K Number
K993303
Date Cleared
2000-03-23

(171 days)

Product Code
Regulation Number
876.5010
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SC 35 Balloon Dilatation Catheter is indicated for stent deployment/optimization of the J&J Palmaz Biliary Stent and for the treatment of biliary strictures.

Device Description

The proposed SC 35 Balloon Dilatation catheter is an over-the-wire catheter designed to be placed over guidewires which have outer diameters of .035" or smaller.

AI/ML Overview

Here's an analysis of the provided text regarding the SC 35 Balloon Dilatation Catheter's acceptance criteria and the supporting study:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Criteria (Implicit)Reported Device PerformanceComments / Supporting Evidence
SafetyBiocompatibilityMet ISO-10993 requirements"biocompatibility testing (according to the FDA guidance document, ODE Blue Book Memorandum #G95-1, May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing") were performed"
PerformanceFunctional integrityDemonstrated safety and effectiveness"Functional and integrity bench testing were performed, and the data supported the substantial equivalence..."
Substantial EquivalenceDemonstrated substantial equivalence to predicate devices"data supported the substantial equivalence of the SC 35 Balloon Dilatation Catheter to the Ultra-thin Diamond Balloon Dilatation Catheter and the Courier ST Balloon Dilatation Catheter."

2. Sample size used for the test set and the data provenance

The document does not specify a "test set" in the context of clinical data or human subjects. The testing described is bench testing for functional, integrity, and biocompatibility. Therefore, details about sample size (in terms of human subjects or cases) and data provenance (country, retrospective/prospective) are not applicable or provided for this type of testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not applicable as the "test set" was not based on expert-adjudicated clinical cases. The ground truth for bench testing is typically defined by engineering standards, specifications, and established biocompatibility protocols.

4. Adjudication method for the test set

This information is not applicable as there was no clinical test set requiring adjudication.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No, an MRMC comparative effectiveness study was not performed, nor is it relevant to this device. This device is a physical medical instrument (a balloon catheter), not an AI-powered diagnostic or decision support tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

No, a standalone algorithm performance study was not performed as this is not an AI/algorithm-based device.

7. The type of ground truth used

The "ground truth" for the safety and performance evaluation largely relied on:

  • Engineering specifications and standards: For functional and integrity bench testing.
  • International biological evaluation standards: Specifically ISO-10993 for biocompatibility.
  • Comparison to predicate devices: The "substantial equivalence" claim is a form of ground truth established by demonstrating similar performance to already approved devices.

8. The sample size for the training set

This information is not applicable as this device is not an AI/machine learning model that requires a training set.

9. How the ground truth for the training set was established

This information is not applicable as this device is not an AI/machine learning model.


Summary of the Study that Proves the Device Meets Acceptance Criteria:

The study proving the SC 35 Balloon Dilatation Catheter meets its acceptance criteria was primarily a bench testing and biocompatibility testing program.

  • Study Type: Functional and integrity bench testing, and biocompatibility testing.
  • Methodology:
    • Functional and Integrity Bench Testing: The document states that "Functional and integrity bench testing... were performed." While specific parameters (e.g., burst pressure, inflation/deflation times, guidewire compatibility) are not detailed, these tests would assess the physical properties and operational reliability of the catheter.
    • Biocompatibility Testing: This was conducted "according to the FDA guidance document, ODE Blue Book Memorandum #G95-1, May 1, 1995, Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing'." This standard outlines a series of tests to evaluate biological responses to medical devices, such as cytotoxicity, sensitization, irritation, and hemocompatibility.
  • Conclusion: The results of these tests "supported the substantial equivalence of the SC 35 Balloon Dilatation Catheter to the Ultra-thin Diamond Balloon Dilatation Catheter and the Courier ST Balloon Dilatation Catheter." This means the device demonstrated comparable safety and performance characteristics to previously approved predicate devices, satisfying the regulatory requirement for substantial equivalence for a Class II device.

The FDA's 510(k) clearance letter (K993303) confirms that based on this information, the device was deemed "substantially equivalent" and could be marketed.

{0}------------------------------------------------

SC 35 Balloon Dilatation Catheter

K993303

Page 1 of 2 September 2, 1999

SUMMARY OF SAFETY AND EFFECTIVENESS

Pursuant to §513(i)(3)(A) of the Food, Drug, and Cosmetic Act, Boston Scientific Corporation is required to submit with this Premarket Notification either an "... adequate summary of any information respecting safety and effectiveness or state that such information will be made available upon request of any person." Boston Scientific Corporation chooses to submit a summary of information respecting safety and effectiveness. According to §513(i)(3)(B), "Any summary under subparagraph (A) respecting a device shall contain detailed information regarding data concerning adverse health effects ... "

The summary regarding the adverse health effects of the proposed SC 35 Balloon Dilatation Catheter is as follows:

Trade Name:SC 35 Balloon Dilatation Catheter
Manufacturer:Boston Scientific CorporationOne Boston Scientific PlaceNatick, MA 01760
Device Generic Name:Balloon Dilatation Catheter
Classification:According to Section 13 of the Federal Food, Drug and Cosmetic Act, thedevice classification is Class II, Performance Standards.
Predicate Devices:The following device is referenced in this premarket notification as predicatedevices for the SC 35 Balloon Dilatation Catheter:Boston Scientific - Ultra-thin Diamond Dilatation CatheterThe devices mentioned above has been determined substantially equivalent byFDA.
Device Description:The proposed SC 35 Balloon Dilatation catheter is an over-the-wire catheterdesigned to be placed over guidewires which have outer diameters of .035" orsmaller.
Indications for Use:The SC 35 Balloon Dilatation Catheter is indicated for stentdeployment/optimization of the J&J Palmaz Biliary Stent and for the treatmentof biliary strictures.

{1}------------------------------------------------

MAR 23

SC 35 Balloon Dilatation Catheter

K993303

Page 2 of 2 September 2, 1999

Safety and Performance:

Functional and integrity bench testing and biocompatibility testing (according to the FDA guidance document, ODE Blue Book Memorandum #G95-1, May 1, 1995, Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing") were performed, and the data supported the substantial equivalence of the SC 35 Balloon Dilatation Catheter to the Ultra-thin Diamond Balloon Dilatation Catheter and the Courier ST Balloon Dilatation Catheter.

Conclusion:

Based on the Indication for Use, technological characteristics and safety and performance testing, the SC 35 Balloon Dilatation Catheter has been shown to be safe and effective for its intended use.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem that appears to be a stylized bird or eagle with three wing-like shapes.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR 2 3 2000

Ms. Terry A. McGovern Senior Regulatory Affairs Specialist Boston Scientific Corporation MEDI-TECH One Boston Scientific Place Natick, MA 01760-1537

Re: K993303 Semi Compliant (SC) 35 Balloon Dilatation Catheter Dated: January 6, 2000 Received: January 11, 2000 Regulatory Class: II 21 CFR §876.5010/Procode: 78 FGE

Dear Ms. McGovern:

We have reviewed your Section 510(k) notification of intent to market the device and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce pror to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act (Act, You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class III (Premarket Aproval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you micht have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predication for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compiance at (301) 594-4591. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-5597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours.

Daniel G. Schultz, M.D. Captain, USPHS Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure(s)

{3}------------------------------------------------

510(k) Number (if known):__K993303

Device Name:__Semi Compliant 35 Balloon Dilatation Catheter

Indications For Use:

The Semi Compliant 35 Balloon Dilatation Catheter is indicated for stent deployment/optimization of the J&J Palmaz Biliary Stent and for the treatment of biliary strictures.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Optional Format 3-10-98)

. '

H

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number_1993303

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.