K Number
K991880
Manufacturer
Date Cleared
1999-11-09

(160 days)

Product Code
Regulation Number
866.3305
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Light Diagnostics SimulFluor® HSV1/2 Immunofluorescence Assay is a direct immunofluorescence test intended for the detection and identification of herpes simplex virus type 1 (HSV-1) or herpes simplex virus type 2 (HSV-2) following amplification in cell culture or by direct examination of clinical specimens prepared by cytocentrifugation. Specimens found to be negative on direct specimen examination should be tested by cell culture.

For in vitro diagnostic use.

Device Description

Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay utilizes a single reagent for the simultaneous detection and identification of HSV-1 and HSV-2. The SimulFluor® HSV 1/2 Reagent consists of two components; the primary component specific for HSV-1 will bind to the glycoprotein C and a capsid-associated protein in HSV-1 infected cells, while the secondary component, specific for HSV-2, will bind to the glycoprotein G in HSV-2 infected cells. Unbound reagent is removed by rinsing with phosphate-buffered saline (PBS). Illumination with ultraviolet light allows visualization of the antigenantibody complexes by fluorescence microscopy. When an FITC filter set is used, HSV-1- infected cells will exhibit apple-green fluorescence and HSV-2infected cells will exhibit yellow-gold fluorescence. The uninfected cells will stain a dull red due to the presence of Evans blue in the SimulFluor® HSV 1/2 reagent.

A blend of monoclonal antibodies directed against HSV-1 and HSV-2 is used in the Light Diagnostics SimulFluor® HSV 1/2 reagent. The use of monoclonal antibodies ensures increased specificity of reagent and reduces the risk of nonspecific background or interference.

AI/ML Overview

This document describes the acceptance criteria and study details for the Light Diagnostics SimulFluor® HSV1/2 Immunofluorescence Assay.

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria are implied by the comparison to predicate devices and the confidence intervals reported for sensitivity, specificity, and percent agreement, indicating that the device's performance should be comparable to or ideally exceed these established methods. The "Clinical Study: Site 1" and "Clinical Study: Site 2" sections provide the reported device performance. Given the type of device (an immunofluorescence assay comparing to similar assays), the primary performance metrics are sensitivity, specificity, and percent agreement.

Performance MetricAcceptance Criteria (Implied)Reported Device Performance (Site 1 - Direct Specimen vs. Culture)Reported Device Performance (Site 1 - vs. Comparative Device)Reported Device Performance (Site 2 - Shell Vials vs. Comparative Device)Reported Device Performance (Site 2 - Culture Plates vs. Comparative Device)
HSV-1 DetectionHigh sensitivity, specificity, and agreement compared to predicate.Sensitivity: 89.5% (17/19), Specificity: 100% (128/128), Agreement: 98.6%Percent Agreement: 100%Percent Agreement: 96%Percent Agreement: 100%
HSV-2 DetectionHigh sensitivity, specificity, and agreement compared to predicate.Sensitivity: 92.6% (25/27), Specificity: 100% (120/120), Agreement: 98.6%Percent Agreement: 100%Percent Agreement: 100%Percent Agreement: 100%

Note on Acceptance Criteria: The document does not explicitly state numerical acceptance criteria in a dedicated section. Instead, the "Conclusions drawn from evaluations" section states that the device's performance characteristics "were shown to be substantially equivalent to those of Bartels HSV Typing Test and the DPC PathoDx® Herpes Typing kit." Therefore, the reported performance metrics (sensitivity, specificity, and percent agreement with tight confidence intervals) meeting or exceeding those of the predicate devices are the implicit acceptance criteria.

2. Sample Size Used for the Test Set and Data Provenance

  • Clinical Study: Site 1

    • Test Set Sample Size: 191 specimens were submitted. After exclusions (40 insufficient cells, 1 contaminated culture, 3 VZV positive), the evaluable sample size for direct specimen testing was 147 (19+128 for HSV-1, 27+120 for HSV-2). For comparison against a comparative device, 33 HSV-1 and 29 HSV-2 isolates were identified by both reagents.
    • Data Provenance: North-central United States, retrospective (specimens "were submitted" indicating existing samples).
  • Clinical Study: Site 2

    • Test Set Sample Size:
      • Shell vials: 214 specimens. 24 HSV-1 and 37 HSV-2 isolates were detected.
      • Culture plates: 227 specimens. 32 HSV-1 and 46 HSV-2 isolates were detected.
    • Data Provenance: Southwestern United States, retrospective (specimens "were submitted" indicating existing samples).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

The document does not specify the number or qualifications of experts used to establish the ground truth. It refers to "clinical virology laboratory" and "reference laboratory" and implies standard laboratory procedures for culture confirmation.

4. Adjudication Method for the Test Set

The document does not describe an adjudication method for reconciling disagreements. For Site 1, the "culture confirmation" was used as the reference against which the direct specimen testing was compared. For the comparison against predicate devices, it appears a straightforward comparison of results was performed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve With AI vs Without AI Assistance

This section is not applicable as the described device is an immunofluorescence assay for detecting viruses, not an AI-assisted diagnostic tool interpreted by human readers. It's a laboratory test where a technician observes fluorescence, not a system that improves human diagnostic performance via AI.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

This section is not applicable. The device is an immunofluorescence assay that requires manual preparation, staining, and microscopic observation by a trained laboratory technician. It is not an automated algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The ground truth used was cell culture (viral culture), which is a gold standard for detecting and identifying HSV-1 and HSV-2. In Site 1, "culture confirmation" was used. In Site 2, "standard cultures" and "spin-amplified shell vials" were used as reference methods.

8. The Sample Size for the Training Set

The document does not explicitly describe a separate "training set" for the clinical evaluation. The non-clinical evaluation section mentions that antibodies were "characterized for their ability to detect HSV types 1 and 2" using "reference viral strains and clinical isolates," but this is more akin to initial assay development and validation rather than a distinct training set for an algorithm.

9. How the Ground Truth for the Training Set Was Established

Given the absence of an explicit "training set" as understood for machine learning algorithms, this question is not applicable in the context of this immunofluorescence assay submission. The "characterization" of the antibodies relied on testing with "reference viral strains and clinical isolates" whose identities would have been established through standard virological methods.

{0}------------------------------------------------

510(k) Summary

Submitter:Light Diagnostics
28835 Single Oak Drive
Temecula, CA 92590
Tel: 909/676-8080
Fax: 909/676-9209

Contact Person: Cindv Pennv

Product Name

Trade Name: Light Diagnostics SimulFluor® HSV1/2 Common Name: Immunofluorescence Assay Classification Name: Herpes simplex virus Classification Number: 866.3305-83LKC

Intended Use:

Light Diagnostics SimulFluor® HSV1/2 Immunofluorescence Assay is a direct immunofluorescence test intended for the detection and identification of herpes simplex virus type 1 (HSV-1) or herpes simplex virus type 2 (HSV-2) following amplification in cell culture or by direct examination of clinical specimens prepared by cytocentrifugation. Specimens found to be negative on direct specimen examination should be tested by cell culture.

For in vitro diagnostic use.

Predicate Devices:

  • Bartels Herpes Simplex Virus Type-Specific Fluorescent Monoclonal 1) Antibody Test (HSV Typing) The Bartels HSV Typing Test is intended for use in the qualitative identification and differentiation of herpes simplex virus (HSV) type 1 and type 2 in inoculated cell cultures. For in vitro diagnostic use.
  • DPC PathoDx® Herpes Typing Kit 2)

The DPC PathoDx Herpes Typing Kit is an immunofluorescence test designed for the detection and typing of herpes simplex virus types I and II (HSV-I and HSV-II) in direct clinical specimens and following growth in tissue culture. All direct clinical specimens which are negative or have inadequate numbers of cells must be reevaluated by cell culture. For in vitro diagnostic use.

{1}------------------------------------------------

Device description:

Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay utilizes a single reagent for the simultaneous detection and identification of HSV-1 and HSV-2. The SimulFluor® HSV 1/2 Reagent consists of two components; the primary component specific for HSV-1 will bind to the glycoprotein C and a capsid-associated protein in HSV-1 infected cells, while the secondary component, specific for HSV-2, will bind to the glycoprotein G in HSV-2 infected cells. Unbound reagent is removed by rinsing with phosphate-buffered saline (PBS). Illumination with ultraviolet light allows visualization of the antigenantibody complexes by fluorescence microscopy. When an FITC filter set is used, HSV-1- infected cells will exhibit apple-green fluorescence and HSV-2infected cells will exhibit yellow-gold fluorescence. The uninfected cells will stain a dull red due to the presence of Evans blue in the SimulFluor® HSV 1/2 reagent.

A blend of monoclonal antibodies directed against HSV-1 and HSV-2 is used in the Light Diagnostics SimulFluor® HSV 1/2 reagent. The use of monoclonal antibodies ensures increased specificity of reagent and reduces the risk of nonspecific background or interference.

Technological Comparison of Methods:

The Light Diagnostics SimulFluor® HSV1/2 Immunofluorescence Assay is substantially equivalent to Bartels HSV Typing and the DPC PathoDx® Herpes Typing DFA reagents:

  • A. All three methods are intended for use in the detection of HSV-1 and HSV-2 antigens in infected cells.
  • B. All three methods are in vitro test methods.
  • C. All three methods use a direct immunofluorescence assay procedure for staining of slides using FITC filter sets.
  • D. Two of the three methods (PathoDx® and SimulFluor®) are intended for the detection of HSV-1 and HSV-2 in direct patient specimens.

{2}------------------------------------------------

The methods differ in that:

  • A. The Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay consists of one reagent which contains specific monoclonal antibodies directed against HSV-1 and HSV-2 and tagged with two different fluorescent labels. This allows simultaneous visualization and identification of both HSV-1 and HSV-2-infected cells in one well. Both PathoDx® and Bartels kits consist of two separate reagents, each of which contains FITC-labeled monoclonal antibodies directed against either HSV-1 or HSV-2. Two separate wells are necessary to detect and identify both viruses in one sample.

SimulFluor® HSV 1/2 Diagnostics Data for Light Performance Immunofluorescence Assay:

    1. Non-clinical evaluation:
      The conjugated monoclonal antibodies used in the Light Diagnostic SimulFluor® HSV 1/2 reagent were characterized for their ability to detect HSV types 1 and 2. These antibodies reacted with either HSV-1 or HSV-2 when tested with reference viral strains and clinical isolates. The conjugated monoclonal antibodies were also tested against a variety of viruses and bacteria including those that present as herpetic-like lesions, and cell lines commonly used to isolate HSV-1 and HSV-2. No crossreactivity was observed.
    1. Clinical evaluation:
      Clinical studies were conducted at two clinical sites. At site one (Site 1) in the north-central United States, the staining of the Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay on direct specimens was compared to culture confirmation using the Predicate Device for the detection and identification of HSV-1 and HSV-2. At a second site in the southwestern United States (Site 2), the Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay was compared to the Predicate Device in spin-amplified shell vials and standard culture for the detection and identification of HSV-1 and HSV-2. The Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay was compared to the Bartels HSV Typing Test at Site 1 and the DPC PathoDx® Herpes Typing kit at Site 2.

{3}------------------------------------------------

Clinical Study: Site 1

One hundred and ninety-one specimens were submitted to a clinical virology laboratory for the detection of HSV. Slides from patient specimens were prepared by cytocentrifugation for direct specimen testing and inoculated into standard tube culture for the isolation of HSV. The direct specimen and culture confirmation results obtained by the SimulFluor® HSV 1/2 reagent was compared to the culture confirmation stained with Bartel's HSV Typing Reagent. Of the 191 specimens submitted, forty samples had insufficient cells for direct evaluation, one culture was contaminated, and three samples were found positive for varicella zoster virus.

Compared to culture, direct specimen testing using the SimulFluor® HSV 1/2 reagent had a sensitivity of 89.5% (17/19) (95% Confidence Interval of 66.6% to 98.7%, Exact) and specificity of 100% (128/128) (95% Confidence Interval of 97.2% to 100%, Exact) with a percent agreement of 98.6% (95% Confidence Interval of 95.2% to 99.8%, Exact) for the detection of HSV-1; and a sensitivity of 92.6% (25/27) (95% Confidence Interval of 75.7% to 99.1%, Exact) and specificity of 100% (120/120) (95% Confidence Interval of 97% to 100%. Exact) with a percent agreement of 98.6% (95% Confidence Interval of 95.2% to 99.8%. Exact) for the detection of HSV-2.

Thirty-three HSV-1 and 29 HSV-2 isolates were identified with both reagents. When compared to the Comparative Device, the SimulFluor® HSV 1/2 reagent had 100% correlation for the detection of both HSV-1 and HSV-2 for a percent agreement of 100% for both HSV-1 and HSV-2. The 95% confidence intervals were 98% to 100%. Exact.

Clinical Study: Site 2

Two hundred and twenty-eight specimens were submitted to a reference laboratory for HSV identification. Two hundred and twenty-seven specimens were cultured in standard cultures using 24-well cluster plates and 214 specimens were cultured in spin-amplified shell vials. The SimulFluor® HSV 1/2 reagent was compared to DPC's PathoDx® Herpes Typing Kit for both culture methods.

Two hundred and fourteen shell vials were inoculated and stained with both the SimulFluor® HSV 1/2 and the Comparative Device. Twenty-four HSV-1 and thirty-seven HSV-2 isolates were detected with the Comparative Device. The SimulFluor® HSV 1/2 reagent identified an additional isolate which was not detected in standard culture. The percent agreement for the detection of HSV-1 by the SimulFluor® HSV 1/2 reagent is 96% (95% Confidence Interval of 79.6% to 99.9%. Exact). For the detection of HSV-2 the percent agreement was 100%.

{4}------------------------------------------------

Two hundred and twenty-seven specimens were inoculated in 24-well culture plates for isolation of herpes, and stained with SimulFluor® HSV 1/2 and the Comparative Device. Thirty-two isolates of HSV-1 and 46 isolates of HSV-2 were detected in culture by both reagents. There was 100% agreement between the two reagents for the detection of HSV-1 and HSV-2 with a percent agreement of 100% for the detection of HSV-1 and HSV-2 by the SimulFluor® HSV1/2 reagent. Overall Percent Agreement was 100% (227/227) [95% Confidence Interval of 98.4 to 100%. Exactl.

    1. Conclusions drawn from evaluations:
      Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay uses a standard direct immunofluorescence assay procedure for the detection of HSV-1 and HSV-2 in patient specimens and in cell culture. The monoclonal antibodies used in the reagent have been characterized so as to ensure specificity and reliability of the product. In clinical evaluations, the performance characteristics of the reagent were shown to be substantially equivalent to those of Bartels HSV Tvping Test and the DPC PathoDx® Herpes Typing kit.

The characterization and clinical evaluation of the Light Diagnostics SimulFluor® HSV 1/2 Immunofluorescence Assay demonstrates the safety and effectiveness of this product when used as intended in the product insert.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles three stylized human profiles facing to the right, with flowing lines beneath them.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

NOV - 9 1999

Ms. Cindy Penny Manager, Quality Assurance Chemicon International, Inc. 28835 Single Oak Drive Temecula, California 92590

Re: K991880 Trade Name: Light Diagnostics SimulFluor™ HSV 1/2 Immunofluorescence Regulatory Class: III Product Code: GQN Dated: August 31, 1999 Received: September 2, 1999

Dear Ms. Penny:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours.

Steven Butman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Number (if known):

Light Diagnostics Simulfluor™ HSV 1 / 2 Immunofluorescence Device Name: Assay

Light Diagnostics SimulFluor™ HSV1/2 Indications For Use: Immunofluorescence Assay is a direct immunofluorescence test intended for the detection and identification of herpes simplex virus type 1 (HSV-1) and herpes simplex virus type 2 (HSV-2) following amplification in cell culture or by direct examination of clinical specimens prepared by cytospin. Specimens found to be negative on direct specimen examination should be tested by cell culture. For in vitro diagnostic use.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Woody Dubbs (Division Sign-Off)
-------------------------------------------------------------------------

Division of Clinical Laboratory Devices

510(k) NumberK991880
------------------------
Prescription UseXOROver-The-Counter-Use
(Per 21 CFR 801.109)

§ 866.3305 Herpes simplex virus serological assays.

(a)
Identification. Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.(b)
Classification. Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).