K Number
K993208
Date Cleared
2000-04-03

(192 days)

Product Code
Regulation Number
862.3610
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The STC Methamphetamine Intercept™ MICRO-PLATE EIA is intended for use in the qualitative determination of methamphetamine in oral fluid collected with the Intercept™ Drugs of Abuse (DOA) Oral Specimen Collection Device. For In Vitro Diagnostic Use.

Device Description

The STC Methamphetamine Intercept™ MICRO-PLATE EIA is a competitive micro-plate immunoassay for the detection of methamphetamine in oral fluid collected with the Intercept™ DOA Oral Specimen Collection Device. Specimen or standard is added to an EIA well in combination with an enzyme-labeled hapten derivative. In an EIA well containing an oral fluid specimen positive for methamphetamine, there is a competition between the drug and the enzyme-labeled hapten to bind the antibody fixed onto the EIA wells are then washed, substrate is added, and color is produced. The absorbance measured for each well at 450 mm is inversely proportional to the amount of methamphetamine present in the specimen or calibrator/control.

AI/ML Overview

Here's a breakdown of the acceptance criteria and the study details for the STC Methamphetamine Intercept™ MICRO-PLATE EIA, based on the provided text:

Acceptance Criteria and Device Performance

Acceptance Criteria CategorySpecific Criteria/MetricReported Device Performance (STC Methamphetamine Intercept™ MICRO-PLATE EIA)
Analytical Sensitivity/Limit of Detection (LOD)LOD (ng/mL)8.0 ng/mL
PrecisionIntra-Assay %CV Range6.2 - 7.8%
Inter-Assay %CV Range7.5 - 8.4%
Analytical Specificity/Cross-Reactivity% Cross-Reactivity for d-Methamphetamine100% (used as kit standard)
% Cross-Reactivity for MDMA (Ecstasy)288.471% (at 12.5 ng/mL spiked concentration)
Absence of signal for a defined list of common substances (e.g., Acetylsalicylic Acid, Cocaine, Ibuprofen, Morphine) at 10,000 µg/mLNone of the tested compounds produced a signal less than or equal to the Oral Fluid Cutoff Calibrator.
Clinical Performance (Agreement with GC/MS)% Agreement with Oral Fluid GC/MS for positive and negative samples98% agreement (24 positive, 25 negative, 0 false positive, 1 false negative)
Clinical Performance (Agreement with Urine Assay)% Agreement with Urine Assay (EIA and GC/MS) for positive and negative samples94% agreement (20 positive, 54 negative, 3 false positive, 2 false negative)
Sample pH EffectNo false positives at specific pH rangesFalse positives at pH ≤ 6.0
Effect of Common MaterialsIdentified interfering substancesOrange Juice (pH effect), Cough Syrup (Pseudoephedrine), Antiseptic (pH effect)

Study Information

2. Sample Size and Data Provenance for the Test Set:
* Sample Size: A total of 80 matching oral fluid and urine specimens were collected. An additional 6 random oral fluid samples were screened and confirmed for methamphetamine.
* Data Provenance: The specimens were collected from "known or suspected drug abusers." The country of origin is not explicitly stated, but given the US FDA regulatory submission, it is likely the study was conducted in the United States. The data appears to be retrospective, as specimens were collected and then tested.

3. Number of Experts and Qualifications for Ground Truth - Test Set:
* The document does not specify the number of experts used to establish the ground truth or their qualifications.
* Ground truth was established by GC/MS (Gas Chromatography/Mass Spectrometry), which is a gold standard analytical method, not by human expert consensus or interpretation of images. Therefore, the concept of "experts" in the context of visual interpretation of results is not directly applicable here for establishing the ground truth of methamphetamine presence. The experts involved would be laboratory personnel skilled in GC/MS operation and data interpretation.

4. Adjudication Method for the Test Set:
* The document does not describe an adjudication method in the traditional sense of resolving discrepancies between human readers. For the clinical performance, samples that tested positive by EIA and 20 samples that tested negative were confirmed by GC/MS. This acts as a confirmation step rather than an adjudication method between multiple interpretations of the device's output.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:
* No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical and clinical performance of the device itself against established methods (GC/MS, predicate urine assay), not on how human readers' performance improves with or without the device's assistance.

6. Standalone Performance Study:
* Yes, a standalone performance study was done. The entire clinical and analytical evaluation describes the performance of the "STC Methamphetamine Intercept™ MICRO-PLATE EIA" as a standalone device. The "clinical data" section directly compares the device's results to GC/MS for oral fluid and a commercial immunoassay/GC/MS for urine samples, demonstrating the algorithm/device's performance without specific human interpretation or intervention in the primary measurement.

7. Type of Ground Truth Used:
* The primary ground truth for both oral fluid and urine samples was established by GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate chemical analytical method, often considered a definitive method for drug detection.

8. Sample Size for the Training Set:
* The document does not explicitly state a training set size or describe a distinct training phase for the device's development. This is typical for immunoassay devices, where the "training" analogous to machine learning models is inherent in the assay design and optimization process (e.g., antibody selection, reagent concentrations, cutoff determination) rather than a separate data-driven training set in the AI sense.

9. How Ground Truth for the Training Set Was Established:
* Since no explicit training set or AI-like "training" is described, the question of how its ground truth was established is not applicable in the context of this document. The assay's performance characteristics (e.g., cutoff) were determined through analytical studies rather than by training on a labeled dataset. The recommended cutoff of 40 ng/mL was explicitly stated to be "within the limit of detection by the STC Methamphetamine Intercept™ MICRO-PLATE EIA," implying it was determined through internal analytical validation.

{0}------------------------------------------------

PREMARKET NOTIFICATION 510(k) SUMMARY (As Required By 21 CFR 807.92)

807.92 (a):

1.Submitter's Name:STC Technologies, Inc.
Address:1745 Eaton Avenue, Bethlehem, PA 18018
Telephone Number:(610) 882-1820
Contact Person:R. Sam Niedbala, Ph.D., BCFE
Date Prepared:02/10/00
    1. Device Name: Methamphetamine Intercept™ MICRO-PLATE EIA Proprietary Name: Methamphetamine Intercept™ System Usual Name: Enzyme Immunoassay, Amphetamines Classification Name:
  • Device to Which Substantial Equivalence Is Claimed. 3. Roche Diagnostic Systems, Abuscreen ONLINE® kit for amphetamines (urine); K933052
  • Description of Device: 4.

Principle of the Assay

The STC Methamphetamine Intercept™ MICRO-PLATE EIA is a competitive micro-plate immunoassay for the detection of methamphetamine in oral fluid collected with the Intercept™ DOA Oral Specimen Collection Device. Specimen or standard is added to an EIA well in combination with an enzyme-labeled hapten derivative. In an EIA well containing an oral fluid specimen positive for methamphetamine, there is a competition between the drug and the enzyme-labeled hapten to bind the antibody fixed onto the EIA wells are then washed, substrate is added, and color is produced. The absorbance measured for each well at 450 mm is inversely proportional to the amount of methamphetamine present in the specimen or calibrator/control. Because currently there are no SAMHSA assigned cutoffs for methamphetamine testing using oral fluid, STC recommends a cutoff of 40 ng/mL when testing oral fluid collected with the Intercept™ DOA Oral Specimen Collection Device. This cutoff is within the limit of detection by the STC Methamphetamine Intercept™ MICRO-PLATE EIA.

Anti-Methamphetamine Coated Plate -- Rabbit anti-methamphetamine polyclonal antibody immobilized on a polystyrene plate supplied in dry form.

Methamphetamine Enzyme Conjugate -- Horseradish peroxidase labeled with a

methamphetamine hapten diluted in a protein matrix of bovine serum with protein stabilizers.

Substrate Reagent -- One bottle containing 3,3', 5,5' tetramethylbenzidine.

Stopping Reagent -- Each bottle contains 2 N sulfuric acid.

Oral Fluid Negative Calibrator -- Oral Fluid Diluent negative for methamphetamine.

Oral Fluid Negative Control - Oral Fluid Diluent containing 20 ng/mL (v/v) methamphetamine.

Oral Fluid Cutoff Calibrator -- Oral Fluid Diluent containing 40 ng/mL (v/v) methainphetamine.

Oral Fluid Positive Control - Oral Fluid Diluent containing 80 ng/mL (v/v) methamphetamine.

Principle of the Intercept™ DOA Oral Specimen Collection Device

Saliva is a complex mixture of parotid, submandibular, sublingual and minor salivary gland secretions mixed with mucin, bacteria, leukocytes, sloughed epithelial cells and gingival crevicular fluid. The fact that methamphetamine is present in oral fluid following human use is well documented.

{1}------------------------------------------------

The Intercent™ DOA Oral Fluid Collection Device was developed for the purpose of collecting oral fluid for diagnostic testing. The collection device consists of a treated absorbent cotton fiber pad affixed to a nylon stick (Collection Pad) and a preservative solution in a plastic container (Specimen Vial). The Collection Pad is impregnated with a mixture of common salts and gelatin which creates a hypertonic environment and an increased osmotic pressure wherever it contacts oral mucosal cells. The pad is placed in contact with the gingival mucosa (between the lower gum and cheek) which enhances the flow of mucosal transudate across the mucosal surfaces onto the absorptive cotton fibers of the pad. Following the collection period, the Collection Pad is placed into a vial containing a preservative solution which serves to inhibit the growth of oral micro-organisms recovered on the Collection Pad. The vial is sealed with a plastic cap and transported to a laboratory for processing and testing. Following processing, a fluid containing a mixture of saliva components and the preservative solution is recovered which is suitable for testing for the presence of methamphetamine in the Methamphetamine Intercept™ MICRO-PLATE EIA manufactured by STC Technologies, Bethlehem, PA. Refer to the Intercept™ DOA Oral Specimen Collection Device product insert for specific instructions on the proper collection, handling, and adequacy of oral fluid samples.

  • Intended Use Statement: રે.
    The STC Methamphetamine Intercept" MICRO-PLATE EIA is intended for use by clinical laboratories in the qualitative determination of methamphetamine in oral fluid collected with the Intercept™ DOA Oral Specimen Collection Device. For In Vitro Diagnostic Use.

    1. Summary of Technological Characteristics:
      The STC Methamphetamine Intercopt™ MICRO-PLATE EIA is based on the principle of solid phase competitive enzyme immunoassay . This application is for the STC Methamphetamine EIA as a screening tool for the detection of methamphetamine using specimens collected with the Intercept™ DOA Oral Specimen Collection Device manufactured by Epitope, Inc., Beaverton, Oregon.

807.92 (b):

  • Non Clinical Data: .

Analytical Sensitivity/Limit Of Detection - The Limit of Detection (LOD) was defined from the signal-to-noise ratio at the zero-drug concentration as the mean zero absorbance (An) minus the noise times three (LOD = An - 3SD). The LOD was determined by obtaining the average absorbance value for 80 readings of Oral Fluid Diluent and calculating the standard deviation (SD) and 3SD of the absorbance. The absorbance value minus 3SD was then extrapolated from the curve and represents the sensitivity of the assay. The LOD was calculated to be 8.0 ng/mL.

Precision - The precision of the STC Methamphetamine Intercept™ MICRO-PLATE EIA was assessed by testing Oral Fluid Diluent containing 0, 20, 40, 60 and 80 ng/mL methamphetamine. The intra-assay precision was determined by analyzing each level 16 times per run for 4 runs. Inter-assay precision was determined by analyzing 2 samples at each level twice per day for 20 days. The results of this testing are described in the following table:

METHAMPHETAMINE(ng/mL)MEAN(O.D.)MEAN(%CV)INTER-ASSAY(n=4/Bay, 20 days)
02.0567.87.5
200.9787.07.7
400.7106.27.9
600.5547.87.5
800.4806.48.4

Analytical Specificity/Cross-Reactivity - The analytical specificity of an immunoassay is defined as the cross-reactivity of substances in the assay which are structurally related to the target compound. The percent cross-reactivity of a compound in the STC Methamphetamine Intercept™ MICRO-

{2}------------------------------------------------

PLATE EIA is defined as the apparent methamphetamine concentration divided by the spiked concentration times 100.

The cross-reactivities of structurally related compounds were calculated at several spiked concentrations in Oral Fluid Diluent. The following table indicates the apparent concentration of methamphetamine for each substance at a concentration which cross-reacted in the assay. Note: D-Methamphetamine was used as the kit standard and, therefore, will exhibit 100% cross-reactivity.

β-Phenethylamine10,0007.9950.080
Diphenhydramine10,0006.5030.065
D-Amphetamine5,00032.1600.643
Fenfluramine10025.91725.917
Isoxsuprine10,0005.9940.060
1-Ephedrine2,50019.2920.772
1-Methamphetamine1,0002.8390.284
1-Phenylalanine10,0005.3430.053
MDA5,00033.8170.676
MDMA12.536.059288.471
Mephentermine2,50031.3611.254
Phentermine10,00015.9350.159
Phenylephrine10,00013.2990.133
Phenylpropanolamine10,0009.7900.098
Procaine10,0009.9170.099
Pseudoephedrine5,00028.7480.575
  • out of range: not detectable

The user should be aware that the determination of methamphetamine equivalents for each compound is only to calculate the % cross-reactivity of these compounds in the assay. For many of these compounds, the absorbance readings obtained were below the limit of detection of 8.0 ng/mL for the assay. As a result, the % cross-reactivities for these compounds at the levels tested are considered estimates only.

The following compounds were spiked into Oral Fluid Control Matrix at a target concentration of 10,000 ug/mL and tested for cross-reactivity. None were found to produce a signal less than or equal to that of the Oral Fluid Cutoff Calibrator.

Acetylsalicylic AcidCocaethyleneIbuprofenNordiazepam
AlprazolamCocaineImipraminePenicillin
AmobarbitalCodeineLidocainePentobarbital
AmpicillinCotinineMedazepamPhencyclidine
BenzoylecgonineDextromethorphanMeperidinePhenobarbital
ButabarbitalDiacetylmorphineMethadoneProcainamide
ButalbitalFenoprofenMetoprololQuinidine
CaffeineGemfibrozilMorphineTemazepam
ChlordiazepoxideGentisic AcidNalorphineTheophylline
ChlorpromazineGlipizideNaproxenΔ9-THC
ClonazepamHydrocodoneNiacinamideZomepirac
ClorazepateHydromorphoneNorchlordiazepoxide

It is possible that other substances and/or factors not listed above may interfere with the test and cause false results, e.g., technical or procedural errors.

2. Clinical Data:

A total of 80 matching oral fluid and urine specimens were collected from known or suspected drug abusers. An additional six (6) random oral fluid samples were screened and confirmed for the presence of methamphetamine. All oral fluid specimens were tested using the STC Methamphetamine Intercept™ MICRO-PLATE EIA. All samples that tested positive and 20 samples

{3}------------------------------------------------

that tested negative by EIA were confirmed by GC/MS. All urine samples were screened using a commercial immunoassay kit for amphetamines. Positive samples were confirmed using GC/MS.

For oral fluid testing, a 40 ng/mL cutoff was used for EIA and GC/MS. For urine testing, the cutoffs were 1,000 ng/mL and 500 ng/mL respectively, for the initial screen and GC/MS confirmation based on SAMHSA guidelines.(1) The % agreement of the STC EIA results to the urine EIA and GC/MS results are shown below:

Image /page/3/Figure/2 description: The image shows two tables comparing STC Oral Fluid EIA results with Urine Assay and Oral Fluid GC/MS results. In the first table, with Urine Assay, the agreement is 94%, with values of 20 and 54 for matching positive and negative results, respectively, and 3 and 2 for mismatched results. The second table compares STC Oral Fluid EIA with Oral Fluid GC/MS, showing a 98% agreement, with 24 and 25 for matching positive and negative results, respectively, and 0 and 1 for mismatched results.

3. Conclusions:

A comparison of the performance data for the new device vs. the predicate device is given below:

1. Limit of Detection

a managar mana mana mana mana mana da maga cagama a mana mana mana mana mana mana mana mana mana man
STC Methamphetamine MICRO-PLATE EIA8.0 ng/mL
Roche Abuscreen ONLINE®< 30 ng/mL

2. Precision

TestIntra-Assay %CV RangeInter-Assay %CV Range
STC Methamphetamine MICRO-PLATE EIA6.2 - 7.87.5 - 8.4
Roche Abuscreen ONLINE®3 - 64 - 7

3. Sample pH Effect

and Holen Research Comments of the Property of Child Property of Children
I STC Methamphetamine MICRO-PLATE EIAFalse positives at pH ≤ 6.0
Roche Abuscreen ONLINE®Concession of Concell College of Children Company of Concession, A Location A Lasteroped ofCompany of the country of the country of the country ofNot Tested

4. Effect of Common Materials

AssayAdulterants Which Affect Assay
STC Methamphetamine MICRO-PLATE EIAOrange Juice (pH effect)Cough Syrup (Pseudoephedrine)Antiseptic (pH effect)
Roche Abuscreen ONLINE®Not Tested

{4}------------------------------------------------

5. Cross-Reactivity

% Cross Reactivity
CompoundOral FluidUrine
β-Phenethylamine0.0802
3-HydroxytyramineNot Tested<0.2
d,l-EphedrineNot Tested<0.2
d,l-MethamphetamineNot Tested0.2
d-EphedrineNot Tested<0.2
d-Methamphetamine1000.5
d-PhenylpropanolamineNot Tested0.2
d-PseudoephedrineNot Tested<0.2
Diphenhydramine0.065Not Tested
dl-AmphetamineNot Tested51
Fenfluramine25.917Not Tested
Isoxsuprine0.060Not Tested
l-AmphetamineNot Tested2
l-Ephedrine0.772<0.2
l-Methamphetamine0.2840.2
l-NorpseudoephedrineNot Tested<0.2
l-Phenylalanine0.053Not Tested
l-PhenylpropanolamineNot Tested1
l-PseudoephedrineNot Tested<0.2
MDA0.67632
MDMA288.4710.2
Mephentermine1.254<0.2
p-HydroxyamphetamineNot Tested14
p-HydroxymethamphetamineNot Tested0.3
PhendimetrazineNot Tested<0.1
Phentermine0.159<0.1
Phenylephrine0.133Not Tested
Phenylpropanolamine0.0980.7
Procaine0.099Not Tested
PropylhexidineNot Tested0.5
Pseudoephedrine0.575Not Tested
TyramineNot Tested0.3

6. References

  • (1) "Urine Testing for Drugs of Abuse", National Institute on Drug Abuse (NIDA), Research Monograph 73, 1986.
    RS- Nellik

R. Sam Niedbala, Ph.D., BCFE Chief Science Officer

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

APR 3 2000

R. Sam Niedbala, Ph.D., BCFE Executive Vice President STC Technologies, Inc. 1745 Eaton Avenue Bethlehem, Pennsylvania 18018-1799

Re: K993208

Trade Name: STC Methamphetamine-Intercept™ MICRO-PLATE EIA Regulatory Class: II Product Code: LAF Dated: February 11, 2000 Received: February 16, 2000

Dear Dr. Niedbala:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Autman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

STATEMENT OF INDICATIONS FOR USE 16993208 510(k) Number (if known):

Device Name: STC Methamphetamine Intercept™ MICRO-PLATE EIA

Indications For Use: The STC Methamphetamine Intercept™ MICRO-PLATE EIA is intended for use in the qualitative determination of methamphetamine in oral fluid collected with the Intercept™ Drugs of Abuse (DOA) Oral Specimen Collection Device. For In Vitro Diagnostic Use.

Jal Cooper

(Division Sign-Off)

Division of Clinical Laboratory Devices

510(k) Number K993208

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use ___________

c:\fda\indica.doc

§ 862.3610 Methamphetamine test system.

(a)
Identification. A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.(b)
Classification. Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).