AFFIRM ONE-STEP HOME PREGNANCY TEST

K962888 · Athena Medical Corp. · LCX · Sep 23, 1996 · Clinical Chemistry

Device Facts

Record IDK962888
Device NameAFFIRM ONE-STEP HOME PREGNANCY TEST
ApplicantAthena Medical Corp.
Product CodeLCX · Clinical Chemistry
Decision DateSep 23, 1996
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1155
Device ClassClass 2

Indications for Use

Affirm One-Step home pregnancy test is a rapid one-step visual test for the qualitative detection of human chorionic gonadotropin in human urine to aid in the diagnosis of pregnancy. This test is intended for over-the-counter home use.

Device Story

Lateral flow immunoassay for qualitative hCG detection in human urine; utilizes polyclonal and monoclonal antibodies on reagent pad and test strip. User applies urine to absorbent wick; capillary action transports sample to reagent pad; hCG reacts with conjugate. Positive result indicated by color lines in test and control windows; negative result by control line only. Intended for over-the-counter home use; provides rapid visual results to aid pregnancy diagnosis. Device features pen-like housing with moisture-resistant clear shield over result windows.

Clinical Evidence

Clinical study of 100 specimens compared urine stream and dip methods against SAS™ Pregnancy Strip. Results showed 100% sensitivity (54 true positives, 46 true negatives). Bench testing confirmed 25mIU/ml sensitivity and specificity against LH, FSH, and TSH. Interference testing showed no cross-reactivity with common substances (acetaminophen, caffeine, etc.).

Technological Characteristics

Lateral flow immunoassay; utilizes polyclonal goat anti-alpha hCG, monoclonal anti-beta hCG, and polyclonal goat anti-mouse antibody. Pen-like plastic housing with moisture-resistant clear shield. Qualitative visual readout. No electronic components or software.

Indications for Use

Indicated for individuals seeking to detect pregnancy via qualitative hCG measurement in urine. Suitable for home use starting from the first day of a missed period.

Regulatory Classification

Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

Predicate Devices

Reference Devices

Related Devices

Submission Summary (Full Text)

{0} K962888 # 510(k) Summary Submitted by: Athena Medical Corporation 10180 SW Nimbus Ave. Suite J-5 Portland, OR 97223 SEP 23 1996 Contact: Krista Lopardo Phone: 503/968-8800 ext. 19 Fax: 503/639-3674 Product Name: Proprietary Name: Affirm One-Step home pregnancy test Classification Name: Human chorionic gonadotropin test systems Intended Use: Affirm One-Step home pregnancy test is a rapid one-step visual test for the qualitative detection of human chorionic gonadotropin in human urine to aid in the diagnosis of pregnancy. This test is intended for over-the-counter home use. Description of Device: Affirm detects the presence of hCG in urine by using a combination of polyclonal and monoclonal antibodies on the reagent pad and the test strip. The assay begins functioning with the addition of urine onto the absorbent wick. Through capillary action the urine is pulled along the membrane to the reagent pad where the hCG reacts with the conjugate. Positive results are determined by one color line in the test window and one color line in the control window. The absence of hCG produces only one color line in the control window. Predicate device for substantial equivalence comparison: OneStep Be Sure® Plus Midstream Pregnancy Home Test Predicate device used as reference test for clinical data: SAS™ Pregnancy Strip Date Prepared: July 19, 1996 25 {1} # Performance characteristics: ## Specificity: Affirm One-Step home pregnancy test was evaluated for specificity against LH, FSH and TSH. At concentrations of 300mIU/ml LH, and 1000mIU/ml FSH and TSH. Affirm did not cross-react with these substances at these levels. Only when Affirm was compared to these three substances and a 25mIU/ml concentration of hCG simultaneously did the tests respond and show a positive reading for the presence of hCG. ## Sensitivity and Accuracy: The sensitivity and accuracy of the Affirm One-Step home pregnancy test was determined using 20 negative urine samples spiked with hCG at varying levels. The following results are described in the table on the following page. Sensitivity of Affirm One-Step home pregnancy test | Concentration Spiked at: | Results: Positive | Results: Negative | | --- | --- | --- | | 10mIU/ml | 5 | 15 | | 25mIU/ml | 20 | 0 | | 50mIU/ml | 20 | 0 | | 100mIU/ml | 20 | 0 | | 0mIU/ml | 0 | 20 | Out of the 20 samples that were tested at the 10mIU/ml concentration, 5 of them did generate a positive reading after about 15 minutes from the point of saturation. The 20 samples tested at the 25mIU/ml concentration all generated a positive reading well within the 20 minutes, therefore our claim is a 25mIU/ml sensitivity with the Affirm One-Step home pregnancy test. Even though the test is sensitive enough to read concentrations lower than 25mIU/ml this does not occur regularly for us to claim a lower sensitivity for the test. {2} # Interfering Substances: The following potentially interfering substances were added to hCG negative and 25 mIU/ml hCG spiked urine samples. Interfering test samples were compared directly to the control result. The following results are described in the table: Interference of the Affirm One-Step home pregnancy test | Substance: | Concentration: | Interference: | | --- | --- | --- | | Acetaminophen | 20 mg/dI | NONE | | Acetylsalicylic Acid | 20 mg/dI | NONE | | Ascorbic Acid | 20 mg/dI | NONE | | Atropine | 20 mg/dI | NONE | | Caffeine | 20 mg/dI | NONE | | Gentisic Acid | 20 mg/dI | NONE | | Glucose | 2 mg/dI | NONE | | Hemoglobin | 200 mg/dI | NONE | # Predicate Device Result Correlation: One hundred specimens were evaluated by the urine stream method and the dip method. The women participating in the study were solicited by a flyer appearing in clinics around town, and asked to participate at one of the following locations: Athena Medical Corporation, Planned Parenthood in either Southeast Portland or Beaverton. Women who participated in the urine stream method answered a questionnaire about the product. Both the Affirm One-Step home pregnancy test and the SAS™ Pregnancy Strip identified 54 true positives and 46 true negatives generating a 100% sensitivity for the Affirm One-Step home pregnancy test. {3} Substantial Equivalence: Comparison Product: Be Sure* Plus Manufactured by: Syntron Bioresearch Inc. Carlsbad, CA 92008 Similarities to Be Sure* Plus: | Items to be compared | Affirm One-Step home pregnancy test | Be Sure* Plus | | --- | --- | --- | | • Detection method: | hCG in urine specimens reacts with the colored-conjugate on the pad producing a solid color (red) line in the test region. Mouse conjugate binds to goat anti-mouse antibody in the control region on the strip. | hCG in urine specimens reacts with the colored-conjugate on the pad producing a solid color (red) line in the test region. Mouse conjugate binds to goat anti-mouse antibody in the control region on the strip. | | • Usage: | Anytime of the day. | Anytime of the day. | | • Pregnancy determination: | As early as the first day of a women’s missed period. | As early as the first day of a women’s missed period. | | • Antibodies: | Polyclonal goat anti-alpha antibody hCG is on the test strip, monoclonal anti-beta antibody hCG is on the conjugate pad, and polyclonal goat anti-mouse antibody is on the control line. | Polyclonal goat anti-alpha antibody hCG is on the test strip, monoclonal anti-beta antibody hCG is on the conjugate pad, and polyclonal goat anti-mouse antibody is on the control line. | | • Results: | Positive = color line in test window and control window. Negative = color line in control window only. | Positive = color line in test window and control window. Negative = color line in control window only. | | • Format: | Lateral Flow assay | Lateral Flow assay | | • Sensitivity: | Detects level of hCG as low as 25mIU/ml. | Detects level of hCG as low as 25mIU/ml. | | • Device design: | Slender, pen-like shape, removable cap, absorbent tip partially enclosed, two window result view. | Slender, pen-like shape, removable cap, absorbent tip partially enclosed, two window result view. | | • Visual detection: | A solid color line appears in control and test window in the presence of hCG. A solid color line appears in the control window only in the absence of hCG. | A solid color line appears in control and test window in the presence of hCG. A solid color line appears in the control window only in the absence of hCG. | {4} Differences between Affirm and Be Sure* Plus: | Items to be compared | Affirm One-Step home pregnancy test | Be Sure* Plus | | --- | --- | --- | | Device engineering: | Both the result windows and the “tear drops” exposing the absorbent wick appear on the same side of the device. Area covering wick is smaller producing a smaller cap. Result window includes a moisture resistant clear barrier. | Both the result windows and the 2 square windows exposing the absorbent wick are on opposite sides of the device. Area covering the absorbent wick is larger producing a larger cap. | | Developing time: | Less than 4 minutes for the results to appear. | Less than 5 minutes for the results to appear. | | Absorbent wick: | Partially enclosed except for 2 “tear drops” carved out of the plastic housing. | Partially enclosed except for 2 squares cut out of the plastic housing. | | Results windows: | Covered by a clear plastic shield which acts as splash guard to protect the results. | Fully exposed result window with no plastic covering. | | Test window/Control window: | Test window shape is: “+”. control window shape is: “o”. | Test window is a rectangle while the control window is a square. | | Color of holder: | Ivory | Blue/White | The differences between the Affirm One-Step home pregnancy test and Be Sure* Plus are insignificant and do not effect the safety and effectiveness of the device in any significant way.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...