The UltraEzAir® is a topical anesthesia applicator used to apply topical anesthetic to a patient's oropharynx and upper airway region through the working channel of a flexible nasal laryngoscope using air flow. The device is designed for and intended to be used by physicians trained and experienced in flexible endoscopic techniques for elective outpatient procedures. The safety and effectiveness of this device for use in the performance of topical anesthesia of the lower airways (below the level of the trachea) have not been established.
Device Story
Electrically driven topical anesthesia applicator; delivers nebulized 4% lidocaine solution to oropharynx/upper airway. System comprises reusable base, single-use non-sterile mist assembly, and single-use delivery line. Operates via piezoelectric nebulizer (1.66 MHz) and internal air pump; pushes lidocaine mist through working channel of flexible nasal laryngoscope. Physician visualizes target tissue via laryngoscope during application. Features three delivery rates, safety timer (9-minute limit), and temperature sensor/indicator. Used by trained physicians in outpatient settings. Benefits include precise, visualized topical anesthesia delivery for procedures like transnasal esophagoscopy, FEES/ST, vocal fold injections, biopsies, and stroboscopy.
Clinical Evidence
No clinical or animal data provided. Substantial equivalence supported by non-clinical bench testing, including functional verification, biocompatibility (ISO 10993-1), gas pathway emissions (ISO 18562), human factors engineering (IEC 62366), and electrical safety (IEC 60601-1/1-2).
Technological Characteristics
Electrically powered topical anesthesia applicator. Piezoelectric transducer (1.66 MHz ± 50 kHz) generates lidocaine mist. Components: reusable base, single-use mist assembly, single-use delivery line. Connectivity: standalone. Biocompatibility per ISO 10993-1; gas pathway testing per ISO 18562. Safety features: 9-minute timer, temperature sensor. Dimensions/form factor: base unit with external power source.
Indications for Use
Indicated for adult patients (≥80 lbs) requiring topical anesthesia of the oropharynx and upper airway prior to endoscopic examination in outpatient clinical facilities or offices.
Regulatory Classification
Identification
A laryngotracheal topical anesthesia applicator is a device used to apply topical anesthetics to a patient's laryngotracheal area.
K093584 — EZ-MISTER · N.M. Beale Co., Inc. · Mar 4, 2011
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 16, 2024
% Eric Bannon Vice President of Regulatory and Clinical Affairs AlvaMed, Inc. 1116 Great Plain Avenue, Suite 1 Needham, Massachusetts 02495
Re: K240114
Trade/Device Name: UltraEzAir® (UEA1A) Regulation Number: 21 CFR 868.5170 Regulation Name: Laryngotracheal Topical Anesthesia Applicator Regulatory Class: Class II Product Code: CCT Dated: September 13, 2024 Received: September 13, 2024
Dear Eric Bannon:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rue"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-device-advicecomprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic.
{2}------------------------------------------------
See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Bradley Q. Quinn -S
Bradley Quinn Assistant Director DHT1C: Division of Anesthesia, Respiratory, and Sleep Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT, and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K240114
Device Name
UltraEzAir® (UEA1A)
Indications for Use (Describe)
The UltraEzAir® is a topical anesthesia apply topical anesthetic to a patient's oropharynx and upper airway region through the working channel of a flexible nasal laryngoscope using air flow. The device is designed for and intended to be used by physicians trained and experienced in flexible endoscopic techniques for elective outpatient procedures.
The safety and effectiveness of this device for use in the performance of topical anesthesia of the lower airways (below the level of the trachea) have not been established.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
#### 1.1 Name and Address of Submitter
DUALAMS, Inc., dba AIRKOR® 7139 Azalea Lane Dallas, Texas 75081 John Reid President Phone: +1 (972) 327 3345 johnfreid@airkor.com
#### 1.2 Correspondent/Primary Contact Person
Eric Bannon Vice President of Regulatory and Clinical Affairs AlvaMed, Inc. ebannon@alvamed.com Phone: +1 (781) 710-8243 Fax: +1 (617) 249-0955
#### 1.3 Submission Information
#### Date Summary Prepared:
October 16, 2024
| Subject Device: | Trade/Device Name: | UltraEzAir® |
|-------------------|--------------------|------------------------------------------------------------|
| | Manufacturer: | Airkor® |
| | Common Name: | Laryngotracheal topical anesthesia applicator |
| | Regulation Number: | 21 CFR 868.5170 |
| | Regulation Name: | Laryngotracheal topical anesthesia applicator |
| | Regulation Class | Class II |
| | Product Code: | CCT - Applicator (Laryngo-Tracheal), Topical<br>Anesthesia |
| | Review Panel: | Anesthesiology |
| Predicate Device: | Clearance: | K153470 dated August 19, 2016 |
| | Trade/Device Name: | MADgic Laryngo-Tracheal Mucosal Atomization<br>Device |
| | Manufacturer: | TELEFLEX MEDICAL |
| | Regulation Number: | 21 CFR 868.5170 |
{5}------------------------------------------------
| | Regulation Name: | Laryngotracheal topical anesthesia applicator | |
|--|------------------------------|------------------------------------------------------------|--------------------------------|
| | Regulation Class | Class II | |
| | Product Code: | CCT - Applicator (Laryngo-Tracheal), Topical<br>Anesthesia | |
| | Reference Device: Clearance: | | K201167 dated December 3, 2020 |
| | Trade/Device Name: | AireHealthTM Nebulizer | |
| | Manufacturer: | AireHealth Inc. | |
| | Regulation Number: | 21 CFR 868.5630 | |
| | Regulation Name: | Nebulizer | |
| | Regulation Class | Class II | |
| | Product Code: | CAF Nebulizer (Direct Patient Interface) | |
#### Valid Predicate Discussion
MADgic Laryngo-Tracheal Mucosal Atomization Device was selected as the valid predicate device used to support the 510(k) submission because it was cleared using well-established methods. The predicate device meets the expected predicate performance due to its duration of use. After conducting a search on the Manufacturer and User Facility Device Experience (MAUDE) Database, Medical Device Reporting (MDR), MedSun Reports Database, Medical Device Safety & Availability (Biologics) and Recall Database websites it was found that there are no known unmitigated use-related or design safety issues and the predicate device has not been subject to a design-related recall.
| Valid Predicate<br>Device | A - Well<br>established<br>methods | B - Meets or<br>exceeds expected<br>predicate<br>performance | C - Unmitigated<br>use-related or<br>design-related<br>safety issues | D - Associated<br>design-related<br>recall |
|--------------------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------------------------------------------------------|--------------------------------------------|
| MADgic Laryngo-<br>Tracheal Mucosal<br>Atomization<br>Device | Used relevant<br>methods that were<br>published in the<br>public domain | History of safe use,<br>established due to<br>duration of device<br>on the market | No known<br>unmitigated use-<br>related or design<br>related safety<br>issues | No design-<br>related recall<br>identified |
#### 1.4 Device Description
The UltraEzAir® is a prescription medical device intended for application of nebulized 4% topical lidocaine solution to the oropharynx and upper airway region of patients. The subject device is an electrically driven device. The subject device is intended to be used only in an outpatient clinical facility or office for anesthetic procedures by a healthcare professional who is trained by Airkor® on its use.
{6}------------------------------------------------
The UltraEzAir® is intended to be used on adult patients, weighing at least 80 lbs. requiring topical anesthesia of oropharynx and upper airway region prior to endoscopic examination in an outpatient clinical facility or office.
The UltraEzAir® consists of a reusable Base, a single-use non-sterile Mist Assembly placed on the Base, and a single-use nonsterile Delivery Line connected to the Mist Assembly. The Delivery Line is then connected to the working channel of an Olympus or Pentax flexible nasal laryngoscope and the lidocaine mist is applied using the distal end of the attached laryngoscope to adjacent target tissue. The application process is visualized using the attached laryngoscope.
## 1.5 Indications for Use
The UltraEzAir® is a topical anesthesia applicator used to apply topical anesthetic to a patient's oropharynx and upper airway region through the working channel of a flexible nasal laryngoscope using air flow. The device is designed for and intended to be used by physicians trained and experienced in flexible endoscopic techniques for elective outpatient procedures.
The safety and effectiveness of this device for use in the performance of topical anesthesia of the lower airways (below the level of the trachea) have not been established.
#### 1.6 Summary of Technological Characteristics
The principle of operation is the same as the primary predicate device: the subject device is a topical anesthetic applicator. Both devices use tubing for the anesthetic solution to the same mucosal membrane target region. For both the subject and the predicate devices, anesthetic solution is sprayed on the intended patient anatomy. The subject device mist of lidocaine solution is generated in the Mist Assembly, using piezoelectric nebulizer technology. The built-in air pump of the UltraEzAir® pushes the lidocaine mist out of the Mist Assembly, to be applied to the target. Users have three lidocaine mist delivery rates to choose from. The UltraEzAir® is equipped with a safety timer, which stops the generation of mist at 9 minutes of continuous use, and a temperature sensor and indicator, which also stops operation to prevent thermal damage to the internal components.
The following table provides an overview of general technological characteristics in comparison to the predicate and reference devices. Any differences in the technological characteristics are minor and reflect market strategy and/or perceived user preferences and do not impact the safety, effectiveness, or substantial equivalence of the device. Biocompatibility evaluation testing, reprocessing method evaluation, and performance testing have been performed on the subject device in order to establish substantial equivalence to the predicate devices.
{7}------------------------------------------------
# Substantial Equivalence Comparison Table
| Device | Subject Device | Predicate Device | Reference Device |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | UltraEzAir | MADgic™ | AireHealth™ Nebulizer |
| K Number | K240114 | K153470 | K201167 |
| Regulatory | | | |
| Indication for Use | The UltraEzAir® is a<br>topical anesthesia<br>applicator used to<br>apply topical anesthetic<br>to a patient's<br>oropharynx and upper<br>airway region through<br>the working channel of<br>a flexible nasal<br>laryngoscope using air<br>flow. The device is<br>designed for and<br>intended to be used by<br>physicians trained and<br>experienced in flexible<br>endoscopic techniques<br>for elective outpatient<br>procedures such as:<br>- Transnasal<br>esophagoscopy<br>- Fiberoptic<br>endoscopic evaluation<br>of swallowing with<br>sensory testing<br>(FEES/ST)<br>- Vocal fold injections<br>- Laryngeal biopsies<br>- Laryngeal mass<br>excisions<br>- Laser excisions<br>- Stroboscopy Dynamic<br>voice assessment<br>- Dynamic voice<br>assessment<br>- Peritonsillar abscess<br>drainage | The MADgic™<br>LaryngoTracheal<br>Mucosal Atomization<br>Device is intended for<br>the application of<br>topical anesthetics to<br>the oropharynx and<br>upper airway region. | The AireHealth™<br>Nebulizer electronic<br>vibrating mesh<br>nebulizer is designed to<br>nebulize liquid<br>medications for<br>inhalation by a patient.<br>The AireHealth™<br>Nebulizer may be used<br>in adults or children 5<br>years of age and older.<br>The AireHealth™<br>Nebulizer is a portable<br>Nebulizer for use in and<br>out of the home<br>environment.<br><br>The AireHealth™<br>Nebulizer is not<br>intended as a life<br>sustaining or life<br>supporting device. The<br>AireHealth™ Nebulizer<br>is not intended for use<br>with Pentamidine. |
| Device | Subject Device | Predicate Device | Reference Device |
| | The safety and<br>effectiveness of this<br>device for use in the<br>performance of topical<br>anesthesia of the lower<br>airways (below the<br>level of the trachea)<br>have not been<br>established. | | |
| Regulation Number | 21 CFR 868.5170 | 21 CFR 868.5170 | 21 CFR 868.5630 |
| Regulation Name | Laryngotracheal Topical<br>Anesthesia Applicator | Laryngotracheal Topical<br>Anesthesia Applicator | Nebulizer |
| Device Class | Class II | Class II | Class II |
| Product Code | CCT | CCT | CAF |
| Product Classification | Applicator (Laryngo-<br>Tracheal), Topical<br>Anesthesia | Applicator (Laryngo-<br>Tracheal), Topical<br>Anesthesia | Nebulizer (Direct<br>Patient Interface) |
| Prescription Use | Yes | Yes | Yes |
| Intended Use | | | |
| Method of Operation | Application of topical<br>anesthetic to a<br>patient's oropharynx<br>and upper airway<br>region through the<br>working channel of the<br>flexible nasal<br>laryngoscope using air<br>flow | Application of topical<br>anesthetics to the<br>oropharynx and upper<br>airway region | Nebulization of liquid<br>medications for<br>inhalation by a patient<br>though the nasal or<br>oral cavity |
| Environment of Use | Professional healthcare<br>setting | Professional healthcare<br>setting | In Home or Out of<br>Home Use |
| Patient Population | Adult patients requiring<br>topical anesthesia of<br>oropharynx and upper<br>airway region prior to<br>endoscopic<br>examination | Patients requiring<br>topical anesthesia<br>before intubation | Children 5 years of age<br>or older and adults<br>requiring nebulized<br>liquid medication for<br>inhalation |
| Intended Medications | 4% topical lidocaine<br>solution | Topical anesthetic<br>solution (including 4%<br>topical lidocaine) | Liquid medication for<br>inhalation |
| Dosage Amount | User Controlled –<br>Maximum of 0.4 ml (16<br>mg) of 4% topical<br>lidocaine solution at<br>High Mist Density<br>setting during typical 5-<br>minute delivery time | User Controlled –<br>2.8 ml lidocaine from a<br>3ml Syringe | Unknown |
| Device | Subject Device | Predicate Device | Reference Device |
| Device Features | | | |
| Typical Particle Size | 70% of particles are<br>greater than 11.72<br>microns and average<br>particle size is 12<br>microns. Particle size<br>range is 0.54 to > 14.9<br>microns.- | 30-100 microns | 2.0-6.0 microns |
| Piezoelectric<br>Transducer Operating<br>Frequency | 1.66 MHz ± 50 kHz | N/A | 115kHz |
| Method of Operation | Previous Submission<br>Delivery form is a fine<br>mist using piezoelectric<br>technology and air flow<br>through an endoscope.<br>Device does not come<br>into direct contact with<br>patient. | Delivery form is a fine<br>spray mist generated<br>by piston syringe.<br>Device comes into<br>direct contact with<br>patient | Inhalation by a patient.<br>Nebulizes liquid<br>medications using<br>vibrating mesh<br>nebulizer incorporated<br>with a piezoelectric<br>transducer. |
| Visualization of Target<br>Anatomy | Flexible Nasal<br>Laryngoscope | Laryngoscope | N/A |
| Rate of Anesthetic<br>Delivery | Variable between 0.062<br>and 0.077 mL/min | Unknown | Minimum nebulization<br>rate of 0.25 mL/min |
| Flow Rate | 1.5 L/min | Unknown | Unknown |
| Hardware | | | |
| Sterilization | Non-sterile; Single use | Non-sterile; Single use | N/A |
| Shelf Life of Tubing | 2 years from date of<br>manufacture | Three (3) years from<br>date of manufacture | N/A |
| Use Life of Unit | 3 years or 2,500 5-<br>minute uses whichever<br>comes first | N/A | 3 Years |
| Energy Source | External | Manual | Internal |
| Operating Temperature | 15 to 30 °C | Unknown | 5 to 38 °C |
| Operating Humidity | 30 to 75% Relative<br>Humidity | Unknown | 15 to 90% Relative<br>Humidity |
| Biocompatibility | Per ISO 10993-1 | Per ISO 10993-1 | Per ISO 10993-1 |
{8}------------------------------------------------
{9}------------------------------------------------
## 1.7 Performance Testing Summary
Summary of Biocompatibility Testing:
The UltraEZAir is considered to be an externally communicating device with tissue contact for a limited duration (< 24 hrs) with indirect gas pathway contact. The biocompatibility evaluation for the UltraEZAir device was conducted in accordance with the guidance document "Use of International Standard ISO 10993-1, 'Biological Evaluation of Medical Devices Part 1: Evaluation
{10}------------------------------------------------
and Testing Within a Risk Management Process," The testing included the following:
- Cytotoxicity
- Sensitization
- Irritation
- Acute Systemic Toxicity
- Material Mediated Pyrogenicity
Per ISO 18562-1 "Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 1: Evaluation and testing within a risk management process", Gas Pathway Testing using the methods in ISO 18562-2 "Biocompatibility evaluation of breathing gas pathways in healthcare applications – Part 2: Tests for emissions of particulate matter" and ISO 18562-3 "Biocompatibility evaluation of breathing gas pathways in healthcare applications - Part 3: Tests for emissions of volatile organic compounds (VOCs)" was completed as part of the biocompatibility testing.
## Summary of Non-Clinical Functional and Performance Testing
Reprocessing instructions for the reusable Base were evaluated in the Human Factors testing.
Non-clinical bench performance testing was performed on the subject device including verification/validation testing to functional specifications, which demonstrated that the device is as safe and effective as the predicate device.
The following Functional and Performance Bench testing has been completed:
Testing to IEC 60601-1, IEC 60601-1-2 Testing to AIM 7351712 for RFID Interference Device Characterization for: Flow Rate, Delivery Rate, Particle Size, Back Pressure, Pressure Over Time, Air Flow Rate Over Time, and Filter Efficiency Human Factors Engineering per IEC 62366 Small Bore Connections Evaluation System Verification: Inspection and Non-Functional System Labeling Inspection Packaging/Shipping Validation
Together, these verification/validation activities successfully demonstrated that the device correctly performs as designed, has been validated for its intended use, and raises no new
{11}------------------------------------------------
questions regarding either safety or effectiveness when compared to the predicate device. Therefore, the verification/validation testing conducted supports a determination of substantial equivalence for the subject device.
## Summary of Animal and Clinical Data
Animal and Clinical testing were not required for a determination of substantial equivalence of the UltraEzAir®.
## 1.8 Shelf Life
The useful life of the Base was determined to be 3 years after purchase or 2,500 5-minute uses, whichever comes first.
The shelf life of the non-sterile Mist Assembly and Delivery Lines were evaluated by AirKor® and it has been concluded that the Mist Assembly and Delivery Lines are not susceptible to degradation that would lead to functional failure. However, the shelf life of the Mist Assembly and Delivery Lines were determined to be two years from date of manufacture.
{12}------------------------------------------------
## 1.9 Conclusion
AirKor® considers the UltraEzAir® to be substantially equivalent to the The MADgic Laryngo-Tracheal Mucosal Atomization Device. The information presented in the 510(k) supports that the UltraEzAir® is as safe, as effective, and performs as well as the predicate devices and is substantially equivalent to the identified predicate and reference devices in design rationale, methodology of use, and performance.
Panel 1
/
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.