(41 days)
Not Found
Not Found
No
The device description details a standard automated immunoassay process based on chemical reactions and light detection. There is no mention of AI, ML, or any computational analysis beyond reporting results based on a direct relationship between light units and ferritin concentration.
No.
The "Intended Use" section states that the device is for "quantitative determination of Ferritin in serum or plasma" and "to be used to aid in the diagnosis of iron deficiency anemia and iron overload." This indicates a diagnostic purpose rather than a therapeutic one.
Yes
The intended use explicitly states, "It is to be used to aid in the diagnosis of iron deficiency anemia and iron overload." This direct contribution to diagnosis classifies it as a diagnostic device.
No
The device description clearly outlines a physical automated chemiluminescence system (ACS:180) that performs chemical reactions and measurements on biological samples (serum or plasma). This involves hardware components for dispensing, incubating, separating, washing, and detecting light units. The software component is part of this larger hardware system, not a standalone software-only device.
Yes, this device is an IVD (In Vitro Diagnostic).
Here's why:
- Intended Use: The intended use explicitly states it's for the "quantitative determination of Ferritin in serum or plasma." Serum and plasma are biological specimens taken from the human body.
- Purpose: The intended use also states it is "to be used to aid in the diagnosis of iron deficiency anemia and iron overload." This indicates the test is performed in vitro (outside the body) on a sample to provide information for diagnostic purposes.
- Device Description: The description details a laboratory assay process involving the analysis of a sample (serum or plasma) using reagents and an automated system to measure a specific analyte (ferritin). This is characteristic of an in vitro diagnostic test.
- Performance Studies: The performance studies describe analyzing samples from different patient populations (healthy, iron deficiency, iron overload, etc.) and comparing results to an alternate method, which are standard practices for validating an IVD.
The definition of an IVD is a medical device that is used to examine specimens taken from the body, such as blood, urine, or tissue, to help diagnose diseases or conditions. This device clearly fits that description.
N/A
Intended Use / Indications for Use
The intended use of Chiron Diagnostics ACS:180 Ferritin Assay is for the quantitative determination of Ferritin in serum or plasma using the Chiron Diagnostics ACS:180® Automated Chemiluminescence Systems; to aid in the diagnosis of iron deficiency anemia and iron overload.
Product codes (comma separated list FDA assigned to the subject device)
DBF
Device Description
Ferritin plays a significant role in the absorption, storage, and release of iron. As the storage form of iron, ferritin remains in the body tissues until it is needed for erythropoiesis. When needed, the iron molecules are released from the apoferritin shell and bind to transferrin, the circulating plasma protein that transports iron to the erythropoietic cells.
Although dietary iron is poorly absorbed, the body conserves its iron stores carefully, reabsorbing most of the iron released from the breakdown of red blood cells. As a result, the body normally loses only 1 to 2 mg of iron per day, which is generally restored by the iron absorbed in the small intestine from dietary sources.
Ferritin is found in serum in low concentrations and is directly proportional to the body's iron stores. Serum ferritin concentration, when analyzed with other factors such as serum iron, iron-binding capacity, and tissue iron stores, is valuable in the diagnosis of iron-deficiency anemias, anemias of chronic infection, and conditions such as thalassemia and hemochromatosis that are associated with iron overload. Measurement of serum ferritin is particularly valuable in distinguishing iron-deficiency anemias caused by low iron stores from those resulting from inadequate iron utilization.
The Chiron Diagnostics ACS:180 Ferritin assay is a two-site sandwich immunoassay using direct, chemiluminometric technology, which uses constant amounts of two anti-ferritin antibodies. The first antibody, in the Reagent, is a polyclonal goat anti-ferritin antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-ferritin antibody, which is covalently coupled to paramagnetic particles. The ACS:180 system automatically performs the following steps:
- dispenses 25 uL of sample into a cuvette .
- . dispenses 100 µL of Lite Reagent and 450 µL of Solid Phase and incubates for 7.5 minutes at 37°C
- . separates, aspirates, and washes the cuvettes with reagent water
- . dispenses 300 uL each of Reagent 1 and Reagent 2 to initiate the chemiluminescent reaction
- . reports results according to the selected option, as described in the system operating instructions or in the online help system
A direct relationship exists between the amount of ferritin present in the patient sample and the amount of relative light units (RLUs) detected by the system.
Mentions image processing
Not Found
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies (study type, sample size, AUC, MRMC, standalone performance, key results)
Expected Results
In clinical studies, the following values for apparently healthy male and female subjects with normal liver function enzyme tests, bilirubin, and serum iron tests, were determined:
Normal Males: N=142, Geo. Mean (ng/mL) = 94, Geo. Mean (pmol/L) = 207, 95th Percentile Range (ng/mL) = 22-322, 95th Percentile Range (pmol/L) = 48-708
Normal Females: N=134, Geo. Mean (ng/mL) = 46, Geo. Mean (pmol/L) = 101, 95th Percentile Range (ng/mL) = 10-291, 95th Percentile Range (pmol/L) = 22-640
The following values for patients with several diagnosed conditions were determined:
Iron Deficiency: N=60, Geo. Mean (ng/mL) = 11.6, (pmol/L) = 26, Total Observed Range (ng/mL) = 0.68-34.5, (pmol/L) = 1.5-76
Other Anemias: N=7, Geo. Mean (ng/mL) = 610.8, (pmol/L) = 1344, Total Observed Range (ng/mL) = 13.0-1390.8, (pmol/L) = 29-3060
Iron Overload: N=44, Geo. Mean (ng/mL) = 1899.6, (pmol/L) = 4178, Total Observed Range (ng/mL) = 334.6-8573.0, (pmol/L) = 736-18861
Renal Dialysis: N=31, Geo. Mean (ng/mL) = 312.3, (pmol/L) = 687, Total Observed Range (ng/mL) = 31.3-1321.2, (pmol/L) = 68.9-2907
Chronic Liver Disease: N=34, Geo. Mean (ng/mL) = 1967.1, (pmol/L) = 4328, Total Observed Range (ng/mL) = 7.9-12826.0, (pmol/L) = 17-28217
Sensitivity and Assay Reportable Range
The ACS:180 Ferritin assay measures ferritin concentrations up to 1650 ng/mL (3630 pmol/L) with a minimum detectable concentration of 0.5 ng/mL (1 pmol/L).
Method Comparison
For 276 samples in the range of 3.1 to 1621 ng/mL (6.8 to 3566.2 pmol/L), the relationship between the ACS:180 Ferritin assay and an alternate chemiluminescent method is described by the equation:
ACS:180 Ferritin = 1.01 (alternate chemiluminescent method) + 1.84 ng/mL
Correlation coefficient (r) = 1.00
Precision
Four samples were assayed 3 times in 8 assays. The following results were obtained:
Mean (ng/mL) = 13.1, Mean (pmol/L) = 29, Within-run % CV = 2.76, Total % CV = 4.98
Mean (ng/mL) = 54.8, Mean (pmol/L) = 121, Within-run % CV = 2.64, Total % CV = 6.07
Mean (ng/mL) = 162.7, Mean (pmol/L) = 358, Within-run % CV = 2.73, Total % CV = 4.68
Mean (ng/mL) = 359.5, Mean (pmol/L) = 791, Within-run % CV = 3.62, Total % CV = 5.08
Key Metrics (Sensitivity, Specificity, PPV, NPV, etc.)
Sensitivity: minimum detectable concentration of 0.5 ng/mL (1 pmol/L).
Precision: Within-run % CV ranged from 2.64 to 3.62; Total % CV ranged from 4.68 to 6.07.
Predicate Device(s): If the device was cleared using the 510(k) pathway, identify the Predicate Device(s) K/DEN number used to claim substantial equivalence and list them here in a comma separated list exactly as they appear in the text. List the primary predicate first in the list.
Not Found
Reference Device(s): Identify the Reference Device(s) K/DEN number and list them here in a comma separated list exactly as they appear in the text.
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information for the subject device only (e.g. presence / absence, what scope was granted / cleared under the PCCP, any restrictions, etc).
Not Found
§ 866.5340 Ferritin immunological test system.
(a)
Identification. A ferritin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the ferritin (an iron-storing protein) in serum and other body fluids. Measurements of ferritin aid in the diagnosis of diseases affecting iron metabolism, such as hemochromatosis (iron overload) and iron deficiency amemia.(b)
Classification. Class II (performance standards).
0
AUG - 5 1999
Summary of Safety and Effectiveness
As required by.21 CFR 807.92, the following 510(k) Summary is provided:
1. Submitters Information
| Contact person: | William J. Pignato
Director of Regulatory Affairs |
|-------------------------|-------------------------------------------------------------------------|
| Address: | Chiron Diagnostics Corporation
63 North Street
Medfield, MA 02052 |
| Phone:
Fax
e-mail | (508) 359-3825
(508) 359-3356
william.pignato.b@bayer.com |
| Date Summary Prepared: | June 22, 1999 |
2. Device Information
Proprietary Name: | Chiron Diagnostics ACS: 180 Ferritin |
---|---|
Common Name: | Ferritin Immunological Test System |
Device Classification: | Class II: 21 CFR 866.5350 |
3. Predicate Device Information
Name: | Chiron Diagnostics ACS: 180 Ferritin Immunoassay |
---|---|
Manufacturer: | Chiron Diagnostics Corporation |
4. Device Description
Ferritin plays a significant role in the absorption, storage, and release of iron. As the storage form of iron, ferritin remains in the body tissues until it is needed for erythropoiesis. When needed, the iron molecules are released from the apoferritin shell and bind to transferrin, the circulating plasma protein that transports iron to the erythropoietic cells.
Although dietary iron is poorly absorbed, the body conserves its iron stores carefully, reabsorbing most of the iron released from the breakdown of red blood cells. As a result, the body normally loses only 1 to 2 mg of iron per day, which is generally restored by the iron absorbed in the small intestine from dietary sources.
Ferritin is found in serum in low concentrations and is directly proportional to the body's iron stores. Serum ferritin concentration, when analyzed with other factors such as serum iron, iron-binding capacity, and tissue iron stores, is valuable in the diagnosis of iron-deficiency anemias, anemias of chronic infection, and conditions such as thalassemia and hemochromatosis that are associated with iron overload. Measurement of serum ferritin is
1
particularly valuable in distinguishing iron-deficiency anemias caused by low iron stores from those resulting from inadequate iron utilization.
5. Statement of Intended Use
The intended use of Chiron Diagnostics ACS:180 Ferritin Assay is for the quantitative determination of Ferritin in serum or plasma using the Chiron Diagnostics ACS:180® Automated Chemiluminescence Systems; to aid in the diagnosis of iron deficiency anemia and iron overload.
6. Summary of Technological Characteristics
The Chiron Diagnostics ACS:180 Ferritin assay is a two-site sandwich immunoassay using direct, chemiluminometric technology, which uses constant amounts of two anti-ferritin antibodies. The first antibody, in the Reagent, is a polyclonal goat anti-ferritin antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-ferritin antibody, which is covalently coupled to paramagnetic particles. The ACS:180 system automatically performs the following steps:
- dispenses 25 uL of sample into a cuvette .
- . dispenses 100 µL of Lite Reagent and 450 µL of Solid Phase and incubates for 7.5 minutes at 37°C
- . separates, aspirates, and washes the cuvettes with reagent water4
- . dispenses 300 uL each of Reagent 1 and Reagent 2 to initiate the chemiluminescent reaction
- . reports results according to the selected option, as described in the system operating instructions or in the online help system
A direct relationship exists between the amount of ferritin present in the patient sample and the amount of relative light units (RLUs) detected by the system.
6. Performance Characteristics
Expected Results
In clinical studies, the following values for apparently healthy male and female subjects with normal liver function enzyme tests, bilirubin, and serum iron tests, were determined:
Category | N | Geo. Mean (ng/mL) | Geo. Mean (pmol/L) | 95th Percentile Range (ng/mL) | 95th Percentile Range (pmol/L) |
---|---|---|---|---|---|
Normal Males | 142 | 94 | 207 | 22-322 | 48-708 |
Normal Females | 134 | 46 | 101 | 10-291 | 22-640 |
The following values for patients with several diagnosed conditions were determined:
| Category | N | Geo. Mean
(ng/mL) | (pmol/L) | Total Observed Range
(ng/mL) | (pmol/L) |
|-----------------|----|----------------------|----------|---------------------------------|----------|
| Iron Deficiency | 60 | 11.6 | 26 | 0.68-34.5 | 1.5-76 |
| Other Anemias | 7 | 610.8 | 1344 | 13.0-1390.8 | 29-3060 |
2
Iron Overload | 44 | 1899.6 | 4178 | 334.6-8573.0 | 736-18861 |
---|---|---|---|---|---|
Renal Dialysis | 31 | 312.3 | 687 | 31.3-1321.2 | 68.9-2907 |
Chronic Liver Disease | 34 | 1967.1 | 4328 | 7.9-12826.0 | 17-28217 |
As with all diagnostic assays, each laboratory should determine its own reference range(s) for the diagnostic evaluation of patient results.
Sensitivity and Assay Reportable Range
The ACS:180 Ferritin assay measures ferritin concentrations up to 1650 ng/mL (3630 pmol/L) with a minimum detectable concentration of 0.5 ng/mL (1 pmol/L).
Method Comparison
For 276 samples in the range of 3.1 to 1621 ng/mL (6.8 to 3566.2 pmol/L), the relationship between the ACS:180 Ferritin assay and an alternate chemiluminescent method is described by the equation:
ACS:180 Ferritin = 1.01 (alternate chemiluminescent method) + 1.84 ng/mL Correlation coefficient (r) = 1.00
Precision
Four samples were assayed 3 times in 8 assays. The following results were obtained:
| Mean
(ng/mL) | Mean
(pmol/L) | Within-run
% CV | Total
% CV |
|-----------------|------------------|--------------------|---------------|
| 13.1 | 29 | 2.76 | 4.98 |
| 54.8 | 121 | 2.64 | 6.07 |
| 162.7 | 358 | 2.73 | 4.68 |
| 359.5 | 791 | 3.62 | 5.08 |
3
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
AUG - 5 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Mr. William J. Pignato Director of Regulatory Affairs Chiron Diagnostics Corporation (A Bayer Diagnostics Corporation) 63 North Street Medfield, Massachusetts 02052
Re: K992157
Trade Name: Chiron Diagnostics ACS: 180 FERRITIN Assay Regulatory Class: II Product Code: DBF Dated: June 22, 1999 Received: June 25, 1999
Dear Mr. Pignato:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
4
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
5
Page of
510(k) Number (if known): _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
Device Name: Chiron Diagnostics ACS:180 FERRITIN Assay
Indications for Use:
The intended use of Chiron Diagnostics ACS:180 Ferritin Assay is for the quantitative determination of Ferritin in serum or plasma using the Chiron Diagnostics ACS:180® Automated Chemiluminescence Systems. It is to be used to aid in the diagnosis of iron deficiency anemia and iron overload.
(PLEASE DO NOT WRITE BELOW THIS LINE--CONTINUE ON ANOTHER PAGE, IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) | |
---|---|
Division Sign-Off | |
Division of Clinical Laboratory Devices | |
510(k) Number | K992157 |
| Prescription Use
(Per 21 CFR 801.109) | OR | Over-The-Counter Use
(Optional Format 1-2-96) |
------------------------------------------ | ---- | -------------------------------------------------- |
---|