K Number
K994226
Date Cleared
2000-03-22

(98 days)

Product Code
Regulation Number
876.5665
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

SteriChek ™ Total Chlorine Reagent Strips provide a quick convenient means of testing for low levels of total chlorine (i.e. total chloramines plus free chlorine) in water used to prepare dialysate. The color that develops in the pad after exposure to the sample according to the directions is compared to a color chart to determine the concentration of total chlorine present in the sample.

Device Description

The device is made up of a 0.20 inch square off-white reagent pad that has been chemically treated and affixed to one end of a 3.25 inch by 0.20 inch white opaque polystyrene strip. The reagent pad is activated by exposing it to the sample. The color of the pad is visually compared to a color chart to determine the amount of total chlorine present in the sample.

AI/ML Overview

The provided text describes the SteriChek™ Total Chlorine Reagent Strips, a device for testing total chlorine levels in water for dialysate preparation. Here's a breakdown of the requested information based on the provided document:

Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated or Implied)Reported Device Performance
Detection Range: Detect total chlorine concentrations between 0 and 3 ppm.The device is used as a quantitative method to detect total chlorine concentrations between 0 and 3 ppm.
Low-Level Detection: Reliably measure total chlorine concentrations as low as 0.1 ppm.The device will reliably measure total chlorine concentrations as low as 0.1 ppm total chlorine.
Equivalence to Predicate Device: Performance characteristics should be equivalent to the predicate device (Serim™ HiSense Test Kit)."The performance characteristics of SteriChek™ Total Chlorine reagent Strips and Serim™ HiSense Test Kit were analyzed with water samples... The performance was equivalent."

Study Information

  1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: The document states "water samples in which either sodium hypochlorite or chloramines were added to give a range of free chlorine or combined chlorine levels." However, it does not specify the exact number of samples or the range of concentrations tested beyond 0-3 ppm and 0.1 ppm low-level detection.
    • Data Provenance: The document does not explicitly state the country of origin. Given the submission is to the FDA (USA), it's highly probable the testing was conducted in the USA. It's a retrospective analysis of the device's performance against controlled samples.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • The document does not mention the use of human experts to establish ground truth for visual comparison. The "color blocks are calibrated in terms of chlorine concentration," implying a predefined standard. The "color of the pad is visually compared to a color chart" by the user.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • None is explicitly stated. The device relies on a direct visual comparison by the user to a color chart.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC study was not done. This device is a simple reagent strip with visual comparison, not an AI-powered system or one that requires human readers in the traditional sense of medical image interpretation.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, a standalone performance assessment was done. The performance of the SteriChek™ strips was directly compared to the predicate device's performance using water samples with known chlorine concentrations. The "algorithm" here is the chemical reaction on the strip leading to a color change, which is then interpreted by visual comparison. The study assesses the inherent ability of the strip to accurately reflect chlorine concentration.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth was established by known, precisely prepared chemical concentrations of free chlorine and chloramines in water samples. These were "water samples in which either sodium hypochlorite or chloramines were added to give a range of free chlorine or combined chlorine levels."
  7. The sample size for the training set:

    • The document does not specify a separate training set. This is a chemical reagent strip, not a machine learning model that typically involves training data. The "calibration" of the color chart would have been done during the device's development, but details on that process or sample sizes are not provided within this summary.
  8. How the ground truth for the training set was established:

    • As there's no explicitly mentioned training set in the context of machine learning, this question is not fully applicable. However, the ground truth for calibrating the color chart (which informs the "training" aspect of the device's functionality) would have been established through controlled laboratory experiments using precisely measured concentrations of chlorine to correlate color intensity with ppm values. This process is implied by the statement "The color blocks are calibrated in terms of chlorine concentration."

{0}------------------------------------------------

994226 p. i/2

MAR 2 2 2000

SteriChek™ Total Chlorine Reagent Strips 510(k) Submission Environmental Test Systems, Inc.

510(k) SAFETY AND EFFECTIVENESS SUMMARY

Prepared:December 10, 1999
Submitter:Environmental Test Systems, Inc.
Address:23575 County Road 106Elkhart, IN 46514U.S.A.(219) 262-2060
Contact:David A. Morris, Ph.D.Vice President, Technology
Device Trade/Proprietary Name:SteriChek ™ Chlorine Reagent Strips
Device Common Name:Total Chlorine Reagent Strips
Classification Name:Class II CH
Predicate Device:Serim ™ HiSense Test Kit
Device Description:The device is made up of a 0.20 inch square off-white reagent pad that hasbeen chemically treated and affixed to one end of a 3.25 inch by 0.20 inchwhite opaque polystyrene strip. The reagent pad is activated by exposingit to the sample. The color of the pad is visually compared to a color chartto determine the amount of total chlorine present in the sample.
Intended Use:SteriChek ™ Total Chlorine Reagent Strips provide a quick convenientmeans of testing for low levels of total chlorine (i.e. total chloramines plusfree chlorine) in water used to prepare dialysate. The color that developsin the pad after exposure to the sample according to the directions iscompared to a color chart to determine the concentration of total chlorinepresent in the sample.
Technological Characteristics:The concentration of chlorine in rinse water is obtained by comparing the

{1}------------------------------------------------

K994226 p.2/2

MAR 2 2 2000

SteriChek™ Total Chlorine Reagent Strips 510(k) Submission Environmental Test Systems, Inc.

color of the reagent pad with color blocks on the label. The color blocks are calibrated in terms of chlorine concentration in parts per million (ppm). The device is used as a quantitative method to detect total chlorine concentrations between 0 and 3 ppm. The device will reliably measure total chlorine concentrations as low as 0.1 ppm total chlorine.

SteriChek™ Chlorine Reagent Strips contains an indicator, Michler's thioketone, an activating surfactant, and other non-reactive ingredients. Free chlorine and combined chlorine (chloramines) react with the indicator to form a blue complex. The amount of the blue complex is dependent on the concentration of free chlorine and combined chlorine in the sample.

Assessment of Performance:

The performance characteristics of SteriChek™ Total Chlorine reagent Strips Serim™ and HiSense Test Kit were analyzed with water samples in which either sodium hypochlorite or chloramines were added to give a range of free chlorine or combined chlorine levels. The performance was equivalent.

Conclusion: The SteriChek™ Total Chlorine Reagent Strips have the same intended use as the predicate device. The predicate device's indicator system (qualitative combined dry and liquid colorimetric method) is different than that of the Total Chlorine Reagent Strips. However, both systems effectively measure the total free and combined chlorine levels in water. The SteriChek™ Total Chlorine Reagent Strips have no technological characteristics that raise new types of safety or effectiveness questions.

{2}------------------------------------------------

Public Health Service

MAR 2 2 2000

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

David A. Morris, Ph.D. Vice President, Technology Environmental Test Systems, Inc. 23575 County Road, #106 Elkhart, IN 46514

Re: K994226 SteriChek™ Total Chlorine Reagent Strips Dated: March 1, 2000 Received: March 6, 2000 Requiatory Class: II 21 CFR $876.5665/Procode: 78 MSY

Dear Dr. Morris:

We have reviewed your Section 510K) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassfied in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket not affect any obligation you might have under sections 531 through 542 of the Act for devices under Radiation Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally market predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591. Additionally, for question and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Daniel G. Schultz, M.D. Captain, USPHS Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

SteriChek™ Total Chlorine Reagent Strips 510(k) Submission Environmental Test Systems, Inc.

510(k) Number (if known) K994336

Device Name: SteriChek™ Total Chlorine Reagent Strips.

Indications for Use:

:

SteriChek™ Total Chlorine Reagent Strips provide a quick convenient means of testing for low levels of total chlorine (i.e. total chloramines and free chlorine) in water used to prepare dialysate.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED).

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109)

(Division Sign-Off) Division of Reproductive, Abdominal, El and Radiological Dev 510(k) Number

Over-The-Counter Use

Page 6

OR

§ 876.5665 Water purification system for hemodialysis.

(a)
Identification. A water purification system for hemodialysis is a device that is intended for use with a hemodialysis system and that is intended to remove organic and inorganic substances and microbial contaminants from water used to dilute dialysate concentrate to form dialysate. This generic type of device may include a water softener, sediment filter, carbon filter, and water distillation system.(b)
Classification. Class II (special controls). The device, when it is a water purification subsystem disinfectant, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.