K Number
K992405

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-09-22

(64 days)

Product Code
Regulation Number
878.5010
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include:

Shoulder procedures(e.g. Rotator cuff repair)
Knee procedures(e.g. Patellar ligament / tendon avulsion repair)
Ankle procedures(e.g. Achilles tendon reconstruction and repair)
Foot procedures(e.g. Hallux valgus reconstruction)
Elbow procedures(e.g. Biceps tendon reattachment)
Bladder neck suspension(female urinary incontinence due to urethral hypermobility)

The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures.

Device Description

The Axya Bone Anchor System consists of an electronic control module and a reusable welding activator with an integral connector cable. The Kit (BAK) consists of a disposable suture welding sleeve packaged together with appropriate and currently approved bone anchors and USP polypropylene monofilament suture. The Kit is intended for use with the reusable system components in order to complete the suture welding process for the attachment of soft tissues to bone anchors in surgical procedures where bone anchors are indicated

AI/ML Overview

The Axya Bone Anchor System and Kit (BAK) is a Class II device intended to secure soft tissues to bone anchors in surgical procedures. The device uses an ultrasonic suture welding process.

Here's an analysis of the acceptance criteria and supporting study information:

1. Acceptance Criteria and Reported Device Performance

The provided document does not explicitly state specific numerical acceptance criteria for the device's performance. Instead, it focuses on demonstrating that the BAK is substantially equivalent to predicate devices and that its performance (specifically suture strength and safety) meets established standards and is comparable to traditional methods.

The reported device performance is presented as:

Acceptance Criteria (Inferred from study objectives)Reported Device Performance
Suture Knot Strength: Ensure sutures placed by the welding process exhibit strength comparable to or exceeding USP requirements for respective sizes and types of suture materials.- "Knot strength" characteristics substantially above the USP requirements for the respective sizes and types of suture materials tested. - USP size 2 polypropylene monofilament sutures secured by a weld formed with the AxyaWeld Tips exceed the USP knot strength requirements. - Strength requirements are met regardless of the angle of applied tension between the suture and the bone anchor stem.
Suture Fatigue and Elongation: Demonstrate acceptable long-term performance under cyclic loading.- Evaluated in bench-test models. - Welded sutures found to meet suggested criteria following cyclic testing.
Safety - Histopathologic Change: Confirm no significant difference in safety or efficacy compared to traditional suture placement, specifically regarding tissue reactions to the weld.- No histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy in both dorsal skin and bowel tissue of New Zealand rabbits. - Conclusion: No significant difference in safety or efficacy between traditional manual knot-tying and suture welding.
Safety - Thermal Injury: Ensure no risk of thermal injury to the patient from the ultrasonic energy source.- No portion of the ultrasonic generator comes into contact with human tissues due to design features. - Virtually no risk of causing a thermal injury to the patient.
Sterility: Achieve a specified sterility assurance level (SAL).- Sterilized by exposure to gamma irradiation at 2.5 Mrads, providing a SAL of at least 10^-6. - Alternatively, by ethylene oxide gas sterilization, also providing a SAL of 10^-6. - Sterility processes, manufacturing process, and packaging process are validated by the manufacturer.
Biocompatibility: Fabricated from materials with a substantial history of use in medical devices.- Materials from which the Axya device is fabricated have an established history of use in medical applications. - Specific materials used by Axya have been thoroughly tested in accordance with applicable FDA guidelines.

2. Sample Size Used for the Test Set and Data Provenance

The document describes two main types of studies:

  • In Vitro (Bench-test) Studies:

    • Sample Size: Not explicitly stated with specific numbers for each test (e.g., number of sutures tested for knot strength, number of samples for fatigue). It generally refers to "data presented" and "tests" covering "respective sizes and types of suture materials" and "USP size 2 polypropylene monofilament sutures."
    • Data Provenance: Conducted "in vitro" and described as "bench-test models." No country of origin is specified, but it's implied to be part of the manufacturer's testing in the US. The data is retrospective in the sense that it refers back to studies conducted for prior 510(k) clearances (ASLS and SWK).
  • Animal Studies:

    • Sample Size: Not explicitly stated as a number of animals or implant sites. It mentions "New Zealand rabbits" for dorsal skin and bowel tissue studies. Without specific numbers, it's impossible to quantify the sample size.
    • Data Provenance: Animal models (New Zealand rabbits). The studies were previously conducted and referenced from a prior 510(k) (K980988), making them retrospective in the context of this current submission.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This information is not provided in the document.

  • For the in vitro studies, "ground truth" would be objective measurements (e.g., force required for failure, cycle count to failure), not expert consensus.
  • For the animal studies, the "investigators" concluded on safety and efficacy based on histopathology. However, the number of investigators/pathologists, their specific qualifications, or how their consensus was reached is not mentioned.

4. Adjudication Method

This information is not provided in the document. Given the nature of the studies (in vitro measurements and animal histopathology from earlier submissions), a formal human-reader adjudication method as often seen in imaging studies (e.g., 2+1) is unlikely to apply directly. The "investigators" formed conclusions, but the method of their consensus/adjudication is not detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not done.

This type of study is primarily relevant for diagnostic imaging devices where human readers interpret medical images with and without AI assistance to assess changes in diagnostic accuracy or efficiency. The Axya Bone Anchor System is a surgical tool, not a diagnostic imaging device. The comparative effectiveness assessment focused on mechanical properties (suture strength) and biological safety (tissue reaction) compared to traditional methods, not human reader performance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, in essence, standalone performance was evaluated for the device's core function.

The in vitro studies on suture strength, fatigue, and elongation represent the standalone mechanical performance of the welded suture. The animal safety studies evaluated the biological response to the device's output (welded suture) without direct human intervention impacting the "welding" function itself during the test. The "algorithm" here isn't a software algorithm but rather the mechanical and ultrasonic welding process. The efficacy of this process was demonstrated independently of varied human surgical technique (e.g., how a human would tie a knot).

7. The Type of Ground Truth Used

  • Quantitative Bench-Test Data: For suture strength and fatigue, the ground truth was derived from objective, measurable outcomes obtained from bench-tests (e.g., force measurements for knot strength, cycle counts for fatigue). These are "physical performance metrics."
  • Histopathology: For the animal safety studies, the ground truth for tissue reaction and safety was established by detailed microscopic examination of tissue samples (histopathology) at the implant sites. This is a "pathology" based ground truth.

8. The Sample Size for the Training Set

Not applicable.

The Axya Bone Anchor System is a mechanical device, not a machine learning or AI-driven system that requires a "training set" in the conventional sense. The device's function is deterministic based on its design and physical principles. Its development would involve engineering design, prototyping, and testing, not statistical training on a dataset.

9. How the Ground Truth for the Training Set Was Established

Not applicable.

As there is no training set as understood in AI/ML, there is no ground truth established for a training set. The device's performance is demonstrated through its inherent mechanical and biological properties, not learned patterns from data.

{0}------------------------------------------------

SUMMARY OF SAFETY AND EFFECTIVENESS

General Company Information

  • Axya Medical, Inc. Name:
  • 100 Cummings Center Address: Suite 444C Beverly, MA 01915

(978) 232 - 9997 Telephone: (978) 232 - 9998 Fax:

General Device Information

  • Product Name: Bone Anchor System and Kit (BAK)
  • Classification: Accessory to an Endoscope - Class II

Predicate Devices

Axya Suturing and Ligating System (ASLS) [510(k) K980988]

Axya Suture Welding System and Kit (SWK) [510(k) K983108]

Description

The Axya Bone Anchor System consists of an electronic control module and a reusable welding activator with an integral connector cable. The Kit (BAK) consists of a disposable suture welding sleeve packaged together with appropriate and currently approved bone anchors and USP polypropylene monofilament suture. The Kit is intended for use with the reusable system components in order to complete the suture welding process for the attachment of soft tissues to bone anchors in surgical procedures where bone anchors are indicated

{1}------------------------------------------------

K992405
2 of 3

Intended Use

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include:

Shoulder procedures(e.g. Rotator cuff repair)
Knee procedures(e.g. Patellar ligament / tendon avulsion repair)
Ankle procedures(e.g. Achilles tendon reconstruction and repair)
Foot procedures(e.g. Hallux valgus reconstruction)
Elbow procedures(e.g. Biceps tendon reattachment)
Bladder neck suspension(female urinary incontinence due to urethral hypermobility)

The BAK is indicated for use in both traditional open surgery and endoscopic surgical proceduras.

Substantial Equivalence

This submission supports the position that the Axya Bone Anchor System and Kit (BAR) is substantially equivalent in design and function to the Axya Suturing and Ligating System (ASLS) [510(k) K980988], the Axya Suture Welding System and Kit (SWK) [510(k) K983108] and the Endo-Judge disposable suture placement system developed by Synergistic Medical Technologies and marketed by Ethicon Endosurgery [510(K) K932591], and that it is appropriate for its intended application. Suture applicators and suture placement devices which may be used in both endoscopic or traditional open surgical procedures have been classified under 21 CFR 876.1500. These devices are indicated for the placement of sutures to close elther traumatic or surgically produced wounds. The BAK is fabricated from materials with a substantial history of use in medical devices. Both the BAK and the predicate devices secure the suture loop with an ultrasonic weld.

The 510(k) Notice for the predicate ASLS system contains summaries of both in yitro studies which were conducted to evaluate the safety, efficacy and appropriateness of the suture welding system. These data are applicable to the Bone Anchor System because the design and components of the welding mechanism are the same as those used in both the ASLS and the SWK that were previously cleared for marketing. Data were presented which demonstrate that sutures placed by means of the suture welding process exhibit "knot strength" characteristics substantially above the USP requirements for the respective sizes and types of suture materials tested. These tests confirm that the polypropylene monofilament sutures placed with the ultrasonic suture welding technology are equivalent in holding strength (efficacy) to sutures placed with conventional knotting techniques.

Additional data presented in the Notice for the Bone Anchor System and Kit confirm that USP size 2 polypropylene monofilament sutures secured by a weld formed with the AxyaWeld Tips exceed the USP knot strength requirements, Data have been presented that demonstrate that strength requirements are met regardless of the angle of applied tension between the suture

{2}------------------------------------------------

and the bone anchor stem. Suture fatigue and elongation over the functional life have been evaluated in bench-test models and the welded sutures have been found to meet the suggested criteria following cyclic testing.

The safety of the suture delivery system was previously evaluated by placing both welded sutures and manually knotted sutures in the dorsal skin and in bowel tissue of New Zealand rabbits [510(k) K980988]. Positive (electrocautery contact) and negative (normal tissue) controls were used in the bowel study. In both animal models there was no histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy. The investigators concluded that there was no significant difference in safety of efficacy between the traditional method of suture placement and the technique which includes replacement of manual knot-tying with suture welding. The design of the AxyaWeld Sleeve also prevents the heated area of suture from coming into direct contact with tissues.

The ultrasonic energy source used to weld and secure the suture loop is the same generator used for the Automatic Suturing and Ligating System and that used for the Suture Welding System and Kit, and is similar to the energy source used in the UltraCision Harmonic Scalpel [510(k) K895252].

Because of design features of the BAK, no portion of the ultrasonic generator comes into contact with human tissues. Because of this, there is virtually no risk of causing a thermal injury to the patient. The suture material is heated and welded by friction and the weld is formed by melting and fusing the polymer. No "flux" or "welding rod" is employed and no new chemical entities are introduced or produced in the welding process.

The Axya Bone Anchor System and Kit are sterilized by exposure to gamma irradiation at 2.5 Mrads which provides a sterility assurance level of at least 10° or, alternatively, by means of an ethylene oxide gas sterilization process which also provides a SAL of 10°. The sterility process selected is dictated by the type of suture included with a given product configuration. These sterlity processes, the manufacturing process, and the packaging process are validated by the manufacturer.

Axya Medical, Inc. believes that the information provided establishes that similar legally marketed and pre-enactment devices have been used for the same types of clinical applications as the Axya Bone Anchor System and Kit. The materials from which the Axya device is fabricated have an established history of use in medical applications, and the specific materials used by Axya have been thoroughly tested in accordance with applicable FDA quidelines.

{3}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

SEP 2 2 1999

Mr. Howard L. Schrayer Axya Medical, Inc. 100 Cummings Center, Suite 444C Beverly, Massachusetts 01915

Re: K992405

Trade Name: Axya Bone Anchor System and Kit Regulatory Class: II Product Code: GAW and MBI Dated: July 19, 1999 Received: July 20, 1999

Dear Mr. Schrayer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit/tray. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

{4}------------------------------------------------

Page 2-Mr. Howard L. Schrayer

for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.

FDA notes that your device will contain sutures for which you have provided evidence that the suture characteristics are not altered by the sterilization process used in the device. However, you should be aware of the following additional information regarding the inclusion of a suture as a component of your device:

  1. The labeling, packaging and method of sterilization of the suture cannot be changed without prior notification, review and clearance by FDA.

  2. The supplier of the sutures used in your device cannot be changed without prior notification, review and clearance by FDA.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on the labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Number (if known): 510(k) K992405

Device Name: Axya, Bone Anchor System and Kit, (BAK)

Indications For Use:

The Bone Anchor System and Kit is indicated for securing soft tissue structures to bone anchors in surgical procedures where bone anchors are indicated. Examples of these procedures include:

Shoulder procedures(e.g. Rotator cuff repair)
Knee procedures(e.g. Patellar ligament / tendon avulsion repair)
Ankle procedures(e.g. Achilles tendon reconstruction and repair)
Foot procedures(e.g. Hallux valgus reconstruction)
Elbow procedures(e.g. Biceps tendon reattachment)
Bladder neck suspension(female urinary incontinence due to urethral hypermobility)

The BAK is indicated for use in both traditional open surgery and endoscopic surgical procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription UseX
Use
(Per 21 CFR 801.109)

OR

Over-The-Counter

(Optional Format 1-2-96)

Division of General Restorative Devices
510(k) Number K992405

§ 878.5010 Nonabsorbable polypropylene surgical suture.

(a)
Identification. Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.(b)
Classification. Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.