QUANTSTUDIO DX REAL-TIME PCR INSTRUMENT

K123955 · Life Technologies Corporation · OOI · Mar 8, 2013 · Clinical Chemistry

Device Facts

Record IDK123955
Device NameQUANTSTUDIO DX REAL-TIME PCR INSTRUMENT
ApplicantLife Technologies Corporation
Product CodeOOI · Clinical Chemistry
Decision DateMar 8, 2013
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.2570
Device ClassClass 2

Indications for Use

The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument.

Device Story

Benchtop real-time PCR instrument; utilizes fluorescent-based PCR reagents for qualitative/quantitative detection of target nucleic acid sequences. System includes instrument with eGUI touchscreen, thermal block (Peltier-based), heated cover, plate adaptor, and computer workstation with QuantStudio™ Dx software. Operated by trained laboratory technologists in clinical settings. Workflow involves manual transfer of amplification mixture to 96-well plates (10-30 µL volume), followed by automated thermal cycling, optical detection of stimulated fluorescence, and data analysis. Output provides diagnostic results for nucleic acid targets. Benefits include standardized, automated amplification and detection for clinical molecular diagnostics.

Clinical Evidence

Bench testing and clinical performance evaluation conducted using Quidel Molecular Direct C. difficile Assay. Clinical study (n=792) compared device against tissue culture cytotoxin assay and enhanced toxigenic culture. Sensitivity 93.3% (vs cytotoxin) and 87.3% (vs culture); Specificity 93.4% (vs cytotoxin) and 98.7% (vs culture). Precision/reproducibility studies confirmed performance across sites and operators.

Technological Characteristics

Benchtop real-time PCR instrument; Peltier-based thermal cycling; microvolume fluorimeter for optical detection of stimulated fluorescence. Supports 96-well Fast PCR plates (10-30 µL reaction volume). Connectivity via computer workstation. Software-controlled automated amplification and detection.

Indications for Use

Indicated for in vitro diagnostic use by trained laboratory technologists to detect FDA-cleared/approved nucleic acid sequences in human-derived specimens using fluorescence-based PCR.

Regulatory Classification

Identification

Instrumentation for clinical multiplex test systems is a device intended to measure and sort multiple signals generated by an assay from a clinical sample. This instrumentation is used with a specific assay to measure multiple similar analytes that establish a single indicator to aid in diagnosis. Such instrumentation may be compatible with more than one specific assay. The device includes a signal reader unit, and may also integrate reagent handling, hybridization, washing, dedicated instrument control, and other hardware components, as well as raw data storage mechanisms, data acquisition software, and software to process detected signals.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9. The special control is FDA's guidance document entitled “Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems.” See § 862.1(d) for the availability of this guidance document.

The special control is FDA's guidance document entitled "Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems."

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K123955 ### MAR 0 8 2013 ### 510(k) Summary ### Summary of Safety and Effectiveness ### Submitter Information - 21 CFR 807.92(a)(1): - Life Technologies Corporation Submitter: 5791 Van Allen Way Carlsbad, CA 92008 - Manufacturer: Life Technologies Holdings Pte Ltd Blk 33, #07-06, Marsiling Industrial Estate, Road 3 Singapore 739256 Establishment Registration No: 3003673482 - Contact: Deanna Vella, Regulatory Affairs Manager Phone: 760-918-3000 760-476-6934 Fax: E-mail: deanna.vella@lifetech.com Alternate Contact: Nikki Arora, Engineer, Regulatory Affairs Phone: 650-554-2268 Fax: 650-638-6786 nikki.arora@lifetech.com E-mail: Date Prepared: December 20, 2012 ### Name of Device and Classification - 21 CFR 807.92(a)(2): QuantStudio™ Dx Real-Time PCR Instrument Product Name: Device Classification: Class II Product Code: OOI, Real-Time Nucleic Acid Amplification System for Real Time Instruments. {1}------------------------------------------------ # ذابة ( ) - ) - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - - Predicate: Abbott m2000™ System consisting of m2000sp and m2000rt instruments, K092705 ### Device Description - 21 CFR 807.92(a)(4): The QuantStudio™ Dx Real-Time PCR Instrument is a bench top Real-Time PCR instrument that uses fluorescent-based polymerase chain reaction (PCR) reagents to provide qualitative or quantitative detection of target nucleic acid sequences (targets) using real-time analysis. The QuantStudio™ Dx Real-Time PCR Instrument system includes the following components: - QuantStudio™ Dx Real-Time PCR instrument with embedded graphical user . interface (eGUI) Touchscreen - Thermal Block, also referred to as the sample block, with associated Heated Cover . and Plate Adaptor - Calibration and verification materials for instrument qualification . - . Computer workstation with a monitor, keyboard and mouse - QuantStudio™ Dx instrument software . ### Intended Use/Indications for Use - 21 CFR 807.92(a)(5): The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument. ### Summary of technological characteristics of the device compared to the predicate devices- 21 CFR 807.92(a)(6): The Life Technologies QuantStudio™ Dx Real-Time PCR Instrument ("Subject Device") and the legally marketed devices, Abbott m2000™ System ("Predicate Device") is described in the table below: {2}------------------------------------------------ Predicates Comparison – Life Technologies QuantStudio™ Dx Real-Time PCR Instrument vs. Abbott m2000™ System | Item | Subject Device<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument | Predicate Device<br>Abbott m2000™ System | | |------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------| | Similarities | | | | | 510(k) | N/A | K092705 | | | Regulation | 862.2570<br>Instrumentation for<br>clinical multiplex test<br>systems. | Same | | | Product Code | OOI: Real-Time<br>Nucleic Acid<br>Amplification System<br>for Real Time<br>Instruments. | OOI: Real-Time Nucleic<br>Acid Amplification<br>System for Real Time<br>Instruments. | | | Device Class | Class II | Same | | | Intended Use | The QuantStudio™ Dx Real-Time PCR<br>Instrument with<br>QuantStudio™ Dx<br>Software is intended to<br>perform fluorescence-<br>based PCR to provide<br>detection of FDA<br>cleared and approved<br>nucleic acid sequences<br>in human-derived<br>specimens. The<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument with<br>QuantStudio™ Dx<br>Software is intended<br>for in vitro diagnostic<br>use by trained<br>laboratory<br>technologists in<br>combination with<br>nucleic acid reagent<br>kits/tests manufactured<br>and labeled for<br>diagnostic purposes on | The Abbott m2000™<br>System is intended for in<br>vitro diagnostic use in<br>performing FDA cleared<br>and approved nucleic acid<br>testing in clinical<br>laboratories. It comprises<br>the Abbott m2000sp and<br>the Abbott m2000rt<br>instruments. The Abbott<br>m2000sp is an automated<br>system for performing<br>sample preparation for<br>nucleic acid testing. The<br>Abbott m2000rt is an<br>automated system for<br>performing fluorescence-<br>based PCR to provide<br>quantitative and qualitative<br>detection of nucleic acid<br>sequences. | | | Item | Subject Device<br>QuantStudio™ Dx<br>Real-Time PCR<br>Instrument | Predicate Device<br>Abbott m2000™ System | | | Similarities | | | | | | this instrument. | | | | Technology/<br>Detection | Real-Time PCR | Same | | | | Specimen Types | Nucleic acid | Same | | Assay Format | Homogeneous, closed<br>tube PCR | Same | | | Degree of<br>Automation | Requires manual<br>transfer of<br>amplification mixture<br>to<br>amplification/detection<br>instrument.<br>Automated control of<br>amplification,<br>detection, and data<br>analysis. | Same | | | Primary<br>Operational<br>Amplification<br>and Detection<br>Components | Integrated thermal<br>cycler and<br>microvolume<br>fluorimeter for walk<br>away PCR<br>amplification and<br>detection | Same | | | | Heating Method<br>for Amplification | Peltier device with<br>sample block | Same | | Detection<br>Procedure | Optical detection of<br>stimulated<br>fluorescence | Same | | | Detection<br>Chemistries | Fluorescence labeled<br>target-specific probes | Same | | | Item | Subject Device<br>QuantStudioTM Dx Real-Time PCR Instrument | Predicate Device<br>Abbott m2000TM System | | | Differences | | | | | Product Code | OOI: Real-Time Nucleic Acid Amplification System for Real Time Instruments. | OOI: Real-Time Nucleic Acid Amplification System for Real Time Instruments. | | | User Interface | PC with instrument-specific software. Instrument has touchscreen console. | PC with instrument-specific software | | | Amplification Reaction Volume | 10-30 $\mu$ L in 96-well Fast PCR plates | 25-100 $\mu$ L in 96-well PCR plates | | | Sample Preparation | No automated sample processing instrument offered in conjunction with the QuantStudioTM Dx Instrument. | Pairing with the m2000sp instrument provides automated sample processing. | | ・ · : . {3}------------------------------------------------ and the control control control of the control of the control of and the state of the state {4}------------------------------------------------ Our analysis of the differences in user interface, amplification reaction volumes, and sample preparation between the Subject Device and the Predicate Device indicates that these differences do not impact performance or raise new/different questions of safety and effectiveness, and therefore render the Subject Device as Substantially Equivalent. ### Special Control/Guidance Document Referenced (if applicable): Class II Special Controls Guidance Document: Instrumentation for Clinical Multiplex Test Systems: http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocum ents/ucm077819.htm ### Performance Data - 21 CFR 807.92(b): As noted in the Cover Letter in this 510(k), Quidel® Corporation will submit a traditional 510(k) for the Molecular Real-Time PCR Direct C. Difficile Tox A/B that will be used with the QuantStudio™ Dx Real-Time PCR Instrument. To that end, testing to demonstrate non-clinical performance of the OuantStudio™ Dx Real-Time PCR Instrument was led by Quidel® Corporation as part of a collaboration agreement between {5}------------------------------------------------ the two companies. This section provides a brief summary of the non-clinical performance studies and conclusions that demonstrate instrument performance when testing the Quidel Molecular Real-Time PCR Direct C. Difficile Tox A/B.. Complete non-clinical performance data can be found in Quidel's Molecular Real-Time PCR Direct C. Difficile Tox A/B traditional 510(k) submission. ### Non-Clinical Performance Data- 21 CFR 807.92(b)(1): ### Analytical performance: #### Precision/Reproducibility: a. Precision: For the Precision/Within Laboratory Repeatability study, a blinded fourmember panel consisting of C. difficile positive and negative sample was tested by two operators, twice a day using a single assay lot of Quidel Molecular Direct C. difficile Assay for twelve (12) days. | QuantStudio™ Dx Real-Time PCR Instrument | | | | | |------------------------------------------|--------|--------|----------|----------| | <i>C. difficile</i> | 5X LoD | 2X LoD | 0.3X LoD | Negative | | % Detection | 100% | 100% | 88% | 0% | | Average Ct | 16.51 | 17.70 | 21.13 | N/A | | STDEV | 0.42 | 0.76 | 1.37 | N/A | | %CV | 2.6% | 4.3% | 6.5% | N/A | Reproducibility: In order to confirm the reproducibility of the Quidel Molecular Direct C. difficile Assay a blinded and randomized study panel containing Clostridium difficile negative and positive samples were tested at three (3) test sites, two of which were clinical sites. Each site tested a reproducibility panel and assay controls for five (5) days in triplicate on each instrument. The testing was done by two operators at each site. Each operator ran the panel once a day using one lot of Quidel Molecular Direct C. difficile Assay. {6}------------------------------------------------ | Reproducibility Results – QuantStudio™ Dx Instrument | | | | | | | | | | | |------------------------------------------------------|---------|-----------|-----|---------|-----------|-----|---------|-----------|-----|------------------| | Panel<br>Member<br>ID | Site 1 | | | Site 2 | | | Site 3 | | | Total<br>Results | | | Results | AVE<br>Ct | %CV | Results | AVE<br>Ct | %CV | Results | AVE<br>Ct | %CV | | | High<br>Negative<br>0.3x LoD | 8/30 | 22.9 | 5.0 | 15/30 | 22.5 | 1.3 | 15/30 | 22.5 | 1.5 | 38/90 | | Low<br>Positive<br>2x LoD | 30/30 | 20.4 | 5.9 | 30/30 | 19.0 | 5.1 | 30/30 | 19.2 | 0.8 | 90/90 | | Med<br>Positive<br>5x LoD | 30/30 | 18.4 | 4.2 | 30/30 | 17.5 | 0.4 | 30/30 | 17.9 | 0.7 | 90/90 | | Negative<br>Specimen | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/90 | | Negative<br>Control | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/30 | N/A | N/A | 0/90 | | Positive<br>Control | 30/30 | 15.7 | 0.6 | 30/30 | 15.7 | 0.1 | 30/30 | 15.5 | 0.1 | 90/90 | #### b. Detection limit: The analytical sensitivity (limit of detection or LoD) of the Quidel Molecular Direct C. difficile Assay was determined on QuantStudio™ Dx instrument using quantified (CFU/mL) cultures of two C. difficile strains (ATCC BAA-1870 and ATCC BAA-1872) serially diluted in a negative fecal matrix. Analytical sensitivity (LoD) is defined as the lowest concentration at which 95% of all replicates tested positive. | Instrument | Strain | | |--------------------------------------|--------------------------------------|---------| | ATCC BAA-1870 LoD<br>(CFU per assay) | ATCC BAA-1872 LoD<br>(CFU per assay) | | | QuantStudio™<br>Dx | 4.2 E-01 | 4.0E-02 | The final assay LoD is defined as the higher of the two strain concentrations where 95% positivity was observed. The final assay LoD is 4.2E-01 CFU/assay. {7}------------------------------------------------ ### Clinical Performance Data- 21 CFR 807.92(b)(2): Comparison studies: Method comparison with predicate device: a. Performance characteristics of the Quidel Molecular Direct C. difficile Assay were established during a prospective study conducted August to November 2012. Seven hundred and ninety-two (792) samples used for this study were collected from patients suspected of having Clostridium difficile-associated disease (CDAD) at four (4) distinct geographical sites across the United States. These specimens were tested with the Quidel Assay on the Life Technologies QuantStudio™ Dx Real-Time PCR Instrument at one of three (3) facilities. | Combined Sites – Age and Gender Distribution | | | | | |----------------------------------------------|--------|--------|---------------------------------------------------------------------------------------------------------------------------------------|---------------| | Age | Gender | | Total prevalence of the<br>Quidel Molecular<br>Direct C. difficile<br>Assay on the<br>QuantStudioTM Dx<br>Real-Time PCR<br>Instrument | | | | Male | Female | Total | | | Unknown | | | 2 | 50% (1/2) | | ≤ 2 years | 5 | 5 | 10 | 10% (1/10) | | 2 to <12 years | 28 | 21 | 49 | 24% (12/49) | | 12 to <18 years | 10 | 14 | 24 | 21% (5/24) | | 18 to 21 years | 6 | 7 | 13 | 8% (1/13) | | >21 to 59 years | 158 | 170 | 328 | 18% (60/328) | | ≥ 60 years | 163 | 203 | 366 | 18% (65/366) | | Total | 370 | 420 | 792 | 18% (145/792) | Patient age and gender for the combined sites are presented below. * includes two (2) patient samples with unknown age and gender. ### Tissue Culture Cytotoxicity Assay Comparison Seven hundred and ninety-two (792) samples were tested by both the Quidel Molecular Direct C. difficile Assay and the tissue culture cytotoxin assay. Three (3) specimens (0.4%) were indeterminate in the cytotoxin assay due to toxicity in the antitoxin well. One (1) specimen (0.1%) was invalid in the Quidel Molecular Direct C. difficile Assay {8}------------------------------------------------ when initially tested. The specimen yielded a valid result (it was negative) when retested according to the Quidel Molecular Direct C. difficile Assay draft instructions for use. We elected to calculate clinical performance based on the initial test result obtained for each specimen. Therefore, the data below is for the remaining seven hundred and eighty-eight (788) specimens. | Combined Sites – Combined Ages | | | | | | | | | |-------------------------------------------------------------------------------------------|-------|-----|-----|-------|-------------|--------|-------|-------| | Tissue Culture Cytotoxin | | | | | | 95% CI | | | | | | POS | NEG | Total | Sensitivity | | | | | Quidel<br>Molecular<br>Real-<br>Time<br>PCR<br>Direct C.<br>difficile<br>Tox A/B<br>Assay | POS | 98 | 45* | 143 | 93.3% | 86.9% | 96.7% | | | | NEG | 7** | 638 | 645 | Specificity | 93.4% | 91.3% | 95.0% | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | | Total | 105 | 683 | 788 | | | | | * Of the forty-five (45) discordant specimens (Quidel Molecular Positive/Tissue Culture Cytotoxin Negative) reported, forty-four (44) were tested with a FDA-cleared molecular device. Thirty-five (35) of these specimens were positive for C. difficile, and nine (9) were negative. The remaining specimen was unavailable for testing. **Seven (7) discordant specimens (Quidel Negative/Tissue Culture Cytotoxin Positive) reported were tested with a FDA-cleared molecular device. Two (2) of these specimens were found positive for C. difficile, and five (5) were negative. {9}------------------------------------------------ ### Enhanced Toxigenic Culture Comparison Seven hundred and ninety-two (792) samples were tested by both the Ouidel Molecular Direct C. difficile Assay and enhanced toxigenic culture. One (1) specimen (0.1%) was invalid in the Quidel Molecular Direct C. difficile Assay when initially tested. The specimen yielded a valid result (it was negative) when retested according to the Quidel Molecular Direct C. difficile Assay draft instructions for use. We elected to calculate clinical performance based on the initial test result obtained for each specimen. Therefore, the data below is for the remaining seven hundred and nimety-one (791) specimens. | Combined Sites – Combined Ages | | | | | | | | | |--------------------------------------------------------|-------|------|--------|-------|-------------|-------|-------|-------| | Enhanced Toxigenic Culture | | | 95% CI | | | | | | | | | POS | NEG | Total | Sensitivity | 87.3% | 81.1% | 91.6% | | Quidel<br>Molecular<br>Direct C.<br>difficile<br>Assay | POS | 137 | 8* | 145 | Specificity | 98.7% | 97.5% | 99.4% | | | NEG | 20** | 626 | 646 | | | | | | | Total | 157 | 634 | 791 | | | | | * Eight (8) discordant specimens (Quidel Molecular Positive/Enhanced Toxigenic Culture Negative) reported were tested with a FDA-cleared molecular device. Two (2) of these specimens were positive for C. difficile, and six (6) were negative. ** Seventeen (17) out of twenty (20) discordant specimens (Quidel Negative/ Enhanced Toxigenic Culture Positive) reported, were tested with a FDA-cleared molecular device. Three (3) specimens were unavailable for testing. Eleven (11) of these specimens were found negative for C. difficile, and six (6) were positive. ### Conclusion This summary of safety and effectiveness information provides the necessary detail for a determination of substantial equivalence for the QuantStudio™ Dx Real-Time PCR Instrument. {10}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 March 8, 2013 Life Technologies Deanna Vella 5791 Van Allen Way Carlsbad, California 92008 Re: K123955 Trade/Device Name: Life Technologies Regulation Number: 21 CFR §862.2570 Regulation Name: QuantStudio™ DX Real-Time PCR Instrument Regulatory Class: Class II Product Code: OOI Dated: December 20, 2012 Received: December 21, 2012 Dear Ms. Vella: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may; therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {11}------------------------------------------------ Page 2 - Ms. Vella device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807:97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Sincerely yours, ## Uwerschierf -S for Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {12}------------------------------------------------ ### Indications for Use Statement 510(k) Number (if known): k123955 Device Name: QuantStudio™ Dx Real-Time PCR Instrument ### Indications for Use: The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended to perform fluorescence-based PCR to provide detection of FDA cleared and approved nucleic acid sequences in human-derived specimens. The QuantStudio™ Dx Real-Time PCR Instrument with QuantStudio™ Dx Software is intended for in vitro diagnostic use by trained laboratory technologists in combination with nucleic acid reagent kits/tests manufactured and labeled for diagnostic purposes on this instrument. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Center for Devices and Radiological Health Raquel Aृङ्घिeat -- 3 2013.03.03.07 -05.00 -05.00' Division Sign-Off CDRH, Center for Devices and Radiological Health 510(k) K123955 Page _1_ of _1__
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...