FUBUKI XF-R Neurovascular Long Sheath

K251560 · Asahi Intecc Co., Ltd. · QJP · Jun 20, 2025 · Cardiovascular

Device Facts

Record IDK251560
Device NameFUBUKI XF-R Neurovascular Long Sheath
ApplicantAsahi Intecc Co., Ltd.
Product CodeQJP · Cardiovascular
Decision DateJun 20, 2025
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature.

Device Story

Device is a single-lumen neurovascular long sheath with a dilator and rotating hemostasis valve (RHV). Input: manual insertion by trained physician into neurovasculature. Transformation: sheath provides a conduit for interventional devices (e.g., guidewires) and contrast media delivery. Output: physical access to neurovascular lesion sites. Used in clinical settings by neurointerventionalists. Benefits: facilitates navigation of therapeutic devices to target sites in the neurovasculature. Device features a hydrophilic coating on the outer surface and a fluoropolymer inner lumen to reduce friction during device advancement.

Clinical Evidence

Bench testing only. Testing performed per FDA guidance for peripheral PTA and specialty catheters and ISO 10555-1. Includes dimensional verification, simulated use, lubricity, coating integrity, and appearance. Biocompatibility testing conducted per ISO 10993 series, including cytotoxicity, sensitization, reactivity, systemic toxicity, pyrogenicity, and hemocompatibility. In vivo thrombogenicity study demonstrated performance comparable to the predicate.

Technological Characteristics

Single-lumen catheter; 8Fr outer diameter; 6.8Fr inner diameter. Materials: hydrophilic polymer outer coating, fluoropolymer (PTFE) inner lumen. Sterilization: Ethylene Oxide (SAL 10^-6). Supplied with dilator and rotating hemostasis valve. Shelf life: 3 years. Complies with ISO 10555-1 and ISO 10993.

Indications for Use

Indicated for patients requiring percutaneous intravascular procedures in the neurovasculature, specifically to guide interventional devices to a lesion or procedural site and for the injection of contrast media.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 20, 2025 ASAHI INTECC Co., Ltd. % Esther Kim Quality System and Regulatory Affairs Associate ASAHI INTECC USA, Inc. 3002 Dow Avenue, Suite 212 Tustin, California 92780 Re: K251560 Trade/Device Name: FUBUKI XF-R Neurovascular Long Sheath Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: QJP Dated: May 21, 2025 Received: May 21, 2025 Dear Esther Kim: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K251560 - Esther Kim Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K251560 - Esther Kim Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Naira Muradyan -S Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional, and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K251560 | | | Device Name FUBUKI XF-R Neurovascular Long Sheath | | | Indications for Use (Describe) The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} FUBUKI XF-R Neurovascular Long Sheath # 510(k) Summary (as required by 21 CFR § 807.92) # ASAHINTECC CO.,LTD. Global Headquarters and R&D Center 3-100 Akatsuki-cho, Seto-shi, Aichi 489-0071 Japan TEL: +81-561-48-5551 FAX: +81-561-48-5552 http://www.asahi-intecc.co.jp/ # FUBUKI XF-R Neurovascular Long Sheath 510(k) K251560 | Date Prepared: | 21 MAY, 2025 | | --- | --- | | Applicant: | ASAHI INTECC CO., LTD. 3-100 Akatsuki-cho, Seto, Aichi 489-0071 Japan | | Contact: | Esther Kim Quality System and Regulatory Affairs Associate ASAHI INTECC USA, Inc. 3002 Dow Avenue, Suite 212 Tustin, CA 92780 Office: (949) 756-8901 ext. 508 e-mail: aiu_ra@asahi-intecc-us.com | | Device Name: | FUBUKI XF-R Neurovascular Long Sheath | | Device Classification: | Class II, 21 CFR § 870.1250 | | Regulation Name: | Percutaneous Catheter | | Product Code: | QJP – Catheter, Percutaneous, Neurovasculature | | Predicate Device: | FUBUKI XF Neurovascular Long Sheath (K213589) | # Intended Use/Indications For Use The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature. # Device Description The FUBUKI XF-R Neurovascular Long Sheath (FUBUKI XF-R) consists of a long sheath and a dilator. The long sheath is a single lumen neurovascular catheter designed for introduction of interventional devices, such as guidewires and other therapeutic devices. The long sheath consists of three sections: (1) a shaft, (2) a protector and (3) a connector. The distal portion of the shaft consists of a soft tip and a soft tube. The proximal part of ASAHI INTECC CO., LTD. {5} FUBUKI XF-R Neurovascular Long Sheath the shaft is covered by the protector (strain relief) and the connector is bonded to the proximal end of the shaft. The subject device is provided sterile, by ethylene oxide, and is intended for single use only by physicians who have been adequately trained in neurointerventional procedures. The outer surface of the long sheath is coated with a hydrophilic polymer and the inner lumen of the shaft (excluding the connector portion) is lined with a fluoropolymer layer to facilitate movement of the guidewire and other devices. The dilator consists of two parts: (1) a shaft and (2) a connector. The FUBUKI XF-R is supplied with a dilator and rotating hemostasis valve (RHV) packed in a sterile package. ## Comparison with Predicate Device The subject FUBUKI XF-R and the predicate FUBUKI XF have the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, shelf life and operating principle. The following technological differences exist between the subject and predicate devices: - Hydrophilic coating length of the long sheath - Addition of a 95cm model and exclusion of the 110cm model for the long sheath - Exclusion of the Angled tip long sheath model - Addition of a 117cm dilator These differences do not raise new or different questions of safety and effectiveness. The results of the non-clinical testing provide reasonable assurance of substantial equivalence to the predicate device. ASAHI INTECC CO., LTD. Page 2 of 5 {6} FUBUKI XF-R Neurovascular Long Sheath Comparison of Intended Use and Technological Characteristics to Predicate Device | Device Name | Subject Device FUBUKI XF-R Neurovascular Long Sheath | Predicate Device FUBUKI XF Neurovascular Long Sheath | | --- | --- | --- | | 510(k) Number | K251560 | K213589 | | Manufacturer | ASAHI INTECC CO., LTD. | | | Regulation Number | 21 CFR § 870.1250 | | | Regulation Name | Catheter, Percutaneous, Neurovasculature | | | Regulatory Class | II | | | Product Code | QJP | | | Intended Use and Indications for Use | The FUBUKI XF-R Neurovascular Long Sheath is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. The FUBUKI XF-R Neurovascular Long Sheath is also intended to be used for injection of contrast media. The FUBUKI XF-R Neurovascular Long Sheath is intended for use only in the neurovasculature. | This product is intended to be used to guide interventional devices for neurovascular therapy to a lesion or a procedural site for a percutaneous intravascular procedure in the neurovasculature. This product is also intended to be used for injection of contrast media. This product is intended for use only in the neurovasculature. | | Long Sheath | | | | Outer Diameter | 2.70mm (0.106", 8Fr) | | | Inner Diameter | 2.28mm (0.090", 6.8Fr) | | | Effective Length | 80cm, 90cm, 95cm, 100cm | 80cm, 90cm, 100cm, 110cm | | Tip Shape | Straight | Straight, Angled | | Coating | Hydrophilic | | | Construction | Soft tip, soft tube and shaft, protector, connector | | | Coating Length | 20cm | 8cm | | Inner Lumen | PTFE | | | Dilator | | | | Effective Length | 102cm, 112cm, 117cm, 122cm | 102cm, 112cm, 122cm | | Outer Diameter | 2.24mm (0.088") | | | Inner Diameter | 0.91mm (0.036") | | | Other Information | | | ASAHI INTECC CO., LTD. {7} FUBUKI XF-R Neurovascular Long Sheath | Accessory | Hemostasis Valve (Rotating Hemostasis Valve) | | --- | --- | | Sterilization Method | Ethylene Oxide Gas | | Sterilization Level | SAL 10^{-6} | | Shelf Life | 3 years | | Single Use | Yes | ## Non-Clinical Bench Testing / Performance Data The following non-clinical bench testing was performed for the FUBUKI XF-R Neurovascular Long Sheath to support the substantial equivalence determination. The FUBUKI XF-R met all acceptance criteria and performed similarly to the predicate FUBUKI XF. Testing was performed per FDA guidance “Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions” and the recommendations in ISO 10555-1 “Intravascular Catheters – Sterile and Single-Use Catheters – Part 1: General Requirements”. - Dimensional Verification - Simulated Use - Lubricity - Coating Integrity / Particulate Evaluation - Appearance ## Biocompatibility The FUBUKI XF-R was tested per ISO 10993 series and FDA guidance, “Use of International Standard ISO 10993-1, “Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process”.” All testing performed met the requirements as specified within the applicable standard. | Test | Test Method Summary | Conclusions | | --- | --- | --- | | Cytotoxicity MEM Elution Test | Per ISO 10993-5 | Non-Cytotoxic | | Sensitization Kligman Maximization Test | Per ISO 10993-10 | Non-Sensitizing | | Intracutaneous Reactivity Intracutaneous Injection Test | Per ISO 10993-23 | Non-Irritant | | Acute Systemic Toxicity System Injection Test | Per ISO 10993-11 | Non-Toxic | | Material Mediated Pyrogenicity Rabbit Pyrogen Test | Per ISO 10993-11 | Non-Pyrogenic | | Hemocompatibility Rabbit Blood Hemolysis Test (Direct and Indirect Contact) | Per ISO 10993-4 | Non-Hemolytic | | Hemocompatibility Complement Activation Test (SC5b-9) (Direct Contact) | Per ISO 10993-4 | Non-Activator | ASAHI INTECC CO., LTD. {8} FUBUKI XF-R Neurovascular Long Sheath | Hemocompatibility Partial Thromboplastin Time Test (PTT) (Direct Contact) | Per ISO 10993-4 | Non-Activator | | --- | --- | --- | | Thrombogenicity In Vivo Thrombogenicity Study | Per ISO 10993-4 | Comparable thromboresistance to the predicate device. | Sterilization and Shelf Life The changes to the subject device do not impact sterilization and shelf life as compared to the predicate device. The shelf-life of the subject device was supported by device performance testing following 3-year accelerated aging. Conclusion The FUBUKI XF-R Neurovascular Long Sheath has the same intended use and the same or similar technological characteristics such as components, design, materials, sterilization method, and operating principles as the predicate device. The differences between the subject and predicate device do not raise different or new questions of safety or effectiveness. The non-clinical tests demonstrate that the subject device performs as intended and that the FUBUKI XF-R Neurovascular Long Sheath is substantially equivalent to the predicate FUBUKI XF Neurovascular Long Sheath (K213589). ASAHI INTECC CO., LTD.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...