K Number
K250117

Validate with FDA (Live)

Manufacturer
Date Cleared
2025-02-13

(28 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

FaStep Pregnancy Rapid Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. For over-the-counter use.

The Fastep HCG Rapid Test Cassette is a rapid visual immunoassay for the qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. The test kits are for health care professionals use including professionals at physician's office labs (POLs).

Device Description

The FaStep Pregnancy Rapid Test Cassette and the Fastep HCG Rapid Test Cassette are the same device with the first intended for over-the counter use and the second for prescription use. The devices contain individually wrapped pouches containing the device, instructions for use, and pipette droppers.

The devices utilize a combination of antibodies to detect hCG in urine as well as to serve as a run control. The result is displayed to the user in the test window as two lines for a 'Pregnant' positive result and one line for a 'Not Pregnant' negative result.

AI/ML Overview

Here's an analysis of the acceptance criteria and study details for the FaStep Pregnancy Rapid Test Cassette and Fastep HCG Rapid Test Cassette, based on the provided document:

Acceptance Criteria and Device Performance

The document doesn't explicitly state "acceptance criteria" for all performance aspects in a separate table. However, the reported device performance implicitly defines the criteria that were met for substantial equivalence. The key performance indicators for a qualitative pregnancy test are its sensitivity (ability to detect low levels of hCG), specificity (absence of false positives), and agreement with a predicate device/professional results.

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
Analytical SensitivityDetect hCG at or below 20 mIU/mL with high positivity.50% Positivity (Cut-off Sensitivity): 15 mIU/mL (50.7% positive)100% Positivity: 20 mIU/mL (100% positive)Devices consistently detected 20 mIU/mL hCG with 100% positivity across 3 lots and 3 operators.
High Dose Hook EffectNo hook effect observed at high hCG concentrations.No hook effect observed at hCG concentrations up to 2,000,000 mIU/mL.
Analytical Specificity (Interfering Substances)No interference from common substances at specified concentrations.No interference observed from 32 listed substances (e.g., Acetaminophen, Ibuprofen, Albumin, Glucose, Hemoglobin, etc.) at their tested concentrations.
Analytical Specificity (Cross-reactivity)No cross-reactivity with hLH, hFSH, hTSH at high concentrations.No cross-reactivity observed with 500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH.
Analytical Specificity (hCG ß-core fragment)Performance not affected by hCG ß-core fragment.Performance not affected by hCG ß-core fragment concentrations up to 2,000,000 pmol/L.
Analytical Specificity (Urine pH)Performance not affected by urine pH in the physiological range.Urine pH ranges between 4 and 9 do not affect the performance.
Analytical Specificity (Urine Density)Performance not affected by urine density in the physiological range.Urine with a relative density range of 1.000 to 1.035 do not affect the performance.
Method Comparison with Predicate100% concordance with the predicate device.The conformity between the test devices and the predicate device is 100% (59 positive, 53 negative, total 112 samples with 0 discordance).
Lay Person Study AgreementHigh agreement between lay user and professional results.Agreement between lay user results and professional user results was 100% (59 positive, 53 negative, total 112 samples with 0 discordance). The device was also determined to be "easy to use" via questionnaire.
False Positive RateMinimal to no false positive results in non-pregnant women.0 false positive results were observed among 300 non-pregnant females across pre-menopausal, peri-menopausal, and post-menopausal age groups (0+/100- for each group).

2. Sample Sizes and Data Provenance

  • Analytical Performance (Precision/Reproducibility/Sensitivity):
    • Test Set Size: 9 hCG concentrations (0, 10, 12.5, 15, 17.5, 20, 50, 100, 200 mIU/mL) x 10 replicates/day for 5 days x 3 device lots x 3 operators = 1350 data points (for positive/negative determination).
    • Data Provenance: Not explicitly stated, but typically these studies are conducted in a controlled lab environment. No country of origin is mentioned, and it's a prospective lab study.
  • Hook Effect Test:
    • Test Set Size: 7 hCG concentrations (25000 to 2000000 mIU/mL) x 5 replicates x 3 lots x 3 operators = 315 data points.
    • Data Provenance: Prospective lab study.
  • Interfering Substances:
    • Test Set Size: 3 hCG concentrations (0, 10, 20 mIU/mL) x 3 replicates x 3 lots x 3 operators x 32 substances = 864 data points.
    • Data Provenance: Prospective lab study.
  • Cross-reactivity:
    • Test Set Size: 3 hCG concentrations (0, 10, 20 mIU/mL) x 3 replicates x 3 lots x 3 operators x 3 cross-reactants = 81 data points.
    • Data Provenance: Prospective lab study.
  • hCG ß-core fragment:
    • Test Set Size: 4 hCG concentrations (0, 10, 20, 20000 mIU/mL) x 3 replicates x 3 lots x 3 operators x 4 hCGBcf concentrations = 108 data points.
    • Data Provenance: Prospective lab study.
  • Urine pH:
    • Test Set Size: 3 hCG concentrations (0, 10, 20 mIU/mL) x 3 replicates x 3 lots x 3 operators x 6 pH values = 162 data points.
    • Data Provenance: Prospective lab study.
  • Urine Density:
    • Test Set Size: 3 hCG concentrations (0, 10, 20 mIU/mL) x 3 replicates x 3 lots x 3 operators x 8 density values = 216 data points.
    • Data Provenance: Prospective lab study.
  • Method Comparison Study:
    • Test Set Size: 112 urine samples from women.
    • Data Provenance: Samples collected from 112 women aged 18 to 50 at "three clinical sites." Approximately half were early-stage pregnant. Implies prospective collection for this study, likely within the US given the FDA submission.
  • Lay Person Study:
    • Test Set Size: 112 female subjects.
    • Data Provenance: Conducted at "three sites" with females aged 18-50. Implies prospective collection for this study, likely within the US.
  • Specificity Study (False Positives):
    • Test Set Size: 300 urine samples from non-pregnant females (100 per age group: pre-menopausal, peri-menopausal, post-menopausal).
    • Data Provenance: Samples collected from non-pregnant females at "three sites." Implies prospective collection for this study, likely within the US.

3. Number of Experts and Qualifications for Ground Truth

  • Analytical Performance Studies (Sensitivity, Hook Effect, Interfering Substances, etc.): Ground truth is established by the known, spiked concentrations of hCG or other substances. The "experts" performing the tests were "operators" (3 different operators for each study involving replicates across lots), but their specific qualifications are not detailed beyond "operator." These are typically trained lab personnel.
  • Method Comparison Study: The ground truth for comparative purposes was established by the predicate device results, as read by a "separate professional" at each of the three test sites. Their specific qualifications are not detailed beyond "professional."
  • Lay Person Study: The ground truth was established by professional testing of the same urine samples as used by the lay users. The specific qualifications of these "professionals" are not detailed.
  • Specificity Study (False Positives): The ground truth (non-pregnant status) was based on the source of the samples (non-pregnant women across different age groups). The results were read by the device itself; no explicit external expert ground truthing beyond sample collection criteria is mentioned.

4. Adjudication Method for the Test Set

  • Analytical Performance Studies: No formal adjudication method like 2+1 or 3+1 is mentioned. Each replicate/lot/operator result was recorded. Discrepancies (e.g., in the sensitivity study where partial positivity was seen) were simply reported as the percentage of positive/negative readings.
  • Method Comparison Study: Three professional operators read the candidate device, with a different operator at each of the three test sites. A separate professional tested samples using the predicate. It's unclear if there was any adjudication process if different professionals at the same site (if there were multiple for one device) read differently, or if discrepancies between devices were adjudicated. The results table shows perfect agreement, suggesting either no discrepancies or an unstated adjudication process.
  • Lay Person Study: Lay user results were compared to "professional test results." The agreement was 100%, implying robust ground truth or an unstated adjudication process if disagreements occurred.
  • Specificity Study (False Positives): No false positives were observed. Therefore, no adjudication was necessary.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No MRMC comparative effectiveness study was explicitly performed to quantify human reader improvement with AI vs. without AI assistance. The device is a rapid visual immunoassay for qualitative detection of hCG and does not involve AI.
  • The "Method Comparison Study" and "Lay Person Study" compare device performance to a predicate or to professional results, but not with an AI component or a quantifiable "improvement" of human readers with AI.

6. Standalone Performance

  • Yes, standalone performance was extensively done. All analytical performance studies (sensitivity, hook effect, interfering substances, cross-reactivity, pH, density) evaluate the device's inherent performance.
  • The "Lay Person Study" also assesses the standalone performance of the device when used by the intended over-the-counter users.
  • The device is designed for qualitative visual interpretation, so its standalone performance is its primary mode of operation.

7. Type of Ground Truth Used

  • Analytical Performance Studies (Precision/Sensitivity, Hook Effect, Interfering Substances, Cross-reactivity, hCG ß-core fragment, pH, Density): Ground truth was established using spiked samples with known concentrations of hCG or other substances, traceable to WHO International Standard.
  • Method Comparison Study: Ground truth for comparison was the results of the predicate device.
  • Lay Person Study: Ground truth was the results of professional testing of the same urine samples.
  • Specificity Study (False Positives): Ground truth was the known non-pregnant status of the study participants.

8. Sample Size for the Training Set

  • The document describes performance studies, which are analogous to a "test set" for demonstrating effectiveness.
  • The document does not provide details about a "training set" for the device itself. This is a qualitative immunoassay, not a machine learning or AI-driven device that requires a training set in the typical sense. The manufacturing process and quality control would be "trained" or optimized during development, but this is not typically disclosed in a 510(k) summary as a "training set."

9. How the Ground Truth for the Training Set Was Established

  • Since there is no explicit "training set" identified for an AI/ML context, this question is not applicable to the information provided in the document. The device's underlying chemistry and biology are its "training" or design principles.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.

February 13, 2025

Assure Tech LLC % Jenny Xia Director LSI International Inc 504E Diamond Ave., Suite H Gaithersburg, Maryland 20877

Re: K250117

Trade/Device Name: FaStep Pregnancy Rapid Test Cassette: Fastep HCG Rapid Test Cassette Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: Class II Product Code: LCX, JHI Dated: January 9, 2025 Received: January 16, 2025

Dear Jenny Xia:

We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).

Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safetyreporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.

All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory

{2}------------------------------------------------

assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Joseph A. Digitally signed by
Joseph A. Joseph A. Joseph A. Kotarek -S Kotarek -S Date: 2025.02.13
16:11:02 -05'00' Joseph Kotarek, PhD Toxicology Branch Chief Division of Chemistry and Toxicology Devices OHT7: Office of In Vitro Diagnostics Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Indications for Use

510(k) Number (if known) K250117

Device Name FaStep Pregnancy Rapid Test Cassette

Indications for Use (Describe)

FaStep Pregnancy Rapid Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. For over-the-counter use.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

Indications for Use

510(k) Number (if known) K250117

Device Name Fastep HCG Rapid Test Cassette

Indications for Use (Describe)

The Fastep HCG Rapid Test Cassette is a rapid visual immunoassay for the qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. The test kits are for health care professionals use including professionals at physician's office labs (POLs).

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{5}------------------------------------------------

510(k) SUMMARY

K250117

1.Date:January 15, 2025
2.Submitter:Assure Tech LLC1521 Concord Pike, Suite 201Wilmington, Delaware, 19803
3.Contact person:Jenny XiaLSI International Inc.504 East Diamond Ave., Suite HGaithersburg, MD 20877Telephone: 301-525-6856Email: jxa@lsi-consulting.org
4.Device Name:FaStep Pregnancy Rapid Test CassetteFastep HCG Rapid Test Cassette
Classification:Class II
Product Code:LCX; JHI
CFR:862.1155
5.Predicate Devices:Assure Tech hCG Pregnancy Serum/Urine Combo TestCassette, (K152768)

Intended Use 6.

FaStep Pregnancy Rapid Test Cassette is used for qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. For over-the-counter use.

The Fastep HCG Rapid Test Cassette is a rapid visual immunoassay for the qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. This test is intended for use as an aid in the early detection of pregnancy. The test kits are for health care professionals use including professionals at physician's office labs (POLs).

7. Device Description

The FaStep Pregnancy Rapid Test Cassette and the Fastep HCG Rapid Test Cassette are the same device with the first intended for over-the counter use and the second for prescription use. The devices contain individually wrapped pouches containing the device, instructions for use, and pipette droppers.

{6}------------------------------------------------

The devices utilize a combination of antibodies to detect hCG in urine as well as to serve as a run control. The result is displayed to the user in the test window as two lines for a 'Pregnant' positive result and one line for a 'Not Pregnant' negative result.

Similarities
ItemCandidate device:FaStep Pregnancy Rapid TestCassette and Fastep HCGRapid Test Cassette(K250117)Predicate device:Assure Tech hCG PregnancySerum/Urine Combo Test(K152768)
Intended useRapid qualitative detectionof hCG to aid in the earlydetection of pregnancy.Same
SpecimenUrineUrine or serum
Assay technicalImmunochromatographicassaySame
Sensitivity20 mIU/mLSame
ResultsQualitativeSame
Device formatCassetteSame
Differences
ItemDevicePredicate
Target userFor over-the-counter use;For prescription usePrescription use
Read time3-10 minutes5 minutes

Substantial Equivalence Information 8.

9. Test Principle

FaStep Pregnancy Rapid Test Cassette and Fastep HCG Rapid Test Cassette use lateral flow immunoassay for in vitro qualitative detection of Human Chorionic Gonadotropin (HCG) in human urine. Each device contains mouse monoclonal anti-ßhCG antibody colloidal gold conjugate pre-dried on the sample pad. Mouse monoclonal anti-a-hCG antibody (on the Test Line) and goat anti mouse IgG polyclonal antibody (on the Control Line) are coated and immobilized on a nitrocellulose membrane. When the specimen is applied, the gold-antibody conjugate is rehydrated and if there is sufficient HCG in the specimen, the HCG interacts with the gold-conjugated antibodies. The antigen-antibody-gold complex then migrates towards the test window until the Test Zone (T) where it gets captured by immobilized antibodies, forming a visible line (Test line), indicating a positive result. If there is not sufficient HCG in the specimen, no line will be visible in the Test Zone (T), indicating a negative result.

As an internal process control, a colored line will always appear in the Control

{7}------------------------------------------------

Zone (C). Absence of a colored line in the Control Zone(C) even in the presence of a colored line in the Test Zone (T) indicates an invalid test result.

10. Performance Characteristics

A. Analytical nerformance

a. Precision/Reproducibility/Sensitivity

Negative female urine was spiked with hCG standard (Traceable to the 5th WHO) to hCG concentrations of 0, 10, 12.5, 15, 17.5, 20, 50, 100 and 200 mIU/mL. Each sample was tested in 10 replicates per day for 5 days for each device lot. Total of three device lots were tested and each operator tested one lot separately. Tests were performed by three different operators for each sample concentration.

The results are summarized in the table below:

hCGConcentration(mIU/mL)Operator 1Lot 1Operator 2Lot 2Operator 3Lot 3Total result%Negative%Positive
-+-+-+-+
05005005001500100%0%
105005005001500100%0%
12.54554464461331788.7%11.3%
15262423272525747649.3%50.7%
17.56444466441613410.7%89.3%
2005005005001500%100%
5005005005001500%100%
10005005005001500%100%
20005005005001500%100%

The test devices exhibited reproducibility of results.

Based on the above results, the sensitivity of the test devices is demonstrated to be 20 mIU/mL.

b. Linearity/assay reportable range:

Linearity is not applicable since this is a qualitative test. The test devices were evaluated for high dose or hook effect and none were observed.

Hook effect test:

Negative urine samples were spiked with varying hCG concentrations (25000 mIU/mL, 50000 mIU/mL, 100000 mIU/mL, 200000 mIU/mL, 500000 mIU/mL, 1000000 mIU/mL and 2000000 mIU/mL). Samples were tested in replicates of 5 with 3 lots of the test devices by 3 different operators. The results demonstrated that no hook effect was observed at hCG concentrations up to 2,000,000 mIU/mL.

c. Traceability, Stability, Expected values (controls, calibrators, or methods): Traceability:

The test devices are calibrated against reference material traceable to WHO International Standard 5th edition, NIBSC code 07/364.

{8}------------------------------------------------

Stability:

The shelf-life of the test devices at 2~30°C is 24 months based on real time stability data.

d. Analytical specificity

Interfering substance:

To evaluate potential interfering substances of the test devices, urine samples containing 0 mIU/mL, 10 mIU/mL and 20 mIU/mL hCG were spiked with interfering substances to the concentrations listed below. Samples were tested in replicates of 3 with 3 lots of the test devices by 3 different operators. No interference effect was observed at the tested concentrations.

SubstanceTested Concentration
Acetaminophen20 mg/dL
Acetoacetic Acid2000mg/dL
Acetylsalicylic Acid20 mg/dL
Albumin2000 mg/dL
Amoxicillin20 mg/dL
Ascorbic acid20 mg/dL
Atropine20 mg/dL
Ampicillin20 mg/dL
Aspirin80 mg/dL
Benzoylecgonine10 mg/dL
Bilirubin40 mg/dL
Caffeine20 mg/dL
Cannabinol10 mg/dL
Cholesterol250 mg/dL
Codeine6 $\mu$ g/dL
EDTA80 mg/dL
Ephedrine20 mg/dL
Estriol-17-beta1.4 mg/dL
Ethanol1%
Folic acid0.03 mg/dL
Gentensic Acid20 mg/dL
Glucose2000 mg/dL
Gentisic acid20 mg/dL
Hemoglobin2000 mg/dL
Ibuprofen40 mg/dL
Ketone20 mg/dL
Methanol10%
Phenothiazine20 mg/dL
Phenylpropanolamine20 mg/dL
Pregnanediol1.5 mg/dL

{9}------------------------------------------------

Salicylic Acid20 mg/dL
Tetracycline20 mg/dL
Thiophene20 mg/dL
Triglyceride500 mg/dL
β-hydroxybutyrate2000 mg/dL
Uric acid23.5 mg/dL
Vitamin B180 mg/dL
Lipoprotein70 mg/dL

Cross-reactivity :

To evaluate cross-reactivity, negative and positive urine samples (0 mIU/mL, 10 mIU/mL and 20 mIU/mL hCG) were spiked with potential cross reactants (500 mIU/mL hLH, 1000 mIU/mL hFSH, 1000 µIU/mL hTSH). Samples were tested in replicates of 3 with 3 lots of the test devices by 3 different operators. No crossreactivity was observed at tested concentration.

Effects of hCG B-core fragment:

To evaluate the effect of the hCG ß-core fragment, negative urine samples (0 mIU/mL and 10 mIU/mL hCG) and positive urine samples (20 mIU/mL and 20,000 mIU/mL hCG) were spiked with hCG ß-core fragment (hCGBcf) at concentrations of 50.000 pmol/L, 125.000 pmol/L, 250.000 pmol/L and 2.000.000 pmol/L. Samples were tested in replicates of 3 with 3 lots of the test devices by 3 different operators. The performance of the test devices were not affected by hCG B-core fragment concentrations up to 2,000,000 pmol/L.

Effects of urine pH:

To evaluate the effect of urine pH on the results of the test devices, urine samples containing 0 mIU/mL, 10 mIU/mL and 20 mIU/mL hCG were tested at pH values of 4, 5, 6, 7, 8 and 9. Samples were tested in replicates of 3 with 3 lots of the test devices by 3 different operators. The results indicated that urine pH ranges between 4 and 9 do not affect the performance of the test devices.

Effects of urine density:

To evaluate the effect of urine density on the results of the test devices, urine samples containing 0 mIU/mL, 10 mIU/mL and 20 mIU/mL hCG were tested at density values of 1.000, 1.005, 1.011, 1.015, 1.019, 1.024, 1.029 and 1.035. Samples were tested in replicates of 3 with 3 lots of the test devices by 3 different operators. The results indicated that urine with a relative density range of 1.000 to 1.035 do not affect the performance of the test devices.

B. Method_comnarison_study

Method comparison with predicate device:

The performance of the candidate device was compared to the predicate device.

{10}------------------------------------------------

Urine samples were collected from 112 women aged 18 to 50 at three clinical sites. Approximately half of the subjects were pregnant in the early stage of less than 5 weeks. Samples were randomly collected at various time throughout the day. A single lot was used for the study, and 3 professional operators read the results of the candidate device, with a different operator at each of the three test sites. A separate professional tested the samples using 1 lot of the predicate device, with a different operator at each of the three test sites. The samples were blinded and randomized before being tested by professionals. All results are summarized in the table below.

FaStep Pregnancy RapidTest Cassette/ FastepHCG Rapid Test CassettePredicate Test(Assure Tech hCG Pregnancy Serum/Urine Combo TestCassette)Total
Positive (+)Negative (-)
Positive (+)59059
Negative (-)05353
Total5953112

The conformity between the test devices and the predicate device is 100%.

C. Clinical studies

    1. Clinical Sensitivity:
      Not applicable.
    1. Clinical Specificity:
      Not applicable.
    1. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
  • 3.1 Lay person study

A lay user study was performed at three sites with a total of 112 females with diverse educational and occupational backgrounds and ages ranging from 18 to 50 years old. Each subject tested her own urine sample using the FaStep Pregnancy Rapid Test Cassette according to the instructions for use and provided a sample for professional testing. Their results were compared to the results obtained from professionals. Ease of use of the FaStep Pregnancy Rapid Test Cassette was assessed through a questionnaire that was completed at the end of the study and the device was determined to be easy to use. The data demonstrated that the agreement between lay user results and professional user results was 100%. The results are summarized in the table below.

FaStep Pregnancy Rapid TestCassetteProfessional ResultTotal
PositiveNegative
Positive59059

{11}------------------------------------------------

Lay userResultNegativeこつﮐﮯ ﻟﯿﮯ ﮐﮧ ﺍﯾﮏ ﺭﮨﺎﺋﺸﯽ ﮐﮯ ﺍﻭﺭ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﮐﮯ ﺷﮩﺮ ﺍﺱ ﮐﯽ ﻣﺠﻤﻮﻋﯽ ﺁﺑﺎﺩﯼ ﻣﯿﮟ ﻭﺍﻗﻊ ﮨﮯ۔50ﮟ ﻟﯽ ﺍ
Total50こつن ل112

The results above show that the FaStep Pregnancy Rapid Test Cassette can be used by laypersons to obtain correct results.

3.2 Specificity Study to Determine False-Positive Results Rate

300 urine samples were collected from non-pregnant females in pre-menopausal (18-40 years old), peri-menopausal (41-55 years old) and post-menopausal (>55 years old) age groups at three sites. 100 participants for each age group. One device lot was used in the study. No false positive results were observed for all age groups.

GroupTotal result
Pre-menopausal0+/100-
Peri-menopausal0+/100-
Post-menopausal0+/100-

11. Conclusion

Based on the test principle and performance characteristics of the device including precision, interference, specificity, method comparison and lay-user study, it can be concluded that FaStep Pregnancy Rapid Test Cassette/ Fastep HCG Rapid Test Cassette is substantially equivalent to the predicate device.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.