K Number
K101926

Validate with FDA (Live)

Date Cleared
2011-12-20

(529 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended to be used by a single person and should not be shared.

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The FORA D40d Test Strips are for use with the FORA D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use auto-disabling lancing devices.

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, and upper arm should be done only during steady – state times (when glucose is not changing rapidly).

The FORA Wisdom D40d Test Strips are for use with the FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

Device Description

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System consists of one monitor, test strip, control solution, and a pressure cuff. The proposed device has been designed, tested, and demonstrated to work together as a system to produce accurate blood glucose test results and blood pressure measurements.

The proposed device contains speaking function which offers step-by-step audio instructions to help users perform a glucose test or take a blood pressure measurement.

AI/ML Overview

The provided text describes the acceptance criteria and performance studies for the FORA D40d and FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaDevice Performance (Study Results)
Blood Glucose AccuracyMet the requirements of ISO 15197
Blood Pressure AccuracyMet the requirements of ANSI/AAMI SP10
Electrical SafetyMet the requirements of IEC/EN 60601-1:1995
Electromagnetic Compatibility (EMC)Met the requirements of IEC/EN 60601-1-2:2001 and EN 61326-1:2006
Safety for Lab EquipmentMet the requirements of IEC/EN 61010-1:2001 and IEC/EN 61010-2-101:2002
Electromagnetic Interference (FCC)Complies with Part 15 of FCC rules
Software FunctionalitySoftware validation was performed to verify and validate the system works functionally.

2. Sample Size Used for the Test Set and Data Provenance

  • Blood Glucose (Consumer Study): "consumer study was performed to evaluate the accuracy at fingertip and alternative sites including palm, forearm and upper arm." The specific sample size for this consumer study is not explicitly stated in the provided text.
  • Blood Pressure (Clinical Study): "For blood pressure, a clinical study was performed compared to the reference method according to the requirements of ANSI/AAMI SP10." The specific sample size for this clinical study is not explicitly stated in the provided text.
  • Data Provenance: The text does not explicitly state the country of origin of the data. It mentions "Taiwan" as the company's address, but not where the studies were conducted. The studies appear to be prospective as they were conducted to evaluate the accuracy of the new device against reference methods and established standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number of experts used or their qualifications for establishing ground truth for either the blood glucose or blood pressure studies.

4. Adjudication Method for the Test Set

The document does not specify any adjudication method used for either the blood glucose or blood pressure test sets.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This section is not applicable. The device described is a blood glucose and blood pressure monitoring system, not an AI-assisted diagnostic tool that would involve human "readers" or a comparative effectiveness study of AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

This section is not applicable. The device is a physical monitoring system that directly measures glucose and blood pressure, not an algorithm that performs standalone analysis. The performance studies evaluate the system's accuracy against reference methods.

7. The Type of Ground Truth Used

  • Blood Glucose: The ground truth for blood glucose accuracy was established by comparing the device's measurements to results from a YSI-2300, which is a laboratory reference instrument for glucose measurement.
  • Blood Pressure: The ground truth for blood pressure accuracy was established using a "reference method according to the requirements of ANSI/AAMI SP10." This standard typically involves comparison against sphygmomanometry measurements performed by trained observers.

8. The Sample Size for the Training Set

The document does not specify a separate training set or its sample size. The studies described are performance validation/testing studies, not machine learning model training.

9. How the Ground Truth for the Training Set was Established

As no training set is mentioned in the context of machine learning, this question is not applicable. The device's underlying technology and calibration would have been developed and validated during its design and manufacturing process, but this document focuses on the final performance validation.

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains a sequence of handwritten digits and letters. The sequence appears to be 'K101926'. The characters are written in a simple, slightly irregular style, typical of handwriting.

DEC 2 0 2011

Attachment B2

510(k) Summary

1. Submitter Information
Company nameTaiDoc Technology Corporation
Contact personLinda Ko
Address3F, 5F, No. 127, Wugong 2nd Rd.,Wugu Dist., New Taipei City,24888, Taiwan
Phone(+886-2) 6625-8188 ext.1158
FAX(+886-2) 6625-0288
E-mailkolinda@taidoc.com.tw
Date PreparedJuly 7th, 2010
2. Name of Device
Trade/Proprietary NameFORA D40d Blood Glucose plus BloodPressure Monitoring System and FORAWisdom D40d Blood Glucose plusBlood Pressure Monitoring System
Common NamesBlood glucose and Blood Pressure testsystem
Product CodeNBW, CGA, DXN
Classification PanelClinical Chemistry (75) ; Cardiovascular
RegulationsClass II
  1. Predicate Device Trade/Proprietary Name:

Common/Usual Name:

Submitter 510 (k) Number

TD-3252 Blood Glucose plus Blood Pressure Monitoring System Blood glucose and blood pressure test system TaiDoc Technology Corporation K091555

21 CFR 862.1345 / 21 CFR 870.1130

{1}------------------------------------------------

4. Device Description

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System consists of one monitor, test strip, control solution, and a pressure cuff. The proposed device has been designed, tested, and demonstrated to work together as a system to produce accurate blood glucose test results and blood pressure measurements.

The proposed device contains speaking function which offers step-by-step audio instructions to help users perform a glucose test or take a blood pressure measurement.

5. Intended Use

FORA D40d for single patient use

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended to be used by a single person and should not be shared.

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The FORA D40d Test Strips are for use with the FORA D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

FORA Wisdom D40d for multiple patients use

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary

{2}------------------------------------------------

whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use auto-disabling lancing devices.

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, and upper arm should be done only during steady - state times (when glucose is not changing rapidly).

The FORA Wisdom D40d Test Strips are for use with the FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

6. Comparison to Predicate Device

For blood glucose, consumer study was performed to evaluate the accuracy at fingertip and alternative sites including palm, forearm and upper arm compared to YSI-2300. For blood pressure, a clinical study was performed compared to the reference method according to the requirements of ANSI/AAMI SP10.

The FORA D40d and FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems are substantially equivalent to the predicate device, TD-3252 Blood Glucose plus Blood Pressure Monitoring System (K091555).

7. Performance Studies

To evaluate the system accuracy of FORA D40d and FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System, the consumer study was performed according to ISO 15197 and other tests performed according to ANSI/AAMI SP10. Results show that the device met the requirements of ISO 15197 and ANSI/AAMI SP10, and is safe during

{3}------------------------------------------------

the use. It is substantially equivalent to the predicate device with the same effectiveness and safety.

The proposed device met the requirements of IEC/EN 60601-1:1995, IEC/EN 60601-1-2:2001, EN 61326-1:2006, IEC/EN 61010-1:2001 and IEC/EN 61010-2-101:2002. It complies with Part 15 of FCC rules. Software validation was also performed to verify and validate the system works functionally.

8. Conclusion

The FORA D40d and FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems demonstrate satisfactory performance and are suitable for its intended use. The FORA D40d and FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring Systems are substantially equivalent to the predicate device.

{4}------------------------------------------------

Public Health Service

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes extending from its back, representing the department's mission to promote health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle symbol. The logo is presented in black and white.

TAIDOC TECHNOLOGY CORPORATION C/O LINDA KO 3F-5F NO. 127 WUGONG 2ª0 RD. WUGU DISTRICT NEW TAIPEI CITY CHINA (TAIWAN) 24888

ンディ

Food & Drug Administration 10903 New Hampshire Avenue Building 66 Silver Spring, MD 20993

DEC 2 0 2019

Re: K101926

Trade Name: FORA D40d Blood Glucose plus Blood Pressure Monitoring System FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System

Regulation Number: 21 CFR §862.1345 Regulation Name: Glucose Test System Regulatory Class: Class II Product Codes: NBW. CGA. DXN Dated: December 2, 2011 Received: December 12, 2011

Dear Ms. Ko:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).

{5}------------------------------------------------

Page 2

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical

Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm

Sincerely yours,

$\mathcal{N}$

Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety

Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Indications for Use

510(k) Number: K101926 Device Name: FORA D40d Blood Glucose plus Blood Pressure Monitoring System

Indications for Use:

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended to be used by a single person and should not be shared.

The FORA D40d Blood Glucose plus Blood Pressure Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. This system should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).

The FORA D40d Test Strips are for use with the FORA D40d Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (21 CFR Part 801 Subpart D)

Over the Counter Use ___X (21 CFR Part 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K101926

Page 1 of 2

{7}------------------------------------------------

Indications for Use .

510(k) Number: k101926_:

510(K) Number. K101726.

Device Name: FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System

Indications for Use:

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm. This system is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control. This system is only used with single-use auto-disabling lancing devices.

The FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System should not be used for the diagnosis of or screening for diabetes, nor for neonatal use. Alternative site testing such as the palm, forearm, and upper arm should be done only during steady – state times (when glucose is not changing rapidly).

The FORA Wisdom D40d Test Strips are for use with the FORA Wisdom D40d Blood Glucose plus Blood Pressure Monitoring System to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, and upper arm.

This system is also intended to be used to measure non-invasively the systolic and diastolic blood pressure and pulse rate of an adult individual at home. The blood pressure is measured by using an inflatable cuff wrapped around the arm.

This system contains speaking function, but is not intended for use by the visually impaired.

Over the Counter Use _________________________________________________________________________________________________________________________________________________________ Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (21 CFR Part 801 Subpart C) (21 CFR Part 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD)

Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) 1 = 10/926

Page 2 of

4

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.