K Number
K170287

Validate with FDA (Live)

Date Cleared
2017-05-24

(114 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandibular arches to provide support for prosthetic restorations in edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.

Device Description

The Dental Implants OKTAGON® Bone Level Tapered Design are an addition to the currently distributed Dental Implant System® OKTAGON Bone Level. The implants can be used for immediate or early implantation after loss or extraction of natural teeth. The endosteal implant body of the subject device has a conical shape in the coronal region and is cylindrical in the apical region. The apical cylindrical part of the implant is relevant for the given endosteal diameter of 3.3 to 4.1 mm. The implants are made of commercially pure Titanium Grade 4 conforming to ASTM- F67. The surface is micro-structured in the endosted part and has been blasted with high-grade corundum and afterwards acid-etched. The implant shoulder is polished. The prosthetic connection is achieved with the help of an inner cone with an additional octagonal antirotation device. A sterile cover screw is enclosed with the implant so that an immediate occlusion of the internal thread is possible after successful insertion.

AI/ML Overview

This document is a 510(k) Pre-Market Notification for a medical device, specifically Dental Implants OKTAGON® Bone Level Tapered Design. It does not describe an AI medical device or the results of a study proving a device meets acceptance criteria related to AI/software performance.

Therefore, I cannot provide the requested information, which includes:

  1. A table of acceptance criteria and the reported device performance.
  2. Sample size used for the test set and the data provenance.
  3. Number of experts used to establish the ground truth for the test set and their qualifications.
  4. Adjudication method for the test set.
  5. Multi-reader multi-case (MRMC) comparative effectiveness study results.
  6. Standalone (algorithm only) performance.
  7. Type of ground truth used.
  8. Sample size for the training set.
  9. How the ground truth for the training set was established.

This document focuses on demonstrating substantial equivalence to predicate implant devices based on material composition, design, intended use, and mechanical performance (fatigue and breakage tests for physical implants, not AI performance).

{0}------------------------------------------------

Image /page/0/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The caduceus is a symbol of medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 24, 2017

Hager & Meisinger GmbH Melanie May Regulatory Affairs Hansemannstrasse 10 Neuss. 41468 GERMANY

Re: K170287

Trade/Device Name: Dental Implants OKTAGON® Bone Level Tapered Design Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: May 4, 2017 Received: May 8, 2017

Dear Melanie May:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely.

Lori A. Wiggins -S

for

Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K170287

Device Name

Dental Implants OKTAGON® Bone level Tapered Design

Indications for Use (Describe)

The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandbular arches to provide support for prosthetic restorations in edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Dental Implants OKTAGON® Bone Level Tapered Design

510(k) Summary

1. Applicant's Name and Address

Hager & Meisinger GmbH
Hansemannstrasse 10
41468 Neuss
Germany
Phone: 0049 2131 2012 292
Fax: 0049 2131 2012 223
Contact Person: Dr. Melanie MayRegulatory Affairs

2. Date prepared

Date prepared:

05/23/2017

3. Name of the device

Trade Name:Dental Implants OKTAGON® Bone level Tapered Design
Common Name:Endosseous dental implants
Classification Name:Endosseous dental implants
Product Code:DZE
Regulation No:872.3640
Class:II
Panel:Dental

4. Predicate Devices:

510(k) No.ManufacturerTrade Name
K143539Hager & Meisinger GmbHDental Implant System OKTAGON® Bone Level
K140878Straumann USA,LLCStraumann® Bone Level Tapered Implants

5. Device Description:

The Dental Implants OKTAGON® Bone Level Tapered Design are an addition to the currently distributed Dental Implant System® OKTAGON Bone Level. The implants can be used for immediate or early implantation after loss or extraction of natural teeth.

The endosteal implant body of the subject device has a conical shape in the coronal region and is cylindrical in the apical region. The apical cylindrical part of the implant is relevant for the given endosteal diameter of 3.3 to 4.1 mm. The implants are made of commercially pure Titanium Grade 4 conforming to ASTM- F67. The surface is micro-structured in the endosted part and has been blasted with high-grade corundum and afterwards acid-etched. The implant shoulder is polished.

{4}------------------------------------------------

Dental Implants OKTAGON® Bone Level Tapered Design

510(k) Summary

The prosthetic connection is achieved with the help of an inner cone with an additional octagonal antirotation device. A sterile cover screw is enclosed with the implant so that an immediate occlusion of the internal thread is possible after successful insertion.

The submission contains the following tapered design implant variations:

BLT NC Ø 3.3 mm, L 8.0 mm BLT NC Ø 3.3 mm, L 10.0 mm BLT NC Ø 3.3 mm, L 12.0 mm

BLT NC Ø 3.75 mm, L 8.0 mm BLT NC Ø 3.75 mm, L 10.0 mm BLT NC Ø 3.75 mm, L 12.0 mm

BLT NC Ø 4.1 mm, L 8.0 mm BLT NC Ø 4.1 mm, L 10.0 mm BLT NC Ø 4.1 mm, L 12.0 mm

6. Indications for Use:

The Dental Implants OKTAGON® Bone Level Tapered Design are surgically placed in the maxillary and/or mandibular arches to provide support for prosthetic restorations or partially edentulous patients. The implants are intended to be used with OKTAGON® abutments and prosthetic parts. The implants are intended for delayed loading in a two-stage surgery or for immediate loading when good primary stability is achieved and with appropriate occlusal load.

7. Performance tests and used standards

Information regarding biocompatibility testing, sterility validation and shelf life was leveraged from the predicate device KI43539.

The following standards have been followed for the development, production, performance and safety testing of the subject devices.

  • Fatigue Testing according to ISO 14801 ●
  • Evaluation of Biocompatibility according to ISO 10993-1, ISO 7405
  • Cytotoxicity Testing according to ISO 10993-5
  • Sterilization Validation for implants and accessories according to ISO I I I 37-1, ISO I I I I 37-2,

{5}------------------------------------------------

Dental Implants OKTAGON® Bone Level Tapered Design

510(k) Summary

  • Sterile Barrier Testing according to ASTM F88/F88M-09 and ASTM F1929-98 ●
  • Packaging of sterilized implants according to ISO I 1607-1
  • . Material Specification according to ASTM F67
  • Risk Management according to ISO I 497 I

In order to demonstrate substantial equivalence performance tests (breakage and fatigue tests) have been conducted fulfilling the requirements of ISO 14801 and the FDA's "Guidance for Industry and FDA Staff – Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments".

8. Basis for substantial equivalence

The intended use for Dental Implant OKTAGON® Bone Level Tapered Design is identical to the named predicated devices. The implants have the same material composition and the same surface treatments. In addition, the implants are of the same, root-form design

Since the Dental Implants OKTAGON® Bone Level Tapered Design and the predicate implants have the same design, are made of the same materials (see table), exhibit the same performance characteristics (e.g. strength and function), and are furthermore intended for the same use, no new performance or safety issues are raised.

ManufacturerHager & Meisinger GmbHHager & Meisinger GmbHStraumann USA, LLCManufacturerHager & Meisinger GmbHHager & Meisinger GmbHStraumann USA, LLC
510(k) NumberunknownK143539 (Primary Predicate Device)K140878 (Reference Predicate Device)MaterialTitanium Grade 4Titanium Grade 4TitaniumGrade4Titanium-13Zirconiumalloy (Roxolid®)
Intended UseThe Dental ImplantsOKTAGON® Bone LevelTapered Design are surgicallyplaced in the maxillary and/ormandibular arches to providesupport for prostheticrestorations in edentulous orpartially edentulous patients.The implants are intended tobe used with OKTAGON®abutments and prostheticparts. The implants areintended for delayed loading ina two-stage surgery or forimmediate loading when goodprimary stability is achievedand with appropriate occlusalload.The implants are surgicallyplaced in the maxillary and/or mandibular arches toprovide support forprosthetic restorations inedentulous or partiallyedentulous patients. Theparts are intended to beused with OKTAGON®Bone Level abutments andprosthetic parts. TheDental Implant systemOKTAGON® Bone Level isintended for delayedloading, or for immediateloading when good primarystability is achieved andwith appropriate occlusalload.Straumann® dentalimplants are indicated fororal endosteal implantationin the upper and lower jawand for the functional andesthetic oral rehabilitationof edentulous and partiallydentate patients.Straumann dental implantscan also be used forimmediate or earlyimplantation followingextraction or loss of naturalteeth. Implants can beplaced with immediatefunction on single-toothand/or multiple toothapplications when goodprimary stability is achievedand with appropriateSurfaceendosseousMicro-structured (grit blastedand acid etched)Micro-structured(gritblasted and acid etched)SLA (grit blasted then acidetched), SLA active
ManufacturerHager & Meisinger GmbHHager & Meisinger GmbHStraumann USA, LLCConnectionInner cone with octagonalanti-rotation deviceInner cone with octagonalanti-rotation deviceInner cone with octagonalanti-rotation device
occlusal loading, to restorechewing function. Theprosthetic restorations usedare single crowns, bridgesand partial or full dentures,which are connected to theimplants by thecorresponding elements(abutments). In cases offully edentulous patients, 4or more implants must beused in immediately loadedcases.Length8 to 12 mm8 to 14mm8 to 16 mm
DescriptionThe endosteal implant body ofthe subject device has aconical shape in the coronalregion and is cylindrical in theapical region. The apicalcylindrical part of the implantis relevant for the givenendosteal diameter of 4.1mm. The implants are madeof commercially pure TitaniumGrade 4 conforming to ASTM-F67. The surface is micro-structured in the endostealpart and has been blastedwith high-grade corundum andafterwards acid-etched. Theimplant shoulder is polished.The prosthetic connection isachieved with the help of aninner cone with an additionaloctagonal anti-rotation device.A sterile cover screw isenclosed with the implant sothat an immediate occlusionof the internal thread ispossible after successfulinsertion.Subject to this submission is atapered design implantvariation of 4.1 mm, which isavailable in a range ofendosseous lengths.The root-form designedimplant is made ofcommercially pure TitaniumGrade 4 conforming toASTM StandardSpecification F67. Thesurface is micro-structuredin the endosteal section andthe implant shoulder ispolished. The implantsurface has been blastedwith high-grade corundumand afterwards acid-etched.The prosthetic connection isachieved with the help of aninner cone with anadditional octagonal anti-rotation device.A sterile locking screw isenclosed with the implantso that an immediateocclusion of the internalthread is allowed aftersuccessful insertion.The endosseous dentalimplants are available in arange of endosseousdiameters and lengths.The devices represent a lineextension of the previouslycleared Bone Level implantsof the Straumann DentalImplant System (K062129,K083550 and K121131).The apical part of theimplants has a taperingthread form and threecutting flutes.The implants are made ofTitanium Grade 4respectively Titanium-13Zirconium alloy(Roxolid®). The surfacefinishes are either SLA (gritblasted then acid etched)or SLAactiveThe implant-to-abutmentconnection (NC, RC) isachieved by an inner conewith octagonal anti-rotation device.The devices are available inthe endosseous diameters3.3, 4.1 and 4.8 mm.EndosseousDiameter3.3 to 4.1 mm3.3 to 4.8 mm3.3 to 4.8 mm

{6}------------------------------------------------

Dental Implants OKTAGON® Bone Level Tapered Design

510(k) Summary

{7}------------------------------------------------

Dental Implants OKTAGON® Bone Level Tapered Design

510(k) Summary

Based on these observations, we conclude that the Dental Implant OKTAGON® Bone Level Tapered Design are substantially equivalent to the legally marketed predicate devices described.

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.