K Number
K994278
Date Cleared
2000-03-28

(99 days)

Product Code
Regulation Number
862.1050
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Beckman Coulter, Inc's Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

Device Description

The Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma.

AI/ML Overview
{
  "acceptance_criteria_and_device_performance": {
    "title": "Acceptance Criteria and Reported Device Performance for Access® Ostase® Assay",
    "table": {
      "headers": ["Performance Characteristic", "Acceptance Criteria (Implicit via predicate comparison)", "Reported Device Performance"],
      "rows": [
        ["Method Comparison (Slope vs. Predicate)", "Close to 1", "0.9756"],
        ["Method Comparison (Intercept vs. Predicate)", "Close to 0", "-0.5987"],
        ["Method Comparison (Correlation r vs. Predicate)", "Close to 1", "0.9895"],
        ["Mean % Difference (Osteoporosis vs. Predicate)", "Low Percentage Difference", "-5.96%"],
        ["Mean % Difference (Paget's vs. Predicate)", "Low Percentage Difference", "-7.01%"],
        ["Between-Run %CV (Imprecision)", "Low Percentage (e.g., <10%)", "3.3% to 5.9%"],
        ["Within-Run %CV (Imprecision)", "Low Percentage (e.g., <10%)", "1.5% to 2.6%"],
        ["Total %CV (Imprecision)", "Low Percentage (e.g., <10%)", "3.6% to 6.4%"],
        ["Recovery (Spike and Recovery)", "Close to 100%", "Average 92.4% (range 89.9% to 95.3%)"],
        ["Recovery (Specimen Dilution Average)", "Close to 100%", "Average 92.2% (range 78.1% to 106.1%)"],
        ["Interference (Absence of significant interference)", "No significant interference", "No significant interference from tested substances"]
      ]
    },
    "notes": "The acceptance criteria for the Access® Ostase® Assay are implicitly defined by its substantial equivalence claim to the Tandem®-R Ostase® Assay predicate device. The reported performance data demonstrate close agreement, high correlation, and low variability, indicating that the device performs comparably to the predicate and meets general expectations for diagnostic assays."
  },
  "study_details": {
    "sample_size_test_set": {
      "method_comparison": "172 samples (88 osteoporosis, 84 Paget's)",
      "imprecision": "Across various concentrations (exact number of individual samples not specified, but multiple runs and concentrations were evaluated).",
      "recovery_spike": "15 samples",
      "recovery_dilution": "10 samples",
      "interfering_substances": "Not explicitly stated for the test set, but substances were tested at specified concentrations.",
      "data_provenance": "Not explicitly stated for the test set. However, a crucial statement indicates that all clinical studies were generated using the Tandem-R Ostase Assay (the predicate device), implying that the clinical utility data, rather than the analytical performance data, came from studies on the predicate."
    },
    "number_of_experts_ground_truth_test_set": "Not applicable; this is a quantitative immunoassay comparing a new device to a predicate device, not a qualitative diagnostic interpretation requiring expert consensus on images or similar data.",
    "qualifications_of_experts": "Not applicable.",
    "adjudication_method": "Not applicable.",
    "multi_reader_multi_case_comparative_effectiveness_study": "No, this is an in vitro diagnostic (IVD) device for quantitative measurement of a biomarker, not an imaging or interpretive device requiring human-in-the-loop assessment.",
    "standalone_performance": "Yes, the study presents the standalone performance of the Access® Ostase® Assay by comparing its measurements directly to those of the predicate device (Tandem®-R Ostase® Assay) and evaluating its analytical characteristics (imprecision, recovery, interference).",
    "type_of_ground_truth_test_set": "The 'ground truth' for the analytical performance studies (method comparison, imprecision, etc.) is the quantitative measurement provided by the established predicate device (Tandem®-R Ostase® Assay) and established analytical methodologies. For clinical utility, the 'ground truth' for indications (managing osteoporosis and Paget's disease) was established by studies performed using the predicate device, which were then referenced for the new device through the substantial equivalence pathway.",
    "sample_size_training_set": "Not applicable for this type of IVD device. The device is not based on a machine learning algorithm that requires a dedicated training set in the typical sense. Performance is established through analytical validation against a predicate.",
    "ground_truth_for_training_set_established": "Not applicable."
  }
}

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

1.0 SUBMITTED BY:

Jill Kull, RAC, Staff Regulatory Specialist Beckman Coulter Inc., 7330 Carroll Rd. / P.O. Box 269006, San Diego, CA 92196-9006 Telephone: (858) 621-4584 FAX: (858) 621-4752 e-mail: jfkull@beckman.com

2.0 DATE SUBMITTED:

March 14, 2000

DEVICE NAME(S) 3.0

  • 3.1 Proprietary Names Access® Ostase® Assay on the Access Immunoassay Analyzer 3.2
    • Classification Names Bone alkaline phosphatase (BAP) test system.

PREDICATE DEVICE 4.0

Beckman CoulterProductPredicatePredicate CompanyDocketNumber
Access® Ostase®AssayTandem®-ROstase® AssayBeckman Coulter, Inc.K961573

5.0 DESCRIPTION OF THE DEVICE

The Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma.

Beckman Coulter Inc., Section 510(k) Notification Access® Ostase® Assay - K994278 Additional Information

Page 26

{1}------------------------------------------------

6.0 INTENDED USE OF THE DEVICE

This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

7.0 COMPARISON TO THE PREDICATE

TechnologicalCharacteristicAccess OstaseTandem-ROstase(Predicate)
AnalyteMeasuredHuman bone alkaline phosphataseSame
Intended UseAid in the management ofpostmenopausal osteoporosis andPaget's diseaseSame
Solid PhaseAntibodyMonoclonal antibody to BAPSame
Specimen MatrixHuman serum and plasmaHuman serum
Assay SignalEnzymetric/LuminometerRadiometric/Gamma Counter
Assay SolidPhaseMagnetic particlesBead
System MethodAutomatedManual

{2}------------------------------------------------

8.0 SUMMARY OF PERFORMANCE DATA

8.1 Method Comparison Study Results

Access Ostase vs. Tandem-R Ostase

SlopeInterceptrN
0.9756-0.59870.9895172

Mean % Differences Between Access Ostase and Tandem-R Ostase

PopulationsN(# pairs)Mean%Difference*StandardDeviation
Osteoporosis88-5.96%12.37%
Paget's84-7.01%14.25%
  • Mean % Difference = [(Access -Tandem )/(Access+ Tandem/2)) X 100

8.2 Imprecision Summary

The between-run %CV observed across the concentrations tested ranged from 3.3% to 5.9%. The within-run %CV observed across the concentrations tested ranged from 1.5% to 2.6%. The total %CV observed across the concentrations tested ranged from 3.6% to 6.4%.

8.3 Recovery and Specimen Dilution

The results of spike and recovery across 15 samples ranged from 89.9% to 95.3% with an average recovery of 92.4%.

The results for ten (10) samples in the dilution study ranged from 78.1% to 106.1% with an average recovery of 92.2%. Regression analysis of these data yielded slopes ranging from 0.9280 to 1.0075 with an overall average slope of 0.9682. Correlation coefficients (r) ranged from 0.9999 to 0.9983 with an overall average correlation coefficient of 0.9994.

{3}------------------------------------------------

8.4 Interfering Substances

The following substances and concentrations were evaluated for interference in the Access Ostase Assay. There is no significant interference from any of the substances tested at these concentrations.

SubstanceConcentrationTestedSubstanceConcentrationTested
Acetaminophen35 mg/dLIbuprofen40 mg/dL
alendronate8 mg/dLPamidronate18 mg/dL
Aspirin50 mg/dLprogesterone25 mg/dL
bilirubin –unconjugatedconjugated40 mg/dL20 mg/dLprotein3.8 and15.6 g/dL
calcitonin-salmon112 IU/dLraloxifene12 mg/dL
Calcium40 mg/dLrisedronate6 mg/dL
estrogen10 mg/dLtriglycerides2000 mg/dL
etidronate105 mg/dLvitamin D80,500 IU/dL
hemoglobin500 mg/dL

In summary, there is no significant interference from any of the substances tested at the concentrations listed above.

  • 8.5 Reactivity with Intestinal, Placental and Liver Alkaline Phosphatase lsoenzymes
    In these studies, 100 U/L of intestinal alkaline phosphatase yielded a result of 1.8 µg/L in the Access Ostase Assay. 100 U/L of placental alkaline phosphatase activity yielded a result of 0.3 µg/L in the Access Ostase Assay.

This summary of safety and effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and the implementing regulation 21 CFR 807.92.

{4}------------------------------------------------

  • Question 4. Please add a statement at the beginning of the clinical trial section indicating that all the clinical trials were performed with the Tandem-R.
  • The following statement was added to page 5 (page 95 of the original Answer: submission) of the directional insert.

CLINICAL STUDIES

In the following section the studies were generated using the Tandem-R Ostase Assay.

Question 5: Please send a copy of the revised labeling for the File.

Answer: A copy of the revised directional insert is attached.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three birds in flight, arranged in a row.

MAR 28 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Jill Kull, RAC Staff Regulatory Specialist Beckman Coulter, Inc. 7330 Carroll Road P.O. Box 269006 San Diego, California 92196-9006

K994278 Re: Trade Name: Access® Ostase® Assay Regulatory Class: II Product Code: CIN, JIS Dated: March 14, 2000 Received: March 15, 2000

Dear Ms. Kull:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

INDICATIONS FOR USE STATEMENTS

510(k) Number (if known): Not yot assigned

Access® Ostase® Assay Device Name:

Indications for Use:

Beckman Coulter, Inc's Access Ostase Assay is a paramagnetic particle, chemiluminescent immunoassay for use with the Access Immunoassay System for the quantitative measurement of bone alkaline phosphatase (BAP), an indicator of osteoblastic activity, in human serum and plasma. This device is intended to be used as an aid in the management of postmenopausal osteoporosis and Paget's disease.

21 CFR 862.1050 Alkaline Phosphatase or isoenzymes test system

  • (a) Identification. An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.
    (b) Classification. Class II.

Division of Clinical Laboratory Devices

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(per 21 CFR 801.109)

OR

Over-the-Counter Use Optional Format 1-2-96

Beckman Coulter Inc., Section 510(k) Notification Access® Ostase® Assay AccessOstase\Section_1.doc

Page 14 Section 1

§ 862.1050 Alkaline phosphatase or isoenzymes test system.

(a)
Identification. An alkaline phosphatase or isoenzymes test system is a device intended to measure alkaline phosphatase or its isoenzymes (a group of enzymes with similar biological activity) in serum or plasma. Measurements of alkaline phosphatase or its isoenzymes are used in the diagnosis and treatment of liver, bone, parathyroid, and intestinal diseases.(b)
Classification. Class II.