K252090 · JJGC Indústria e Comércio de Materiais Dentários S.A. · DZE · Dec 12, 2025 · Dental
Device Facts
Record ID
K252090
Device Name
Pterygoid Indication for GM Helix Implants
Applicant
JJGC Indústria e Comércio de Materiais Dentários S.A.
Product Code
DZE · Dental
Decision Date
Dec 12, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
GM Helix Implant: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Helix Implant System with 3.75 and 4.0 diameter and 18 mm length are indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Implants in the pterygoid indication should be used in splinted applications that utilize at least two implants. GM Helix LG Implant: The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. The GM Helix LG Implant System is indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. GM Mini Conical Abutment 45°: The Mini Conical Abutments 45° and 60° are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. GM Mini Conical Abutment 45° Slim: The Mini Conical Abutments 45° and 52° are indicated for surgical procedures in Zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
Device Story
Device consists of threaded root-form dental implants (GM Helix and GM Helix LG) and associated 45° conical abutments. Implants are surgically placed in the pterygoid region of the upper jaw by dental surgeons to provide anchorage for prosthetic teeth in patients with severe jaw resorption. Implants are used in splinted, multiple-unit configurations. The system utilizes a Grand Morse (GM) taper connection. The device provides mechanical support for prosthetic restorations, restoring chewing function and esthetics. Clinical benefit is derived from the ability to anchor prostheses in atrophic bone without extensive grafting. The device is used in a clinical setting by a dentist or oral surgeon.
Clinical Evidence
Clinical literature review provided to support pterygoid placement. Studies (retrospective and cohort) evaluated survival rates of pterygoid implants (diameters 3.3-5.0mm, lengths 6-28mm) in atrophic maxilla. Reported survival rates ranged from 88.2% to 100% over follow-up periods up to 10 years. Bench testing included dynamic fatigue testing per ISO 14801 and insertion torque testing, demonstrating performance equal to or better than the predicate.
Technological Characteristics
Implants: Titanium grade 4. Abutments: Titanium alloy. Connection: Grand Morse (GM) taper with internal hex. Implant diameters: 3.75 and 4.0 mm. Implant lengths: 18-25 mm. Sterilization: Gamma irradiation (implants), Ethylene Oxide (abutments). Standards: ISO 14801 (fatigue), ISO 10993-1/5/18 (biocompatibility), ISO 11137 (gamma sterilization), ISO 11135 (EO sterilization).
Indications for Use
Indicated for edentulous or partially edentulous patients requiring oral rehabilitation in the upper or lower jaw. Specifically indicated for surgical installation in the pterygoid region (upper jaw) in cases of severe jaw resorption for multiple unit restorations using splinted applications (minimum two implants) with 45° angulation abutments.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
December 12, 2025
JJGC Indústria e Comércio de Materiais Dentários S.A.
% Jennifer Jackson
Sr. Director, Regulatory Affairs and Quality
Straumann USA, LLC
60 Minuteman Road
Andover, Massachusetts 01810
Re: K252090
Trade/Device Name: Pterygoid Indication for GM Helix Implants
Regulation Number: 21 CFR 872.3640
Regulation Name: Endosseous Dental Implant
Regulatory Class: Class II
Product Code: DZE, NHA
Dated: November 12, 2025
Received: November 12, 2025
Dear Jennifer Jackson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K252090 - Jennifer Jackson
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-
{2}
K252090 - Jennifer Jackson
Page 3
assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Sherrill Lathrop Elitzer
for Andrew Steen
Assistant Director
DHT1B: Division of Dental and ENT Devices
OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120
Expiration Date: 07/31/2026
See PRA Statement below.
510(k) Number (if known)
K252090
Device Name
Pterygoid Indication for GM Helix Implants
Indications for Use (Describe)
GM Helix Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Helix Implant System with 3.75 and 4.0 diameter and 18 mm length are indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Implants in the pterygoid indication should be used in splinted applications that utilize at least two implants.
GM Helix LG Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. The GM Helix LG Implant System is indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
GM Mini Conical Abutment 45°:
The Mini Conical Abutments 45° and 60° are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
GM Mini Conical Abutment 45° Slim:
The Mini Conical Abutments 45° and 52° are indicated for surgical procedures in Zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
FORM FDA 3881 (8/23)
Page 1 of 2
PSC Publishing Services (301) 443-6740
{4}
This section applies only to requirements of the Paperwork Reduction Act of 1995.
**DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.**
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/23)
Page 2 of 2
{5}
JJGC Indústria e Comércio de Materiais Dentários S.A.
13-1
# Traditional 510(k) Submission
# Pterygoid Indication for GM Helix Implants
K252090
# 13. 510(k) Summary
## Submitter's Contact Information
Submitter:
Straumann USA, LLC
60 Minuteman Road
Andover, MA 01810, USA
Registration No.: 1222315
Owner/Operator No.: 9005052
On the behalf of:
JJGC Indústria e Comércio de Materiais Dentários S.A (dba Neodent)
Av. Juscelino Kubitschek de Oliveira, 3291
Curitiba, Paraná, Brazil 81270-200
Registration No.: 3008261720
Owner/Operator No.: 10031702
Contact Person:
Jennifer M. Jackson, MS, RAC
Sr. Director, Regulatory Affairs and Quality
Phone Number: +1 978 747-2509
Fax Number: +1 978 747-0023
Email: jennifer.jackson@straumann.com
Prepared By &
Alternate Contact:
Leticia Milani
Senior Regulatory Affairs Analyst
Date of Submission:
December 12, 2025
## Name of the Device
Trade Names: Pterygoid Indication for GM Helix Implants
Common Name: Pterygoid Indication for GM Helix Implants
Classification Name: Implant, Endosseous, Root-Form
Regulation Number: 21 CFR 872.3640
Device Classification: 2
Product Code(s): DZE; NHA
Classification Panel: Dental
{6}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
# Predicate Device(s)
## Primary Predicate:
- K212785 – Blue Sky Bio Dental Implant System
## Reference Devices:
- K163194 – Neodent Implant System – GM Line
- K190958 – Neodent Implant System
- K190718 – Neodent Implant System – GM Zygomatic Implant System
- K232099 – Neodent Implant System – GM Zygomatic Implant System
- K182620 – MRI Compatibility for Existing Neodent Implant System
## Device Description
This premarket notification seeks to expand the indications for use of the subject implants and abutments, in order to include anchorage in the pterygoid region. These devices were previously cleared under 510(k) K163194, K190958, K190718 and K232099 for the functional and esthetic oral rehabilitation of the upper or lower jaw in edentulous or partially edentulous patients. The proposed implants and abutments remain with the same design and features already cleared in the original submission.
## Indications for Use
GM Helix Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Helix Implant System with 3.75 and 4.0 diameter and 18 mm length are indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Implants in the pterygoid indication should be used in splinted applications that utilize at least two implants.
JJGC Indústria e Comércio de Materiais Dentários S.A.
{7}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
GM Helix LG Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. The GM Helix LG Implant System is indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function.
GM Mini Conical Abutment 45°:
The Mini Conical Abutments 45° and 60°* are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
GM Mini Conical Abutment 45° Slim:
The Mini Conical Abutments 45° and 52° are indicated for surgical procedures in Zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly.
* The GM Mini Conical Abutment 60° was cleared for use with zygomatic implants in K203542.
JJGC Indústria e Comércio de Materiais Dentários S.A.
{8}
JJGC Indústria e Comércio de Materiais Dentários S.A.
13-4
# Traditional 510(k) Submission
## Pterygoid Indication for GM Helix Implants
K252090
## Technological Characteristics
The technological characteristics of the subject devices are compared to the primary predicate and reference devices in the Table 1.
The purpose of this submission is addition of the pterygoid indication. The subject implants and abutments have equivalent indications for use statements to the primary predicate (K212785). Specifically, when used in the pterygoid region the devices are all indicated for multiple unit restorations in splinted applications that utilize at least two implants in cases of severe jaw resorption in order to restore patient esthetics and chewing function. The proposed implant diameters (3.75 and 4.0mm) and implant lengths (GM Helix Implant -18mm and GM Helix LG Implants - 20, 22.5 and 25mm) for use in the pterygoid region are consistent with those specified for use in the pterygoid region in K212785. Additionally, the abutments specified for use in the pterygoid region (GM Mini Conical Abutment 45° and GM Mini Conical Abutment 45° Slim) have an angulation of 45° consistent those specified for use in the pterygoid region in K212785. The subject Neodent devices were originally developed and cleared with separate indication for use statements, so each device type statement is modified to add the pterygoid indication.
This submission is focused on labeling, and no changes were made to the existing cleared device designs. Therefore, the existing device clearances are included as reference devices in Table 1 to support the device designs.
{9}
# Traditional 510(k) Submission
# Pterygoid Indication for GM Helix Implants
K252090
Table 1– Table of Substantial Equivalence for proposed devices.
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | | |
| --- | --- | --- | --- | --- | --- |
| K Number | - | K212785
Blue Sky Bio Dental Implant System | K163194
Neodent Implant System – GM Line | K190958
Neodent Implant System | K232099
GM Zygomatic Implant System |
| Implant Indications for Use | GM Helix Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The GM Helix Implant System with 3.75 and 4.0 diameter and 18 mm length are indicated for surgical installation in the pterygoid region (upper jaw), for multiple unit restorations, when combined with 45° angulation abutment, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. Implants in the pterygoid indication should be used in splinted applications that utilize at least two implants.
GM Helix LG Implant:
The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. | Blue Sky Bio Long Implant System is intended for surgical placement in the bone of the upper jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. Implants may be used with single-stage or two-stage procedures, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. Blue Sky Bio Long implants can be placed bicortically in cases of reduced bone density. Blue Sky Bio Long Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Blue Sky Bio Long Implant System with a 45° angulation are indicated for surgical installation in the pterygoid region only, in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading | The Neodent Implant System is intended to be surgically placed in the bone of the upper or lower jaw to provide support for prosthetic devices, such as artificial teeth, to restore chewing function. It may be used with single-stage or two-stage procedures, for single or multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading. The Neodent GM Helix LG implants can be placed bicortically in cases of reduced bone density. The Neodent GM Helix LG implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. | |
JJGC Indústria e Comércio de Materiais Dentários S.A.
{10}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | | |
| --- | --- | --- | --- | --- | --- |
| K Number | - | K212785
Blue Sky Bio Dental Implant System | K163194
Neodent Implant System – GM Line | K190958
Neodent Implant System | K232099
GM Zygomatic Implant System |
| Abutment Indication for Use | GM Mini Conical Abutment 45°: The Mini Conical Abutments 45° and 60°^{1} are indicated for use with Zygomatic Implants, in cases of severe jaw resorption, in order to restore patient aesthetics and chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used with single stage or two-stage procedures, for multiple unit restorations, and may be loaded immediately when good primary stability is achieved and with appropriate occlusal loading.
GM Mini Conical Abutment 45° Slim: The Mini Conical Abutments 45° and 52° are indicated for surgical procedures in Zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. The Mini Conical Abutments 45° are also indicated for use with Pterygoid implants in cases of severe jaw resorption, in order to restore patient esthetics and chewing function. The Mini Conical Abutments may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly. | | | | GM Mini Conical Abutment 52° and 45°: Product indicated for surgical procedures in zygomatic bones, making possible the rehabilitation with screw-retained abutments over the implant, thus restoring the chewing function. It may be used in one- or two-stage procedures, multiple unit restorations, and immediate loading when there is primary stability and adequate occlusal load. Multiple rehabilitations may be splinted rigidly. |
| Material | Implant: Titanium grade 4
Abutment: Titanium alloy. | Implant: Titanium alloy. | Implant: Titanium grade 4
Abutment: Titanium alloy. | Implant: Titanium grade 4
Abutment: Titanium alloy. | Abutment: Titanium alloy. |
| Implant Abutment Connection | Grand Morse (GM) | Internal Hex | Grand Morse (GM) | Grand Morse (GM) | Grand Morse (GM) |
1 The GM Mini Conical Abutment 60° was cleared for use with zygomatic implants in K203542.
JJGC Indústria e Comércio de Materiais Dentários S.A.
{11}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | | |
| --- | --- | --- | --- | --- | --- |
| K Number | - | K212785
Blue Sky Bio Dental Implant System | K163194
Neodent Implant System – GM Line | K190958
Neodent Implant System | K232099
GM Zygomatic Implant System |
| Design | Threaded root-form implant with internal Grand Morse (GM) taper connection with internal hex. | Threaded root-form implant with internal hex taper connection. | Threaded root-form implant with internal Grand Morse (GM) taper connection with internal hex. | Threaded root-form implant with internal Grand Morse (GM) taper connection with internal hex. | |
| Implant Diameter | 3.75 and 4.0 mm | 3.7, 4.3 and 5.0 mm | 3.75 and 4.0 mm | 3.75 and 4.0 mm | |
| Implant Length | GM Helix Implant: 18mm
GM Helix LG Implant:20; 22.5 and 25 mm | 18²; 20; 22.5 and 25 mm³ | 8, 10; 11.5; 13; 16 and 18 mm | 20; 22.5 and 25 mm | |
| Abutment Angle | 45° | Straight, 12°; 17°; 24°; 30° and 45°(45° used with BIO MAX LONG in pterygoid placement only) | Straight, 17° and 30° | Straight, 17° and 30° | 45° and 52° |
| Abutment Gingival Height | 1.5 and 2.5 mm | | | | 1.5 and 2.5 mm |
| Implant Surface | Neoporos and Acqua. | Grit blasted and acid etched | Neoporos and Acqua. | Neoporos. | |
| Single use | Yes | Yes | Yes | Yes | Yes |
| Sterilization Method | Implant: Gamma Irradiation.
Abutment: Ethylene Oxide | Gamma Irradiation. | Implant: Gamma Irradiation.
Abutment: Ethylene Oxide | Implant: Gamma Irradiation.
Abutment: Ethylene Oxide | Implant: Gamma Irradiation.
Abutment: Ethylene Oxide |
² Per GUDID entry for article IPJN3718 Pterygoid NP BlueSky Implant 3.7x18
³ Per K212785 510(k) Summary equivalence table for the two-piece BIO LONG Implants
JJGC Indústria e Comércio de Materiais Dentários S.A.
13-7
{12}
# Traditional 510(k) Submission
# Pterygoid Indication for GM Helix Implants
K252090
| FEATURE | PROPOSED DEVICE | PRIMARY PREDICATE DEVICE | REFERENCE DEVICE | | |
| --- | --- | --- | --- | --- | --- |
| K Number | - | K212785
Blue Sky Bio Dental Implant System | K163194
Neodent Implant System – GM Line | K190958
Neodent Implant System | K232099
GM Zygomatic Implant System |
| Literature review to support pterygoid | Clinical literature was submitted in support of the subject implant placement in the pterygoid region and implant dimensions. Clinical literature detailed in clinical data section reported on implant diameters covering those of the subject devices (Ø3.3-4.2mm with lengths of 10-28mm) for pterygoid placement. | Clinical literature was submitted in support of the subject implant placement in the pterygoid region and implant dimensions. Clinical literature in the summary reported on implant diameters covering those of the subject devices (Ø3.5-5.0mm with lengths of 10-25mm) for pterygoid placement. | | | |
JJGC Indústria e Comércio de Materiais Dentários S.A.
{13}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
# Performance Testing
## Bench Testing
Assessments regarding dynamic fatigue testing were conducted according to the FDA guidance document “Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments” and ISO 14801 “Dentistry — Implants — Dynamic loading test for endosseous dental implants”. For dynamic fatigue tests, the results demonstrated that in identical conditions, the subject devices exhibit a level of performance equal or better than the primary predicate device. Insertion tests were also performed and it could be proven that there is an adequate insertion torque in different bone classes when the implant is inserted according to the surgical procedure.
## MRI Compatibility Testing
An assessment was made to demonstrate that the MR conditional labeling from K182620 is applicable to the subject devices, and a patient treated with them can be safely scanned observing the parameters previously established per reference devices.
## Biocompatibility Testing
Assessments regarding biological compatibility were performed according to ISO 10993-1 and to the FDA Guidance document “Use of International Standard ISO 10993-1, ‘Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process’, Guidance for Industry and Food and Drug Administration Staff, Document issued on: June 16, 2016”.
Representative samples of the worst-case devices were subjected to the following:
- Biological Safety Assessment guided by ISO 10993-1.
- Chemical characterization was performed per ISO 10993-18.
- Cytotoxicity testing was performed per ISO 10993-5.
The subject devices are equivalent in material and manufacturing processes to the reference devices, therefore, no new issues regarding biocompatibility were raised and no additional biocompatibility testing was required.
JJGC Indústria e Comércio de Materiais Dentários S.A.
{14}
JJGC Indústria e Comércio de Materiais Dentários S.A.
13-10
# Traditional 510(k) Submission
## Pterygoid Indication for GM Helix Implants
K252090
## Sterilization validation
The proposed implants are sterilized by Gamma Irradiation, according to ISO 11137-1 and ISO 11137-2. A minimum Sterility Assurance Level (SAL) of 1 x 10⁻⁶ has been validated for the method. The proposed abutments are sterilized by Ethylene Oxide (EO), the method was validated to a sterility assurance level (SAL) of 1×10⁻⁶ in accordance with ISO 11135:2014, "Ethylene Oxide – Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices". EO sterilization residuals have been verified to be less than the maximum allowable limits as defined in ISO 10993-7 "Biological Evaluation of Medical Devices - Part 7: Ethylene Oxide Sterilization Residuals". The sterilization of the subject devices are identical to the sterilization already cleared for the reference devices.
## Shelf Life
The expiration date of the devices was determined considering the integrity of the product and the packaging tests after shelf life testing. The types of packaging of the subject devices are identical to the packaging of the reference devices. The shelf life for the NeoPoros Implants is 5 years, while for the Acqua Implants it is 4 years, which is the same from the reference devices (K163194 and K190958). The shelf life for the abutments is 5 years, which is the same from reference devices (K190718 and K232099).
## Clinical Literature
Clinical literature review was conducted to support the placement of long implants in the pterygoid region, combined with angled abutments (including 45°), for the oral rehabilitation of patients. Collected data showed that these implants exhibited high survival and success rate (88.2% to 100%), over follow-up periods up to 10 years. In addition, the subject devices are within the range of dimensions reported for pterygoid implants in the literature. Collectively, these characteristics, supported by clinical evidence and technological comparisons, demonstrate that the GM Helix and GM Helix LG Implants are safe, effective, and substantially equivalent to the predicate device and the devices cited in the literature for pterygomaxillary rehabilitation.
{15}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
| Article (LRS) | Implant Diameter | Implant Length | Implant Angulation | Abutment Angle | Comment | Reference |
| --- | --- | --- | --- | --- | --- | --- |
| 8 | 3.75 – 5 mm | 6 – 20mm | Not stated | 0°, 17° and 30° | A total of 189 patients were rehabilitated with complete fixed maxillary prostheses supported by 1.817 implants, and 356 of these implants placed in pterygomaxillary sites. In total, 314 of the 356 implants placed in pterygomaxillary sites were osseointegrated, corresponding to a cumulative survival rate of 88.2% after a mean follow-up period of 4.68 years. The authors concluded that implants in this anatomical site provide predictable posterior support and reduce the need for sinus augmentation or bone grafting procedures | Balshi, T.J, Wolfinger, G.J e Balshi, S.F. Analysis of 356 pterygomaxillary implants in edentulous arches for fixed prosthesis anchorage. Int. J. Oral Maxillofac. Implant. . 1999. |
| 34 | 4mm | 7 – 18 mm | Not stated | Not stated | This retrospective study evaluated 992 implants placed in the pterygomaxillary region, and reported a survival rate of 93.75% up to 10 years after loading. According to the authors, this outcome suggests that longer implants, which are more likely to engage both cortical plates of the pterygomaxillary region, may provide enhanced primary stability and improved long-term success. | Balshi, Thomas, Slauch, Robert e Balshi, Stephen F. Brånemark System Implant Lengths in the Pterygomaxillary Region: A Retrospective Comparison. Implant Dentistry. 2013. |
| 35 | 3.75 – 4mm | 18 – 20 mm | 15°, 45° and 60° | 0° and 17° | This retrospective study evaluated the success of pterygoid implants and associated prostheses in patients with an atrophic posterior maxilla. Implant and prosthesis success were considered main outcomes and were assessed for at least 36 months after implant loading. A total of 66 implants were placed in the pterygoid region and 172 implants in the anterior maxilla. Of the 66 pterygoid implants placed into function, 62 survived over the 36-month follow-up period, resulting in a success rate of 99% and mean bone loss of 1.21 mm. These authors' findings indicated that implant placement in the pterygoid region is a viable alternative for rehabilitation of the atrophic posterior maxilla, providing excellent stabilization for bone-anchored prostheses in both partially and completely edentulous patients. | Curi, Marcos Martins, Cardoso, Camila Lopes and Ribeiro, Karina de Cássia Braga. Retrospective Study of Pterygoid Implants in the Atrophic Posterior Maxilla.: The International journal of oral & maxillofacial implants. 2015. |
| 39 | 3.75 – 4.2 mm | 18 – 25mm | 25° and 45° | 8° – 40° | The study evaluated the success rates of implants placed in the pterygoid region and prostheses placed using the TTPHIL technique in the posterior atrophic maxilla. An analysis was conducted on 75 patients, and the findings suggest that implants placed using the TTPHIL demonstrate high survival rates (96.8%), minimal marginal bone loss, and reliable stability in the rehabilitation of patients with posterior maxillary atrophy over two years. | Nag, Venkat P. Ratna, et al. Minimally Invasive Reconstruction of Atrophic Maxilla Using Novel TTPHIL Technique - A Retrospective Evaluation of Pterygoid Implants and Prosthesis Success in 75 Patients. Dentistry: Advanced Research. 2022. |
| 40 | 3.75 mm | 20 mm | Not stated | 45° | This case report describes a graftless approach for full-arch immediate maxillary rehabilitation using the TTPHIL technique. The mouth opening was assessed and found to be adequate for implant placement in the pterygoid region. This case report demonstrated that the TTPHIL technique is a clinically effective for full-arch | Nag, Venkat Ratna, et al. Tall and Tilted Pin Hole Immediately Loaded Implants (TTPHIL) Technique for Maxillary Arch Rehabilitation. International Journal of Research & Review. 2018. |
JJGC Industria e Comércio de Materiais Dentários S.A.
{16}
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
| | | | | | maxillary rehabilitation, particularly in cases with limited bone availability. | |
| --- | --- | --- | --- | --- | --- | --- |
| 42 | 3.75 – 4.2 mm | 13 – 20 mm | 60° to 90° | Not stated | This study evaluated the clinical outcomes of placing implants in pterygoid region with a more vertical inclination than the conventional 45°, measured relative to the maxillary (occlusal) plane, in a follow-up over 14 years. In total, 454 implants were placed in 392 patients, and 438 (96.5%) achieved successful osseointegration, while 16 (3.5%) failed. Thirteen of these failures occurred at stage-two surgery, before loading. Among the failed implants, seven occurred in men and eight in women; ten failures were in smokers (60%) and four in non-smokers (40%). Two implants fractured after five years of loading in a bruxism patient. | Rodríguez, Xavier, et al. Modified Surgical Protocol for Placing Implants in the Pterygomaxillary Region: Clinical and Radiologic Study of 454 Implants. The International Journal of Oral & Maxillofacial Implants. 2012. |
| 45 | Not stated | 16 – 28 mm | Not stated | Not stated | This retrospective multicenter study aimed to evaluate the surgical and prosthetic success rates of pterygoid implants placed with a minimum insertion torque of 45 Ncm to support fixed partial or full-arch rehabilitations without cantilevers. A total of 92 pterygoid implants were placed in 56 patients and followed up for one year. The pterygoid 92 implants used in this study measured between 16 and 28 mm in length. The study reported a 100% surgical success rate, with all implants achieving a torque value equal to or greater than 45 Ncm. Overall, this retrospective study demonstrated that pterygoid implants provide a reliable treatment alternative for patients with severe posterior maxillary atrophy. | Tealdo, T., et al. A retrospective multicentric study of 56 patients treated with 92 pterygoid implants for partial/full arch implant supported fixed rehabilitation: implant and prosthesis success rate. Eur j musculoskel dis. 2023. |
| 46 | 4.0 mm | 11.5 – 18 mm | 72° to 80° | Not stated | This study investigated the 1-year survival and success rates of pterygoid implants and their corresponding prostheses in individuals with severe atrophy of the posterior maxilla who required a complete-arch, immediately loaded fixed prosthesis. The study included 15 participants, with a total of 27 implants placed in the pterygoid region, with final angulations of 25° to 40° relative to the maxillary plane. All prostheses remained stable, and all implants showed complete osseointegration without pain, radiolucency, or peri-implant pathology. Overall, both survival and success rates were 100% with 1 year of follow-up period, with no failures observed in either pterygoid or conventional implants. | Signorini, Luca, et al. Immediate fixed rehabilitation supported by pterygoid implants for participants with severe maxillary atrophy: 1-Year postloading results from a prospective cohort study. The journal of prosthetic dentistry. 2021. |
JJGC Industria e Comércio de Materiais Dentários S.A.
{17}
JJGC Indústria e Comércio de Materiais Dentários S.A.
13-13
Traditional 510(k) Submission
Pterygoid Indication for GM Helix Implants
K252090
Conclusion
The data included in this submission demonstrate that the subject devices are substantially equivalent to the predicate devices in terms of functional and mechanical properties, sterilization method, technological characteristics, design, indications for use and raw material.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.