K Number
K980988

Validate with FDA (Live)

Manufacturer
Date Cleared
1998-05-11

(55 days)

Product Code
Regulation Number
878.4800
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ASLS is indicated for the placement of synthetic monofilament sutures in soft tissues and for the ligation of tubular structures wherever conventional monofilament sutures or surgical staples are appropriate. The ASLS is indicated for use in both traditional open surgery and ondoscopic surgical procedures.

Device Description

The Axya Automatic Suturing and Ligating System (ASLS) is a device for the placement of sutures in soft tissue to close either traumatic or surgically produced wounds. It uses curved, hollow needles to place the suture through soft tissue structures and secures the suture loop with an ultrasonic weld. The ultrasonic energy source used to weld and secure the suture loop is similar to the energy source used in the UltraCision Harmonic Scalpel. The device is sterilized by exposure to gamma irradiation.

AI/ML Overview

Here's an analysis of the provided text regarding the Axya Automatic Suturing and Ligating System (ASLS), broken down into the requested categories:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific CriteriaReported Device Performance
Suture Strength (Efficacy)"Knot strength" substantially above USP requirements for respective suture sizes/types.Sutures placed by ASLS exhibit "knot strength" characteristics substantially above the USP requirements for the respective sizes and types of suture materials tested.
Holding strength equivalent to conventional knotting techniques for synthetic monofilament sutures.Synthetic monofilament sutures placed with ultrasonic welding are equivalent in holding strength (efficacy) to sutures placed with conventional knotting techniques.
Strength Retention (Efficacy for Absorbable Sutures)Strength loss curves for synthetic absorbable sutures follow manufacturer's profile and exceed predicted requirements based on USP for knot strength.Strength loss curves for synthetic absorbable sutures (Maxon™) follow the profile described in the suture manufacturer's package insert and exceed the predicted strength requirements based on USP requirements for knot strength of synthetic absorbable suture.
Biocompatibility/Tissue Reaction (Safety)No histopathologic change at suture implant sites due to ultrasonic sealing (compared to traditional methods).No histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy, in both dorsal skin and bowel tissue of New Zealand rabbits.
No significant difference in safety/efficacy between traditional suture placement and ASLS.Investigators concluded there was no significant difference in safety or efficacy between the traditional method of suture placement and the technique which includes replacement of manual knot-tying with suture welding.
Thermal Injury (Safety)No portion of the ultrasonic generator comes into contact with human tissues, enabling virtually no risk of thermal injury."Because of design features of the ASLS, no portion of the ultrasonic generator comes into contact with human tissues. There is virtually no risk of causing a thermal injury to the patient."
SterilizationSterility Assurance Level (SAL) of at least 10^-6 for gamma irradiation at 2.5 Mrads or 2.0 Mrads (depending on suture type).Sterilized by exposure to gamma irradiation at 2.5 Mrads (or 2.0 Mrads) which provides a sterility assurance level of at least 10^-6.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state the numerical sample size for the test sets in either the in vivo or in vitro studies. It mentions:

  • In Vitro Studies: "respective sizes and types of suture materials tested" and "one-week intervals over seven weeks."
  • In Vivo Studies: "dorsal skin and in bowel tissue of New Zealand rabbits." This indicates rabbit models were used.
  • Data Provenance: The studies were conducted by Axya Medical, Incorporated. The location is not specified, but the submission is to the U.S. FDA, implying the data would be expected to adhere to U.S. regulatory standards. The studies described are prospective as they were conducted specifically to evaluate the ASLS.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document mentions "investigators" who concluded on the safety and efficacy of the in vivo studies. However, it does not specify the number of experts, their qualifications, or how they established the ground truth beyond their conclusions based on histopathologic analysis.

4. Adjudication Method for the Test Set

The document does not describe a formal adjudication method (e.g., 2+1, 3+1). The "investigators" reached conclusions, implying a collegiate or consensus-based approach, but no specific process is detailed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

No, an MRMC comparative effectiveness study involving human readers is not described. The study compared the device's performance to conventional knotting or to USP standards.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

Yes, the studies described are standalone in the sense that they evaluate the performance of the ASLS device itself (suture strength, strength retention, tissue reaction) without a human operator's diagnostic or interpretative input being part of the primary outcome measure. The device replaces manual knot-tying, so its performance is measured against the output of that manual process.

7. The Type of Ground Truth Used

  • Suture Strength: Based on USP requirements and comparisons to "conventional knotting techniques."
  • Strength Retention: Based on the suture manufacturer's package insert profile and USP requirements.
  • Biocompatibility/Tissue Reaction: Established through histopathologic examination of tissue samples from the animal models, with "investigators" providing the conclusions.
  • Thermal Injury: Implied from the design features of the device (no tissue contact) and possibly verified by temperature monitoring, though not explicitly detailed.

8. The Sample Size for the Training Set

The document does not mention a training set because this device is a mechanical/energy-based surgical instrument, not an AI/algorithm-based diagnostic system that requires machine learning training.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set, the establishment of ground truth for a training set is not applicable.

{0}------------------------------------------------

KA80988

SUMMARY OF SAFETY AND EFFECTIVENESS

This summary supports the position that the Axya Automatic Suturing and Ligating System (ASLS) is substantially equivalent in design and function to the Endo-Judge disposable suture placement system developed by Synergistic Medical Technologies and marketed by Ethicon Endosurgery [510(k) K932591], and that it is appropriate for its intended application. Suture applicators and suture placement devices which may be used in both endoscopic or traditional open surgical procedures have been classified under 21 CFR 876.1500. These devices are indicated for the placement of sutures in soft tissue to close either traumatic or surgically produced wounds. The ASLS is fabricated from materials with a substantial history of use in medical devices. Both the ASLS and the predicate device use curved, hollow needles to place the suture through soft tissue structures. The ASLS secures the suture loop with an ultrasonic weld while the predicate device requires that the suture be secured with manually placed knots.

The 510/k) Notice contains summaries of both in vivo and in vitro studies which were conducted to evaluate the safety, efficacy and appropriatness of the ASLS. Data are presented which demonstrate that sutures placed by means of the ASLS exhibit "knot strength" characteristics substantially above the USP requirements for the respective sizes and types of suture materials tested. These tests confirm that the synthetic monofilament sutures placed with the ultrasonic suture welding technology are equivalent in holding strenath (efficacy) to sutures placed with conventional knotting techniques. In further efficacy studies of synthetic absorbable sutures (Maxon™) these materials were tested for strength at baseline and at one-week intervals over seven weeks following immersion in a constant temperature bath maintained at 98.5 degrees F. The strength loss curves for this synthetic absorbable suture follow the profile described in the suture manufacturer's package insert and exceed the predicted strength requirements based on USP requirements for knot strength of synthetic absorbable suture. This demonstrates that the welding process had no deleterious effect on the strength retention profile of the synthetic absorbable sutures. The safety of the ASLS suture delivery system was evaluated by placing both welded sutures and manually knotted sutures in the dorsal skin and in bowel tissue of New Zealand rabbits. Positive (electrocautery contact) and negative (normal tissue) controls were used in the bowel study. In both animal models there was no histopathologic change seen at the suture implant sites where the suture was sealed with ultrasonic energy. The investigators concluded that there was no significant difference in safety of efficacy between the traditional method of suture placement and the technique which includes replacement of manual knot-tying with suture welding.

The ultrasonic energy source used to weld and secure the suture loop is similar to the energy source used in the UltraCision Harmonic Scalpel (510(k) K895252). Because of design features of the ASLS, no portion of the ultrasonic generator comes into contact with human tissues. There is virtually no risk of causing a thermal injury to the patient. The suture material is heated and welded by friction and the weld is formed by melting and fusing the polymer. No "flux" or "welding rod" is empolyed and no new chemical entitles are introduced or produced in the welding process.

{1}------------------------------------------------

The Axya Automatic Suturing and Ligating System is sterilized by exposure to gamma irradiation at 2.5 Mrads which provides a sterility assurance level of at least 10° or by in addition at 2.0 Mirado which process which also provides a SAL of 10°. The sterility process selected is dictated by the type of suture included with a given product configuration. This sterility processes, the manufacturing process, and the packaging process are validated by the manufacturer.

Axya LLC believes that the information provided establishes that similar legally marketed and pre-enactment devices have been used historically for the same types of clinical applications as the Axya Automatic Suturing and Ligating System. The materials from which the Axya device is fabricated have an established history of use in medical applications, and the specific materials used by Axya have been thoroughly tested in accordance with applicable FDA guidelines.

{2}------------------------------------------------

Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 1 1998

Mr. Howard L. Schrayer ·Axya Medical, Incorporated 12 South Street Marblehead, Massachusetts 01945

Re: K980988 Axya Medical Automatic Suturing and Trade Name: Liqating System Regulatory Class: II GAB and KOG Product Code: Dated: March 16, 1998 Received: March 17, 1998

Dear Mr. Schrayer:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate ase scared in the encreases, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set ... forth in the Quality System Regulation (QS) for Medical General regulation (21 CFR Part 820) and that, Devices: through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Mr. Schrayer

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.

Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K980988

Device Name: Axya, Automatic Suturing and Ligating System, (ASLS)

Indications For Usc:

The ASLS is indicated for the placement of synthetic monofilament sutures in soft tissues and for the ligation of tubular structures wherever conventional monofilament sutures or surgical staples are appropriate. The ASLS is indicated for use in both traditional open surgery and ondoscopic surgical procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K990988

Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use

§ 878.4800 Manual surgical instrument for general use.

(a)
Identification. A manual surgical instrument for general use is a nonpowered, hand-held, or hand manipulated device, either reusable or disposable, intended to be used in various general surgical procedures. The device includes the applicator, clip applier, biopsy brush, manual dermabrasion brush, scrub brush, cannula, ligature carrier, chisel, clamp, contractor, curette, cutter, dissector, elevator, skin graft expander, file, forceps, gouge, instrument guide, needle guide, hammer, hemostat, amputation hook, ligature passing and knot-tying instrument, knife, mallet, disposable or reusable aspiration and injection needle, disposable or reusable suturing needle, osteotome, pliers, rasp, retainer, retractor, saw, scalpel blade, scalpel handle, one-piece scalpel, snare, spatula, stapler, disposable or reusable stripper, stylet, suturing apparatus for the stomach and intestine, measuring tape, and calipers. A surgical instrument that has specialized uses in a specific medical specialty is classified in separate regulations in parts 868 through 892 of this chapter.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 878.9.