(63 days)
The Piccolo MetLyte 7 Reagent Disc, used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of creatine kinase, creatinine, glucose, potassium, total carbon dioxide and urea nitrogen in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
Indications for Use:
Creatine Kinase The creatine kinase assay is used for the quantitation of creatine kinase in human heparinized whole blood, heparinized plasma, or serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction, progressive muscular dystrophy, dermatomyositis, rhabdomyolysis due to drug ingestion, hyperosmolality, autoimmune disease, delirium tremens, convulsions, Crush syndrome, hypothyroidism, surgery, severe exercise, intramuscular injection, physical inactivity, decreased muscle mass.
Creatinine The creatinine assay is used for the quantitation of creatinine in human heparinized whole blood, heparinized plasma, or serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases and monitoring of renal dialysis.
Glucose The glucose assay is used for the quantitation of glucose in human heparinized whole blood, heparinized plasma, or serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including adult and juvenile diabetes mellitus and hypoglycemia.
Potassium The potassium assay is used for the quantitation of potassium in human heparinized whole blood, heparinized plasma, or serum. Potassium measurements are used in the diagnosis and treatment of renal glomerular or tubular disease, adrenocortical insufficiency, diabetic ketoacidosis, excessive intravenous potassium therapy, sepsis, panhypopituitarism, in vitro hemolysis, hyperaldosteronism, malnutrition, hyperinsulinism, metabolic alkalosis and gastrointestinal loss.
Total Carbon Dioxide The total carbon dioxide assay is used for the quantitation of total carbon dioxide in human heparinized whole blood, heparinized plasma, or serum. Total carbon dioxide measurements are used in the diagnosis and treatment of primary MetLyte 7 alkalosis and primary respiratory alkalosis and acidosis.
Urea Nitrogen The urea nitrogen assay is used for the quantitation of urea nitrogen in human heparinized whole blood, heparinized plasma, or serum. Urea nitrogen measurements are used in the diagnosis and treatment of renal and MetLyte 7 diseases.
The Piccolo MetLyte 7 Reagent Disc is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.
The provided text describes the acceptance criteria and study results for the Piccolo MetLyte 7 Reagent Disc for Creatine Kinase, Potassium, and Total Carbon Dioxide tests.
Here's the breakdown of the information requested:
1. Table of Acceptance Criteria and Reported Device Performance
The acceptance criteria are implied by the "Performance Characteristics" sections in Tables 1-3, where the new device's performance (Piccolo® Point-of-Care Chemistry Analyzer) is compared to legally marketed predicate devices. The studies aim to show substantial equivalence.
| Analyte | Performance Characteristic | Acceptance Criteria (Implied by Predicate Device) | Reported Device Performance (Piccolo) |
|---|---|---|---|
| Creatine Kinase | Intended Use | quantitative determination of creatine kinase | quantitative analysis of creatine kinase |
| Methodology | isoenzymes | isoenzymes | |
| Sample Type | serum | whole blood, plasma, serum | |
| Sensitivity | 0.0001 A/min/U/L | 0.0001 A/min/U/L | |
| Reagents | liquid mono reagent | dry test-specific reagent beads | |
| Temperature | 37°C | 37°C | |
| Calibration | pre-set calibration | bar code with disc-specific calibration data | |
| Assay Range | 0 - 2,000 U/L | 5 - 5,000 U/L | |
| Accuracy: Corr. Coeff. | (Not explicitly stated for predicate) | 0.967 | |
| Accuracy: Slope | (Not explicitly stated for predicate) | 1.194 | |
| Accuracy: Intercept | (Not explicitly stated for predicate) | -24.983 | |
| Accuracy: SEE | (Not explicitly stated for predicate) | 9.050 | |
| Accuracy: R-Square | (Not explicitly stated for predicate) | 0.934 | |
| Potassium | Intended Use | quantitative determination of potassium | quantitative analysis of potassium |
| Methodology | ion-selective electrodes | enzymatic activation | |
| Sample Type | whole blood, plasma, serum, urine | whole blood, plasma, serum | |
| Sensitivity | 0.58 (mmol/L)/(mV) | 0.024 A/min/mmol/L | |
| Reagents | N/A (for KNA™ 2) | dry, test specific reagent beads | |
| Temperature | 37° C | 37° C | |
| Calibration | automatic 1 point calibration every 2 hrs; 2 point calibration every 8 hrs | bar code with disc-specific calibration data | |
| Assay Range | 1.0 - 99.9 mmol/L | 1.5 - 8.5 mmol/L | |
| Accuracy: Corr. Coeff. | (Not explicitly stated for predicate) | 0.969 | |
| Accuracy: Slope | (Not explicitly stated for predicate) | 0.863 | |
| Accuracy: Intercept | (Not explicitly stated for predicate) | 0.573 | |
| Accuracy: SEE | (Not explicitly stated for predicate) | 0.141 | |
| Accuracy: R-Square | (Not explicitly stated for predicate) | 0.939 | |
| Total Carbon Dioxide | Intended Use | quantitative determination of total carbon dioxide | quantitative analysis of total carbon dioxide |
| Methodology | enzymatic using phosphoenol pyruvate carbo-xylase and malate dehydrogenase | enzymatic | |
| Sample Type | serum and plasma | whole blood, plasma, serum | |
| Sensitivity | 0.0427 A/mmol/L | 0.0037 A/min/mmol/L | |
| Reagents | liquid substrate reagent 1 and liquid enzyme reagent 2 | dry test-specific reagent beads | |
| Temperature | 37° C | 37° C | |
| Calibration | Roche calibrator serum every 6 months | bar code with disc-specific calibration data | |
| Assay Range | 0 - 40 mmol/L | 5 - 40 mmol/L | |
| Accuracy: Corr. Coeff. | (Not explicitly stated for predicate) | 0.947 | |
| Accuracy: Slope | (Not explicitly stated for predicate) | 0.903 | |
| Accuracy: Intercept | (Not explicitly stated for predicate) | 2.444 | |
| Accuracy: SEE | (Not explicitly stated for predicate) | 0.837 | |
| Accuracy: R-Square | (Not explicitly stated for predicate) | 0.900 | |
| Linearity | Creatine Kinase | Statistically linear at 99% significance level | F-Ratio: 0.11, Slope: 1.00, Intercept: -7.45, Corr. Coeff.: 1.00 |
| Potassium | Statistically linear at 99% significance level | F-Ratio: 0.37, Slope: 1.05, Intercept: 0.03, Corr. Coeff.: 1.00 | |
| Total Carbon Dioxide | Statistically linear at 99% significance level | F-Ratio: 0.88, Slope: 1.09, Intercept: -0.71, Corr. Coeff.: 0.97 | |
| Precision | Within-Run & Total CV Ratio | (Not explicitly stated, but expected to be low) | Creatine Kinase: CV 1.5-2.0%, Potassium: CV 5.2-6.3%, Total Carbon Dioxide: CV 8.6-10.7% |
2. Sample Size Used for the Test Set and Data Provenance
- Creatine Kinase (Accuracy): n = 47 samples
- Potassium (Accuracy): n = 58 samples
- Total Carbon Dioxide (Accuracy): n = 60 samples
- Linearity: The number of data points for the linearity analysis is not explicitly stated, only the F-ratio, slope, intercept, and correlation coefficient.
- Precision: n = 120 (results pooled from 6 instruments each running 20 discs) for both within-run and total precision for each analyte.
- Sample Type Comparison: A study was conducted, but the sample size is not specified.
Data Provenance: The document implies these are prospective studies conducted by Abaxis, Inc. No information on the country of origin of the data is provided, but the company is based in Sunnyvale, CA, USA. The data is for assessing the performance of the new device against predicate devices (retrospective comparison to predicate characteristics, but prospective data collection for the new device).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
Not applicable. This study involves chemical analysis on a diagnostic device, not interpretation of images or patient data by human experts to establish ground truth. The "ground truth" for accuracy studies is typically the established reference method or the predicate device's results.
4. Adjudication Method for the Test Set
Not applicable, as this is a device performance study relying on quantitative measurements, not subjective interpretation requiring adjudication.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, an MRMC comparative effectiveness study was not done. This study focuses on the analytical performance of a diagnostic device, not human reader performance with or without AI assistance.
6. Standalone Performance
Yes, the studies describe the standalone performance of the Piccolo MetLyte 7 Reagent Disc when run on the Piccolo Point-of-Care Chemistry Analyzer. The accuracy, linearity, and precision results presented are for the algorithm (device) only.
7. Type of Ground Truth Used
The ground truth for the accuracy studies appears to be the measurements obtained from the existing legally marketed predicate devices (COBAS FARATM Chemistry System for Creatine Kinase and Total Carbon Dioxide, and Radiometer KNA™ 2 Sodium-Potassium Analyzer for Potassium). For linearity and precision, the ground truth would be ideal linear performance and expected variability of control materials, respectively.
8. Sample Size for the Training Set
The document does not explicitly mention a "training set" as it would for a machine learning algorithm. For an in vitro diagnostic device like this, the "development" or "calibration" data is implicitly used to establish the disc-specific calibration data that is embedded in a barcode. The accuracy, linearity, and precision studies represent independent validation or testing.
9. How the Ground Truth for the Training Set Was Established
Not explicitly stated in terms of a "training set" ground truth. The device relies on "bar code with disc-specific calibration data." This implies the calibration data (which serves a similar role to a training set for the device's operational parameters) would have been established through a process of running known concentration standards. The details of this process and the ground truth used for it are not provided in this summary.
{0}------------------------------------------------
AUG 26 1000
510(K) SUMMARY OF SAFETY AND EFFECTIVENESS APPENDIX E
{1}------------------------------------------------
1320 Chesapeake Terrace. Sunnyvale, Phone 408 . 734-0200 Fax 408 . 73
862.1600
Image /page/1/Picture/1 description: The image shows the word "ABAXIS" in bold, sans-serif font. A curved line starts from the top right of the letter "X", loops around the letters "I" and "S", and ends at the bottom right of the letter "S". The word and the curved line together form a logo.
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: __
1. Applicant Information:
| Date Prepared:Name:Address: | June 22, 1999Abaxis, Inc.1320 Chesapeake DriveSunnyvale, CA 94089 | |
|---|---|---|
| Contact Person:Phone Number:Fax Number: | Lisa G. McGrath(408) 745-6880(408) 734-2874 | |
| Device Information: | ||
| ClassificationTrade Name: | Class IICreatine Kinase, Potassium and Total Carbon Dioxidetest systems included on the Piccolo MetLyte 7 Reagent Disc | |
| Classification Name: | Bicarbonate/Bicarbonate test systemCreatinine Kinase test system | 862.1160862.1215 |
(Substantial equivalence for creatinine kinase, glucose and urea nitrogen, which are test systems included on the Piccolo MetLyte 7 Reagent Disc, were established previously in K934592 and K942782, therefore these test systems are not included in this summary.)
Potassium test system
{2}------------------------------------------------
ldentification of legally marketed device to which the submitter claims 3. equivalence:
The following table identifies the legally marketed device to which Abaxis claims equivalence:
| Subject Device | Predicate Device | |||
|---|---|---|---|---|
| PredicateDevice | Manufacturer | 510(k) Number | Date of SEDetermination | |
| Piccolo®Potassium TestSystem | KNA™ 2Sodium-PotassiumAnalyzer | RadiometerAmerica, Inc. | K830805 | 4/8/83 |
| Piccolo CreatineKinase and TotalCarbon DioxideTest System | RocheReagentsfor carbondioxide andcreatinekinase onthe COBASFARA™ChemistrySystem | RocheDiagnosticSystems, Inc. | CK: K834502CO2 K844987 | 2/27/841/18/85 |
4. Description of the Device:
The Piccolo MetLyte 7 Reagent Disc is designed to separate a heparinized whole blood sample into plasma and blood cells. The disc meters the required quantity of plasma and diluent, mixes the plasma with diluent, and delivers the mixture to the reaction cuvettes along the disc perimeter. The diluted plasma mixes with the reagent beads, initiating the chemical reactions that are then monitored by the analyzer. Alternately, the disc may also be used with serum.
5. Statement of Intended Use:
The Piccolo MetLyte 7 Reagent Disc run on the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of creatine kinase, creatinine, glucose, potassium, total carbon dioxide and urea nitrogen in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
6. Summary of the technological characteristics of the new device in comparison to those of the predicate device:
Tables 1 - 3 outline the technological characteristics of the Piccolo MetLyte 7 Reagent Disc (for creatine kinase, potassium and total carbon dioxide) in comparison to those of the legally marketed predicate devices.
{3}------------------------------------------------
| Table 1: Specification Comparison Creatine Kinase Test System | ||
|---|---|---|
| Piccolo® Point-of-CareChemistry Analyzer | COBAS FARATM ChemistrySystem | |
| Intended Use | quantitative analysis ofcreatine kinase | quantitative determination ofcreatine kinase |
| Methodology | isoenzymes | isoenzymes |
| Sample Type | whole blood, plasma,serum | serum |
| Sensitivity | 0.0001 A/min/U/L | 0.0001 A/min/U/L |
| Reagents | dry test-specific reagentbeads | liquid mono reagent |
| Temperature | 37°C | 37°C |
| Calibration | bar code with disc-specific calibration data | pre-set calibration |
| Assay Range | 5 - 5,000 U/L | 0 - 2,000 U/L |
| Performance Characteristics | ||
| Accuracy: | ||
| Sample size (n) | 47 | |
| Range of Samples Tested | 6.2 - 813.2 | |
| Corr. Coefficient | 0.967 | |
| Slope | 1.194 | |
| Intercept | -24.983 | |
| SEE | 9.050 | |
| R-Square | 0.934 |
Table 1: Specification Comparison Creatine Kinase Test System
Table 2: Specification Comparison: Potassium Test System
| Table 2. Specification Comparison: Potassium Test System | ||
|---|---|---|
| Piccolo Point-of-CareChemistry Analyzer | Radiometer KNA™ 2Sodium-Potassium Analyzer | |
| Intended Use | quantitative analysis ofpotassium | quantitative determination ofpotassium |
| Methodology | enzymatic activation | ion-selective electrodes |
| Sample Type | whole blood, plasma,serum | whole blood, plasma, serum,urine |
| Sensitivity | 0.024 A/min/mmol/L | 0.58 (mmol/L)/(mV) |
| Reagents | dry, test specific reagentbeads | N/A |
| Temperature | 37° C | 37° C |
| Calibration | bar code with disc-specific calibration data | automatic 1 point calibrationevery 2 hours; 2 pointcalibration every 8 hours |
| Performance Characteristics | ||
| Assay Range | 1.5 - 8.5 mmol/L | 1.0 - 99.9 mmol/L |
| Accuracy: | ||
| Sample size (n) | 58 | |
| Range of Samples Tested | 2.0 - 6.8 | |
| Corr. Coefficient | 0.969 | |
| Slope | 0.863 | |
| Intercept | 0.573 | |
| SEE | 0.141 | |
| R-Square | 0.939 |
{4}------------------------------------------------
| Piccolo® Point-of-CareChemistry Analyzer | COBAS FARA™ ChemistrySystem | |
|---|---|---|
| Intended Use | quantitative analysis oftotal carbon dioxide | quantitative determination oftotal carbon dioxide |
| Methodology | enzymatic | enzymatic using phos-phoenol pyruvate carbo-xylase and malate de-hydrogenase |
| Sample Type | wholeblood, plasma, serum | serum and plasma |
| Sensitivity | 0.0037 A/min/mmol/L | 0.0427 A/mmol/L |
| Reagents | dry test-specific reagentbeads | liquid substrate reagent 1 andliquid enzyme reagent 2 |
| Temperature | 37° C | 37° C |
| Calibration | bar code with disc-specific calibration data | Roche calibrator serumevery 6 months |
| Performance Characteristics | ||
| Assay Range | 5 - 40 mmol/L | 0 - 40 mmol/L |
| Accuracy: | ||
| Sample size (n) | 60 | |
| Range of Samples Tested | 6.1 -38.5 | |
| Corr. Coefficient | 0.947 | |
| Slope | 0.903 | |
| Intercept | 2.444 | |
| SEE | 0.837 | |
| R-Square | 0.900 |
Specification Comparison: Total Carbon Dioxide Test System Tahle 3-
7. Brief discussion of the clinical and nonclinical tests relied on for a determination of substantial equivalence.
Tables 4 - 5 summarize the results of clinical and non-clinical tests performed using the Piccolo MetLyte 7 Reagent Disc.
Linearity:
Data for creatine kinase, potassium and total carbon dioxide were found to be statistically linear at the 99% significance level by the F-test.
Table 4: Summary of Linearity
| Creatine Kinase | Potassium | Total Carbon Dioxide | |
|---|---|---|---|
| F-Ratio | 0.11 | 0.37 | 0.88 |
| Slope | 1.00 | 1.05 | 1.09 |
| Intercept | -7.45 | 0.03 | -0.71 |
| Corr. Coefficient | 1.00 | 1.00 | 0.97 |
(99% Critical F 2.99)
{5}------------------------------------------------
Precision:
Precision studies were designed to evaluate within-run and total precision of the analytes included on the Piccolo® MetLyte 7 Reagent Disc when run on the Piccolo Point-of-Care Chemistry Analyzer.
Table 5: Within-Run and Total Precision for Creatine Kinase, Potassium and Total Carbon Dioxide Run on the Piccolo Point-of-Care Chemistry Analyzer ^
| Analyte | Within-Run(n = 120) | Total(n= 120) |
|---|---|---|
| Creatine Kinase (U/L) | ||
| Moni-Trol 1Mean | 134 | 134 |
| SD | 2.7 | 2.7 |
| CV | 2.0 | 2.0 |
| Moni-Trol 2Mean | 526 | 526 |
| SD | 7.7 | 7.7 |
| CV | 1.5 | 1.5 |
| Potassium (mmol/L) | ||
| Moni-Trol 1Mean | 6.1 | 6.1 |
| SD | 0.32 | 0.35 |
| CV | 5.2 | 5.7 |
| Moni-Trol 2Mean | 4.1 | 4.1 |
| SD | 0.24 | 0.26 |
| CV | 5.9 | 6.3 |
| Total Carbon Dioxide (mmol/L) | ||
| Moni-Trol 1Mean | 21 | 21 |
| SD | 2.29 | 2.29 |
| CV | 10.7 | 10.7 |
| Moni-Trol 2Mean | 10 | 10 |
| SD | 0.90 | 0.90 |
| CV | 8.6 | 8.6 |
A Results pooled from 6 instruments each running 20 discs.
{6}------------------------------------------------
Sample Type Comparison:
A study was conducted to examine to compare venous whole blood. finger puncture whole blood and serum on the Piccolo® Point-of-Care Chemistry Analyzer.
Serum and whole blood comparability were established for each analyte and serum, whole blood and finger stick comparability were established for creatine kinase when run on the Piccolo Point-of-Care Chemistry Analyzer.
8. Conclusions
The clinical and non-clinical tests performed for creatine kinase, potassium and total carbon dioxide when run on the Piccolo Point-of-Care Chemistry Analyzer demonstrate that the test systems are as safe, effective and performs as well as the legally marketed devices identified above.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, symbolizing the department's mission to protect the health of all Americans.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
AUG 26 1999
Ms. Lisa G. McGrath Regulatory Affairs Specialist Abaxis, Inc. 1320 Chesapeake Terrace Sunnyvale, California 94089
Re: K992140
Trade Name: Piccolo® MetLyte 7 Reagent Disc Regulatory Class: I reserved Product Code: JLB Dated: June 22, 1999 Received: June 24, 1999
Dear Ms. McGrath:
.. .
We have reviewed your Section 510/k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{8}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrlv/dsma/dsmamain.html".
Sincerely yours,
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
Indications for Use
| 510(k) Number (if known): | K992140 |
|---|---|
| --------------------------- | --------- |
Device Name:
Piccolo® MetLyte 7 Reagent Disc
Intended Use:
The Piccolo MetLyte 7 Reagent Disc, used with the Piccolo Point-of-Care Chemistry Analyzer is intended to be used for the in vitro quantitative determination of creatine kinase, creatinine, glucose, potassium, total carbon dioxide and urea nitrogen in heparinized whole blood, heparinized plasma, or serum in a clinical laboratory setting or point-of-care location.
Indications for Use:
Creatine Kinase The creatine kinase assay is used for the quantitation of creatine kinase in human heparinized whole blood, heparinized plasma, or serum. Measurements of creatine kinase and its isoenzymes are used in the diagnosis and treatment of myocardial infarction, progressive muscular dystrophy, dermatomyositis, rhabdomyolysis due to drug ingestion, hyperosmolality, autoimmune disease, delirium tremens, convulsions, Crush syndrome, hypothyroidism, surgery, severe exercise, intramuscular injection, physical inactivity, decreased muscle mass.
The creatinine assay is used for the quantitation of creatinine in Creatinine human heparinized whole blood, heparinized plasma, or serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases and monitoring of renal dialysis.
Glucose The glucose assay is used for the quantitation of glucose in human heparinized whole blood, heparinized plasma, or serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders, including adult and juvenile diabetes mellitus and hypoglycemia.
The potassium assay is used for the quantitation of potassium in Potassium human heparinized whole blood, heparinized plasma, or serum. Potassium measurements are used in the diagnosis and treatment of renal glomerular or tubular disease, adrenocortical insufficiency, diabetic ketoacidosis, excessive intravenous potassium therapy, sepsis, panhypopituitarism, in vitro hemolysis, hyperaldosteronism, malnutrition, hyperinsulinism, metabolic alkalosis and gastrointestinal loss.
Total Carbon Dioxide The total carbon dioxide assay is used for the quantitation of total carbon dioxide in human heparinized whole blood, heparinized plasma, or serum. Total carbon dioxide measurements are used in the diagnosis and treatment of primary MetLyte 7 alkalosis and primary respiratory alkalosis and acidosis.
{10}------------------------------------------------
Indications for Use, Page 2
The urea nitrogen assay is used for the quantitation of urea Urea Nitrogen nitrogen in human heparinized whole blood, heparinized plasma, or serum. Urea nitrogen measurements are used in the diagnosis and treatment of renal and MetLyte 7 diseases.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use __
(Per 21 CFR 801.109)
OR
Over- The Counter Use __
(Optional Format 1-2-96)
Jev Harcey
§ 862.1210 Creatine test system.
(a)
Identification. A creatine test system is a device intended to measure creatine (a substance synthesized in the liver and pancreas and found in biological fluids) in plasma, serum, and urine. Measurements of creatine are used in the diagnosis and treatment of muscle diseases and endocrine disorders including hyperthyroidism.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.