K152986 · Nova Biomedical Corporation · PZI · Jan 27, 2016 · Clinical Chemistry
Device Facts
Record ID
K152986
Device Name
StatStrip Xpress 2 Glucose Hospital Meter System
Applicant
Nova Biomedical Corporation
Product Code
PZI · Clinical Chemistry
Decision Date
Jan 27, 2016
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The device system is intended for multiple-patient use.
Device Story
StatStrip Xpress 2 Glucose Hospital Meter System; point-of-care glucose monitoring device for hospital use. Input: blood sample via test strip. Operation: electrochemical glucose measurement. Output: glucose concentration displayed on 2.2-inch color graphics screen. Modifications from predicate: color display, AAA battery power, ergonomic flat-surface design, relocated test strip port, updated interface software. Operated by healthcare professionals in clinical settings. Benefits: facilitates rapid glucose monitoring for multiple patients; validated for disinfection with Clorox Germicidal Disposable Wipes to prevent cross-contamination.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence is based on identical sensor technology, measurement electronics, and algorithms to the predicate device.
Technological Characteristics
Hand-held glucose meter; 2.2-inch color graphics display; 3-button keypad. Powered by 2 AAA batteries. Electrochemical sensing principle using glucose oxidase (Aspergillus sp.) test strips. Capillary action sample intake. Non-volatile data storage for 400 records. Connectivity: standalone. Sterilization: N/A (device is not sterile).
Indications for Use
Indicated for use as a glucose hospital meter system for multiple-patient use in a clinical setting.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
StatStrip Xpress Glucose Hospital Meter System (K150461)
Related Devices
K161856 — StatStrip Xpress Glucose Hospital Meter System · Nova Biomedical Corporation · Nov 15, 2016
K150461 — StatStrip Xpress Glucose Hospital Meter System · Nova Biomedical Corporation · May 20, 2015
K182549 — StatStrip Xpress 2 Glucose Hospital Meter System · Nova Biomedical Corporation · Dec 13, 2018
K163490 — StatStrip Xpress 2 Glucose Hospital Meter System · Nova Biomedical Corporation · Jan 6, 2017
Submission Summary (Full Text)
{0}
SPECIAL 510(k): Device Modification OIR Decision Memorandum
To: THE FILE
RE: DOCUMENT NUMBER K152986
This 510(k) submission contains information/data on modifications made to the SUBMITTER'S own Class II, Class III or Class I devices requiring 510(k). The following items are present and acceptable (delete/add items as necessary):
1. The name and 510(k) number of the SUBMITTER'S previously cleared device. (For a preamendments device, a statement to this effect has been provided.) K150461, StatStrip Xpress Glucose Hospital Meter System.
2. Submitter's statement that the INDICATION/INTENDED USE of the modified device as described in its labeling HAS NOT CHANGED along with the proposed labeling which includes instructions for use, package labeling, and, if available, advertisements or promotional materials (labeling changes are permitted as long as they do not affect the intended use).
3. A description of the device MODIFICATION(S), including clearly labeled diagrams, engineering drawings, photographs, user's and/or service manuals in sufficient detail to demonstrate that the FUNDAMENTAL SCIENTIFIC TECHNOLOGY of the modified device has not changed.
This change was for:
1) Meter screen display has changed from a segmented display to a 2.2 inches color graphics display.
2) The battery used in the device has changed from a single Li 2450 coin cell battery to two AAA batteries.
3) Modified ergonomic design to a flat, top surface design that eliminates ridges, recessed corners and raised edges of the predicate device.
4) Modified button shape and position.
5) Moved the test strip port from the top of the meter to the bottom of the meter.
6) Minor change in the software to incorporate color into the interface of the StatStrip Xpress 2 Glucose Hospital Meter.
7) Name change of the system name from StatStrip Xpress Glucose Hospital Meter System to StatStrip Xpress 2 Glucose Hospital Meter System.
4. Comparison Information (similarities and differences) to applicant's legally marketed predicate device including, labeling, intended use, physical characteristics. Studies performed include meter robustness studies, human factor / usability study, and the CLIA waiver Precision and Method Comparison studies.
5. A Design Control Activities Summary which includes:
a) Identification of Risk Analysis method(s) used to assess the impact of the modification on the device and its components, and the results of the analysis
b) Based on the Risk Analysis, an identification of the verification and/or validation activities required, including methods or tests used and acceptance criteria to be applied
The labeling for this modified subject device has been reviewed to verify that the indication/intended use for the device is unaffected by the modification. In addition, the submitter's description of the particular modification(s) and the comparative information between the modified and unmodified devices demonstrate that the fundamental scientific technology has not changed. The submitter has provided the design control information as specified in The New 510(k) Paradigm and on this basis, I recommend the device be determined substantially equivalent to the previously cleared (or their preamendment) device.
{1}
Page 2 of 2
The device system is intended for multiple-patient use. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial testing laboratory demonstrating complete inactivation of hepatitis B virus (HBV) with the chosen disinfectant, Clorox Germicidal Disposable Wipes (EPA Registration # 67619-12). Robustness studies were also performed by the sponsor demonstrating that there was no change in performance or external materials of the meter and lancing device after 10950 cleanings and 10950 disinfection steps with the 10905 wipes. The robustness studies were designed to simulate 3 years of multiple-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.