K Number
K994158

Validate with FDA (Live)

Date Cleared
2000-07-24

(228 days)

Product Code
Regulation Number
888.3030
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The MacroPoreOS Protective Sheet is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures. The MacroPoreOS Protective sheet is also indicated for cement restriction in total joint arthroplasty procedures.

Only when used in conjunction with traditional rigid fixation, the MacroPoreOS Protective Sheet is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:

  • · Long bones
  • Flat bones
  • · Short bones
  • · Irregular bones
  • · Appendicular skeleton
  • · Thorax

When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving:

  • · Tumor resections where bone strength has not been compromised
  • · Iliac crest harvests

This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.

Device Description

MacroPoreOS Protective Sheet is a resorbable, macroporous implant in sheet form manufactured from polylactic acid (PLA). MacroPoreOS Protective Sheet can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPoreOS Protective Sheet to the desired shape or size. MacroPoreOS Protective Sheet is fully malleable when heated to approximately 65℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPoreOS Protective Sheet can be rolled into a tube or used as a flat sheet. MacroPoreOS Protective Sheet can be used either alone or in conjunction with internal bone fixation devices such as plates and screws, which also can serve to further stabilize the anatomical region. The MacroPoreOS System includes MacroPoreOS Protective Sheets, a selection of resorbable MacroSorb Screws, MacroSorb Tacks and associated manual instruments.

MacroPoreOS Protective Sheet is provided in sheets of 20 x 20 mm to 120 x 120 mm and will be provided in other sizes as needed for particular surgical procedures. The pore size ranges from 500 microns to 2500 microns in diameter, with pores distributed uniformly throughout the sheet in an offset or aligned pattern. The thickness of the MacroPoreOS Protective Sheet ranges from 0.50 mm to 2.0 mm according to the orthopedic region to be treated, however, not to exceed a total mass of 18 grams.

The MacroPoreOS Protective Sheet is fabricated from polylactic acid (PLA).

AI/ML Overview

Here's an analysis of the acceptance criteria and study information for the MacroPoreOS Protective Sheet (K994158), based on the provided text:

Acceptance Criteria and Device Performance (K994158)

The provided document describes a 510(k) submission, which primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than presenting a performance study with specific acceptance criteria in the same way a clinical trial might. Therefore, the "acceptance criteria" here are largely defined by the characteristics of the predicate device and the mechanical properties deemed equivalent.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Inferred from Predicate Equivalence)Reported Device Performance (MacroPoreOS Protective Sheet)
Intended Use: Similar to predicate (Sofamar Danek Timesh and other MacroPore Protective Sheets) for reinforcing weak bony tissue and cement restriction."The MacroPoreOS Protective Sheet shares indications for use with the predicate device as both the I he Macrol oroon Trotective Sheet and the Timesh predicated for reinforcing weak bony tissue in orthopedic procedures. The MacroPoreOS Protective Sheet and the Timesh predicate also share cement restriction indication for use."
Material: Biocompatible, amorphous polymer."Fabricated from polylactic acid (PLA)." "Poly (L-lactide-co-D,L-lactide) 70:30, amorphous."
Design: Thin semi-rigid sheet, macroporous, cuttable to size. Pore size and spacing comparable to predicate."Resorbable, macroporous implant in sheet form manufactured from polylactic acid (PLA). Can be cut with scissors... fully malleable when heated... Pores are 500 microns to 2500 microns in diameter, distributed uniformly... thickness ranges from 0.50 mm to 2.0 mm." "The pore size and spacing of the predicate device is within the pore size and spacing specifications of the MacroSorbOS Protective Sheet."
Dimensions: Provided in rectangular sheets, comparable to predicate."Sheets of 20 x 20 mm to 120 x 120 mm" and "Dimensions of the predicate device are also operiodions or the MacroSorbOS Protective sheet as both devices are provided in rectangular sheets that oompanders..." (implying similar size formats).
Mechanical Properties: Initial and in-vivo strengths and rigidity substantially equivalent to predicate.In Vitro Testing: - Effect of Heating: Viscosity stayed within appropriate range over 120 minutes at 60°C. - Accelerated Aging: As rigid and strong as the predicate after simulated 6-month in vivo exposure. Retains strength for 9 months, decreases to zero after approximately 18 months. - Mechanical Testing: "Determined the MacroPoreOS Protective Sheet to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions."
Crystallinity: Amorphous (non-crystalline)."The tests ran on the sterile and non-sterile samples revealed no endothermic spikes. From this we verify the implants are amorphous or non-crystalline."
Resorption: Intended to be metabolized by the body."The polymer devices are intended to be metabolized by the body and do not require removal." (Implied by accelerated aging test showing strength decrease over time).

2. Sample Size Used for the Test Set and Data Provenance

The document does not describe a clinical test set in the traditional sense (e.g., patient data for diagnostic accuracy). The "testing" primarily involved in vitro studies and comparative analysis to predicate devices.

  • Sample Size: Not explicitly stated in terms of patient numbers or clinical cases. The samples for in vitro tests would be material samples of the MacroPoreOS Protective Sheet itself.
  • Data Provenance: The 'data' are results from laboratory testing (in vitro) performed by the manufacturer, MacroPore, Inc. There is no mention of country of origin for clinical data or whether it was retrospective or prospective, as no human clinical data is presented for this 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

Not applicable. No "ground truth" based on expert consensus for clinical cases is mentioned, as this submission relies on demonstrating substantial equivalence through mechanical and material properties, not clinical diagnostic or treatment efficacy data from human subjects.

4. Adjudication Method for the Test Set

Not applicable for the same reasons as point 3.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No. This is a 510(k) premarket notification for a Class II bone fixation appliance. It does not involve AI, image interpretation, or a comparative effectiveness study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

No. This device is not an algorithm or AI system.

7. The Type of Ground Truth Used

The "ground truth" in this context is established by the known, accepted properties and performance of the legally marketed predicate devices (Sofamar Danek Timesh K974017, MacroPore Protective Sheet K972913, MacroPore Protective Sheet K983360) and standard scientific/engineering principles for material characterization and mechanical testing. The MacroPoreOS Protective Sheet's characteristics (material, design, mechanical strength, resorption profile) are compared to these established benchmarks.

8. The Sample Size for the Training Set

Not applicable. There is no training set mentioned, as this is not an AI/machine learning device. The "training" for the device, if one were to stretch the analogy, would be the research and development process that led to its design, informed by predicate device characteristics.

9. How the Ground Truth for the Training Set was Established

Not applicable, as there's no training set. The "ground truth" for the device's design parameters was established through comparison to legally marketed predicate devices and physical/chemical characterization tests (e.g., DSC for crystallinity, viscosity testing, accelerated aging).

{0}------------------------------------------------

K994158

K994158 Summary

MacroPoreOS Protective Sheet

Page 1 of 4

ADMINISTRATIVE INFORMATION

Manufacturer Name:

Official Contact:

MacroPore, Inc. 6740 Top Gun Street San Diego, CA 92121

Kenneth K. Kleinhenz Director of Regulatory Affairs Telephone (858) 458-0900 Fax (858) 458-0994

DEVICE NAME

Classification Name:

Trade/Proprietary Name:

MacroPoreOS Protective Sheet

Plate, Bone

ESTABLISHMENT REGISTRATION NUMBER 2031733

DEVICE CLASSIFICATION AND PRODUCT CODE

As shown in 21CFR 888.3030 Bone Fixation Appliances are intended for use in orthopedic procedures and are classified as Class II. Bone Plates have been assigned Product Code HRS.

INTENDED USE

The MacroPoreOS Protective Sheet is intended to maintain the relative position of weak bony tissue such as bone grafts, bone graft substitutes, or bone fragments from comminuted fractures. The MacroPoreOS Protective sheet is also indicated for cement restriction in total joint arthroplasty procedures.

Only when used in conjunction with traditional rigid fixation, the MacroPoreOS Protective Sheet is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:

  • · Long bones
  • Flat bones
  • · Short bones
  • · Irregular bones
  • · Appendicular skeleton
  • · Thorax

When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet is intended to maintain the relative position of bone graft substitutes in reconstructive orthopedic procedures involving:

  • · Tumor resections where bone strength has not been compromised
  • · Iliac crest harvests

This device is not intended for use in the spine. The device is not intended for load bearing indications unless used in conjunction with traditional rigid fixation.

{1}------------------------------------------------

DEVICE DESCRIPTION

Design Characteristics

Design Charactive Sheet is a resorbable, macroporous implant in sheet form manufactured from polylactic acid (PLA). MacroPoreOS Protective Sheet can be cut with scissors to the desired shape and size. The MacroPore Power Pen can also be used to cut or shape the MacroPoreOS Protective Sheet to the desired shape or size. MacroPoreOS Protective Sheet is fully malleable when heated to approximately 65℃ (for example, by the use of sterile hot water), and thus can be conformed three dimensionally to most any anatomical orientation. The MacroPoreOS Protective Sheet can be rolled into a tube or used as a flat sheet. MacroPoreOS Protective Sheet can be used either alone or in conjunction with internal bone fixation devices such as plates and screws, which also can serve to further stabilize the anatomical region. The MacroPoreOS System includes MacroPoreOS Protective Sheets, a selection of resorbable MacroSorb Screws, MacroSorb Tacks and associated manual instruments.

MacroPoreOS Protective Sheet is provided in sheets of 20 x 20 mm to 120 x 120 mm and will be provided in other sizes as needed for particular surgical procedures. The pore size ranges from 500 microns to 2500 microns in diameter, with pores distributed uniformly throughout the sheet in an offset or aligned pattern. The thickness of the MacroPoreOS Protective Sheet ranges from 0.50 mm to 2.0 mm according to the orthopedic region to be treated, however, not to exceed a total mass of 18 grams.

Material Composition

The MacroPoreOS Protective Sheet is fabricated from polylactic acid (PLA).

In Vitro Testing

Because the MacroPoreOS Protective Sheet is intended to be heated in the surgical suite to temperatures above the material's glass transition temperature to facilitate shaping to anatomic structures, testing was performed to determine the effect of prolonged heating in saline at 60°C on inherent viscosity. The testing demonstrates that viscosity stayed within an appropriate range over 120 minutes. Therefore, the relatively brief exposure anticipated during the surgical preparation of MacroPoreOS Protective Sheet is not expected to have a significant effect on its mechanical properties.

Accelerated aging testing was performed on MacroPoreOS Protective Sheet. Testing demonstrated that the MacroPoreOS Protective Sheet is as rigid and as strong as the predicate after a simulated 6 month in vivo exposure. Furthermore, simulated in vivo accelerated testing indicates that the MacroPoreOS Protective Sheet retains all of its strength for the first 9 months and a steadily decrease in strength to zero after approximately 18 months.

Mechanical testing was performed on the MacroPoreOS Protective Sheet which determined the MacroPoreOS Protective Sheet to be substantially equivalent to the mechanical strengths of the predicate devices under indication for use conditions.

{2}------------------------------------------------

K994158 Summary

Crystallinity was tested for by DSC (differential scanning calorimetery). This test measures the amount Of heat energy that is absorbed by a material. A crystalline material will require more energy once it of near it's melting point. This release of heat energy can be seen on a graph as a sharp spike and is referred to as a "melting endotherm". The tests ran on the sterile and non-sterile samples revealed no endothermic spikes. From this we verify the implants are amorphous or non-crystalline.

EQUIVALENCE TO MARKETED PRODUCT

MacroSorbOS Protective Sheet shares indications and design principles with the following predicate device, which has been determined by FDA to be substantially equivalent to the following preamendment devices: Sofamar Danek Timesh (K974017).

Indications For Use

The MacroPoreOS Protective Sheet shares indications for use with the predicate device as both the I he Macrol oroon Trotective Sheet and the Timesh predicated for reinforcing weak bony tissue in orthopedic procedures. The MacroPoreOS Protective Sheet and the Timesh predicate also share cement restriction indication for use.

Design and Materials

Design and MacroSorbOS Protective Sheet and the predicate device (Sofamar Danek Timesh) are similar, consisting of a thin semi-rigid sheet with macroporosity. Both the predicate device I milesh) are similar, consistent ave a semi-rigid construction with pores of similar diameter and spacing. The pore size and spacing of the predicate device is within the pore size and spacing specifications of the MacroSorbOS Protective Sheet. The dimensions of the predicate device are also operiodions or the MacroSorbOS Protective sheet as both devices are provided in rectangular sheets that oompanders to the mechanical characteristics of the MacroSorbOS Protective Sheet are substantially equivalent to the predicate device with respect to initial and in vivo strengths and rigidity. In addition to physical characteristics, both the predicate device and the MacroSorbOS Protective Sheet can be cut to specific shapes and sizes by the end user. The titanium device differs from MacroSorbOS Protective Sheet device in that it may be left in place permanently or must be removed surgically, whereas the polymer devices are intended to be metabolized by the body and do not require removal.

{3}------------------------------------------------

. .- THE STORESSONATION------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
.

:

ત્તા

UMMARY : TABLE OF SUBSTANTIAL EQUIVALENCE

.

Intended UseSubject DevicePredicate DeviceRelated Device
The MacroPoreOS Protective Sheet isintended to maintain the relativeposition of weak bony tissue such asbone grafts, bone graft substitutes, orbone fragments from comminutedfractures. The MacroPoreOSProtective sheet is also indicated forcement restriction in total jointarthroplasty procedures. Only whenused in conjunction with traditionalrigid fixation, the MacroPoreOSProtective Sheet is intended tomaintain the relative position weakbony tissue in trauma andreconstructive orthopedic proceduresinvolving: Long bones, Flat bones,Short bones, Irregular bones,Appendicular skeleton, Thorax. Whenused alone (without traditional rigidfixation), the MacroPoreOS ProtectiveSheet is intended to maintain therelative position of bone grafts orbone graft substitutes in reconstructiveorthopedic procedures involving:Tumor resections where bone strengthhas not been compromised, Iliac crestharvests. This device is not intendedfor use in the spine. The device is notintended for load bearing indicationsunless used in conjunction withtraditional rigid fixation.MacroSorboS ProtectiveSheetTimesh (K974017)Sofamar DanekFor use in any oral-maxillo-cranio-facial surgical reconstructiveprocedure, either orthognathic ortrauma, wherein rigid or semi-rigidinternal fixation is utilized as ameans of holding bone fragmentstogether. Alternatively, the Timeshsystem is also indicated for use inreinforcing weak bony tissues inorthopaedic surgical proceduressuch as pelvic reconstructions,acetabular reconstruction, andcement restriction. This product isnot intended for spinal use.MacroPore Protective Sheet(Protego System)(K972913)MacroPore Protective Sheet isintended for use in trauma andreconstructive procedures in themidface and craniofacial skeleton:1. Comminuted fractures of thenaso-ethmoidal and infraorbitalareas 2. Comminuted fractures ofthe frontal sinus wall 3. Traumaof the midface or craniofacialskeleton 4. Reconstructiveprocedures of the midface orcraniofacial skeleton. The systemis not intended for use in themandible and/or for full loadbearing procedures.MacroPore Protective Sheet(K983360)MacroPore Protective Sheet isintended to facilitate healing intrauma, reconstruction and boneaugmentation procedures of themandible. The following specificindications are included: tomaintain the relative position ofbony fragments in trauma andbone graft porcedures, and tocontain and prevent migration andshifting of autograft, allograftand/or bone graft substitutes thatmay be necessary inreconstructive procedures.
DesignSheets of 0.50 - 2.0 mmthickness, sizes 20 x 20 mm to120 x 120 mm or as required.Sheets of 0.2mm thickness,size 76 x 45 mm.Plates, screws and tacks ofvarious sizes.Sheets of 0.50-2.0 mmthickness, sizes 20 x 20 mm to120 x 120 mm or as required.
MaterialPoly (L-lactide-co-D,L-lactide) 70:30, amorphousTitaniumPoly (L-lactide-co-D,L-lactide)70:30, amorphousPoly (L-lactide-co-D,L-lactide)70:30,amorphous
Product CodeHRSHRSHRS and HWCJEY

.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around it. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 2 4 2000

Mr. Kenneth K. Kleinhenz Director of Regulatory Affairs MacroPore, Inc. 6740 Top Gun Street San Diego, California 92121

Re: K994158

Trade Name: MacroSorbOS Protective Sheet Regulatory Class: II Product Codes: HRS, MAI, HWC Dated: April 24, 2000 Received: April 25, 2000

Dear Mr. Kleinhenz:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general control provisions of the Act. The general control provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. Kenneth K. Kleinhenz

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

ume R. bochner -

Celia M. Witten, Ph.D., M.D. Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Device Name: MacroPoreOS Protective Sheet

Indications for Use:

Indications for Use:
The MacroPoreOS Protective Sheet is intended to maintain the relative fromers, from was The MacrororeOS Protective Sheet is meanse substitutes, or bone fragments from a bony tissue such as bone graits, bone gran-sacething, and indicated for cement restriction in total joint arthroplasty procedures.

Only when used in conjunction with traditional rigid fration, the MacroPoreoSings in traves Only when is intended to maintain the relative position weak bony tissue in trauma and reconstructive orthopedic procedures involving:

  • · Long bones
  • · Flat bones
  • Short bones
  • · Irregular bones
  • · Appendicular skeleton
  • Thorax

When used alone (without traditional rigid fixation), the MacroPoreOS Protective Sheet When used alone (without traditional right and in the grafts or bone grafts substitutes in reconstructive orthopedic procedures involving:

  • · Tumor resections where bone strength has not been compromised
  • · Iliac crest harvests

This device is not intended for use in the spine. The device is not intended for load I mis device is nor meets used in conjunction with traditional rigid fixation.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NECESSARY) 」 2 w = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =

Concurrence of CDRH, Office of Device Evaluation (ODE)

X Prescription Use _

OR

Over-The-Counter Use

Mina R. lochner'
Division Sign Off

(Division Sign-Off) Division of General Restorative Devices K994158 510(k) Number_

§ 888.3030 Single/multiple component metallic bone fixation appliances and accessories.

(a)
Identification. Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.