K Number
K233248

Validate with FDA (Live)

Date Cleared
2023-12-11

(74 days)

Product Code
Regulation Number
870.1280
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FARADRIVE Sheath is intended for percutaneous catheter introduction into the chambers of the heart, including the left side of the heart through the interatrial septum.

Device Description

The FARADRIVE Steerable Sheath is used to facilitate access to the vasculature and into the chambers of the heart. The handle is furnished with a rotating Actuator Knob that enables a single plane unidirectional deflection and has the ability to actively straighten the distal end of the FARADRIVE Steerable Sheath curve when rotated.

The FARADRIVE Steerable Sheath includes an integral hemostasis valve to minimize blood loss or air ingress during catheter and dilator manipulation and a side flush line with three-way stopcock for blood aspiration, fluid infusion, blood sampling and pressure monitoring. The sheath also includes distal vent holes to facilitate aspiration and minimize cavitation. The supplied dilator is shaped to assist in facilitating vascular or chamber access when fully inserted into the sheath. The proximal end of the dilator includes a feature that will lock to the sheath hub when fully inserted.

AI/ML Overview

The provided text is a 510(k) summary for the FARADRIVE™ Steerable Sheath. This document primarily focuses on demonstrating substantial equivalence to a predicate device through comparison of technological characteristics and non-clinical performance testing. It does not describe a study involving an algorithm or AI, nor does it establish ground truth with experts or involve human readers.

Therefore, I cannot provide the requested information about acceptance criteria and the study proving a device meets acceptance criteria using an AI/algorithm-based evaluation. The available information pertains to the regulatory clearance of a medical device (a steerable sheath), not a diagnostic algorithm.

Specifically, the document details physical and performance characteristics of the FARADRIVE™ Steerable Sheath and compares them to a predicate device (POLARSHEATH™ Steerable Sheath 12F) to establish substantial equivalence for regulatory approval.

Here's a breakdown of why the requested information cannot be extracted from the provided text:

  • No AI/Algorithm: The submitted device, FARADRIVE™ Steerable Sheath, is a physical medical device (a catheter sheath), not an AI or algorithm-based diagnostic tool.
  • No "Test Set" or "Training Set": These terms are relevant to machine learning studies. The "test set" mentioned in the document refers to performance testing of the physical device per design specifications, not an image or data set used for evaluating an algorithm.
  • No "Ground Truth" by Experts: The "ground truth" concept is typically used to establish the correct diagnosis or finding in data used for training/testing AI. This document does not describe such a process.
  • No "MRMC Comparative Effectiveness Study": This type of study compares the diagnostic performance of human readers with and without AI assistance. This is not applicable to a physical steerable sheath.
  • No "Standalone Performance": "Standalone performance" refers to the algorithm's performance without human intervention. Again, this is not relevant to a physical device.

What the document does provide is a summary of non-clinical performance testing for a physical device:

The "Summary of Non-Clinical Performance Testing" on page 8 confirms that the testing aimed to show the device "conforms to user needs and its intended use and is thus substantially equivalent to the predicate." The listed tests include:

  • Sterilization validation
  • Biocompatibility assessment
  • Design verification testing (at nominal and accelerated aging timepoints)
  • Design validation testing summary for the FARAPULSE Pulsed Field Ablation System (presumably to show compatibility or a larger system context, not a performance study of the sheath itself)

These tests are standard for physical medical devices to ensure they meet safety and performance requirements, but they do not involve the concepts of AI, algorithms, ground truth, or expert readers as outlined in your request.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

December 11, 2023

Boston Scientific Corporation Nicholas Zurn Regulatory Affairs Specialist II 4100 Hamline Avenue North St. Paul, Minnesota 55112

Re: K233248

Trade/Device Name: FARADRIVE™ Steerable Sheath Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: Class II Product Code: DRA Dated: September 28, 2023 Received: September 28, 2023

Dear Nicholas Zurn:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"

{1}------------------------------------------------

(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Aneesh S. Deoras -S

Aneesh Deoras Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices

{2}------------------------------------------------

Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K233248

Device Name FARADRIVE™ Steerable Sheath

Indications for Use (Describe)

The FARADRIVE Sheath is intended for percutaneous catheter introduction into the chambers of the heart, including the left side of the heart through the interatrial septum.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Boston Scientific. The logo is in a dark blue color and features a serif font. The words "Boston Scientific" are stacked on top of each other.

510(k) Summary in Compliance with 21 CFR 807.92

Submitter Information:

Submitter Name: Boston Scientific Corporation

Submitter Address: 4100 Hamline Avenue North St. Paul. Minnesota 55112-5798 USA

Contact #1: Nicholas Zurn Regulatory Affairs Specialist II Phone: (651) 287 5061 Email: nicholas.zurn(@bsci.com

Contact #2: Richard Kimura Principal Regulatory Affairs Specialist Phone: (651) 582 4000 Email: richard.kimura@bsci.com

Date Prepared: 28-September-2023

Device Information:

Generic Name: Steerable Sheath

Trade Name: FARADRIVE™ Steerable Sheath

Product Code: DRA (Catheter, Steerable)

510(k) Number: K233248

Device Class: Class II

Review Panel: Cardiovascular (DRA)

{5}------------------------------------------------

Classification Regulations: 21 CFR 870.1280

Predicate Device Information:

POLARSHEATH™ Steerable Sheath 12F

Manufacturer: Boston Scientific Corporation 510(k) #: K223824 FDA Clearance Date: March 22nd, 2023 Product Code: DRA

Reference Device Information:

MitraClip G4 Steerable Guide Catheter

Manufacturer: Abbott Medical 510(k) #: K190167 FDA Clearance Date: May 29th, 2019 Product Code: DRA

Indications for Use:

The FARADRIVE Sheath is intended for percutaneous catheter introduction into the vasculature and into the chambers of the heart, including the left side of the interatrial septum.

Device Description:

The FARADRIVE Steerable Sheath is used to facilitate access to the vasculature and into the chambers of the heart. The handle is furnished with a rotating Actuator Knob that enables a single plane unidirectional deflection and has the ability to actively straighten the distal end of the FARADRIVE Steerable Sheath curve when rotated.

The FARADRIVE Steerable Sheath includes an integral hemostasis valve to minimize blood loss or air ingress during catheter and dilator manipulation and a side flush line with three-way stopcock for blood aspiration, fluid infusion, blood sampling and pressure monitoring. The sheath also includes distal vent holes to facilitate aspiration and minimize cavitation. The supplied dilator is shaped to assist in facilitating vascular or chamber access when fully inserted into the sheath. The proximal end of the dilator includes a feature that will lock to the sheath hub when fully inserted.

{6}------------------------------------------------

Substantial Equivalence & Comparison of Technological Characteristics:

CharacteristicPredicate Device (K223824)POLARSHEATH™ Steerable Sheath12F (Boston ScientificCorporation)Subject Device (K233248)FARADRIVE Steerable Sheath(Boston Scientific Corporation)
Indication for UseThe POLARSHEATH™ steerablesheath is indicated for percutaneouscatheter introduction into thevasculature and into the chambers ofthe heart.The FARADRIVE Sheath is intendedfor percutaneous catheterintroduction into the vasculature andinto the chambers of the heart,including the left side of the heartthrough the interatrial septum.
Single Use or ReusableSingle UseSingle Use
Shaft DesignDeflectable shaft with single lumenDeflectable shaft with single lumen
DeflectionUnidirectional to 155°Unidirectional to 180° with activestraightening back to 0°
Hemostasis ValveYesYes
Side Flush Port TubingYesYes
StopcockNoYes
Braided ShaftYesYes
Radiopaque Marker(s)YesYes
Guidewire Compatibility0.032 inches (0.81mm) and 0.035inches (0.89mm)0.035 inches (0.89mm)
SterilizationProvided sterile via Ethylene OxideProvided sterile via Ethylene Oxide
PackagingTyvek nylon/polyethylene pouchTyvek nylon/polyethylene pouch
Shelf Life12 months24 months
Sheath Dimensions
Overall Length82cm91cm
Usable Length68cm74cm
Sheath OD5.3mm (15.9Fr)5.8mm (17.4Fr)
CharacteristicPredicate Device (K223824)POLARSHEATH Steerable Sheath12F (Boston ScientificCorporation)Subject Device (K233248)FARADRIVE Steerable Sheath(Boston Scientific Corporation)
Sheath ID4.2mm (12.7Fr)4.3mm (13Fr)
Dilator Dimensions
Dilator Usable Length85cm94cm
OD0.159 inches0.163 inches
ID0.039 inches0.037 inches
Materials
- Pebax with stainless steel braid andPTFE liner- Silicone Rubber- Polycarbonate- Polyethylene- Pebax, Nylon 12, with stainlesssteel braid and hydrophilic coatedPebax liner- Silicone Rubber- Polycarbonate- Polyethylene

Table 1 - COMPARISON OF THE SUBJECT FARADRIVE STEERABLE SHEATH VERSUS PREDICATE POLARSHEATH STEERABLE SHEATH 12F

{7}------------------------------------------------

Summary of Non-Clinical Performance Testing:

The non-clinical performance testing included in this 510(k) application provides evidence that the FARADRIVE Steerable Sheath conforms to user needs and its intended use and is thus substantially equivalent to the predicate previously cleared under K223824. No new safety or effectiveness issues were raised during device performance testing. The following test data is provided in this 510(k) application:

  • Sterilization validation, including EO Cycle Validation and Product Qualification, for the . sterile FARADRIVE Steerable Sheath
  • Biocompatibility assessment and testing for the patient-contacting components of the ● FARADRIVE Steerable Sheath
  • Design verification testing of the FARADRIVE Steerable Sheath and its packaging, ● demonstrating that the device and its packaging meets the design input requirements as per the product and packaging specifications
    • Design verification testing at nominal (T=0) timepoint O
    • Design verification testing at an accelerated aging timepoint for a labeled shelf life of O T=24 months
  • Design validation testing summary for the FARAPULSE Pulsed Field Ablation System ● demonstrating that the system conforms to user needs and intended use

{8}------------------------------------------------

Conclusion:

Based on the intended use, fundamental technological characteristics, and performance testing, the proposed FARADRIVE Steerable Sheath device has shown to be appropriate for its intended use and is substantially equivalent to the predicate device.

§ 870.1280 Steerable catheter.

(a)
Identification. A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.(b)
Classification. Class II (performance standards).