Globe Introducer (601-01001)

K250529 · Kardium, Inc. · DRA · Jun 3, 2025 · Cardiovascular

Device Facts

Record IDK250529
Device NameGlobe Introducer (601-01001)
ApplicantKardium, Inc.
Product CodeDRA · Cardiovascular
Decision DateJun 3, 2025
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1280
Device ClassClass 2

Indications for Use

The Globe Introducer is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum.

Device Story

The Globe Introducer is a single-use, steerable sheath system designed to facilitate percutaneous access to the vasculature and heart chambers. The device comprises a 16 Fr inner diameter steerable sheath with a hydrophilic coating, a dilator, and a dilator loader. The sheath handle features an integrated steering knob for bidirectional deflection control, a steering indicator, and a slider. The slider includes a hemostatic seal, a loader connection port, and a saline flushing line with a three-way stopcock for air management. The device is operated by a clinician in a clinical setting to guide catheters into the heart. The system provides a stable conduit for catheter placement, aiding in cardiac procedures. It is sterilized via ethylene oxide.

Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility, sterilization, packaging, shelf-life, and mechanical performance (e.g., steering knob torque, sheath tensile strength, lubricity, radiopacity, air egress, and pressure withstand).

Technological Characteristics

Materials: PEBAX, Nylon, PTFE liner, Silicone rubber, Polycarbonate, Polyethylene, PVC, Acetal, Rilsan. Features: 16 Fr ID steerable sheath, hydrophilic coating, bidirectional deflection, hemostatic seal, saline flush line. Sterilization: Ethylene Oxide. Connectivity: None. Software: None.

Indications for Use

Indicated for percutaneous catheter introduction into the vasculature and heart chambers, including the left heart via the interatrial septum.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0} FDA U.S. FOOD & DRUG ADMINISTRATION June 3, 2025 Kardium Inc. Ricardo Romero Vice President, RA&QA 8518 Glenlyon Parkway, Unit 155 Burnaby, BC V5J 0B6 Canada Re: K250529 Trade/Device Name: Globe® Introducer Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: May 5, 2025 Received: May 6, 2025 Dear Ricardo Romero: We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov {1} K250529 - Ricardo Romero Page 2 Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download). Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181). Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050. All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory- {2} K250529 - Ricardo Romero Page 3 assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, # Aneesh S. Deoras -S Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {3} FORM FDA 3881 (8/23) Page 1 of 1 PSC Publishing Services (301) 443-0340 EF | DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use | Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below. | | --- | --- | | 510(k) Number (if known) K250529 | | | Device Name Globe® Introducer | | | Indications for Use (Describe) The Globe Introducer is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum. | | | Type of Use (Select one or both, as applicable) ☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | | | CONTINUE ON A SEPARATE PAGE IF NEEDED. | | | This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* | | | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov | | | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | | {4} Kardium K250529 # 510(k) Summary per 21 CFR §807.92 | Sponsor | Kardium Inc. 155-8518 Glenlyon Parkway Burnaby, British Columbia V5J 0B6 Canada | | --- | --- | | Contact Name and Information | Ricardo Romero Vice President, RA&QA Kardium Inc. Tel: +1 604.248.8891 | | Date Prepared by | February 21st, 2025 | | Proprietary Name | Globe® Introducer | | Common Name | Steerable catheter | | Product Code | DRA | | Classification | Class II, 21 CFR Part 870.1280 | | Predicate Device | FARADRIVE™ Steerable Sheath (K233248) | | Device Description | The Globe® Introducer is a single use, ethylene oxide sterilized medical device used to facilitate percutaneous access to the vasculature and into the heart chambers. The device consists of a 16 Fr (5.3 mm) inner diameter (ID) steerable sheath with a hydrophilic coating, a dilator compatible with 0.89 mm (0.035 in) guidewires, and a dilator loader. The proximal end of the sheath includes a handle with an integrated steering knob for bidirectional deflection control of the sheath tip, a steering indicator, and a slider. The slider contains a loader connection port for the dilator loader and a compatible catheter loader. The slider provides a hemostatic seal and a saline flushing line with a standard Luer fitting stopcock, for air removal and management. | | Indications for Use | The Globe Introducer is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum. | Kardium Inc. Globe® Introducer - 510(k) Summary Page 1 of 5 {5} Kardium K250529 # Predicate Comparison Table 1: Comparison of Technological Characteristics with the Predicate Device | Characteristics | Subject Device: Globe® Introducer | Predicate Device: FARADRIVE™ Steerable Sheath (K233248) | Comparison | | --- | --- | --- | --- | | Product Code | DRA | DRA | Identical | | Intended Use / Indications for Use | The Globe Introducer is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum. | The FARADRIVE Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the heart through the interatrial septum. | Same Intended Use / Indications for Use | | Design Features | Steering knob that rotates in the plane of the handle | Actuator knob that rotates coaxially to the handle | Differences in the steering mechanism do not alter the fundamental scientific technology or pose an added safety risk. | | | Bi-directional shaft deflection clockwise ≥90° counterclockwise ≥45° | Unidirectional shaft deflection from 0° to 180° | The range of motion has been shown to be sufficient for anatomical placement of compatible catheters. The differences do not raise new questions of safety or effectiveness. | | | Integrated hemostatic seal, slider port, and flush line with three-way stopcock | Integrated hemostasis valve and side flush line with three-way stopcock | The differences in design do not alter the fundamental scientific technology or raise new questions about device performance or safety. | Kardium Inc. Globe® Introducer - 510(k) Summary Page 2 of 5 {6} Kardium K250529 | | Distal vent holes | Distal vent holes | Same design characteristics | | --- | --- | --- | --- | | Radiopaque tip marker | Radiopaque tip marker | Same design characteristics | | Dimensions | Sheath size: - Inner diameter: 16 F - Usable length: 71 cm - Sheath outer diameter: 21 F | Sheath size: - Inner diameter: 13 F - Usable length: 74 cm - Sheath outer diameter: 17.4 F | The differences in sheath dimensions ensure compatibility with prescribed catheters under identical intended uses of the sheath. | | Dilator size: - Dilator usable length: 102 cm (40.24 in) - Dilator inner diameter: 0.9 mm (0.035 in) - Dilator outer diameter: 5.2 mm (0.205 in) | Dilator size: - Dilator usable length: 94 cm (36.85 in) - Dilator inner diameter: 0.9 mm (0.037 in) - Dilator outer diameter: 4.4 mm (0.172 in) | The differences in dilator dimensions do not alter the fundamental scientific technology nor do they raise new questions about device performance or safety. | | Guidewire Compatibility | Maximum diameter: 0.89 mm (0.035 in) | Maximum diameter: 0.89 mm (0.035 in) | Same guidewire compatibility. | | Materials | This device contains: - Hydrophilic coated (polyvinylpyrrolidone) polyether block amide (PEBAX) - Nylon - Polytetrafluoroethylene (PTFE) liner - Silicone rubber - Polycarbonate - Polyethylene - Polyvinyl Chloride (PVC) - Acetal - Rilsan | This device contains: - Polyether block amide (PEBAX) - Nylon - Stainless steel - Hydrophilic coated PEBAX liner - Silicone rubber - Polycarbonate - Polyethylene | The subject and predicate devices are constructed of equivalent materials. Any differences have been qualified through biological safety and performance testing. | | Sterilization | Ethylene Oxide (EO) | Ethylene Oxide (EO) | Same sterilization method. | Kardium Inc. Globe® Introducer - 510(k) Summary Page 3 of 5 {7} Kardium K250529 | Single Use | Yes | Yes | Same limitations on use. | | --- | --- | --- | --- | | Packaging | Tyvek nylon/polyethylene pouch with a protective tray | Tyvek nylon/polyethylene pouch | Same sterile barrier system. Minor differences in packaging do not raise new questions about device performance or safety, as demonstrated by pre-clinical and packaging validation testing. | ## Non-clinical Performance Data The following testing was conducted to support the determination of substantial equivalence between the Globe Introducer and the predicate device: - Biocompatibility Testing - Sterilization Validation - Packaging Validation - Shelf-Life Testing - Bench Testing - Particulate Test - Deployment-Retraction Force Test - Flushing Test - Pressure Withstand Test - Corrosion Resistance Test - ISO 80369 Luer Test - Introducer Health Check Test - Dimensional Inspection Test - Steering Knob Torque Test - Sheath Tensile Test - Mechanical Performance Test - Torque Test - Lubricity Test - Radiopacity Test - Dilator Performance Test - Air Egress Test Bench Testing was performed following the applicable principles and recommendations of relevant FDA-recognized consensus standards and guidance documents. The test results demonstrate that the Globe Introducer meets the performance criteria for its intended use and does not raise new questions on safety or effectiveness compared to the predicate device. Kardium Inc. Globe® Introducer - 510(k) Summary Page 4 of 5 {8} K250529 Kardium Statement of Equivalence The Globe Introducer and predicate device have identical or equivalent intended use, indications for use, technological characteristics, and performance. Differences in design do not raise any new or different questions of safety and effectiveness. Based on this and the results of performance testing, the Globe Introducer is substantial equivalent to the predicate device. Kardium Inc. Globe® Introducer - 510(k) Summary Page 5 of 5
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...