K Number
K991709
Manufacturer
Date Cleared
1999-10-08

(142 days)

Product Code
Regulation Number
880.2800
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Browne Cidex™ OPA Indicator is a concentration monitor for use in ortho-phthalaldehyde-containing germicide solutions with a minimum effective concentration of 0.3%.

The Browne Cidex™ OPA Indicator is dedicated for use with Cidex™ OPA Solution.

Device Description

The Browne OPA Indicator and the substantially equivalent device are chemical indicator strips intended to monitor the concentration of liquid chemical germicides. The devices indicate, via a color change, if the concentration of the active ingredient exceeds the MEC (minimum effective concentration) for the solution.

The Browne OPA Indicator consists of a polypropylene strip with an indicator pad on one end. The indicator pad is impregnated with an indicator solution which changes color from light blue to purple in Cidex™ OPA Solution with an ortho-phthaladehyde concentration greater than the MEC for the solution.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study details for the Browne Cidex™ OPA Indicator, based on the provided text:

1. Acceptance Criteria and Device Performance

Acceptance CriteriaReported Device Performance
Indicate "FAIL" (light blue to light blue with purple mottling) when ortho-phthalaldehyde concentration is ≤ 0.3% (MEC).All 90 strips tested in 0.3% solution showed a "FAIL" condition (light blue or light blue with purple mottling).
Indicate "PASS" (purple color) when ortho-phthalaldehyde concentration is > 0.3% (specifically tested at 0.45%).All 90 strips tested in 0.45% solution showed a "PASS" condition (purple color).
Performance under simulated use conditions: Indicate "FAIL" at 0.3% MEC.All 30 indicators tested in 0.3% ortho-phthalaldehyde solution showed a "FAIL" condition.
Performance under simulated use conditions: Indicate "PASS" at 0.45% above MEC.All 30 indicators tested in 0.45% solutions showed a "PASS" condition.
No false positives or false negatives under simulated use conditions.No false positives or false negatives were recorded under simulated use conditions.
Performance under worst-case germicide solution conditions (end-of-use life, contaminated with 5% serum, 5% salt, and 5% sodium dodecyl sulfate): Indicate "FAIL" at 0.3% MEC.For each contaminant tested (serum, salt, SDS), all indicator strips exhibited a "FAIL" condition in the 0.3% Cidex™ OPA solution.
Performance under worst-case germicide solution conditions: Indicate "PASS" at 0.45% above MEC.For each contaminant tested, all indicator strips exhibited a "PASS" condition in the 0.45% solution.

2. Sample Size and Data Provenance

  • Test Set Sample Size:
    • Initial Testing: 180 Browne OPA Indicators (90 for 0.3% solution, 90 for 0.45% solution).
    • Simulated Use Testing: 60 indicators (30 for 0.3% solution, 30 for 0.45% solution).
    • Worst-Case Conditions Testing: Not explicitly stated, but implies a sufficient number of strips for each contaminant at both 0.3% and 0.45% concentrations, and included an unadulterated control. The wording "all of the indicator strips" suggests comprehensive testing for each scenario.
  • Data Provenance: The document does not explicitly state the country of origin for the data. Given the submitter's address in the United Kingdom, it is plausible the testing was conducted there, or at a contract lab. The study is prospective, as it involves testing new batches of the device under defined conditions.

3. Number of Experts and Qualifications for Ground Truth

The device is a chemical indicator that produces a color change based on chemical concentration. The "ground truth" in this context is the actual chemical concentration of ortho-phthalaldehyde in the solutions. Therefore, there are no human experts involved in establishing the ground truth for the test set in the traditional sense of medical image interpretation or clinical diagnosis. The ground truth is established by chemical measurement of the solution concentration.

4. Adjudication Method for the Test Set

Not applicable. The "readings" are objective color changes (light blue/mottling for FAIL, purple for PASS) directly observed from the chemical reaction. There is no subjective interpretation requiring adjudication by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

No. This type of study is typically performed for diagnostic devices where human readers provide interpretations. The Browne OPA Indicator is a chemical concentration monitor, not a diagnostic imaging or interpretive device that would involve human readers in its primary function.

6. Standalone (Algorithm Only) Performance Study

Yes, the performance testing described is a standalone performance study. The device (chemical indicator strip) operates independently to produce a color change based on the chemical environment. There is no human-in-the-loop during the indication process itself, only for observing and recording the resulting color.

7. Type of Ground Truth Used

Chemical Concentration Measurement. The ground truth for the test solutions (0.3% and 0.45% ortho-phthalaldehyde) is based on the precisely controlled preparation and presumed subsequent measurement of these chemical concentrations.

8. Sample Size for the Training Set

Not applicable. The Browne OPA Indicator is a chemical indicator, not a machine learning or AI-based device, so it does not have a "training set" in the context of an algorithm. Its "learning" is inherent in its chemical formulation.

9. How Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device. The chemical formulation of the indicator is developed based on known chemical reactions and thresholds for ortho-phthalaldehyde.

{0}------------------------------------------------

8 1999

OCT

MDCI - NC

18

K 991709

P 1 of 3

510(k) SUMMARY Browne Cidex™ OPA Indicator

August 4, 1999

1. SUBMITTER NAME AND ADDRESS

Mr. Alan Charlton Albert Browne Ltd. Chancerv House 190 Waterside Road Hamilton Industrial Park Leicester LE5 1QZ United Kingdom

2. DEVICE NAME

Browne Cidex™ OPA Indicator Proprietary Name: ortho-Phthalaldehyde Concentration Indicator Common/Usual Name: Physical/Chemical Sterilization Process Indicator Classification Name:

PREDICATE DEVICES 3.

Browne GA Indicator, subject of K922481

4. INTENDED USE

The Browne Cidex™ OPA Indicator (Browne OPA Indicator) is a concentration monitor for use in ortho-phthalaldehyde-containing germicide solutions with a minimum effective concentration of 0.3%.

The Browne Cidex™ OPA Indicator is dedicated for use with Cidex™ OPA Solution.

DEVICE DESCRIPTION ട്.

The Browne OPA Indicator and the substantially equivalent device are chemical indicator strips intended to monitor the concentration of liquid chemical germicides. The devices indicate, via a color change, if the concentration of the

{1}------------------------------------------------

PAGE 19

K991709

p2 of 3

active ingredient exceeds the MEC (minimum effective concentration) for the solution.

The Browne OPA Indicator consists of a polypropylene strip with an indicator pad on one end. The indicator pad is impregnated with an indicator solution which changes color from light blue to purple in Cidex™ OPA Solution with an ortho-phthaladehyde concentration greater than the MEC for the solution.

6. TECHNOLOGICAL CHARACTERISTICS

The technological characteristics of the Browne GA Indicator, subject of K922481, are equivalent to that of the Browne OPA Indicator described in this Both devices are non-sterile, disposable strips containing an submission. indicator pad impregnated with an indicator solution which changes color in a liquid chemical germicide at the appropriate concentration of the active ingredient.

The mechanism of action for the color change is identical for the proposed and predicate devices. The active ingredient of the indicator solution in both devices interacts with the aldehyde moiety of the active ingredient of the germicide solution to produce a color change at the appropriate concentration of the germicide active ingredient. The indicator solution of the proposed device changes color from light blue to purple, while that of the Browne GA Indicator, subject of K922481, changes color from yellow to purple,

7. PERFORMANCE TESTING

Performance testing was conducted on 180 Browne OPA Indicators from 3 production lots in Cidex™ OPA solutions containing 0.3% and 0.45% orthophthaladehyde. Testing was performed according to the Instructions for Use. All 90 strips tested in the 0.3% solution were a light blue or light blue with purple mottling color, indicating a FAIL. All 90 strips tested in the 0.45% solution were a purple color, indicating a PASS.

Additional testing was conducted to evaluate the performance of the Browne OPA Indicator in Cidex™ OPA Solutions containing 0.3% or 0.45% orthophthalaldehyde using three production lots of strips under conditions of simulated

{2}------------------------------------------------

PAGE 20

K 991709

p. 3 of 3

use. All 30 of the indicators tested in the 0.3% ortho-phthaladehyde solution showed a FAIL condition. All 30 of the indicators tested in the 0.45% solutions showed a PASS condition. No false positives or false negatives were recorded.

The performance of the Browne OPA Indicator was also evaluated under conditions selected to simulate a worst case condition for the germicide solution. The indicators were tested in Cidex™ OPA solution containing 0.3% and 0.45% ortho-phthalaldehyde at the end of it's use life and contaminated with 5% serum, 5% salt, and 5% sodium dodecyl sulfate. An unadulterated solution was included as a control. For each of the contaminants tested, all of the indicator strips exhibited a FAIL condition in the 0.3% Cidex™ OPA solution and a PASS in the 0.45% solution.

The data shows that the Browne OPA Indicator is an effective monitor for the 0.3% MEC Cidex™ OPA Solution.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a symbol. The symbol is a stylized caduceus, which is a traditional symbol of medicine, with three intertwined snakes around a staff. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 8 1999

Albert Browne Ltd. c/o Cynthia J.M. Nolte, Ph.D., RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, MA 02760

Re : K991709

Browne Cidex® OPA Indicator Trade Name: Regulatory Class: II Product Code: JOJ September 15, 1999 Dated: Received: September 16, 1999

Dear Dr. Nolte:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any

{4}------------------------------------------------

Page 2 - Dr. Nolte

obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA findinq of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in yitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Director Division of Dental, Infection Control and General Hospital Devices Office of Device Evaluation Center for Devices and Radioloqical Health

Enclosure

{5}------------------------------------------------

K991709 510(k) Number (if known): _

Device Name: Browne Cidex™ OPA Indicator

Indications For Use:

The Browne Cidex™ OPA Indicator is a concentration monitor for use in orthophthalaldehyde-containing germicide solutions with a minimum effective concentration of 0.3%.

The Browne Cidex™ OPA Indicator is dedicated for use with Cidex™ OPA Solution.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qun S. Lin

(Division Sign-Off Division of Dental, Infection Cor and General Hospit 510(k) Numb

Prescription Use (Per 21 CFR 801.109) OR

Over-The-Counter Use X

Albert Browne Ltd. Browne Cidex™ OPA Indicator 510(k) 5/17/99

§ 880.2800 Sterilization process indicator.

(a)
Biological sterilization process indicator —(1)Identification. A biological sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization.(2)
Classification. Class II (performance standards).(b)
Physical/chemical sterilization process indicator —(1)Identification. A physical/chemical sterilization process indicator is a device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.(2)
Classification. Class II (performance standards).