K Number
K992452
Date Cleared
1999-10-18

(88 days)

Product Code
Regulation Number
862.3100
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

'RapidOne' - Amphetamine Test is a one-step, lateral flow immunoassay for the detection of amphetamine in urine. 'RapidOne' - Amphetamine Test is intended for use in the qualitative detection of amphetamine in human urine at 1000 ng/ml. 'RapidOne' - Amphetamine Test is intended for professional use. It is not intended for over the counter sale to non-professionals. The assay is easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS). 'RapidOne' - Amphetamine Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

Device Description

The assay employed in the 'RapidOne' - Amphevamine Test is based on the same principle of highly specific reaction between antigens and antibodies. This assay is a one-step, immunoassay in which a specially labeled drug (drug conjugate) competes with drug which may be present in the sample for the limited number of binding sites on the antibody. The test device consists of a membrane strip onto which anti-amphetamine monoclonal antibody has been immobilized. A colloidal gold-BSA-amphetamine couplex is dried on a reagent pad. In the absence of any drug in the urine sample, the colloidal gold-complex moves with the urine by capillary action to conta the immobilized drug antibody. An autibody-antigen reaction occurs forming a visible line in the 'test' area. The formation of a visible line in the test area occurs when the test is negative. When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody sites on the volloidal gold-antibody complex. If sufficient amount of drug is present, it will fill all of the available binding sites, thus preventing attachment of the labeled antibody to the drug conjugate. An absence of a color band (line) in the test area is indicative of a positive result. A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence of drug in the urine, and therefore, should be present in all reactions. A negative urine will produce two colored bands, and a positive sample will produce only one band.

AI/ML Overview

The acceptance criteria and the study proving the device meets them are described below for the 'RapidOne'-Amphetamine Test.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Intended Use/Performance Claim)Reported Device Performance (Test Set)
Qualitative detection of d-amphetamine in human urine.Compared to EMIT II (predicate device) and confirmed by GC/MS.
Detect 1000 ng/ml of d-amphetamine in urine.At 1000 ng/ml, 40/40 (100%) samples tested positive.
Specificity: No false positives in drug-free urine.50 drug-free samples correctly identified as negative (100% negative agreement).
Sensitivity: Detection of amphetamine in positive samples across a range of concentrations.40 drug-containing specimens (750 ng/ml to 58,000 ng/ml) correctly identified as positive (100% positive agreement).
Reproducibility: Consistent results across tests.Reproducibility study showed consistent results: 40/40 negative at 0 ng/ml, 40/40 positive at 1000 ng/ml and 1250 ng/ml. Partial positives at 500 ng/ml (5/40) and 750 ng/ml (30/40) for samples below the cut-off.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 90 samples were used for the comparative effectiveness study: 50 drug-free, and 40 drug-containing. For the reproducibility study, 40 samples were tested at each of five concentration levels (0, 500, 750, 1000, 1250 ng/ml), with each tested 4 times (twice daily for 5 days), totaling 200 individual tests for reproducibility.
  • Data Provenance: Not specified in the provided text (e.g., country of origin, retrospective or prospective).

3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

Not applicable. The ground truth for the test set was established by established laboratory methods (EMIT II and GC/MS), not by human experts in this context.

4. Adjudication Method for the Test Set

Not applicable. The ground truth was established by laboratory instrumentation (EMIT II and GC/MS), not requiring human adjudication.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

Not applicable. This device is a qualitative immunoassay for drug detection. The study compares the device's performance against a predicate device and a gold standard laboratory method (GC/MS), not human readers.

6. Standalone Performance

Yes, a standalone performance study was conducted. The 'RapidOne'-Amphetamine Test was evaluated independently against 90 samples (50 drug-free, 40 drug-containing) and its results were compared to those of the Syva EMIT-II (predicate device) and confirmed by GC/MS. A reproducibility study also assessed the standalone performance across various concentrations.

7. Type of Ground Truth Used

The primary ground truth for the comparative study was Gas Chromatography/Mass Spectrometry (GC/MS), which is considered a confirmatory method for drug testing and provides quantitative results. The predicate device (EMIT II) was also used for initial comparison.

8. Sample Size for the Training Set

The provided summary does not mention a training set sample size or details about a training process. This type of immunoassay device typically does not involve a "training set" in the same way machine learning models do. Its sensitivity and specificity are determined through the properties of the antibodies and reagents, which are developed and validated during the product development process, rather than "trained" on a specific dataset.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as a distinct "training set" and associated ground truth establishment for a machine learning model is not described for this immunoassay device. The device's inherent analytical performance is based on its chemical and biological components.

{0}------------------------------------------------

K992452

510(k) Summary

Submitter's Name/Address: American Bio Medica Corporation 122 Smith Road Kinderhook, N.Y. 12106

OCT 18 1999

Contact Person: Henry Wells Vice President of Product Development Phone: 518-758-8158 Fax: 518-758-8171

Date of Preparation of this Summary:

September 24, 1999

Device Trade of Proprietary Name:

Device Common/Usual Name or Classification Name:

'RapidOne'-Amphetamine Test

Amphetamine test system

Classification Number/Class

21 C.F.R. § 862.3650 (Class II)

This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92.

The assigned 510(k) is:

Predicate Device: Forefro Amphetamine Test (510(k) No

Test Description:

The assay employed in the 'RapidOne' - Amphevamine Test is based on the same principle of highly specific reaction between antigens and antibodies.

This assay is a one-step, immunoassay in which a specially labeled drug (drug conjugate) competes with drug which may be present in the sample for the limited number of binding sites on the antibody. The test device consists of a membrane strip onto which anti-amphetamine monoclonal antibody has been immobilized. A colloidal gold-BSAamphetamine couplex is dried on a reagent pad. In the absence of any drug in the urine sample, the colloidal gold-complex moves with the urine by capillary action to conta the immobilized drug antibody. An autibody-antigen reaction occurs forming a visible line in the 'test' area. The formation of a visible line in the test area occurs when the test is negative.

When drug is present in the urine sample, the drug or metabolite will compete with the immobilized drug conjugate in the test area for the limited antibody sites on the volloidal gold-antibody complex. If sufficient amount of drug is present, it will fill all of the available binding sites, thus preventing attachment of the labeled antibody to the drug

{1}------------------------------------------------

conjugate. An absence of a color band (line) in the test area is indicative of a positive result.

A control band (line), comprised of a different antibody/antigen reaction, is present on the membrane strip. The control line is not influenced by the presence of drug in the urine, and therefore, should be present in all reactions.

A negative urine will produce two colored bands, and a positive sample will produce only one band.

Intended Use:

'RapidOne' - Amphetamine Test is used for the qualitative detection of d-amphetamine in human urine. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas chromatography/mass spectrometry (GC/MS).

Performance Characteristics:

'RapidOne'-Amphetamine Test will detect 1000 ng/ml of d-amphetamine in urine.

'RapidOne'-Amphetamine Test was compared to EMIT II. Ninety (90) samples were selected for evaluation, fifty (50) of which were found to be drug-free and forty (40) tested as positive by Syva EMIT-II. The forty positive specimens were confirmed and quantified by GC/MS. Both immunoassays correctly identified all of the specimens which contained no drug as negative and determined the 40 drug-containing specimens, ranging in concentration of 750 ng/ml to 58,000 ng/ml, to be positive.

Reproducibility was evaluated using control urines containing concentrations above and below the stated cut-off. Negative urines were also used. Each sample, at each concentration of analyte, was tested 4 times, twice daily, for 5 days,

Conc. (ng/ml)#Result
04040/40 neg
500405/40 pos
7504030/40 pos
10004040/40 pos
12504040/40 pos

Conclusion:

'RapidOne'-Amphetamine Test is substantially equivalent to 'Instacheck' - Drug Screen -Amphetamine Test

{2}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

OCT 18 1999

American Bio Medica c/o Mr. John B. Dubeck KELLER AND HECKMAN LLP 1001 G Street, N.W. Suite 500 West Washington, D.C. 20001

Re: K992452

Trade Name: 'Rapid One' - Amphetamine Test Regulatory Class: II Product Code: DKZ, DOD Dated: July 22, 1999 Received: July 22, 1999

Dear Mr. Dubeck:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrlvdsma/dsmamain.html".

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K993452

'RapidOne' - Amphetamine Test Device Name:

Indications For Use:

'RapidOne' - Amphetamine Test is a one-step, lateral flow immunoassay for the detection of amphetamine in urine. 'RapidOne' - Amphetamine Test is intended for use in the qualitative detection of amphetamine in human urine at 1000 ng/ml.

'RapidOne' - Amphetamine Test is intended for professional use. It is not intended for over the counter sale to non-professionals. The assay is easy to perform, but should not be used without proper supervision. This immunoassay is a simplified qualitative screening method that provides only a preliminary result for use in determining the need for additional or confirmatory testing, i.e., gas-chromatography/mass spectrometry (GC/MS).

'RapidOne' - Amphetamine Test provides only a preliminary analytical test result. A more specific alternate chemical method must be used in order to obtain a more confirmed analytical result. Gas chromatography/mass spectrometry (GC/MS) is the preferred confirmatory method. Clinical and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

tan cooper

Tivision Sign-O---------------------------------------------------------------------------------------------------------------------------------------------------------------510(k) Numbe

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription_Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

§ 862.3100 Amphetamine test system.

(a)
Identification. An amphetamine test system is a device intended to measure amphetamine, a central nervous system stimulating drug, in plasma and urine. Measurements obtained by this device are used in the diagnosis and treatment of amphetamine use or overdose and in monitoring levels of amphetamine to ensure appropriate therapy.(b)
Classification. Class II (special controls). An amphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).