FORA D20 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM; TD-3263 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTE

K091814 · Taidoc Technology Corporation · CGA · Nov 20, 2009 · Clinical Chemistry

Device Facts

Record IDK091814
Device NameFORA D20 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTEM; TD-3263 BLOOD GLUCOSE PLUS BLOOD PRESSURE MONITORING SYSTE
ApplicantTaidoc Technology Corporation
Product CodeCGA · Clinical Chemistry
Decision DateNov 20, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1345
Device ClassClass 2

Indications for Use

The FORA D20/TD-3263 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, or for testing on neonates. The alternative site testing in this system can be used only during steady-state blood glucose conditions. The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4" – 13.8". This meter has some speaking functions but is not intended for use by the visually impaired.

Device Story

FORA D20/TD-3263 is a combination device for blood glucose monitoring and non-invasive blood pressure measurement. Glucose measurement uses capillary whole blood samples; blood pressure measurement uses an inflatable arm cuff (9.4"-13.8" circumference). Device intended for home use by patients with diabetes or by healthcare professionals. Provides quantitative glucose levels, systolic/diastolic blood pressure, and pulse rate. Includes speaking functions for audio output. Assists in monitoring diabetes control programs. Not for diagnostic use or neonates.

Clinical Evidence

No clinical data provided in the document; substantial equivalence is based on technological characteristics and intended use.

Technological Characteristics

Electrochemical biosensor (glucose oxidase) and oscillometric BP monitor. Dimensions: 155x110x69 mm; Weight: 340g. Memory: 450 results. Coding: Manual selection (21-25). Connectivity: Data transmission capability (reviewed under k070941). Standards: ISO 15197:2003, ISO 14971:2000, CLSI EP-9A, CLSI EP5-A.

Indications for Use

Indicated for quantitative glucose measurement in fresh capillary whole blood (finger, palm, forearm, upper-arm, calf, thigh) and non-invasive systolic/diastolic blood pressure and pulse rate measurement in adults. Intended for healthcare professionals and patients with diabetes mellitus. Contraindicated for neonates, diabetes diagnosis/screening, and visually impaired users.

Regulatory Classification

Identification

A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.

Special Controls

*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Related Devices

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE A. 510(k) Number: k091814 B. Purpose for Submission: Modifications to a previously cleared device that include: the addition of speaking functions, change in the type of coding function, an increase in memory, a name change, and a change in size and weight of meter. C. Measurand: Capillary whole blood glucose and blood pressure D. Type of Test: Quantitative, electrochemical biosensor, glucose oxidase Non-invasive Oscillometric E. Applicant: Taidoc Technology Corporation F. Proprietary and Established Names: FORA D20 Blood Glucose Plus Blood Pressure Monitoring System TD-3263 Blood Glucose Plus Blood Pressure Monitoring System G. Regulatory Information: 1. Regulation section: 21 CFR 862.1345; Glucose test system 21 CFR 870.1130; Noninvasive blood pressure measurement system 2. Classification: Class II, Class I (reserved) 3. Product code: {1} Page 2 of 8 NBW, System, Test, Blood Glucose, Over The Counter. CGA, Glucose Oxidase, Glucose DXN, System, Measurement, Blood-Pressure, Non-Invasive 4. Panel: Clinical Chemistry (75) Cardiovascular (74) H. Intended Use: 1. Intended use(s): See Indications for Use below. 2. Indications(s) for use: The FORA D20/TD-3263 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, or for testing on neonates. The alternative site testing in this system can be used only during steady-state blood glucose conditions. The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4" – 13.8". This meter has some speaking functions but is not intended for use by the visually impaired. 3. Special conditions for use statement(s): - For in vitro diagnostic use only - Over-the-Counter and Prescription Use The following limitations apply to the glucose measurement function: - Not intended for use on neonates - Not for the diagnosis of or screening for diabetes mellitus - Not to be used for patients who are dehydrated, hypotensive, in shock, and individuals in hyperglycemic-hyperosmolar state, with or without ketosis. 2 {2} Page 3 of 8 - Critically ill patients should not be tested with a blood glucose meter. - Allows testing on the fingertip, palm, forearm, upper arm, calf, or thigh - Alternate site testing can only be used during steady-state blood glucose conditions - Not intended for use by the visually impaired The following limitations apply to the blood pressure measurement function: - Allows testing only on the arm - Not for use in the presence of common arrhythmia 4. Special instrument requirements: FORA D20/TD-3263 Blood Glucose meter I. Device Description: The FORA D20 Blood Glucose Plus Blood Pressure Monitoring System and the TD-3263 Blood Glucose Plus Blood Pressure Monitoring System are two names for the same meter (same model number). These systems consist of: a glucose meter, lancing device, arm cuff, and Owner's manual. Test strips, lancets, and two levels of control solutions, Normal and High (cleared in k041107) are not included with the meter and are available separately. The changes to the memory function and the addition of voice features affect both the glucose measuring component and the blood pressure measuring component. The modifications to the coding function only affect the blood glucose measuring component. The device has an internal established code, where the user must select the proper code number (code numbers 21-25) from the monitor that matches the one on the test strip vial. The system contains some voice features, but is not intended for use by the visually impaired. J. Substantial Equivalence Information: 1. Predicate device name(s): TaiDoc Technology Corporation; Clever Chek TD-3250 Blood Glucose plus Blood Pressure Monitoring System 2. Predicate 510(k) number(s): k062800 3. Comparison with predicate: {3} Page 4 of 8 Blood glucose measurement function | Similarities | | | | --- | --- | --- | | Item | New Device | Predicate (k062800) | | Indications for use | The FORA D20/TD-3263 Blood Glucose plus Blood Pressure Monitoring System is intended for use in the quantitative measurement of glucose in fresh capillary whole blood from the finger and the following alternative sites: the palm, the forearm, the upper-arm, the calf and the thigh. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, or for testing on neonates. The alternative site testing in this system can be used only during steady-state blood glucose conditions. The system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using a technique in which an inflatable cuff is wrapped around the arm. The cuff circumference is limited to 9.4” – 13.8”. | Same | | Test Principle | Amperometry, measuring a current produced by a chemical for the measurement of blood glucose. | Same | | Sample Type | Fresh capillary whole blood | Same | | Sample Site | Fingertip, the palm, the forearm, the upper-arm, the calf and the thigh | Same | | Sample volume | 0.7 μL | Same | | Measuring time | 7 sec | Same | | Measurement range | 20-600 mg/dL | Same | Blood pressure measurement function | Similarities | | | | --- | --- | --- | | Item | Device | Predicate (k062800) | | Test Principle | Oscillometric method | Same | | Measurement Range | Blood pressure: 0 to 300 mmHg Pulse Rate: 40 to 199 beats/min | Same | {4} Page 5 of 8 | Differences | | | | --- | --- | --- | | Item | Device | Predicate (k062800) | | Memory feature | 450 measurement results with date and time | 352 measurements with day and time | | Speaking functions | Yes | No | | Coding function | Select code number from the meter | Insert code strip into the meter | | Dimensions (mm) | 155x110x69 mm | 137x90x54 mm | | Weight (g) | 340 g without cuff and batteries | 250 g without cuff and batteries | K. Standard/ Guidance Document Referenced (if applicable): - ISO15197:2003 - In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. - ISO 14971:2000 – Medical devices – Application of risk management to medical devices. - CLSI EP-9A: Method Comparison and bias estimation using patient samples; Approved Guideline. - CLSI EP5-A: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline – Second Edition. L. Test Principle: The modifications to the device in this submission are: the addition of speaking functions, change to the type of coding function, an increase in memory, a name change, and a change in size and weight of the meter. Refer to k062800 for a description of the glucose and blood pressure measuring technologies in this device. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Established in the original submission (k062800) b. Linearity/assay reportable range: Established in the original submission (k062800) c. Traceability, Stability, Expected values (controls, calibrators, or methods): 5 {5} Page 6 of 8 The control solutions, Normal and High, with target values of 125 and 300 mg/dL, were originally cleared in k041107. d. Detection limit: Established in the original submission (k062800) e. Analytical specificity: Established in the original submission (k062800) f. Assay cut-off: Established in the original submission (k062800) 2. Comparison studies: a. Method comparison with predicate device: Established in the original submission (k062800) b. Matrix comparison: Established in the original submission (k062800) 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): The sponsor provided a readability study and obtained Flesch-Kincaid Grade Level Scores of 7.9 to 8.0 for the User's Manual and Test Strip insert. 4. Clinical cut-off: {6} Page 7 of 8 Not Applicable 5. Expected values/Reference range: Established in the original submission (k062800) N. Instrument Name: FORA D20 Blood Glucose Plus Blood Pressure Monitoring System or TD-3263 Blood Glucose Plus Blood Pressure Monitoring System O. System Description: 1. Modes of Operation: Each test strip is single use and must be replaced with a new strip for additional readings. Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?: Yes ☐ X (reviewed under k070941) or No ☐ Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?: Yes ☐ or No ☐ X 2. Software: FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types: Yes ☐ X or No ☐ The applicant has provided documentation that indicates the device was designed and developed under good software life-cycle processes. 3. Specimen Identification: There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected, and blood pressure is measured directly. 4. Specimen Sampling and Handling: This device is intended to be used with capillary whole blood from the finger, the palm, the forearm, the upper-arm, the calf and the thigh. Since the whole blood sample is applied directly to the test strip there are no special handling or storage issues. 5. Calibration: The meter uses an internal established code. When the test strip is inserted the user must select the correct code number from the meter (code numbers 21 to 25) that matches the number on the test strip vial. 7 {7} Page 8 of 8 6. Quality Control: Two levels of control solution (previously cleared in k041107), Normal and High, are available for purchase separately, as stated in the labeling. The users are instructed to test the control solutions when using the meter for the first time, when a new vial of test strips is used, when the user suspects the meter or test strips are not working properly, when the blood glucose results are not consistent with how the user feels, when the test strips are exposed to extreme environmental conditions, or if the meter is dropped. When a test strip is inserted into the meter the control mode can be activated by pressing "M". "CTL" will appear on the display. While "CTL" is displayed the results will not be stored in the results memory. An acceptable range for each control level is printed on the test strip vial label. The user is referred to the user manual and customer support for problems and more information. P. Other Supportive Instrument Performance Characteristics Data Not Covered In the "Performance Characteristics" Section above: The new device is a modification of the predicate device. The modifications include the addition of speaking functions, change in the type of coding function, an increase in memory, a name change, and a change in size and weight of meter. The following documentation of these modifications as related to software changes was reviewed and found to be acceptable: level of concern, software description, device hazard analysis, software requirements specifications, software design specification, software development environment description, and verification and validation testing. Q. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. R. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision. 8
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...