K201750 · Biosense Webster, Inc. · MTD · Aug 6, 2020 · Cardiovascular
Device Facts
Record ID
K201750
Device Name
Pentaray Nav eco High-Density Mapping Catheter
Applicant
Biosense Webster, Inc.
Product Code
MTD · Cardiovascular
Decision Date
Aug 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1220
Device Class
Class 2
Indications for Use
The Biosense Webster PENTARAY™ NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY™ NAV eco High-Density Mapping Catheter provides location information when used with compatible CARTO™ 3 EP Navigation Systems. (This catheter is not compatible with CARTO™ 3 EP Navigation Systems prior to Version 3.x).
Device Story
Multi-electrode diagnostic catheter; flower-shaped distal probe with 5 spines; 20 total ring electrodes. Used for electrophysiological mapping (recording/stimulation) of cardiac structures. Operated by physicians in clinical settings. Inputs: electrical signals from heart endocardium. Outputs: electrograms and location data transmitted to compatible CARTO 3 EP Navigation Systems and recording equipment. Enables visualization and analysis of cardiac electrical activity to assist in arrhythmia detection and clinical decision-making.
Clinical Evidence
Bench and animal testing performed. Catheter passed all intended criteria in accordance with appropriate test standards. No human clinical data provided.
Technological Characteristics
Multi-electrode diagnostic catheter; 5-spine flower-shaped distal tip; 20 ring electrodes. Compatible with CARTO 3 EP Navigation Systems (Version 3.x or later).
Indications for Use
Indicated for multiple electrode electrophysiological mapping (recording or stimulation) of atrial and ventricular cardiac structures in patients requiring cardiac electrogram acquisition.
Regulatory Classification
Identification
An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.
K193237 — Carto Octaray Mapping Catheter with Trueref Technology · Biosense Webster, Inc. · Jul 31, 2020
K201806 — Reprocessed PentaRay Nav eco High-Density Mapping Catheter · Sterilmed Inc. (Johnson and Johnson) · Jun 23, 2021
K231312 — PENTARAY® NAV ECO High Density Mapping Catheter, DECANAV® Mapping Catheter, Webster® CS Catheter with Auto ID, Webster® CS Catheter with EZ Steer Technology, Webster® CS Catheter with EZ Steer Technology with Auto ID · Biosense Webster, Inc. · Jul 10, 2023
K211219 — LASSOSTAR NAV Circular Mapping Catheter · Biosense Webster, Inc. · Jan 26, 2022
K211438 — OPTRELL Mapping Catheter with TRUEref Technology · Biosense Webster, Inc. · Mar 18, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 6, 2020
Biosense Webster, Inc Richard Lauhead Sr. Program Lead, RA 33 Technology Dr Irvine, California 92618
Re: K201750
Trade/Device Name: Pentaray Nav eco High-Density Mapping Catheter Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: June 19, 2020 Received: June 26, 2020
Dear Richard Lauhead:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
{3}------------------------------------------------
### 1. 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
| Date Summary<br>Prepared | 06 August 2020 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------|
| Applicant | Biosense Webster, Inc.<br>33 Technology Drive<br>Irvine, CA 92618<br>Establishment Registration Number: 9044811 |
| Official Correspondent | Richard Lauhead<br>Senior Program Lead, Regulatory Affairs<br>Telephone: (949) 789-8583<br>Fax: (949) 450-6886 |
| Trade Name | Pentaray Nav eco High Density Mapping Catheter. |
| Common Name | Electrophysiology Catheter |
| Classification Name | Electrode Recording Catheter |
| Device Classification | Class II, 21 CFR 870.1220<br>Product Code: MTD |
| Model Numbers | D-1282-07-S, D-1282-08-S, D-1282-09-S, D-1282-10-S,<br>D-1282-11-S, D-1282-12-S |
| Predicate device | PENTARAY® Nav eco Mapping Catheter (K123837) |
### Substantially Equivalent To:
The Biosense Webster Inc. PENTARAY® Nav eco Mapping Catheter is substantially equivalent to the Biosense Webster Inc. PENTARAY® Nav eco Mapping Catheter [510(k) K123837 cleared March 05, 2013]. The subject catheter is identical in design to the current catheter. The intended use of the predicate device as a diagnostic catheter designed to facilitate mapping of structures within the heart remains the same in the subject device. The purpose of this 510(k) submission is to add clarification to the precautions. There are no other changes between the subject device and the predicate device approved per K123837.
{4}------------------------------------------------
## Description of the Device Subject to Premarket Notification:
The Biosense Webster PENTARAY® Nav eco Mapping Catheter is a multi-electrode diagnostic catheter designed to facilitate electrophysiological mapping of all structures in the heart (recording and stimulation). The catheter's distal end is a flower-shaped probe with 5 spines that radiate from the center. Each spine has 4 ring electrodes that are used for stimulation and recording within the heart. The flower is available in a 30mm diameter and several ring spacing configurations to achieve optimal mapping and contact with various cardiac structures.
## Indications for Use:
The Biosense Webster PENTARAY™ NAV eco High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY™ NAV eco High-Density Mapping Catheter provides location information when used with compatible CARTO™ 3 EP Navigation Systems. (This catheter is not compatible with CARTO™ 3 EP Navigation Systems prior to Version 3.x).
# Technical Characteristics:
The PENTARAY™ NAV eco High-Density Mapping Catheter has a flower shaped tip containing 20 electrodes arranged across 5 spines Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation system and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias.
### Performance Data:
The PENTARAY™ NAV eco High-Density Mapping Catheter underwent bench and animal testing. The catheter passed all intended criteria in accordance with appropriate test criteria and standards.
### Basis for Determination of Substantial Equivalence:
The subject PENTARAY™ NAV eco High-Density Mapping Catheter is substantially equivalent to the predicate device, based on the identical principles of design, operation, and indications for use. The only change for the proposed device is the addition of an informative precaution
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.