(113 days)
No
The device description and performance studies focus on physical cleaning mechanisms (subsonic agitation, UV irradiation) and standard electronic controls with a timer. While "computer image processing" is mentioned for analyzing lens deposits, this is a separate analysis method and not integrated into the device's core function or decision-making process. There is no mention of AI, ML, or related concepts in the document.
No.
The device is indicated for the cleaning and disinfection of soft contact lenses, not for a therapeutic purpose.
No
The device is described as a cleaning and disinfection unit for soft contact lenses. Its function is to remove lens deposits and microorganisms, and to disinfect the lenses, not to diagnose a medical condition or disease.
No
The device description clearly outlines physical components like a UV light, a vibrating mechanism, an electronic control unit, a cleaning chamber, a lens holder, and a power transformer, indicating it is a hardware device with integrated software for control. While image processing is mentioned for analysis, the core device is not solely software.
Based on the provided text, the Purilense System is not an In Vitro Diagnostic (IVD) device.
Here's why:
- Intended Use: The intended use is for cleaning and disinfecting contact lenses. This is a process applied to a medical device (contact lenses) to maintain their hygiene and usability, not to perform a diagnostic test on a biological sample.
- Device Description: The device description details a system for physically cleaning and disinfecting contact lenses and the surrounding solution. It does not involve analyzing biological samples (like blood, urine, tissue, etc.) to diagnose a condition or provide information about a patient's health.
- Lack of Diagnostic Purpose: The entire description focuses on the cleaning and disinfection process and its effectiveness in removing deposits and microorganisms from the lenses and solution. There is no mention of using the device to detect, measure, or analyze any substance or characteristic in a biological sample for diagnostic purposes.
- Performance Studies: The performance studies focus on the safety of the device components, the stability of contact lenses after cycling, and the microbiological efficacy of the disinfection process. These are all related to the function of a cleaning and disinfection system, not a diagnostic test.
- Predicate Devices: The predicate devices listed are other contact lens cleaning and disinfection systems, further supporting that this device falls within that category, not IVDs.
While the device description mentions "computer image processing quantification of visible light reflections of hydrogel contact lens deposits," this is used to evaluate the effectiveness of the cleaning process on the lenses, not to perform a diagnostic test on a biological sample.
Therefore, the Purilense System is a medical device for cleaning and disinfecting contact lenses, not an In Vitro Diagnostic device.
N/A
Intended Use / Indications for Use
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
Additional Claims: Not Found
Product codes
LYL, LPN
Device Description
The device is a cleaning and disinfection unit for contact lenses. It consists of the Purilens Unit and Purilens Solution. The Purilens Unit is compact, with an upper housing containing a germicidal UV light, and a lower housing which contains a removable, clear cleaning chamber with a lens holder for the contact lenses. A mechanism in the lower portion of the cleaning chamber creates the cleaning turbulence activity. Within the lower housing unit is an electronic control unit which operates the UV light and turbulence vibrating mechanism. The lower unit contains an internal automatic timing device which shuts the unit after its 15 minute cycle.
The lens holder consists of a removable basket which houses the contact lenses in separate chambers and is suspended within the cleaning chamber in the lower housing. The removable basket assembly is shielded from the UV light. The cleaning chamber contains an unpreserved saline medium as a storage solution for the contact lenses.
The Purilens Solution is a preservative free saline solution which is used to fill the cleaning chamber. The turbulence mechanism is a magnetic pedal suspended on the end of a flexible spring located in the bottom of the cleaning chamber. When activated by the electronic connection, the pedal oscillates in a subsonic range of 50 - 120 cycles per second creating a whirlpool effect that swirls the Purilens Solution surrounding the lens baskets and removes debris and microorganisms from the lenses during the 15 minute cycle.
UV radiation energy from the upper housing disinfects loosened microorganisms in the solution. Disinfection is accomplished without heating, preservatives, or chemicals. The unit cannot function unless the electrical interlock is complete between the upper and lower housings.
An electrical transformer is located on the power cord. The transformer converts house current of 110 VAC to 14 VAC for the purpose of operating the device.
Mentions image processing
A lens cleaning analysis using computer image processing quantification of visible light reflections of hydrogel contact lens deposits was performed on all lenses worn in the clinical study.
Mentions AI, DNN, or ML
Not Found
Input Imaging Modality
Not Found
Anatomical Site
Not Found
Indicated Patient Age Range
Not Found
Intended User / Care Setting
Not Found
Description of the training set, sample size, data source, and annotation protocol
Not Found
Description of the test set, sample size, data source, and annotation protocol
Not Found
Summary of Performance Studies
A clinical study was performed with the device and a control with a comparable multipurpose contact lens cleaning and disinfection system. A crossover design study was employed in which all subjects experienced both the Purilens® System and the multipurpose disinfection control for three months each. The results of the clinical study indicate that the Purilens® System performs equally as a multipurpose contact lens solution for the same indication. The effect of the Purilense System indicates that the performance is superior to a comparable multipurpose cleaning and disinfection system.
Key Metrics
Not Found
Predicate Device(s)
ReNu Multipurpose Solution P860023/S6, The Lens Comfort, Inc. Ultrasonic Contact Lens Care Accessory, K962112, Sola/Barnes -Hind Soft Mate Automatic Cleaning Unit, K852386, Clensatron 700CL. K884414
Reference Device(s)
Not Found
Predetermined Change Control Plan (PCCP) - All Relevant Information
Not Found
§ 886.5928 Soft (hydrophilic) contact lens care products.
(a)
Identification. A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.(b)
Classification. Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
0
Image /page/0/Picture/1 description: The image shows a logo for "Purilens COMPLETE CONTACT LENS CARE". The logo features the word "Purilens" in a bold, sans-serif font, with the words "COMPLETE CONTACT LENS CARE" in a smaller font below it. Above the word "Purilens" is a graphic of a circle.
PuriLens, Inc.
12170 Race Track Rd. Tampa, FL 33626
K991965
510(k) Summary
Tel: 813 814-4141 Fax: 813 814-4545
SUBMITTER:
Submitted on behalf of:
Company Name: Address:
Phone: Fax:
Purilens, Inc.. 12170 Race Track Road Tampa, Florida 33626 (813) 814 4141 (813) 814 4545
CONTACT PERSON:
Richard E. Lippman, O.D., F.A.A.O. Official Representative and Correspondent C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway Suite 625 Rockville, MD 20852
DATE SUMMARY PREPARED: August 26, 1999
TRADE NAME: | PurilensR System |
---|---|
COMMON NAME: | contact lens cleaning and disinfection device |
DESCRIPTION of the DEVICE:
The device is a cleaning and disinfection unit for contact lenses. It consists of the Purilens Unit and Purilens Solution. The Purilens Unit is compact, with an upper housing containing a germicidal UV light, and a lower housing which contains a removable, clear cleaning chamber with a lens holder for the contact lenses. A mechanism in the lower portion of the cleaning chamber creates the cleaning turbulence activity. Within the lower housing unit is an electronic control unit which operates the UV light and turbulence vibrating mechanism. The lower unit contains an internal automatic timing device which shuts the unit after its 15 minute cycle.
The lens holder consists of a removable basket which houses the contact lenses in separate chambers and is suspended within the cleaning chamber in the lower housing. The removable basket assembly is shielded from the UV light. The cleaning chamber contains an unpreserved saline medium as a storage solution for the contact lenses.
1
The Purilens Solution is a preservative free saline solution which is used to fill the cleaning chamber. The turbulence mechanism is a magnetic pedal suspended on the end of a flexible spring located in the bottom of the cleaning chamber. When activated by the electronic connection, the pedal osillates in a subsonic range of 50 - 120 cycles per second creating a whirlpool effect that swirls the Purilens Solution surrounding the lens baskets and removes debris and microorganisms from the lenses during the 15 minute cycle.
UV radiation energy from the upper housing disinfects loosened microorganisms in the solution. Disinfection is accomplished without heating, preservatives, or chemicals. The unit cannot function unless the electrical interlock is complete between the upper and lower housings.
An electrical transformer is located on the power cord. The transformer converts house current of 110 VAC to 14 VAC for the purpose of operating the device.
INDICATIONS FOR USE:
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
SUBSTANTIAL EQUIVALENCE:
The Purilense System for Cleaning and Disinfecting Contact Lenses is substantially equivalent to multipurpose contact lens solution accessory products (ReNu Multipurpose Solution P860023/S6, approved January 24, 1994; and similarly approved multipurpose products) which include ingredients for cleaning and disinfecting contact lenses.
The Purilens System is comprised of a two step system. The cleaning step is accomplished by subsonic turbulence created by an agitator wand which is activated by a magnet in close proximity to the wand. This physical agitation of lens solution which aids in removal of lens deposits and microorganisms is substantially equivalent to contact lens ultrasonic cleaners and hand-driven mechanical agitation units (The Lens Comfort, Inc. Ultrasonic Contact Lens Care Accessory, K962112, cleared November 29, 1996; the Sola/Barnes -Hind Soft Mate Automatic Cleaning Unit, K852386, cleared 9/4/85; and the Clensatron 700CL. K884414 cleared 12/29/88). The turbulence created by this action loosens lens deposits and microorganisms from the lens surfaces and places those deposits and microorganisms into the surrounding storage solution, a non-preserved sterile saline. The disinfection step is
2
accomplished by action of ultraviolet irradiation. This activity is created by the presence of an ultraviolet light source in the device's upper housing. As the UV irradiation energy is emitted, it acts on the loosened microorganisms to reduce the organisms viability.
This device is substantially equivalent to ReNu Multipurpose Solution by Bausch & Lomb, approved under Premarket Approval P860023/S1 on July 28, 1988, and other multipurpose soft contact lens solutions with similar indications. ReNu Multipurpose Solution includes in its formulation, ingredients which clean contact lenses with the aid of digital rubbing, and other ingredients which serve as disinfecting agents when lenses are stored for a minimum of four hours or longer.
The equivalence determination is made by the comparison of actions of the relative devices within their indicated uses. The principal difference between the Purilens® System and ReNu Multipurpose Solution for cleaning is the cleaning step for the Purilens® System does not require digital rubbing, but acts by shear forces from subsonic action of the saline solution surrounding the lenses, created by a vibrating agitator. The agitator wand vibrates with sufficient intensity to remove deposits and microorganisms. The cleaning action is equivalent to a series of mechanical cleaning products including The Lens Comfort Contact Lens Care Accessory: the Sola/Barnes-Hind Soft Mate Automatic Cleaning Unit, and the Clensatron, all of which clean lenses by mechanical agitation, either by ultrasonic wave form action of the storage solution, or by mechanical agitation of storage solution in a saline held chamber.
The rinse process for the devices are similar in that the Purilense System rinse is accomplished by the continued agitation of the solution during the complete cycle process of the device. This action acts as a continued rinse as the lens is continually "washed" by the action of the subsonic vibration of the storage medium (saline). The equivalent action of rinsing "washing" lenses occurs with the comparable products using mechanical action including the aforementioned devices. This rinse step is further enhanced by a steady stream of saline rinse prior to setting the lenses in the unit to complete the agitation process.
The disinfection step for the Purilens System is accomplished by ultraviolet irradiation from the UV bulb housed in the upper housing of the device. The energy emitted by the UV bulb acts on the loosened microorganisms as they are circulating about in the saline storage solution in the carrying case chamber due to continued vibrating subsonic action on the solution medium. The effectiveness of disinfection is complete after 15 minutes in that no microorganisms are recovered after the disinfection cycle is complete on the lenses or in the chamber for up to 24 hours after the cycle is completed. The disinfection step for ReNu Multipurpose Solution is a minimum four hour soak after the lenses are stored in a carrying case with the solution added as a storage medium.
3
COMPATIBILITY OF THE DEVICE:
A series of pre-clinical studies were performed on the Purilens® System device in which toxicology studies were performed on the plastic components of the device and determined to be non-toxic and safe for use with contact lenses. The tests conducted included an evaluation of toxicity of residual solution after cycling; a toxicity evaluation by the MEM Elution Method; systemic toxicity in mice; ocular irritation study of extracts in rabbitrs. A cytotoxicity study was also conducted on the plastic coated agitator found at the base of the portable lens carrier. Solution compatibility studies were performed after 180 cycles of the device on Groups I, III, III, and IV soft contact lenses to determine the stability of lens parameters after cycling and found to have no negative affect on the lens stability or compatibility with the storage solution. Microbiological studies were performed on the lenses, storage solution and lens case to determine whether the device performed the disinfection phase of the cycle, and were determined to have performed all tests successfully. Specific tests which were performed included: disinfection efficacy or "D" value test; multi-item challenge test; evaluation of pH and temperature of the Purilens System chamber solution after cycle completion: and an evaluation of microbial accumulation within the unit (biofilm study).
A lens cleaning analysis using computer image processing quantification of visible light reflections of hydrogel contact lens deposits was performed on all lenses worn in the clinical study. The study documented that the lenses were successfully cleaned and free of deposits including proteins, lipids, and mucins.
CLINICAL ANALYSIS
A clinical study was performed with the device and a control with a comparable multipurpose contact lens cleaning and disinfection system. A crossover design study was employed in which all subjects experienced both the Purilens® System and the multipurpose disinfection control for three months each. The results of the clinical study indicate that the Purilens® System performs equally as a multipurpose contact lens solution for the same indication. The effect of the Purilense System indicates that the performance is superior to a comparable multipurpose cleaning and disinfection system.
4
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 1999
Purilens. Inc. c/o Richard E. Lippman, O.D., F.A.A.O. C.L. McIntosh & Associates, Inc. 12300 Twinbrook Parkway, Suite 625 Rockville, MD 20852
Re: K991965
. ……
Trade Name: Purilens® System (consisting of the Purilens Unit and Purilens Solution, a preservative free saline)
Regulatory Class: II Product Codes: LYL and LPN Dated: August 26, 1999 Received: August 27, 1999
Dear Dr. Lippman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
5
This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.
If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its tollfree number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Alvarez L. Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
6
Indications Statement
510(k) Number (if known) K991965_
Device Name: _ Purilens® System
. .
Indications for Use: (Revised: 8/26/99)
The Purilense System is indicated for the cleaning and disinfection of soft (hydrophilic) contact lenses by means of subsonic agitation to remove lens deposits and microorganisms, and ultraviolet irradiation of the surrounding storage solution for disinfection. The lenses may be stored in the disinfected solution for a period of 24 hours following the cleaning and disinfection cycle.
Additional Claims:
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Presription Use
OR
Over-The-Counter Use
(Oprtional Format 1-2-96)
Karin Warburton
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K991965