K Number
K193237

Validate with FDA (Live)

Date Cleared
2020-07-31

(249 days)

Product Code
Regulation Number
870.1220
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CARTO® OCTARAY™ Mapping Catheter with TRUErcf™ Technology is indicated for multiple electrode clectrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.

Device Description

The Biosense Webster CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology is a multi-electrode diagnostic catheter designed to facilitate electrophysiological mapping of all structures in the heart (recording and stimulation). The catheter's distal end is a flower-shaped probe with 8 spines that radiate from the center. Each spine has 6 ring electrodes that are used for stimulation and recording within the heart. The catheter also includes an electrode located at the confluence of the spines that can be used as a close internal unipolar reference. The flower is available in a 30mm or a 40 mm diameter and several ring spacing configurations to achieve optimal mapping and contact with various cardiac structures.

AI/ML Overview

The provided text describes the 510(k) summary for the CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology. It outlines the device's indications for use, technical characteristics, and the performance data used to demonstrate substantial equivalence to a predicate device.

Key Findings from the Document Regarding Acceptance Criteria and Device Performance:

The document primarily focuses on demonstrating substantial equivalence, relying on a combination of bench testing, animal studies, and a clinical study. The acceptance criteria for the clinical study are defined by primary objectives for effectiveness and safety.

1. A table of acceptance criteria and the reported device performance:

Acceptance Criteria (Primary Endpoints)Reported Device Performance
Effectiveness: Completing pre-ablation mapping and clinically indicated mapping without the need for another device.Pre-ablation mapping: 82.1% (23/28) in the Per-Protocol Population had pre-ablation mapping requirements successfully performed without the need of another catheter. (Note: Reasons for failure in the remaining cases were attributed to patient safety, operator's inability to maintain stable contact, or human error, not catheter deficiency).
Clinically indicated mapping: 100.0% (15/15) of cases where clinically indicated mapping was started and finished, it was done without the need for another device.
Safety: No serious adverse events within 7 days of the study procedure resulting from the use of the study catheter.No Serious Adverse Events were reported resulting from the use of the study catheter. All 17 observed adverse events in 12 subjects were classified as unrelated to the study device.

2. Sample sized used for the test set and the data provenance:

  • Test Set Sample Size: The clinical study enrolled 31 subjects.
    • One subject was withdrawn prior to the procedure.
    • Two subjects had mapping performed but not for the study arrhythmia.
    • This resulted in a Safety Population of 30 subjects and a Per-Protocol Population of 28 subjects.
  • Data Provenance: The study was a prospective, single-arm, non-randomized, multi-center study. It was performed outside of the United States.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

The document does not explicitly state the number of experts used to establish ground truth or their specific qualifications for the clinical study. It describes the study as "single arm," implying that the assessment of mapping success and adverse events was likely done by the participating clinicians (physicians/operators) at the study sites. The "Sponsor" (Biosense Webster, Inc.) appears to have reviewed and attributed reasons for protocol deviations and adverse events.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

The document does not describe a formal adjudication method (like 2+1 or 3+1) for the clinical study's outcomes. The reporting of effectiveness and safety endpoints seems to be directly from the study results as interpreted by the sponsor, without explicit mention of independent adjudication of each case by a panel of experts.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

No. This document describes a clinical study for a medical device (mapping catheter) to demonstrate its safety and effectiveness, not an AI/human reader comparative effectiveness study. Therefore, there is no mention of MRMC studies or effect sizes related to human readers improving with AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

Not applicable. The device is a physical mapping catheter, not an AI algorithm. Its performance is inherently linked to its use by a human operator in a clinical setting with the CARTO® 3 EP Navigation System. The study assessed the device in a human-in-the-loop context.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

The ground truth for the clinical study was based on:

  • Clinical assessment by interventional physicians during the procedures: For successful completion of mapping and initiation/completion of clinically indicated mapping without the need for another device.
  • Observation and reporting of adverse events: Documented by study site personnel and reviewed by the sponsor.
  • Protocol compliance: Reviewed against the study protocol by the sponsor to identify deviations.

This is a form of clinical outcome data and expert judgment from the performing physicians.

8. The sample size for the training set:

Not applicable. This device is a medical catheter, not a machine learning model, so there is no concept of a "training set" in the context of an algorithm's development.

9. How the ground truth for the training set was established:

Not applicable, as explained in point 8.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the symbol of the Department of Health & Human Services on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out next to it.

July 31, 2020

Biosense Webster, Inc. Richard Lauhead Sr. Program Lead, RA 33 Technology Dr Irvine. California 92618

Re: K193237

Trade/Device Name: Carto Octaray Mapping Catheter with TRUEref Technology Regulation Number: 21 CFR 870.1220 Regulation Name: Electrode Recording Catheter or Electrode Recording Probe Regulatory Class: Class II Product Code: MTD Dated: June 29, 2020 Received: July 1, 2020

Dear Richard Lauhead:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

{1}------------------------------------------------

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Mark Fellman Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use 1

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationIndications for UseForm Approved: OMB No. 0910-0120Expiration Date: 06/30/2020See PRA Statement below.
510(k) Number (if known)K193237
Device NameCARTO® OCTARAY™ MAPPING CATHETER WITH TRUEREF™ TECHNOLOGY

Indications for Use (Describe)

The CARTO® OCTARAY™ Mapping Catheter with TRUErcf™ Technology is indicated for multiple electrode clectrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.

Type of Use (Select one or both, as applicable) [X] Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (7/17)

Page 1 of 1

PSC Publishing Services (301) 443-6740 EF

{3}------------------------------------------------

1. 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92

Date SummaryPrepared25 June 2020
ApplicantBiosense Webster, Inc.33 Technology DriveIrvine, CA 92618Establishment Registration Number: 9044811
OfficialCorrespondentRichard LauheadSenior Program Lead, Regulatory AffairsTelephone: (949) 789-8583Fax: (949) 450-6886
Trade NameCARTO® OCTARAY™ MAPPING CATHETER WITHTRUEREF™ TECHNOLOGY
Common NameElectrophysiology Catheter
Classification NameElectrode recording Catheter
DeviceClassificationClass II, 21 CFR 870.1220Product Code: MTD
Model NumbersD-1609-01-S, D-1609-02-S, D-1609-03-S, D-1609-04-S,D-1609-05-S, D-1609-06-S
Predicate devicePENTARAY® Nav eco Mapping Catheter (K123837)

Substantially Equivalent To:

The Biosense Webster Inc. CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology is substantially equivalent to the Biosense Webster Inc. PENTARAY® Nav eco Mapping Catheter [510(k) K123837 cleared March 05, 2013]. Like the predicate device, the CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology features a piston like deflection mechanism, allowing for uni-directional deflection of the shaft. The proposed catheter also has similar deflectable curve offerings and a similar flower shaped distal tip. The original intended use of the predicate device as a diagnostic catheter designed to facilitate mapping of structures within the heart remains the same in the proposed device.

Description of the Device Subject to Premarket Notification:

The Biosense Webster CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology is a multi-electrode diagnostic catheter designed to facilitate

{4}------------------------------------------------

electrophysiological mapping of all structures in the heart (recording and stimulation). The catheter's distal end is a flower-shaped probe with 8 spines that radiate from the center. Each spine has 6 ring electrodes that are used for stimulation and recording within the heart. The catheter also includes an electrode located at the confluence of the spines that can be used as a close internal unipolar reference. The flower is available in a 30mm or a 40 mm diameter and several ring spacing configurations to achieve optimal mapping and contact with various cardiac structures.

Indications for Use:

The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The catheter provides location information only when used with compatible versions of the CARTO® 3 EP Navigation System.

Technical Characteristics:

The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ design uses similar technology, has similar intended use, functions, materials and method of operation as the predicate PENTARAY® Nav eco Mapping Catheter. The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology has a flower shaped tip containing 48 electrodes arranged across 8 spines and includes a close uni-polar reference electrode at the confluence of the spines. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation system and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias. The differences in the number of spines and electrodes between the CARTO® OCTARA Y™ Mapping Catheter with TRUEref™ Technology and the PENTARAY® Nav eco Mapping Catheter have been evaluated through bench, animal and biocompatibility testing. Bench, clinical and animal testing demonstrated that the subject device is substantially equivalent to the predicate device. Results from the bench, animal, biocompatibility, and clinical data did not result in new questions with regards to safety and effectiveness of the device.

Performance Data:

The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology underwent bench testing, animal testing, and Clinical testing to demonstrate substantial equivalence.

Testing included Mechanical integrity, electrical leakage, deflection, device functionality, simulated use, biocompatibility, electrical properties, visualization and navigation, sterilization, packaging, shelf life, Animal testing to assess device safety, device maneuverability and signal quality, and clinical testing to assess device effectiveness and safety.

The catheter passed all predetermined acceptance criteria.

{5}------------------------------------------------

The Clinical study was a prospective, single arm, non-randomized, multi-center study. The study intended to enroll 30 subjects with a target of equal numbers of subjects (n=10) among three study arrhythmia subgroups: ischemic ventricular tachycardia, scar related AT (including PAF re-do/repeat procedures), and PsAF. The Study was performed outside of the United States.

The Primary objectives include meeting an effectiveness endpoint of completing preablation mapping and clinically indicated mapping without the need for another device and a safety endpoint of having no serious adverse events within 7 days of the study procedure.

The study population included subjects scheduled to have a clinically-indicated catheter mapping and ablation procedure for management of an arrhythmia in the following 3 subgroups: Ischemic ventricular tachycardia (VT) n=31, Scar-related atrial tachycardia (AT; includes atypical atrial flutter) resulting from previous paroxysmal atrial fibrillation (PAF) ablation procedures, mitral valve repair procedures, or PAF "re-do" procedures n=30, and Persistent Atrial Fibrillation (PsAF) n=30.

Table A summarizes the Subject enrollment.

Table B summarizes subject demographic information

AllEnrolled(N=31)n (%)VT(n=11)n (%)AT/PAF(n=10)n (%)PSAF(n=10)n (%)
Disposition
Enrolled Subjects31(100.0%)11(100.0%)10(100.0%)10(100.0%)
No study catheterinserted1(3.2%)*0 (0.0%)0 (0.0%)1(10.0%)*
Safety Population30(96.8%)11(100.0%)10(100.0%)9(90.0%)
Catheter insertedbut no mapping0 (0.0%)0 (0.0%)0 (0.0%)0 (0.0%)
Mappingperformed but notfor studyarrhythmia2(6.5%) †1(9.1%) †0 (0.0%)1(10.0%) †
Per ProtocolPopulation28(90.3%)10(90.9%)10(100.0%)8(80.0%)

Table A - Subject Enrollment and Accountability (Enrolled Subjects, N=31)

  • Subject 632-003 did not meet eligibility criteria and was withdrawn prior to a study procedure

† Subjects 847-001 (PsAF) & 847-003 (VT) had mapping performed despite not having study arrhythmia

{6}------------------------------------------------

DispositionAllArrhythmias(n=31)OCTARAYFIMAT/PAF(n=10)OCTARAYFIMPsAF(n=10)OCTARAYFIMVT(n=11)
Age at Consent
Mean ± SD (n)67.8 ± 8.52(31)64.5 ± 9.23(10)68.1 ± 7.75(10)70.5 ± 8.25(11)
Median (Q1/Q3)66.0(63.0,74.0)65.5(57.0,73.0)68.0(63.0,72.0)69.0(64.0,79.0)
Min / Max51.0 / 83.051.0 / 76.054.0 / 79.057.0 / 83.0
Sex, n/N (%)
Male27 (87.1%)9 (90.0%)7 (70.0%)11(100.0%)
Female*4 (12.9%)1 (10.0%)3 (30.0%)0 (0.0%)
Race, n/N (%)
Black/AfricanAmerican1 (3.2%)0 (0.0%)1 (10.0%)0 (0.0%)
White25 (80.6%)8 (80.0%)9 (90.0%)8 (72.7%)
Race not reported5(16.1%)2(20.0%)0(0.0%)3(27.3%)
Ethnicity, n/N (%)
Not Hispanic orLatino9 (90.0%)9 (90.0%)10 (100.0%)7 (63.6%)
Not Reported5(16.1%)1(10.0%)0(0.0%)4(36.4%)

Table B - Study Demographic information (Enrolled Subjects, N=31)

Study Results:

The study met Primary endpoints for both safety and effectiveness.

No Serious Adverse events were reported resulting from the use of the study catheter and all physicians were able to perform pre-ablation mapping with the subject device.

In the Per-Protocol Population, 82.1% (23/28) had pre-ablation mapping requirements successfully performed without the need of another catheter.

  • . Three (3) VT subjects did not have LAT mapping performed for reasons of patient safety. It is the position of the Sponsor that these were acceptable minor protocol deviations for patient safety and that the associated endpoint failures do not reflect a

{7}------------------------------------------------

deficiency of the catheter's ability to map but rather a unique unforeseen scenario. Pre-ablation voltage mapping was successfully performed for these subjects.

  • . One (1) VT subject did not have LAT mapping completed due to the operator's inability to maintain stable contact with the area of interest; the Sponsor has made note of this operator's feedback.
  • . One (1) PsAF subject that had CARTOFINDER mapping performed out of sequence was due to human error and was not affected by the catheter design or protocol prescribed workflow.

Clinically indicated mapping was started and finished in 100.0% (15/15) without the need for another device.

The study had 17 adverse events occur across 12 subjects and all were classified as unrelated to the study device. The Adverse Events reported include: Heart Failure, thoracic pain, perifissural nodule, Flu, constipation (two separate subjects), bladder infection, fever, hematuria, AV block 2:1, Dizziness, shoulder bruises, sore throat, non cardiac chest aches, shortness of breath, headaches, and allergic reaction.

Basis for Determination of Substantial Equivalence:

The CARTO® OCTARAY™ Mapping Catheter with TRUEref™ Technology is substantially equivalent to its currently cleared predicate. PENTARA Y Nav eco Mapping Catheter, based on the successful completion of nonclinical bench testing, pre-clinical animal studies, a clinical study, as well as the technological comparison exhibiting similar principles of design, operation, and indications for use.

§ 870.1220 Electrode recording catheter or electrode recording probe.

(a)
Identification. An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.(b)
Classification. Class II (performance standards).