ADVIA Centaur Testosterone II (TSTII), ADVIA Centaur SHBG
K191533 · Siemens Healthcare Diagnostics, Inc. · CDZ · Aug 28, 2019 · Clinical Chemistry
Device Facts
Record ID
K191533
Device Name
ADVIA Centaur Testosterone II (TSTII), ADVIA Centaur SHBG
Applicant
Siemens Healthcare Diagnostics, Inc.
Product Code
CDZ · Clinical Chemistry
Decision Date
Aug 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1680
Device Class
Class 1
Attributes
Pediatric
Indications for Use
The ADVIA Centaur® Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur® XP system. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes. The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur® XP system. The ADVIA Centaur® SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.
Device Story
ADVIA Centaur Testosterone II (TSTII) and SHBG are in vitro diagnostic immunoassays for use on the ADVIA Centaur XP system. TSTII is a competitive chemiluminescent immunoassay; SHBG is a sandwich chemiluminescent immunoassay. Both assays utilize automated processing of human serum or plasma samples. The TSTII assay uses an acridinium ester-labeled hapten and biotinylated anti-testosterone antibody; the SHBG assay uses acridinium ester-labeled anti-SHBG antibody and biotinylated anti-SHBG antibody with streptavidin-coated paramagnetic particles. The system performs quantitative measurements, which are interpreted by clinicians to diagnose and monitor androgen-related disorders. The device provides objective laboratory data to support clinical decision-making regarding hormonal health. Benefits include standardized, automated, and accurate hormone quantification for patient management.
Clinical Evidence
No clinical trials were performed. Evidence is based on analytical bench testing, including LoB/LoD/LoQ determination (CLSI EP17-A2), precision (CLSI EP5-A3), linearity (CLSI EP06-A), and method comparison (CLSI EP9-A3) against ID-LC-MS/MS and predicate devices. Method comparison showed high correlation (r=0.98-1.00). Reference intervals were established using healthy adult and pediatric populations (n≥120 for primary groups) per CLSI EP28-A3c. Biotin interference testing confirmed no significant bias at 30 ng/mL for TSTII and 300 ng/mL for SHBG.
Technological Characteristics
Chemiluminescent immunoassay; competitive (TSTII) or sandwich (SHBG) format. Reagents include acridinium ester-labeled haptens/antibodies and streptavidin-coated magnetic latex particles. Automated processing on ADVIA Centaur XP system. Traceable to ID-LC-MS/MS (TSTII) and WHO 2nd International Standard 08/226 (SHBG). Measuring ranges: 7-1,500 ng/dL (TSTII); 1.6-180 nmol/L (SHBG).
Indications for Use
Indicated for quantitative determination of total testosterone and SHBG in human serum and plasma to aid in the diagnosis and treatment of androgen disorders, including hypogonadism, delayed/precocious puberty, impotence, hirsutism, and virilization in males and females.
Regulatory Classification
Identification
A testosterone test system is a device intended to measure testosterone (a male sex hormone) in serum, plasma, and urine. Measurement of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
K151986 — ADVA Centaur Testosterone II (TSTII), ADVA Centaur Testosterone II (TSTII) Master Curve Material, ADVIA Centaur SHBG, ADVIA Centaur SHBG Calibrator, ADVIA Centaur SHBG Master Curve Material · Siemens Healthcare Diagnostics, Inc. · Jun 29, 2016
K091867 — ADVIA CENTAUR SHBG READYPACK, LOW AND HIGH CALIBRATORS AND MASTER CURVE MATERIAL, MODELS 06520781, 04819274 AND 06520773 · Siemens Healthcare Diagnostics · Jan 14, 2010
K120009 — TESTOSTERONE TEST SYSTEM; CALIBRATOR; AND QUALITY CONTROL MATERIAL (ASSAYED AND UNASSAYED) · Abbott Laboratories · Sep 11, 2012
Submission Summary (Full Text)
{0}
Food and Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993-0002
www.fda.gov
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
ASSAY ONLY
## I Background Information:
A 510(k) Number
K191533
B Applicant
Siemens Healthcare Diagnostics Inc.
C Proprietary and Established Names
ADVIA Centaur® Testosterone (TSTII), ADVIA Centaur® SHBG
D Regulatory Information
| Product Code(s) | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| CDZ | Class I, reserved | 21 CFR 862.1680 - Testosterone Test System | CH - Clinical Chemistry |
## II Submission/Device Overview:
A Purpose for Submission:
Testosterone II (TSTII) - Assay re-standardization to align with Center for Disease Control and Prevention (CDC) HoSt Testosterone Reference Measurement Procedure.
Sex hormone-binding globulin (SHBG) - Updated reference intervals and biotin interference testing.
B Measurand:
Testosterone and Sex Hormone Binding Globulin (SHBG)
K191533 - Page 1 of 14
{1}
# C Type of Test:
Quantitative, chemiluminescent immunoassays
# Intended Use/Indications for Use:
# A Intended Use(s):
See Indications for Use below.
# B Indication(s) for Use:
The ADVIA Centaur® Testosterone II (TSTII) assay is for in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in human serum and plasma using the ADVIA Centaur® XP system. Measurements of testosterone are used in the diagnosis and treatment of disorders involving the male sex hormones (androgens), including primary and secondary hypogonadism, delayed or precocious puberty, impotence in males and, in females, hirsutism (excessive hair) and virilization (masculinization) due to tumors, polycystic ovaries, and adrenogenital syndromes.
The ADVIA Centaur® SHBG assay is an in vitro diagnostic immunoassay for the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur® XP system. The ADVIA Centaur® SHBG assay is intended for use as an aid in the diagnosis of androgen disorders.
# C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only
# D Special Instrument Requirements:
ADVIA Centaur XP System
# IV Device/System Characteristics:
# A Device Description:
The ADVIA Centaur® Testosterone II (TSTII) consist of the following reagents:
| ADVIA Centaur TSTII Primary Reagent ReadyPack (included in assay kit) | | |
| --- | --- | --- |
| Component | Volume | Ingredients |
| ADVIA Centaur TSTII Lite Reagent | 10.0 mL/pack | Acridinium ester-labeled hapten (36 μg/mL) in buffered saline with preservatives |
| ADVIA Centaur TSTII Solid Phase Reagent | 17.0 mL/pack | Streptavidin-coated latex particles (0.33 g/L) in buffered saline with preservatives |
| ADVIA Centaur TSTII Solvent Phase Reagent | 10.0 mL/pack | Streptavidin-coated latex particles (0.33 g/L) in buffered saline with preservatives |
K191533 - Page 2 of 14
{2}
| ADVIA Centaur TSTII Ancillary Reagent ReadyPack (included in assay kit) | | |
| --- | --- | --- |
| Component | Volume | Ingredients |
| ADVIA Centaur TSTII Releasing Agent | 10.0 mL/pack | Steroid releasing agent (0.4 μg/mL) and biotinylated sheep monoclonal anti-testosterone antibody (27 μg/L) in buffered saline and preservatives |
The ADVIA Centaur® SHBG consist of the following reagents:
| ADVIA Centaur SHBG Primary Reagent ReadyPack (included in assay kit) | | |
| --- | --- | --- |
| Component | Volume | Ingredients |
| ADVIA Centaur SHBG Lite Reagent | 3.0 mL/pack | Mouse monoclonal anti-SHBG antibody (~130 ng/mL) labeled with acridinium ester in HEPES-buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, and preservatives |
| ADVIA Centaur SHBG Solid Phase Reagent | 11.0 mL/pack | Streptavidin coupled to latex paramagnetic particles (~150 μg/mL) in HEPES buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, and preservatives |
| ADVIA Centaur SHBG Ancillary Well Reagent | 3.0 mL/pack | Biotinylated monoclonal mouse anti-SHBG antibody (~6 μg/mL) in HEPES buffered saline with bovine serum albumin, mouse serum, sodium azide (< 0.1%), surfactant, and preservatives |
## B Principle of Operation:
ADVIA Centaur® Testosterone (TSTII) assay is a competitive immunoassay using direct chemiluminescent technology. Testosterone in the sample competes with acridinium ester-labeled hapten in the Lite Reagent for binding with the anti-testosterone sheep monoclonal antibody in the ADVIA Centaur TSTII Releasing Agent and streptavidin magnetic latex particles in the Solid Phase Reagent. The ADVIA Centaur TSTII assay uses ADVIA Centaur TSTII Releasing Agent to release bound testosterone from the endogenous binding proteins in the sample.
The ADVIA Centaur® SHBG assay is a sandwich immunoassay that uses three reagent components: Solid Phase, Ancillary Well Reagent, and Lite Reagent. Solid Phase is streptavidin-coupled magnetic latex particles, Ancillary Well Reagent contains biotinylated F(ab)2 monoclonal antibody to SHBG, and Lite Reagent consists of a second F(ab)2 monoclonal antibody to SHBG labeled with acridinium ester (SHBG-AE). The amount of SHBG present in the patient sample is proportional to the amount of relative light units (RLUs) detected by the system. Results are calculated from a master curve adjusted by using high and low calibrators.
K191533 - Page 3 of 14
{3}
V Substantial Equivalence Information:
A Predicate Device Name(s):
ADVIA Centaur Testosterone II (TSTII), ADVIA Centaur SHBG
B Predicate 510(k) Number(s):
K151986
C Comparison with Predicate(s):
TSTII
| Device & Predicate Device(s): | K191533 | K151986 |
| --- | --- | --- |
| Device Trade Name | ADVIA Centaur®
Testosterone II (TSTII) | ADVIA Centaur
Testosterone II (TSTII) |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For in vitro diagnostic use in the quantitative determination of total testosterone (bound and unbound) in serum using the ADVIA Centaur and ADVIA Centaur XP Systems. | Same |
| Methodology | Chemiluminescence | Same |
| Assay Protocol | Competitive immunoassay | Same |
| Traceability/Standardization | ID-LC-MS/MS | Same |
| Specimen Type | Human serum and plasma | Same |
| Sample Volume | 20 μL | Same |
| Measuring Range | 7.0–1500 ng/dL | Same |
| Calibration | 2-point calibration | Same |
| General Device Characteristic Differences | | |
| Reference Interval (ng/dL) | **Adults**
**Males:**
<50 yr: 197.44-669.58
≥50 yr: 187.72-684.19
**Females:**
<50 yr: 8.38-35.01
≥50 yr: <7.00-35.92 | **Adults**
**Males:**
<50 yr: 123.06-813.86
≥50 yr: 86.98-780.10
**Females:**
<50 yr: 9.01-47.94
≥50 yr: <7.00-45.62 |
K191533 - Page 4 of 14
{4}
| Device & Predicate Device(s): | K191533 | K151986 |
| --- | --- | --- |
| | Pediatrics
Males:
Tanner Stage 1: <7.00-13.06
Tanner Stage 2: <7.00-79.13
Tanner Stage 3: <7.00-499.18
Tanner Stage 4: 79.10-747.17
Tanner Stage 5: 224.83-669.65
Age 2-10: <7.00-10.50
Age 11: <7.00-478.50
Age 12: <7.00-487.97
Age 13: 8.28-549.79
Age 14: 8.91-535.34
Age 15: 65.96-756.50
Age 16-21: 228.16-710.74
Females:
Tanner Stage 1: <7.00-10.06
Tanner Stage 2: <7.00-30.11
Tanner Stage 3: <7.00-30.49
Tanner Stage 4: <7.00-35.19
Tanner Stage 5: 11.80-39.30
Age 2-10: <7.00-11.86
Age 11-15: <7.00-27.57
Age 16-21: 11.78-43.34 | Pediatrics
Males:
Tanner Stage 1: <7.00-47.43
Tanner Stage 2: <7.00-174.45
Tanner Stage 3: 10.54-802.75
Tanner Stage 4: 63.89-736.20
Tanner Stage 5: 55.69-897.40
Age 2-10: <7.00-29.44
Age 11: <7.00-353.00
Age 12: <7.00-562.42
Age 13: 8.19-528.87
Age 14: 20.35-777.38
Age 15: 127.21-849.36
Age 16-21: 113.19-882.10
Females:
Tanner Stage 1: <7.00-89.56
Tanner Stage 2: <7.00-38.29
Tanner Stage 3: <7.00-33.86
Tanner Stage 4: <7.00-38.72
Tanner Stage 5: 10.83-50.08
Age 2-10: <7.00-117.76
Age 11-15: <7.00-38.92
Age 16-21: 15.06-42.41 |
SHBG
| Device & Predicate Device(s): | K191533 | K151986 |
| --- | --- | --- |
| Device Trade Name | ADVIA Centaur® SHBG | ADVIA Centaur SHBG |
| General Device Characteristic Similarities | | |
| Intended Use/Indications For Use | For the quantitative determination of sex hormone-binding globulin (SHBG) in human serum and plasma using the ADVIA Centaur XP system. | Same |
| Methodology | Chemiluminescence | Same |
| Assay Protocol | Sandwich immunoassay | Same |
K191533 - Page 5 of 14
{5}
| Device & Predicate Device(s): | K191533 | K151986 |
| --- | --- | --- |
| Traceability/Standardization | WHO 2ndInternational Standard(08/226) | Same |
| Specimen Type | Human serum and plasma | Same |
| Sample Volume | 20 μL | Same |
| Measuring Range | 1.60–180 nmol/L | Same |
| Calibration | 2-point calibration | Same |
| General Device Characteristic Differences | | |
| Reference Interval (nmol/L) | Adults
Males:
<50 yr: 11.54-54.49
≥50 yr: 17.33-71.50
Females:
<50 yr: 17.69-138.26
≥50 yr: 23.65-110.61 | Adults
Males:
<50 yr: 14.55-94.64
≥50 yr: 21.63-113.13
Females:
<50 yr: 10.84->180.00
≥50 yr: 23.15-159.07 |
VI Standards/Guidance Documents Referenced:
- CLSI EP05-A3, 2004 Evaluation of Precision Performance of Quantitative Measurement Procedures; Approved Guideline – Third Edition
- CLSI EP06-A, 2003 Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline – First Edition
- CLSI EP17-A2, 2013; Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline – Second Edition
- CLSI EP28-A3c, 2010; Defining, Establishing and Verifying Reference Intervals in the Clinical Laboratory; Approved Guideline – Third Edition
VII Performance Characteristics (if/when applicable):
A Analytical Performance:
1. Precision/Reproducibility:
TSTII: A precision study was performed in accordance with CLSI EP05-A3 guideline. The panel of 9 samples used in the study included 5 human serum pools (Medical Decision Pools MDP), one female human serum sample pool and three controls. Each sample was tested using one reagent lot over 20 days, with 2 runs per testing day and 2 replicates per run. The results of the precision studies are shown below:
K191533 - Page 6 of 14
{6}
| Sample | N | Mean (ng/dL) | Within-run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| | | | SD | %CV | SD | %CV |
| Control 1 | 80 | 257.38 | 9.06 | 3.5 | 14.38 | 5.6 |
| Control 2 | 80 | 636.57 | 45.14 | 7.1 | 53.12 | 8.3 |
| Control 3 | 80 | 1,021.93 | 62.28 | 6.1 | 81.02 | 7.9 |
| MDP 1 | 80 | 20.92 | 1.3 | 6.2 | 1.86 | 8.9 |
| MDP 2 | 80 | 73.57 | 3.54 | 4.8 | 5.24 | 7.1 |
| MDP 3 | 80 | 312.81 | 13.4 | 4.3 | 23.99 | 7.7 |
| MDP 4 | 80 | 776.64 | 42.59 | 5.5 | 58.26 | 7.5 |
| MDP 5 | 80 | 1,123.83 | 59.22 | 5.3 | 109.57 | 9.7 |
| Female Human Serum | 80 | 15.86 | 1.84 | 11.6 | 2.39 | 15.1 |
SHBG: See previous performance in k091867.
## 2. Linearity:
TSTII: A linearity study was performed in accordance with CLSI EP06-A guideline. A dilution series consisting of 9 levels with concentrations ranging from 6.94 to 1,655.90 ng/dL was prepared by mixing high and low pools of testosterone. For each dilution sample, 3 replicate measurements were made using a single reagent lot, on a single instrument, in a single day. The deviation of the results from the expected linear fit was analyzed. The linear regression formula obtained from the analysis of the results was as follows:
$$
y = 1.004x - 0.015
$$
The results from the linearity study support an analytical measuring range from 7 - 1,500 ng/dL.
SHBG: See previous performance in k151986.
## 3. Analytical Specificity/Interference:
TSTII: Refer to the interference studies reported in k091867.
The sponsor conducted additional testing to evaluate biotin interference on the TSTII assay. Up to 1,500 ng/mL biotin was spiked into two concentrations of TSTII, 83 µg/mL and 289 µg/mL. Triplicate measurements were recorded for each sample (including a negative control without biotin) using a single lot of TSTII assay reagent on the ADVIA Centaur XP Instrument. The sponsor defines no significant interference as <10% change in results, and states that no significant interference was observed at biotin concentrations ≤ 30 ng/mL for both concentrations of TSTII tested.
| % Bias for Samples Containing Various Concentration of Biotin | | | | | | | |
| --- | --- | --- | --- | --- | --- | --- | --- |
| Testosterone Conc. | Biotin Concentration (ng/mL) | | | | | | |
| | 30 | 50 | 100 | 250 | 500 | 1200 | 1500 |
| 83 ng/dL | 4 | 6 | 23 | 84 | 370 | 3981 | 6837 |
| 289 ng/dL | 2 | 7 | 19 | 107 | 441 | >AMR* | >AMR* |
K191533 - Page 7 of 14
{7}
* Interference could not be calculated because the results of the biotin spiked samples were above the claimed analytical measuring range of the assay.
The labeling includes the following limitation: "Specimens that contain biotin at a concentration of 30 ng/mL demonstrate a less than or equal to 10% change in results. Biotin concentrations greater than this may lead to falsely elevated results for patient samples. The recommended adult daily dietary intake for biotin is 30 µg/day. Over the counter dietary supplements promoted for use in hair, skin and nail health may contain 5-10 mg of biotin. Pharmacokinetic studies in healthy adults have shown that ingesting 5 mg of biotin can result in serum levels as high as 73 ng/mL. In rare cases, subjects are prescribed up to 300 mg of biotin per day for therapeutic applications, resulting in serum biotin levels as high as 1,160 ng/mL."
SHBG: Refer to the interference studies reported in k091867.
The sponsor conducted additional testing to evaluate biotin interference on the SHBG assay. Up to 1,200 ng/mL biotin was spiked into two concentrations of SHBG, 19.6 and 44.1 nmol/L. Triplicate measurements were recorded for each sample (including a negative control without biotin) using a single lot of SHBG assay reagent on the ADVIA Centaur XP Instrument. The sponsor defines no significant interference as <10% change in results, and states that no significant interference was observed at biotin concentrations ≤ 600 ng/mL for both concentrations of SHBG tested.
| % Bias for Samples Containing Various Concentration of Biotin | | | | | | |
| --- | --- | --- | --- | --- | --- | --- |
| SHBG Conc. | Biotin Concentration (ng/mL) | | | | | |
| | 38 | 75 | 150 | 3000 | 600 | 1200 |
| 19.6 nmol/L | -3 | -3 | -9 | -7 | -7 | -25 |
| 44.1 nmol/L | -2 | 1 | -3 | -7 | -9 | -26 |
The labeling includes the following limitation: "Specimens that contain biotin at a concentration of 300 ng/mL demonstrate a less than or equal to 10% change in results. Biotin concentrations greater than this may lead to falsely depressed results for patient samples. Over the counter dietary supplements promoted for use in hair, skin and nail health may contain 5-10 mg of biotin. Pharmacokinetic studies in healthy adults have shown that ingesting 10 mg of biotin can result in serum levels as high as 141 ng/mL. In rare cases, subjects are prescribed up to 300 mg of biotin per day for therapeutic applications, resulting in serum biotin levels as high as 1,160 ng/mL."
4. Assay Reportable Range:
TSTII: 7.00 - 1,500.00 ng/dL
SHBG: 1.6 - 180.0 nmol/L
5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):
Traceability:
K191533 - Page 8 of 14
{8}
TSTII: The ADVIA Centaur® Testosterone II (TSTII) assay is standardized using internal standards made from USP-grade testosterone, which are traceable to isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS). The ID-LC-MS/MS is traceable to the primary testosterone standard National Measurement Institute (NMI) M914.
SHBG: The ADVIA Centaur SHBG assay is traceable to the WHO 2nd International Standard for Sex Hormone Binding Globulin (08/226) from National Institute for Biological Standards and Control (NIBSC).
## Standardization:
The ADVIA Centaur® Testosterone II (TSTII) assay was re-standardized to more closely align with the CDC HoSt Testosterone Reference Measurement Procedure (RMP). The ADVIA Centaur® Testosterone II (TSTII) assay is standardized using internal standards manufactured using USP grade testosterone; the assigned values of these standards correlate to the CDC HoSt Testosterone RMP via method comparison. The assigned value of three of the 10 assay gold standards was modified.
The low and high calibrator specifications remain unchanged. Similar to the predicate TSTII assay (K151986), the re-standardized ADVIA Centaur® Testosterone II (TSTII) assay is still standardized using internal standards made from USP-grade testosterone, which are traceable to ID-LC-MS/MS CDC HoST Testosterone RMP. The ADVIA Centaur TSTII Calibrators included in the assay kit for the re-standardized ADVIA Centaur TSTII assay are the same formulation as the calibrators for the predicate ADVIA Centaur TSTII assay.
## 6. Detection Limit:
TSTII: Limit of blank (LoB) study was performed by testing 5 blank samples of human serum pools using 2 reagent lots over 5 days on a single instrument. Each day, 2 runs were conducted. For each sample, 6 replicate measurements were recorded per run (300 replicates total). LoB was calculated non-parametrically at the 95th percentile for each lot. The LoB for each lot was the result at this rank position. If the rank position was a non-integer, the LoB was determined by interpolation between the bracketing values. The largest LoB for both reagent lots was chosen as the assay's LoB.
Limit of detection (LoD) study was performed by testing 10 low level samples (female human serum pools) using 2 reagent lots over 5 days on a single instrument. Each day, 2 runs were performed and for each sample, 6 replicate measurements were recorded per run (1,200 replicates total). LoD was calculated non-parametrically. The higher LoD of the two lots was the assay's LoD.
Limit of quantitation (LoQ) study was performed using 10 female human sera samples with low levels of analytes and 5 blank male and female human serum pools measured over 5 days using 2 reagent lots on 1 instrument per lot. Each day, 2 runs were performed and for each sample, 6 replicate measurements were recorded. For each reagent lot, the within-laboratory precision (expressed as $\% \mathrm{CV}$) was plotted against the mean concentration for each sample. The sponsor defines LoQ as the analyte concentration corresponding to $20\%$ within-lab CV. The largest LoQ for both reagent lots was chosen as the assay's LoQ.
K191533 - Page 9 of 14
{9}
Detection Limits Summary for ADVIA Centaur® Testosterone II (TSTII) assay:
| LoB | LoD | LoQ |
| --- | --- | --- |
| 1.3 ng/dL | 3.57 ng/dL | 4.57 ng/dL |
SHBG: See previous performance in k091867.
7. Assay Cut-Off:
Not applicable.
B Comparison Studies:
1. Method Comparison with Predicate Device:
TSTII: The sponsor conducted a method comparison study to verify that the re-standardization of the assay does not have a significant impact on the performance of the device compared to the previous assay (originally cleared in k151986). The study was designed to compare the performance of the TSTII assay to that of the recognized reference method (CDC HoSt Testosterone Reference Measurement Procedure (RMP) ID-LS-MS/MS). The study evaluated a total of 108 individual male and female adult serum samples spanning the assay range. Of these, 72 samples were individual (non-pooled) human sera from single donors obtained by Siemens, with testosterone concentrations assigned by CDC HoSt Testosterone RMP ID-LS-MS/MS. 30 samples were supplied directly by CDC, and were individual human serum obtained from individual blood bank donors, with testosterone concentrations assigned by CDC HoSt Testosterone RMP ID-LS-MS/MS. Less than 5% of the samples were contrived to cover concentrations at the extreme ends of the measuring range. All samples were run singlicate on one ADVIA Centaur XP system using a single reagent lot over 3 days. The results for ADVIA Centaur® Testosterone (TSTII) versus CDC HoSt Testosterone RMP ID-LS-MS/MS were analyzed using Passing-Bablok regression analysis. The results are summarized below:
| N | 108 |
| --- | --- |
| Slope | 0.97 |
| Intercept | -0.22 |
| R | 0.98 |
| Test Range | 7.27-1,394 ng/dL |
In addition, a method comparison study was performed to evaluate the performance of the re-standardized ADVIA Centaur® Testosterone II (TSTII) assay against the previously cleared assay (k151986). A total of 108 individual male and female adult serum samples spanning the assay range were run on one ADVIA Centaur XP instrument using a single reagent lot. Samples were assayed in singlicate over 3 days. Weighted Deming regression was used for data analysis.
| N | 108 |
| --- | --- |
| Slope | 1.04 |
| Intercept | -4.14 |
| R | 1.00 |
| Test Range | 10.93-1,439.77 ng/dL |
K191533 - Page 10 of 14
{10}
An additional method comparison study was performed to evaluate the performance of the re-standardized ADVIA Centaur® Testosterone II (TSTII) assay against a commercially available testosterone assay, Dimension Vista LOCI Total Testosterone (k151529). A total of 124 serum samples (79 adults and 45 pediatric subjects) spanning the assay range were run on one ADVIA Centaur XP instrument using a single reagent lot and on one Dimension Vista instrument using one Dimension Vista LOCI TSTT reagent lot. Samples were assayed in singlicate over 3 days. Weighted Deming regression was used for data analysis. The results for all specimens are reported below.
| N | 124 |
| --- | --- |
| Slope | 1.01 |
| Intercept | -3.32 |
| R | 0.99 |
| Test Range | 9.00-984 ng/dL |
SHBG: See previous performance in k151986.
2. Matrix Comparison:
TSTII: Following a technical risk assessment of the assay performance attributes that may be impacted by re-standardization of the current TSTII assay, the sponsor determined that assay re-standardization is not expected to impact the results dependent on specimen matrix. The sponsor indicated study control is equally affected by the re-standardization, therefore, there is no effect on the result and the claim is unaffected, this is supported by the data showing similar performance to the predicate device. FDA found this justification to be acceptable. Therefore, the matrix comparison study data from the ADVIA Centaur® Testosterone (TSTII) assay cleared in k151986 remains applicable.
SHBG: See previous matrix comparison study in k151986.
C Clinical Studies:
1. Clinical Sensitivity:
Not applicable.
2. Clinical Specificity:
Not applicable.
3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):
Not applicable.
D Clinical Cut-Off:
Not applicable.
K191533 - Page 11 of 14
{11}
E Expected Values/Reference Range:
TSTII: A reference interval study was performed for the ADVIA Centaur® Testosterone II (TSTII) assay in accordance with CLSI EP28-A3c guideline. A total of 940 pediatrics and 760 adult samples were collected from apparently healthy individuals. The sample groups tested consisted of:
- 250 adult males under 50 years of age
- 135 adult males ages ≥ 50 years old
- 224 adult females under 50 years of age
- 151 adult females ages ≥ 50 years old
- 460 male pediatrics, ages 2-21 across Tanner Stages 1-5
- 483 female pediatrics, ages 2-21 across Tanner Stages 1-5
Testing was performed using the re-standardized assay. Each sample was assayed in singleicate. For adults, the lower and upper reference limits were estimated as the 5th and the 95th percentiles of the distribution of test results using a non-parametric approach in accordance with the recommendation in CLSI EP28-A3c guideline. Due to the limited availability of pediatric samples, the sample analysis method varied depending on the number of samples in each subgroup. Please see the following descriptions of each subgroup reference intervals determination:
- For groups of 120 or more subjects, the lower and upper reference limits were estimated as the 2.5th and the 97.5th percentiles of the distribution of test results using a non-parametric approach in accordance with the recommendation in CLSI guideline EP28-A3.
- For groups of 40 to 119 subjects, the lower and upper reference limits were estimated as the 2.5th and the 97.5th percentiles of the distribution of test results using a robust measure of location and spread, as developed by Horn and Pesce.
- For groups of less than 40 subjects, the lower and upper reference limits were estimated as the 5th and the 95th percentiles of the distribution of test results.
A summary of the median and reference interval values calculated for adult and pediatrics using the re-standardized ADVIA Centaur® Testosterone II (TSTII) assay is shown below:
| | Age | N | ng/dL | | |
| --- | --- | --- | --- | --- | --- |
| | | | Median | Central 90^{th} Ref Interval | |
| | | | | 5^{th} | 95^{th} |
| Adult Male | Under 50 | 250 | 409.72 | 197.44 | 669.58 |
| | 50 and Over | 135 | 377.46 | 187.72 | 684.19 |
| Adult Female | Under 50 | 224 | 18.01 | 8.38 | 35.01 |
| | 50 and Over | 151 | 14.18 | <7.00 | 35.92 |
| Pediatric Male | Tanner Stage 1 | 101 | <7.00 | <7.00 | 13.06 |
| | Tanner Stage 2 | 78 | <7.00 | <7.00 | 79.13 |
| | Tanner Stage 3 | 64 | 59.67 | <7.00 | 499.18 |
| | Tanner Stage 4 | 88 | 376.84 | 79.10 | 747.17 |
K191533 - Page 12 of 14
{12}
| | Age | N | ng/dL | | |
| --- | --- | --- | --- | --- | --- |
| | | | Median | Central 90thRef Interval | |
| | | | | 5th | 95th |
| | Tanner Stage 5 | 129 | 451.17 | 224.83 | 669.65 |
| | Age 2-10 | 147 | <7.00 | <7.00 | 10.50 |
| | Age 11 | 34 | 10.73 | <7.00 | 478.50 |
| | Age 12 | 35 | 132.47 | <7.00 | 487.97 |
| | Age 13 | 34 | 199.02 | 8.28 | 549.79 |
| | Age 14 | 34 | 228.39 | 8.91 | 535.34 |
| | Age 15 | 27 | 327.89 | 65.96 | 756.50 |
| | Age 16-21 | 149 | 453.86 | 228.16 | 710.74 |
| Pediatric Female | Tanner Stage 1 | 138 | <7.00 | <7.00 | 10.06 |
| | Tanner Stage 2 | 60 | 8.17 | <7.00 | 30.11 |
| | Tanner Stage 3 | 49 | 12.98 | <7.00 | 30.49 |
| | Tanner Stage 4 | 98 | 17.37 | <7.00 | 35.19 |
| | Tanner Stage 5 | 133 | 19.76 | 11.80 | 39.30 |
| | Age 2-10 | 159 | <7.00 | <7.00 | 11.86 |
| | Age 11-15 | 174 | 12.95 | <7.00 | 27.57 |
| | Age 16-21 | 145 | 19.81 | 11.78 | 43.34 |
SHBG: A reference interval study was performed for the ADVIA Centaur® SHBG assay in accordance with CLSI EP28-A3c guideline. A total of 760 adult samples were collected from apparently healthy individuals. The sample groups tested consisted of:
250 adult males under 50 years of age
135 adult males ages $\geq 50$ years old
224 adult females under 50 years of age
151 adult females ages $\geq 50$ years old
Each sample was assayed in singlicate. The lower and upper reference limits were estimated as the 5th and the 95th percentiles of the distribution of test results using a non-parametric approach in accordance with the recommendation in CLSI guideline EP28-A3c.
A summary of the median and reference interval values calculated using the re-standardized ADVIA Centaur® SHBG assay is shown below:
| | Age | N | nmol/L | | |
| --- | --- | --- | --- | --- | --- |
| | | | Median | Central 90thRef Interval | |
| | | | | 5th | 95th |
| Adult | < 50 yrs. | 250 | 23.80 | 11.54 | 54.49 |
| Male | ≥ 50 yrs. | 135 | 36.91 | 17.33 | 71.50 |
| Adult | < 50 yrs. | 224 | 46.72 | 17.69 | 138.26 |
| Female | ≥ 50 yrs. | 151 | 47.86 | 23.65 | 110.61 |
K191533 - Page 13 of 14
{13}
VIII Proposed Labeling:
The labeling supports the finding of substantial equivalence for this device.
IX Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
K191533 - Page 14 of 14
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.