K Number
K153470

Validate with FDA (Live)

Manufacturer
Date Cleared
2016-08-19

(261 days)

Product Code
Regulation Number
868.5170
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is intended for the application of topical anesthetics to the oropharynx and upper airway region.

Device Description

The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is a disposable nonsterile device that converts a solution of topical anesthetic into a fine particle spray for application to mucosal surfaces. The device consists of an atomizer tip, a semirigid tubular extension, a standard luer lock adapter and a syringe. The clinician draws up the desired volume of topical anesthetic into the syringe, attaches it to the luer lock fitting of the atomizer, and manipulates the tubing extensions into the desired position. As the syringe plunger is compressed, the anesthetic is forced into the tubular extension and out of the atomizer tip. The tip takes the pressurized fluid column and begins spinning it, allowing the fluid to exit the small hole at the end in a cone shaped spray. The anesthetic mist is gently distributed onto the mucosal surface in front of this tip. Topical anesthesia application can begin in the mouth and pharynx, and then proceed to the hypopharynx, epiglottis and vocal cords, larynx and trachea via direct visualization using a laryngoscope.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the MADgic™ Laryngo-Tracheal Mucosal Atomization Device:

Device: MADgic™ Laryngo-Tracheal Mucosal Atomization Device
FDA 510(k) Number: K153470
Predicate Device: MADgic™ Laryngo-Tracheal Mucosal Atomization Device (K002255)

1. Table of Acceptance Criteria and Reported Device Performance

TestAcceptance CriteriaReported Device Performance
Biocompatibility Testing
Cytotoxicity - L929 MEM Elution AssayGrade of 0, 1, or 2 (not more than 50% of cells are round, devoid of intracytoplasmic granules, and no extensive cell lysis)Acceptable
Sensitization - Kligman Maximization AssayDifference between test article mean score and vehicle control mean score is 1.0 or less (non-irritant).Acceptable
Irritation - Intracutaneous Injection AssayGrade of 1, 0 or less using the Kligman scoring system.Acceptable
Performance Data (Non-Clinical)
Visual InspectionProper assembly, presence of all components, no cracks in tip and luer, no short shots in molded components, solvent presence around bonding surface covering minimum one third total area, no gaps between insert post and tubing.Pass
Flow TestFlow rate ≥ 225 sccm and ≤ 700 sccm.Pass
Leak TestLeak value ≤ 0.05 psi.Pass
Hydrostatic TestDevice must remain assembled after being subjected to a minimum 300 psi hydrostatic pressure.Pass
Atomization Test Post HydrostaticNo streaming, no excessive dripping, or occlusions upon atomization.Pass
TensileTensile strength ≥ 6.0 lb.Pass
Specification Verification (Not direct performance, but conformance)
Typical Particle Size30-100 micronsN/A. Specification Study.
System Dead SpaceMAD600/600OS = 0.24mL; MAD700-730/730OS = 0.18mL; MAD720 = 0.12mLN/A. Specification Study.
Tip Diameter0.19 inches (4.8 mm)N/A. Specification Study.
Applicator LengthMAD600/6000S/700 = 8.9 inches; MAD720/730/730OS = 4.9 inchesN/A. Specification Study.
Conical Fittings with 6% TaperClauses 4.1-4.1 of ISO 540-2:1998Certificate of Conformance from the Supplier
Sterile Hypodermic Syringes of Single UseISO 7886-1:1993Certificate of Conformance from the Supplier

2. Sample size used for the test set and the data provenance

The document does not specify the sample sizes used for any of the biocompatibility or performance tests. The data provenance is not explicitly mentioned, but these are all non-clinical, bench-top tests conducted to verify product specifications and safety, likely performed internally by the manufacturer (Teleflex Medical). No human or animal data is described for these tests; they are related to material properties and device functionality.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This section is not applicable as the described tests are non-clinical and do not involve expert interpretation or human ground truth. They are objective measurements against defined chemical, physical, and engineering criteria.

4. Adjudication method for the test set

This section is not applicable as the described tests are objective measurements against defined criteria, not subjective assessments requiring adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There is no mention of an MRMC comparative effectiveness study, nor is there any AI component to this device. This device is a manual, single-use medical device for topical anesthetic application.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This section is not applicable as the device is not an algorithm or AI-based system. It is a physical medical device.

7. The type of ground truth used

The "ground truth" for the biocompatibility tests is established by recognized international standards (ISO 10993-1) and specific grading scales (e.g., Kligman scoring system). For performance tests, the "ground truth" is the predefined quantitative and qualitative acceptance criteria based on engineering specifications and functional requirements. For specification verification, it's conformance to established international standards (e.g., ISO 540-2, ISO 7886-1) or internal product specifications.

8. The sample size for the training set

This section is not applicable as there is no training set mentioned for this type of device.

9. How the ground truth for the training set was established

This section is not applicable as there is no training set described.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol. The logo is simple and recognizable, and it is often used to represent the U.S. Department of Health & Human Services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 19, 2016

Teleflex Medical Kristen Bisanz Regulatory Affairs Specialist 3015 Carrington Mill Blvd Morrisville, NC 27560

Re: K153470

Trade/Device Name: MADgic™ Laryngo-Tracheal Mucosal Atomization Device Regulation Number: 21 CFR 868.5170 Regulation Name: Laryngotracheal Topical Anesthesia Applicator Regulatory Class: Class II Product Code: CCT Dated: July 22, 2016 Received: July 25, 2015

Dear Kristen Bisanz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, James J. Lee -S

for Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K153470

Device Name

MADgic Laryngo-Tracheal Mucosal Atomization Device

Indications for Use (Describe)

The MADgic Laryngo-Tracheal Mucosal Atomization Device is intended for the application of topical anesthetics to the oropharynx and upper airway region.

Type of Use (Select one or both, as applicable)|X Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

MADgicTM Laryngo-Tracheal Mucosal Atomization Device

Name, Address, Phone and Fax Number of Applicant

Teleflex Medical, Incorporated 3015 Carrington Mill Blvd Morrisville, NC 27560 USA Phone: 919.433.4932 919.433.4996 Fax:

Contact Person

Kristen Bisanz Regulatory Affairs Specialist

Date Prepared

December 01, 2015

Device Name

Trade Name:MADgic™ Laryngo-Tracheal Mucosal AtomizationDevice
Classification Name:Applicator, Laryngo-Tracheal, Topical Anesthesia
Product Code:CCT
Regulation Number:868.5170
Classification:II
Classification Panel:Anesthesiology

Predicate Device

This submission demonstrates substantial equivalence to the predicate device MADgic™ Laryngo-Tracheal Mucosal Atomization Device which was cleared in submission K002255.

Device Description

The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is a disposable nonsterile device that converts a solution of topical anesthetic into a fine particle spray for application to mucosal surfaces. The device consists of an atomizer tip, a semirigid tubular extension, a standard luer lock adapter and a syringe. The clinician draws up the desired volume of topical anesthetic into the syringe, attaches it to the luer lock fitting of the atomizer, and manipulates the tubing extensions into the desired position. As the syringe plunger is compressed, the anesthetic is forced into the tubular extension and out of the atomizer tip. The tip takes the pressurized fluid column and begins spinning it, allowing the fluid to exit the small hole at the end in a

{4}------------------------------------------------

cone shaped spray. The anesthetic mist is gently distributed onto the mucosal surface in front of this tip. Topical anesthesia application can begin in the mouth and pharynx, and then proceed to the hypopharynx, epiglottis and vocal cords, larynx and trachea via direct visualization using a laryngoscope.

Indications for Use

The MADgic™ Laryngo-Tracheal Mucosal Atomization Device is intended for the application of topical anesthetics to the oropharynx and upper airway region.

Patient Population

This device is for patients requiring topical anesthetic before intubation.

Environments of use

The environments of use are hospitals and sub-acute facilities as directed by a physician.

Contraindications

There are no contraindications for this device.

Substantial Equivalence

The proposed device is substantially equivalent to the predicate device:

Predicate DeviceManufacturer510(k) NumberDate ClearedSystem Dead SpaceMAD600/600OS = 0.25mLMAD700-730/730OS =0.19mLMAD720 = 0.13mLMAD600/600OS= 0.24mLMAD700-730/730OS= 0.18mLMAD720 = 0.12mLEquivalent
MADgic™ Laryngo-TrachealMucosal Atomization DeviceWolfe ToryMedicalK002255September19, 2000Tip Diameter0.19 inches (4.8mm)0.19 inches (4.8mm)Identical
PredicateK002255MADgicTM Laryngo-Tracheal MucosalAtomization DeviceProposed MADgicTMLaryngo-TrachealAtomization DeviceEquivalenceApplicator LengthMAD600/600OS= 8.5 inches (21.6cm)MAD720/730/730OS= 4.5 inches (11.4cm)MAD600/600OS= 8.9 inches (22.6cm)MAD720/730/730OS= 4.9 inches (12.4cm)Equivalent
Classification NameApplicator, Laryngo-Tracheal, Topical AnesthesiaApplicator, Laryngo-Tracheal, Topical AnesthesiaIdenticalSterilizationNon-SterileNon-SterileIdentical
Device NameMADgic™ Laryngo-Tracheal MucosalAtomization DeviceMADgic™ Laryngo-Tracheal MucosalAtomization DeviceIdenticalPackagingPackaged individually inpolyethylene pouches withtwenty-five (25) devicesper inner carton, four (4)cartons per case (for a totalof 100 devices)Packaged individually inpolyethylene pouches withtwenty-five (25) devicesper inner carton, four (4)Identical
Common NameAtomizerAtomizerIdenticalBiocompatibilityPer ISO 10993-1Per ISO 10993-1Identical
Product Code73CCT73CCTIdenticalDosage AmountUser ControlledUser ControlledIdentical
ClassificationClass IIClass IIIdenticalMaterial for TubingAlpha Gary 2222R-90Colorite PolymersUnichem, 9088-015, Clear,DEHP-Free PVCEquivalent
Regulation Number868.5170868.5170IdenticalMaterial for LuerAdaptorPolycarbonatePolyvinyl Chloride 2014 G-010Equivalent
Indications for UseThe MADgic™ Laryngo-Tracheal MucosalAtomization Device isintended for the applicationof topical anesthetics to theoropharynx and upperairway region.The MADgic™ Laryngo-Tracheal MucosalAtomization Device isintended for the applicationof topical anesthetics to theoropharynx and upper airwayregion.IdenticalMaterial for Wire302 Stainless Steel302 Stainless SteelIdentical
PrescriptionYesYesIdenticalMaterial for SprayTipDow Calibre 2081-15,PolycarbonateDow Calibre 2081-15,PolycarbonateIdentical
Environment of UseHospitals, Sub-acutefacilities, and emergencymedical servicesHospitals, Sub-acutefacilities, and emergencymedical servicesIdenticalMaterial for SprayTip InsertChevron K-Resin, KR03Chevron K-Resin, KR03Identical
Patient PopulationPatients requiring topicalanesthesia before intubationPatients requiring topicalanesthesia before intubationIdenticalMaterial for SyringePolypropylene and Non-Latex RubberPolypropylene and Non-Latex RubberIdentical
ContraindicationsNoneNoneIdenticalMaterial forAdhesiveCyclohexanoneCyclohexanoneIdentical
Shelf LifeThree (3) years from dateof manufactureThree (3) years from dateof manufactureIdenticalMaterial for FoilHot StampNot present100% Virgin, high claritypolyethylene filmEquivalent
Typical Particle Size30-100 microns30-100 micronsIdentical

{5}------------------------------------------------

Comparison to Predicate Device

The proposed device has the same indications for use, operating principles, classification, function, and general design as the predicate device. Biocompatibility testing and performance testing have been performed on the proposed device in order to establish substantial equivalence to the predicate device. The proposed changes to the device do not impact the safety or effectiveness of the MADgic™ Laryngo-Tracheal Mucosal Atomization Device.

{6}------------------------------------------------

{7}------------------------------------------------

Materials

All patient contacting materials, including those with indirect patient contact, are in compliance with ISO 10993-1. Biocompatibility testing has been performed on the proposed device and meets the acceptance criteria.

TestAcceptance CriteriaResults
Cytotoxicity - L929MEM Elution AssayThe test article will meet the requirements of the test if itobtains a Grade of 0,1,or 2 (not more than 50% of the cellsare round, devoid of intracytoplasmic granules, and noextensive cell lysis)Acceptable
Sensitization -KligmanMaximization AssayThe test article will be considered a non-irritant if thedifference between the test article mean score and thevehicle control mean score is 1.0 or less.Acceptable
Irritation -IntracutaneousInjection AssayThe test article will meet the requirements of the test if itreceives a Grade of 1, 0 or less using the Kligman scoringsystem.Acceptable

Performance Data

Non-clinical performance testing has been conducted in order to support that the proposed device performs as intended and the product conforms to user needs.

TestTest ObjectiveAcceptance CriteriaResults
Visual InspectionTo visually inspect thedevice for defects ofcomponents or assemblyVisually observe for:- Proper assembly and presence of allcomponents- No cracks in the tip and luer- No short shots in the moldedcomponents- The presence of solvent aroundcircumference of bonding surfacecovering a minimum of one third ofthe total area- No gaps between the insert post andtubingPass
Flow TestTo validate the flow rateof the deviceThe flow rate should be greater thanor equal to 225 sccm and less than orequal to 700 sccm.Pass
Leak TestTo validate the leakvalue of the deviceThe leak value should be less than orequal to 0.05psi.Pass
Hydrostatic TestTo validate the effect ofhydrostatic pressure onthe deviceThe device must remain assembledafter being subjected to a minimum300psi hydrostatic pressure.Pass

{8}------------------------------------------------

Atomization TestPost HydrostaticTo validate theatomization function ofthe deviceUpon atomization there can be nostreaming, no excessive dripping, orocclusions.Pass
TensileTo validate the tensilestrength of the deviceThe tensile strength should be greaterthan or equal to 6.0 lb.Pass

The testing below summarizes the specification verification.

TestSpecificationAcceptance Criteria
Typical Particle Size30-100 micronsN/A. Specification Study.
System Dead SpaceMAD600/600OS = 0.24mLMAD700-730/730OS = 0.18mLMAD720 = 0.12mLN/A. Specification Study.
Tip Diameter0.19 inches (4.8 mm)N/A. Specification Study.
Applicator LengthMAD600/6000S/700 =8.9 inchesMAD720/730/730OS =4.9 inchesN/A. Specification Study.
Conical Fittings with6% TaperClauses 4.1-4.1 of ISO 540-2:1998Certificate of Conformancefrom the Supplier
Sterile HypodermicSyringes of Single UseISO 7886-1:1993Certificate of Conformancefrom the Supplier

Conclusion

Based on the performance and comparative test results, the proposed MADgic™ Laryngo-Tracheal Mucosal Atomization Device is substantially equivalent to the predicate device cleared to market in K002255. The modifications made to the MADgic™ Laryngo-Tracheal Mucosal Atomization Device do not introduce any new issues of safety and effectiveness.

§ 868.5170 Laryngotracheal topical anesthesia applicator.

(a)
Identification. A laryngotracheal topical anesthesia applicator is a device used to apply topical anesthetics to a patient's laryngotracheal area.(b)
Classification. Class II (performance standards).