K Number
K992937
Device Name
DENTAL IMPLANT
Date Cleared
2000-02-29

(182 days)

Product Code
Regulation Number
872.3640
Panel
Dental
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Selected Brånemark System® implant products (those identified in the preceding section, "Device Name") are intended to be placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surqical procedure.

If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchoragel can most often be obtained.

Device Description

Brånemark System® dental implants are threaded, root-form implants fabricated from ASTM grade 1 "commercially pure" titanium. They are available in diameters of 3.75 mm and 4.0 mm, and are available in lengths ranging from 10 mm to 21 mm. Research studies have demonstrated that titanium is biocompatible.

AI/ML Overview

The provided text is a 510(k) summary for Nobel Biocare's Brånemark System® Implants, seeking clearance for an immediate loading indication. This document details the device, its intended use, and a comparison to predicate devices, but does not contain any information about specific acceptance criteria or a study proving the device meets those criteria, as typically understood for AI/ML or diagnostic devices.

The 510(k) process for medical devices primarily focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving performance against specific quantitative acceptance criteria in the way a clinical trial for a drug or an AI diagnostic algorithm would.

Therefore, many of the requested items (e.g., sample size for test set, number of experts for ground truth, MRMC studies, standalone performance) are not applicable or cannot be extracted from this type of regulatory submission. The document explicitly states "We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent...".

Here's a breakdown of what can and cannot be answered based on the provided text:


1. A table of acceptance criteria and the reported device performance

  • Not applicable/Not provided. This document is a 510(k) summary for substantial equivalence, not a performance study against specific acceptance criteria. The "performance" assessment is based on demonstrating equivalence to predicate devices already cleared for immediate loading.
Acceptance CriteriaReported Device Performance
Not specified in this documentSubstantially equivalent to predicate devices for immediate loading indication.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Not applicable/Not provided. This document does not present results from a clinical test set with specific sample sizes. The regulatory submission relies on comparison to predicate devices and existing biocompatibility data for titanium.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Not applicable/Not provided. Ground truth establishment, typically for diagnostic accuracy or outcome prediction, is not described in this 510(k) summary.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Not applicable/Not provided. No test set or adjudication method is mentioned.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not applicable/Not provided. MRMC studies are typically for evaluating the impact of AI algorithms on human reader performance, which is not the subject of this 510(k) for dental implants.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable/Not provided. This document is not about an algorithm, but a physical dental implant.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Not applicable/Not provided. This document does not refer to ground truth data in the context of device performance metrics. The implicit "ground truth" for the immediate loading indication is likely established clinical practice and safety/efficacy data already accepted for the predicate devices.

8. The sample size for the training set

  • Not applicable/Not provided. This is not an AI/ML device that requires a training set.

9. How the ground truth for the training set was established

  • Not applicable/Not provided. This is not an AI/ML device.

Summary of Device Acceptance (based on 510(k) Substantial Equivalence):

The device (Brånemark System® Implants with immediate loading indication) achieved acceptance through the FDA's 510(k) clearance process by demonstrating substantial equivalence to legally marketed predicate devices, specifically the Straumann ITI Implant (K984104) and the Sargon Cylindro-Blade Implant (K930071), both of which already had an immediate load indication.

The key aspects considered for substantial equivalence, as detailed in the comparison table, included:

  • Intended Use: Functionally the same (support prosthetic devices, restore chewing function).
  • Indication: Immediate Load.
  • Design: Threaded, root-form implants (similar to Straumann ITI Implant, different but accepted from Sargon).
  • Placement Method: Single stage surgery.
  • Material: Commercially pure titanium (similar to Straumann, different but accepted from Sargon's titanium alloy).
  • Coating: None.
  • Dimensions: Within acceptable ranges relative to predicate devices.

The FDA concluded that the device is substantially equivalent to the predicate devices for the stated indications for use (immediate loading under specific conditions: at least four implants placed, splinted with a bar, predominantly in the anterior mandible, with good initial stability). This substantial equivalence determination allowed the device to be marketed.

{0}------------------------------------------------

FEB 2 9 2000

Nobel Biocare USA, Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999

K992937

Section 7 510(k) Summary

A. Manufacturer Information:

Nobel Biocare USA. Inc. Submitter's Name: Address: 22725 Savi Ranch Parkway U.S. Representative/Distributor Yorba Linda, CA 92887, USA

Contact's Name: Contact's Telephone No .: Date Prepared:

Jeff Hausheer, Ph.D., Regulatory Affairs Specialist 714-282-4800, extension 7832 June 1999

Address-Manufacturer:

Nobelpharma Production AB Dimbovagen 2 Karlskoga S-691-51, Sweden Manufacturer Registration Number: 9611993

B. Device Name:

Common Name:

Dental Implant

Trade Name:

Brånemark System® Implants Indicated For Loading Immediately Following Implant Placement (see table below) (i.e., within 2 to 3 weeks following implant placement)

Brånemark System®: Implant ProductFamily NameDimensions: Diameter &Range of Lengths (mm)
Standard Series Implants (3.75)3.75 x 10-20 mm
Standard Series Implants (4.0)4.0 x 10-18 mm
Mk II Self-Tapping Implants (3.75)3.75 x 10-18 mm
Mk II Self-Tapping Implants (4.0)4.0 x 10-18 mm
Self-Tapping Implants3.75 x 10-18 mm
Conical Self-Tapping Implants3.75 x 13-21 mm
Mk IV Self-Tapping Implants4.0 x 10-18 mm

Classification:

Classification Name:Endosseous Dental Implant
Classification Number:DZE
Classification Citation:21 CFR 872.3640

Page 1 of 3

{1}------------------------------------------------

Section 7 510(k) Summary (continued)

C. Device Description:

Description: Brånemark System® dental implants are threaded, root-form implants fabricated from ASTM grade 1 "commercially pure" titanium. They are available in diameters of 3.75 mm and 4.0 mm, and are available in lengths ranging from 10 mm to 21 mm. Research studies have demonstrated that titanium is biocompatible.

D. Intended Use:

Indications For Use: Selected Brånemark System® implant products (those identified in Table 7.1, below) are intended to be placed in the upper or lower jaw to support prosthetic devices. such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surgical procedure.

If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchorage, can most often be obtained.

Brånemark System®Implant Product FamilyDiameterRange of Lengths510(k)
Mk II Self-Tapping3.75 mm10-18 mmK925762
Mk II Self-Tapping4.0 mm4.0 x 10-18 mmK945398
Standard Series3.75 mm3.75 x 10-20 mmK925765
Standard Series4.0 mm4.0 x 10-18 mmK925764
Self-Tapping3.75 mm3.75 x 10-18 mmK925762
Conical Self-Tapping3.75 mm3.75 x 13-21 mmK925760
Mk IV Self-Tapping4.0 mm4.0 x 10-18 mmK974828
Table 7.1
510(k) Notification:
Brånemark System® Implant Products Indicated For Immediate Loading

{2}------------------------------------------------

Nobel Biocare USA, Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999

Section 7: 510(k) Summary (continued)

E. Comparison to the Predicate Device(s): .

Comparison Of The Predicate Devices And The Branemark System® Implant Products For Which Clearance Of An Immediate Load Indication Is Sought

Predicate ProductPredicate ProductSubmitted Product
CharacteristicStraumann ITI Implant (510(k)K984104)Sargon Cylindro-Blade Implant [510(k)K930071]Brånemark System® Implant Products**
Intended Useintended to be placed in themaxillary and/or mandibulararches to support prostheticrestorations in edentulous orpartially edentulous patients.Intended to act as a replacement formissing teeth by providing the means forfixation of dentures, removable bridgework,or prosthetic teeth.(Functionally, the same): intended to be placedin the upper or lower jaw to support prostheticdevices such as artificial teeth, and to restorepatient's chewing function.
IndicationImmediate LoadImmediate LoadSame
Designthreaded, root-form implanta threaded, tapered, non-solid (hollow),flanged root form implantthreaded, root-form implants
Placement MethodSingle stage surgerySingle stage surgerySame
MaterialCommercially pure titaniumTitanium alloyCommercially pure titanium
CoatingNoneNoneNone
Length (mm, min.-max.)Unknown10mm to 18 mm10mm to 21 mm
Diameter (mm)UnknownAvailable in 1 tapered design only, ranAvailable in one tapered design only; itsdiameter ranges from 3.8 mm to 4.1ranging from 3.8 mm to 4.1 mm3.75 mm and 4.0 mm
Precautions &WarningsUnknownNoneNone
PackagingUnknownVialGlass ampoule in peel-open blister pack
Provided Sterile?UnknownUnknownYes (dry heat = glass ampoule; steam = blisterpack)

= See Table 2.A.1 for identification, dimensions, and 510(k) numbers of each product

Page 3 of 3

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling three overlapping, curved shapes that could be interpreted as abstract representations of human figures or flowing elements.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

FEB 2 9 2000

Dr. Jeff Hausheer Requlatory Affairs Specialist Nobel Biocare USA, Inc. 22895 Eastpark Drive Yorba Linda, CA 92887

Re: K992937 Brånemark System® Implants Trade Name: Requlatory Class: III Product Code: DZE Dated: December 2, 1999 Received: December 6, 1999

Dear Dr. Hausheer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, qood manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of

{4}------------------------------------------------

Page 2 -Dr. Hausheer

the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4690. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Patricio Cavente/for

Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

Nobel Biocare USA. Inc. 510(k) Notification: Immediate Loading Indication June 21, 1999

Section 10 Indications for Use

Page 1 of 1

50

510(k) Number (if known): K ------

Device Name:

Selected Brånemark System® Implant Products:

Mk II Self-TappingK925762
Mk II Self-TappingK945398
Standard SeriesK925765
Standard SeriesK925764
Self-TappingK925762
Conical Self-TappingK925760
Mk IV Self-TappingK974828

Indications For Use: Selected Brånemark System® implant products (those identified in the preceding section, "Device Name") are intended to be placed in the upper or lower jaw to support prosthetic devices, such as artificial teeth, and to restore a patient's chewing function. This may be accomplished using either a two stage surgical procedure or a single stage surqical procedure.

If a single stage surgical procedure is used, these implants may be loaded immediately following insertion - PROVIDED - at least four implants are placed, and are splinted with a bar. These implants must be placed predominantly in the anterior mandible (between the mental foramina) where good initial stability of the implants, with or without bi-cortical anchoragel can most often be obtained.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) n of Dental, Infection
Numbers Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use __ L OR Over-The-Counter Use (Per 21 CFR 801.109) (Optional Format 1-2-96)

§ 872.3640 Endosseous dental implant.

(a)
Identification. An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.(b)
Classification. (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
Classification. Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.