>Air Motors:CX235-3B, CX235-3F, CX235-3C | Dental High-speed Turbine<br>Handpiece (WJ-114, WJ- 112,<br>WJ-124, WJ -122, WJ- 134, WJ<br>-132, WJ - 144, WJ -142, WJ<br>-154, WJ - 152, WJ -164, WJ<br>-162)<br>Dental Low-speed Turbine<br>Handpiece (WJ-414, WJ- 412,<br>WJ-424, WJ-422) | Contra-angle Handpieces: G25, G25L,<br>G23<br>Straight Handpieces: G65, G65L, G63 | -- | | | Product Code | EFB, EGS | EFB, EGS | EFB, EGS | SAME | | | Regulation No. | 21 CFR 872.4200 | 21 CFR 872.4200 | 21 CFR 872.4200 | SAME | | | Class | I | I | I | SAME | | | Indication | High-Speed Air Turbine Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>Straight Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth. | Dental High-speed Turbine<br>Handpiece (WJ-114, WJ- 112,<br>WJ-124, WJ -122, WJ- 134, WJ<br>-132, WJ - 144, WJ -142, WJ<br>-154, WJ - 152, WJ -164, WJ<br>-162) is intended for removing<br>carious material, excess filling<br>material, cavity and crown<br>preparation, finishing tooth<br>preparations and restorations, root | MicroP Dental Low-speed Handpieces<br>and Accessories is intended for removing<br>carious material, excess filling material,<br>cavity and crown preparation, finishing<br>tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>All the devices are designed for use by a<br>trained professional in the field of<br>general dentistry | SAME | | | | Geared Angle Handpieces:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth.<br>Air Motors:<br>It is intended for removing carious material, excess filling material, cavity<br>and crown preparation, finishing tooth preparations and restorations, root<br>canal preparations and polishing teeth. | canal preparations and polishing<br>teeth.<br>Dental Low-speed Turbine<br>Handpiece (WJ-414, WJ- 412,<br>WJ-424, WJ-422) is intended for<br>removing carious material, excess<br>filling material, cavity and crown<br>preparation, finishing tooth<br>preparations and restorations, root<br>canal preparations and polishing<br>teeth. | | | | | Air/water ports | 2/4/6 ports | 2/4 ports | up to four | DIFFERENT 1 | | | Fiberoptic Glass Rod | CX207-G, CX235-1C, CX235C7 | Included in model WJ-162,<br>WJ-164, WJ-424<br>and WJ- 422Disposable | Included in model G25Land G65L | SAME | | | Accessories | NA | NA | NA | SAME | | | Composition of material | Copper (chromium plating), aluminum , rubber and stainless steel | Stainless steel, Brass, Aluminum,<br>Titanium | Stainless steel, Aluminum Alloy | DIFFERENT 2 | | | Chuck | Push button, Screw | Push button, Screw | Push button | SAME | | | Bur extraction force | 22N | 28N | / | DIFFERENT 3 | | | Maximum air pressure | High-speed Air Turbine Handpieces:0.2-0.25Mpa<br><br>Straight handpiece & Geared Angle handpiece & Air Motors:0.245MPa<br>~0.3925MPa | Dental High-speed Turbine<br>Handpiece: 200kPa ~ 250kPa<br>(29.01 psi~ 36.25 psi) Dental<br><br>Low-speed Turbine Handpiece:<br>245 ~ 392 kPa (35.53 psi~ 56.85<br>psi) | 0.248-0.296MPa<br>(36psi~43psi) | SAME | | | Maximum<br>pressure | water | High-Speed Air Turbine Handpieces:<br>0.2-0.25Mpa<br><br>Straight handpiece & Geared Angle handpiece & Air Motors:0.245MPa<br>~0.3925MPa | Dental High-speed Turbine<br>Handpiece: 200kPa ~ 250kPa<br>(29.01 psi~ 36.25 psi)<br><br>Dental Low-speed Turbine<br>Handpiece: 245 ~ 392 kPa (35.53<br>psi~ 56.85 psi) | 0.248-0.296MPa<br>(36psi~43psi) | SAME | | Speed in rpms | | High-speed Air Turbine Handpieces:280000 ~<br>450000<br><br>Straight handpiece &Geared Angle handpiece<br>&Air Motors: <40000rpm | Dental High-speed Turbine<br>Handpiece: 300,000rpm ~<br>400,000rpm<br><br>Dental Low-speed Turbine<br>Handpiece: 18,000 ~ 20,000 rpm | <40000rpm | DIFFERENT 4 | | Conformance<br>with<br>standards for shanks | | Type 3 per ISO 1797-1 | Type 3 per ISO 1797-1 | / | SAME | | Coupling dimensions | | Complied with ISO 3964 | Complied with ISO 3964 | / | SAME | | Hose connections | | Type 1 for 2-hole model / Type 2 for 4- hole<br>model/Type 3 for 6-hole model | Type 1 for 2-hole model / Type 2<br>for 4- hole model | / | SIMILAR | | Lubricant | | The lubricants approved by FDA, such as DO-ALL Dental Handpiece<br>Lubricant<br>(K073353). | The specified lubricant, type<br>"PANA SPRAY Plus"<br>manufactured by NAKANISHI<br>INC (cleared in K163483), must<br>be used during routine<br>maintenance. | MicroP Halley Spray | SIMILAR | | Compliance Standards | | Complied with ISO 10993- 5, ISO 10993-10, ISO 10993-23, ISO<br>10993-11, USP <151> Pyrogen Testing, IEC 60601-1, IEC 60601-1-2, IEC<br>80601- 2-60, ISO14457, ISO 17665-1, ISO 17664, ISO 9168 | Complied with ISO 10993- 5, ISO<br>10993-10, IEC 60601-1, IEC<br>80601- 2-60, ISO14457, ISO<br>17665, ISO 11138, ISO 11607,<br>ISO 17664 | Complied with ISO 10993- 5, ISO<br>10993-10, ISO14457 | SAME | # Table 7-3 General Comparison Page 5 of 8 {8}------------------------------------------------ {9}------------------------------------------------ {10}------------------------------------------------ ### 10. Substantially Equivalent (SE) Conclusion ## DIFFERENT 1 The difference between proposed device and predicate device is Air/water ports, which is use for connection, the difference does not affect the safety and effectiveness of proposed device. Based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device. #### DIFFERENT 2 The Handpiece is tissue-contacting, so it is tested accordance with the requirements of ISO 10993-1 and guidelines. The test results meet the requirements. Based on the biological test results, the proposed device is as safe, and performs as well as the legally marketed predicate device. ## DIFFERENT 3 The difference between proposed device and predicate device is bur extraction force. The ISO 14457 required the bur extraction force shall be at least 22 N for Type 5. Based on the nonclinical tests performed, the proposed device meets the standard and is as safe, as effective, and performs as well as the legally marketed predicate device. ## DIFFERENT 4 The difference between proposed device and predicate device is Speed. The proposed device's speed range is covered by the Predicate Device's speed range. Therefore, the proposed device's speed range can meet the use requirement. We tested as ISO 14457, based on the nonclinical tests performed, the proposed device is as safe, as effective, and performs as well as the legally marketed predicate device. Conclusion: Based on the comparison of technological characteristics, demonstrated through bench testing and intended use, the proposed device is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...