The ATAC Serum Calibrator is intended for use with the ATAC Clinical Systems to establish points of reference that are used in the determination of albumin, calcium, cholesterol, creatinine, direct bilirubin, glucose, iron, magnesium, phosphorus, total bilirubin, total protein, triglycerides, urea and uric acid in human specimens..
Device Story
ATAC Serum Calibrator is a lyophilized human serum-based product containing multiple analytes (albumin, calcium, cholesterol, creatinine, direct bilirubin, glucose, iron, magnesium, phosphorus, total bilirubin, total protein, triglycerides, urea, uric acid). Used with ATAC 8000 Random Access Chemistry Analyzer to establish reference points for clinical chemistry assays. Kit includes 4 vials of calibrator and 4 vials of diluent. Source material tested non-reactive for HIV1/2, HBsAg, and HCV. Healthcare providers use the system to calibrate assays for patient specimen analysis, ensuring accurate quantitative measurement of clinical chemistry markers.
Clinical Evidence
No clinical data. Bench testing only. Stability claims supported by real-time and accelerated stability protocols. Traceability established via NIST SRM reference materials (e.g., SRM 927, 915b, 914, 917b, 937, 929, 916a, 1951b, 912a, 913a).
Technological Characteristics
Serum-based calibrator; liquid or lyophilized form (implied by product type); intended for use with ATAC Clinical Systems; Class II device (21 CFR 862.1150).
Indications for Use
Indicated for use with ATAC Clinical Systems to calibrate assays for albumin, calcium, cholesterol, creatinine, direct bilirubin, glucose, iron, magnesium, phosphorus, total bilirubin, total protein, triglycerides, urea, and uric acid in human specimens.
Regulatory Classification
Identification
A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Related Devices
K030621 — ATAC CALIBRATOR · Elan Diagnostics · Mar 19, 2003
K981706 — AMM CAL, BIL CAL, CO2 CAL, HDL CAL, IRON/MG CAL, LDL CAL, MC CAL, UPRO CAL, AND ISE CALIBRATOR LEVELS 1 AND 2 · Abbott Laboratories · Jul 9, 1998
K024305 — WIENER LAB. CALIBRATOR A PLUS, MODEL 2X3 ML CAT. NR. 1918006 · Wiener Laboratories Saic · Mar 18, 2003
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k110394
B. Purpose for Submission:
New device for adding four new analytes, direct bilirubin, triglycerides, iron, and uric acid to a previously cleared calibrator material (k030621)
C. Measurand:
Analytes: Albumin, Calcium, Cholesterol, Creatinine, Direct Bilirubin, Iron, Glucose, Magnesium, Phosphorus, Total Bilirubin, Total Protein, Triglycerides, Urea, and Uric Acid
D. Type of Test:
Calibrator material
E. Applicant:
Vital Diagnostics
F. Proprietary and Established Names:
ATAC Serum Calibrator (Direct Bilirubin and Iron)
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1150 - Calibrator
2. Classification:
Class II
3. Product code:
JIX - Calibrator, Multi-Analyte Mixture
4. Panel:
Clinical Chemistry - 75
H. Intended Use:
1. Intended use(s):
Refer to indications for use below.
2. Indication(s) for use:
The ATAC Serum Calibrator is intended for use with the ATAC Clinical Systems to establish points of reference that are used in the determination of albumin, calcium, cholesterol, creatinine, direct bilirubin, glucose, iron, magnesium, phosphorus, total bilirubin, total protein, triglycerides, urea and uric acid in human specimens..
{1}
3. Special conditions for use statement(s):
For in vitro diagnostic use only. For prescription use.
4. Special instrument requirements:
ATAC 8000 Random Access Chemistry Analyzer
I. Device Description:
The ATAC Serum Calibrator is a lyophilized human serum with biological materials added as required to obtain desired component levels. The final product contains the following components: albumin, calcium, cholesterol, creatinine, direct bilirubin, glucose, iron, magnesium, phosphorus, total bilirubin, total protein, triglycerides, urea and uric acid in human specimens. The kit consists of 4 vials of the multi calibrators and 4 vials of the ATAC Calibrator Diluents. The volume per vial of the ATAC Calibrator Diluent is 10.0 ml. ATAC Calibrators are manufactured from human serum. The source material in this product have been tested for human immunodeficiency virus (HIV1, HIV2) antibody, hepatitis B surface antigen (HBsAg), and hepatitis C (HCV) antibody and found to be non-reactive.
J. Substantial Equivalence Information:
1. Predicate device name:
k990460
2. Predicate 510(k) number:
Roche Calibrator for Automated Systems (C.f.a.s.)
3. Comparison with predicate:
| Comparison Table | | |
| --- | --- | --- |
| Item | New Device (k110394) | Predicate (k990460) |
| Intended Use | Intended for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. | Same |
| Format | Lyophilized | Same |
| Analytes | Albumin, calcium, cholesterol, creatinine, direct bilirubin, iron, glucose, magnesium, phosphorus, total bilirubin, total protein, urea nitrogen, uric acid | Albumin, calcium, cholesterol, creatinine, direct bilirubin, iron, glucose, magnesium, phosphorus, total bilirubin, total protein, urea (BUN), uric acid, acid phosphatase, alkaline phosphatase, alanine aminotransferase, α-amylase, aspartate aminotransferase, cholinesterase, creatine kinase, γ-glutamyltransferase,, lactase |
{2}
3
| Comparison Table | | |
| --- | --- | --- |
| Item | New Device (k110394) | Predicate (k990460) |
| | | dehydrogenase, lipase, UIBC, lactate, salicylate, HBDH |
| Levels | Single level | Same |
| Stability | Stable 24 months when stored at 2-8°C
Stable 7 days when reconstituted and stored at 2-8°C | Stable until expiration date printed on label when stored at 2-8°C
Stable 2 days when reconstituted and stored at 2-8°C |
K. Standard/Guidance Document Referenced (if applicable):
ISO 17511:2003. In vitro Diagnostic Medical Devices – Measurement of Quantities in Biological Samples
L. Test Principle:
Not applicable.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
Not applicable.
b. Linearity/assay reportable range:
Not applicable.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability: and:
Calibrators are traceable to the reference material shown in the table below:
| Traceability: Reference material | |
| --- | --- |
| Albumin | NIST SRM 927 |
| Calcium | NIST SRM 915b |
| Cholesterol | Cholesterol Reference Method Laboratory Network (CRMLN), NIST SRM 909b/911c |
| Creatinine | NIST SRM 914 |
| Direct Bilirubin | Manual Jendrassik-Grof / Roche Calibrator for Automated Systems |
| Glucose | NIST SRM 917b |
| Iron | NIST SRM 937 |
| Magnesium | NIST SRM 929 |
| Phosphorus | J.T Baker ULTREX, product no. 4921 ** |
{3}
4
| | NIST SRM 3139a now available |
| --- | --- |
| Total Bilirubin | NIST SRM 916a |
| Total Protein | NIST SRM 927 |
| Triglycerides | NIST SRM 1951b / 909b |
| Urea Nitrogen | NIST SRM 912a |
| Uric Acid | NIST SRM 913a, SRM 909b |
**Value Assignment:**
Expected values for the ATAC Serum Calibrator kit are determined using selected commercially available reference materials/standards and at least five vials of candidate calibrators, two analyzers, and two lots of reagents. The target value for each analyte is the mean of the observed 16 replicates. Calibrator assigned values are lot dependent and are listed in the labeling. The mean and range are provided in tabular form in the value assignment sheets.
**Stability:**
Shelf-Life and Open Vial Stability claims were supported by the real time and accelerated stability protocols/testings and acceptance criteria that were described and found to be adequate.
Shelf-Life: ATAC Serum Calibrator Kit is stable for 24 months when stored unopened at $2 - 8^{\circ}\mathrm{C}$.
Open-Vial: The reconstituted vials are stable up to 7 days when stored at $2 - 8^{\circ}\mathrm{C}$.
**d. Detection limit:**
Not applicable.
**e. Analytical specificity:**
Not applicable.
**f. Assay cut-off:**
Not applicable.
**2. Comparison studies:**
**a. Method comparison with predicate device:**
Not applicable.
**b. Matrix comparison:**
Not applicable.
**3. Clinical studies:**
**a. Clinical Sensitivity:**
Not applicable
**b. Clinical specificity:**
Not applicable
**c. Other clinical supportive data (when a. and b. are not applicable):**
Not applicable
{4}
4. Clinical cut-off: Not applicable
5. Expected values/Reference range: The expected values are provided in the labeling, as shown below.
| Albumin | 4.31 g/dL |
| --- | --- |
| Calcium | 10.03 mg/dL |
| Cholesterol | 183.9 mg/dL |
| Creatinine | 4.88 mg/dL |
| Direct Bilirubin | 2.15 mg/dL |
| Glucose | 173.9 mg/dL |
| Iron | 194.9 μg/dL |
| Magnesium | 2.49 mEq/L |
| Phosphorus | 5.85 mg/dL |
| Total Bilirubin | 4.56 mg/dL |
| Total Protein | 6.55 g/dL |
| Triglycerides | 206.8 mg/dL |
| Urea Nitrogen | 34.3 mg/dL |
| Uric Acid | 6.34 mg/dL |
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.