K Number
K992318

Validate with FDA (Live)

Date Cleared
1999-08-02

(21 days)

Product Code
Regulation Number
876.5010
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Medtronic AVE Stent Delivery System - For Use In Biliary Indication device is intended to maintain patency of a bile duct which is occluded by a malignant tumor.

Device Description

The device consists of a balloon expandable intralumenal stent premounted onto the balloon of an over-the-wire delivery catheter. The device has two radiopaque platinum markers imbedded on the inner shaft (at each end of the stent) to aid in the placement of the stent during fluoroscopy. The delivery system is compatible with 0.035" guidewires and has a useable length of 75cm to 90cm. The device is provided in a sterile package.

AI/ML Overview

The Medtronic AVE Stent Delivery System did not provide acceptance criteria in a quantitative format with reported performance. Instead, the submission relies on demonstrating substantial equivalence to a predicate device through performance testing. The study focuses on direct comparisons between the subject device and the predicate device for various physical and mechanical properties.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

The submission does not present acceptance criteria in a quantitative table format. Instead, it describes performance testing conducted to demonstrate substantial equivalence to a predicate device. The general acceptance criterion seems to be that the subject device performs comparably to the predicate device for the listed tests.

TestPurposeReported Performance
Balloon BurstTo compare the stent/delivery system of the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Bond StrengthTo compare the stent/delivery system of the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Balloon Deflation TimeTo compare the stent/delivery system of the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Crossing ProfileTo compare the stent/delivery system of the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Diameter versus Inflation PressureTo create and compare the compliance curves for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Stent Free AreaTo compare the stent dimensional data for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Stent LengthTo compare the stent dimensional data for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Crush StrengthTo compare the stent dimensional data for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Stent RecoilTo compare the stent dimensional data for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
Stent ExpansionTo compare the stent dimensional data for the subject device and the predicate device.Not explicitly stated, but the conclusion is that the two devices are substantially equivalent.
BiocompatibilityTo ensure the material used in the stent is safe for biological contact."The material used in the Medtronic AVE Stent passed all biocompatibility tests."
SterilizationTo ensure the sterilization method is effective."The Medtronic AVE Stent is provided sterile." and validated according to ANSI/AAMI/ISO 11137 - 1994, Method I, with a SAL of 10-6. "Labeled pyrogen free."

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document does not specify sample sizes for each of the performance tests conducted. It refers to "performance testing" being conducted but provides no details on the number of units tested for each parameter. The data provenance is also not mentioned; it's a submission for a medical device by Medtronic AVE, Inc. in Santa Rosa, California, but details on where the tests were conducted or the origin of any 'data' beyond the tests are absent. These appear to be bench tests, not clinical studies involving patients.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This section is not applicable. The study involves demonstrating the physical and mechanical performance of a medical device against a predicate device, not establishing a ground truth for a diagnostic algorithm or interpretation. There are no "experts" involved in judging test results in the clinical sense mentioned here.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This section is not applicable for the same reasons as point 3. There is no adjudication method described as this is not a study requiring human interpretation or consensus.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study done. This submission is for a medical device (stent delivery system), not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This section is not applicable as this is a medical device (stent delivery system), not an algorithm or AI system.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The "ground truth" in this context is the performance and characteristics of the predicate device (Peripheral AVE Stent Delivery System - For Use In Biliary Indication) and established engineering/performance standards for medical devices, coupled with the results from the specific performance tests conducted on the subject device. The goal is to show the subject device performs equivalently.

8. The sample size for the training set

This section is not applicable as the submission describes a physical medical device, not a machine learning model. There is no concept of a "training set" here.

9. How the ground truth for the training set was established

This section is not applicable for the same reasons as point 8.

{0}------------------------------------------------

K992318
Page 1 of 4

510(k) Summary for the Medtronic AVE Stent Delivery System

510(k)SummaryThis summary of 510(k) safety and effectiveness information is beingsubmitted in accordance with the requirements of AMDA 1990 and 21 CFR807.92.
IdentificationThe assigned 510(k) number is
Applicant:Medtronic AVE, Inc.Peripheral Technologies2330A Circadian WaySanta Rosa, California 95407Contact: Susan L. Walton
Phone: (707) 591-7315 FAX: (707) 591-7406e-mail: susan.walton@medtronic.comDate submitted: July 9, 1999
Tradename:Device Name: Medtronic AVE StentModel Numbers: B7080, B8080, B70100, B80100Classification Name: Catheter, Biliary and accessories
Section 513DeviceClassificationClassification: Class IIClassification Panel: 78FGE
EquivalenceMedtronic AVE claims substantial equivalence to the Peripheral AVE StentDelivery System - For Use In Biliary Indication.
Intended UseThe Medtronic AVE Stent Delivery System - For Use In Biliary Indicationdevice is intended to maintain patency of a bile duct which is occluded by amalignant tumor.

Continued on next page

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the number K992318 written in black ink at the top. Below that, the text "Page 2 of 4" is written in cursive, also in black ink. The handwriting appears to be somewhat messy and uneven.

510(k) Summary for the Medtronic AVE Stent Delivery

System, Continued

Description of Device The device consists of a balloon expandable intralumenal stent premounted onto the balloon of an over-the-wire delivery catheter. The device has two radiopaque platinum markers imbedded on the inner shaft (at each end of the stent) to aid in the placement of the stent during fluoroscopy. The delivery system is compatible with 0.035" guidewires and has a useable length of 75cm to 90cm. The device is provided in a sterile package.

Size range:Diameters: 7.0 to 8.0 mmLengths 80 to 100 mm
---------------------------------------------------------------

Comparison Table This device is intended to maintain patency of a biliary duct which is occluded by a malignant tumor.

CharacteristicSubject DevicePredicate Device
Intended UseThis device is intended to maintain patency of a bile duct which is occluded by tumor.This device is intended to maintain patency of a biliary duct which is occluded by tumor.
Physical Characteristics (stent)316L stainless steel balloon expandable premounted stent Diameters 7 - 8mmLengths 80 to 100mm316L stainless steel balloon expandable premounted stent Diameters 6 - 10mmLengths 17 to 60mm
Physical Characteristics (delivery system)Balloon delivery system PTA catheter (PET)5.3 to 5.8 F shafts75 to 90cm lengths0.035" guidewire ØBalloon delivery system PTA catheter (PET)5.3 to 5.8 F shafts75 to 90cm lengths0.035" guidewire Ø
Anatomical siteBiliary ductBiliary duct
Target populationPatients with bile duct obstruction caused by malignant tumor.Patients with biliary duct obstruction caused by malignant tumor.

Continued on next page

{2}------------------------------------------------

510(k) Summary for the Medtronic AVE Stent Delivery

System, Continued

PerformanceTestingThe subject device for this 510(k) is identical to the predicate device with theexception of an alternate stent segment (component) and extended stentlengths. The flexible Medtronic AVE Stent was found to be substantiallyequivalent in K983008 approved November 25, 1998. Performance testingwas conducted on the subject device for the purpose of direct comparison tothe predicate device. The testing was chosen to highlight any differencesbetween the subject device and the predicate device. The balloon materialshave not been changed. The useable catheter length is not changed. Themanufacturing process is not changed.
TestPurpose
• Balloon BurstTo compare the stent/delivery system of thesubject device and the predicate device. The
• Bond Strengthsubject device and the predicate device. The
• Balloon Deflation Timedata will support a premarket notification for
• Crossing Profilethe Medtronic AVE Stent Delivery System -For Use In Biliary Indication.
• Diameter versusInflation PressureTo create and compare the compliance curvesfor the subject device and the predicate device.The data will support a premarket notificationfor the Medtronic AVE Stent Delivery System -For Use In Biliary Indication.
• Stent Free AreaTo compare the stent dimensional data for thesubject device and the predicate device. The
• Stent Lengthsubject device and the predicate device. The
• Crush Strength• Stent Recoil• Stent Expansiondata will support a premarket notification forthe Medtronic AVE Stent Delivery System -For Use In Biliary Indication.
ConclusionsThe performance testing and comparison of the Peripheral AVE StentDelivery System - For Use In Biliary Indication and the Medtronic AVE StentDiameter versus Inflation Pressure prove the two devices are substantiallyequivalent.
AdditionalInformationThe summary includes any other information reasonably deemed necessaryby FDA.
BiocompatibilityThe material used in the Medtronic AVE Stent passed all biocompatibilitytests.

Continued on next page

{3}------------------------------------------------

K992318
Page 4 of 4

510(k) Summary for the Medtronic AVE Stent Delivery System, Continued

SterilizationThe Medtronic AVE Stent is provided sterile.
The Medtronic AVE Stent is not intended for sterilization orreuse/resterilization by the user.
Medtronic AVE validates the sterilization method for its stent deliverysystems according to the ANSI/AAMI/ISO 11137 - 1994, Method I:Sterilization of Healthcare Products - Requirements for Validation andRoutine Control - Radiation Sterilization. The Sterility Assurance Level(SAL) is 10-6.
The Medtronic AVE Stent is labeled pyrogen free. LAL testing is performeddaily in compliance with FDA guidance on Validation of Limulus AmebocyteLysate Test as an End Product Endotoxin Test for Human and AnimalParenteral Drugs, Biological Products and Medical Devices - Section V - 2Inhibition and Enhancement Testing as part of Medtronic AVE's productrelease criteria.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol with three horizontal lines above it, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol. The logo is black and white.

2 1999 Alle

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Susan L. Walton Regulatory Coordinator Medtronic AVE, Peripheral Technologies 2330A Circadian Way Santa Rosa, CA 95407

K992318 Re:

Medtronic AVE Bridge™ Stent System (Hi-Flex) - Biliary Indication Regulatory Class: II 21 CFR 876.5010 Product Code: 78 FGE Dated: July 10, 1999 Received: July 12, 1999

Dear Ms. Walton:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act and the limitations described below. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

The Office of Device Evaluation has determined that there is a reasonable likelihood that this device will be used for an intended use not identified in the proposed labeling and that such use could cause harm. Therefore, in accordance with Section 513(i)(1)(E) of the Act, the following limitation must appear in the Warnings scction of the device's labeling:

The safety and effectiveness of this device for use in the vascular system have not been established.

Furthermore, the indication for biliary use must be prominently displayed in all labeling, including pouch, box, and carton labels, instructions for use, and other promotional materials, in close proximity to the trade name, of a similar point size, and in bold print.

{5}------------------------------------------------

Page 2 -- Ms. Susan L. Walton

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations (1 renamed in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to mspotions) the Food and Diam may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

The FDA finding of substantial equivalence of your device to a legally marketed predicate device The I Dri mainig of sacctantial equice and permits your device to proceed to the market. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification if the limitation statement above is added to your labeling, as described.

Please note that the above labeling limitations are required by Section 513(i)(1)(E) of the Act. Therefore, a new 510(k) is required before these limitations are modified in any way or removed from the device's labeling.

If you desire specific information about the application of other labeling requirements to your device (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4616. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html" .

Sincerely yours,

Sincerely yours,

Susan Alpert, Ph.D., M.D.

Susan Alpert, Ph.D., M.D. Director Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Page _ 1 of 1

510(k) Number (if known): K992318

Device Name: _Medtronic AVE Bridge™ Stent System (Hi-Flex) - Biliary Indication

FDA's Statement of the Indications For Use for device:

The Medtronic AVE Bridge™ Stent System (Hi-Flex) – Biliary Indication is intended to maintain patency of a bile duct which is occluded by a malignant tumor.

Prescription Use」 OR (Per 21 CFR 801.109)

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

Girard W. Seaymore
Division Sign Off

eproductive, Abdominal, E 510(k) Number

§ 876.5010 Biliary catheter and accessories.

(a)
Identification. A biliary catheter and accessories is a tubular flexible device used for temporary or prolonged drainage of the biliary tract, for splinting of the bile duct during healing, or for preventing stricture of the bile duct. This generic type of device may include a bile collecting bag that is attached to the biliary catheter by a connector and fastened to the patient with a strap.(b)
Classification. Class II (special controls). The device, when it is a bile collecting bag or a surgical biliary catheter that does not include a balloon component, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.