(187 days)
The Dade Behring N High Sensitivity CRP assay is an in vitro diagnostic test for the quantitative determination of C-reactive protein in human serum as well as heparin- and EDTA-plasma by means of particle-enhanced immunonephelometry using Dade Behring Nephelometer Systems. In acute phase response, increased levels of a number of plasma proteins, including Creactive protein, are observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders, and associated diseases.
N High Sensitivity CRP is a nephelometric assay for the quantitative measurement of C-reactive protein in human serum, or heparinized or EDTA plasma.
This document is a 510(k) summary for the Dade Behring N High Sensitivity CRP assay, a nephelometric assay for the quantitative measurement of C-reactive protein (CRP). The submission is for a modification in the intended use of the device, changing it to "for the detection and evaluation of tissue injury, inflammatory disorders, and associated diseases." The core assay reagents and test system remain unchanged from the previously cleared device.
Here's an analysis of the provided information:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not specify quantitative acceptance criteria or detailed device performance metrics in the format of a table. It states that the device was previously cleared for a similar intended use ("detection and evaluation of inflammatory disorders, tissue injury and infection") and that the current submission only involves a modification to the intended use.
The justification for the modified intended use is: "This labeling change is supported by a large, published literature base that uses both the Dade Behring methodology as well as other high sensitivity CRP methods." This implies that the device's performance characteristics (accuracy, precision, linearity, etc.) were already established and accepted under the previous 510(k) clearance, and the new intended use is supported by existing clinical evidence of CRP's utility in the expanded conditions.
Therefore, for this specific submission, the "acceptance criteria" appear to be:
- The absence of changes to the assay reagents or test system.
- The existence of a "large, published literature base" supporting the broader clinical utility of high sensitivity CRP methods, including the Dade Behring methodology, for the expanded indications.
2. Sample Size Used for the Test Set and Data Provenance
The document does not describe a new study with a specific test set, sample size, or data provenance. The submission is a labeling change supported by existing literature rather than new clinical trials.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Not applicable. As no new test set or ground truth establishment is described in this 510(k) summary.
4. Adjudication Method for the Test Set
Not applicable. No new test set or adjudication process is described.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
No. This is for an in-vitro diagnostic assay (CRP measurement), not an imaging or diagnostic device requiring human reader interpretation in the context of an MRMC study.
6. If a Standalone Performance Study Was Done
The device is an in vitro diagnostic assay. Its performance (accuracy, precision, analytical sensitivity, specificity, etc.) would have been established in previous 510(k) submissions, as it is stated that "FDA has previously cleared this assay via the 510(k) process for the intended use for 'the detection and evaluation of inflammatory disorders, tissue injury and infection'." The current submission explicitly states, "There are no other changes to the assay reagents or test system." Therefore, the standalone (algorithm only) analytical performance would have been established and accepted under the prior clearance. This document does not present new standalone performance data.
7. The Type of Ground Truth Used
For the initial clearance of the assay, the ground truth for CRP measurement would have likely been established through:
- Reference methods: Comparison to a gold standard or well-established CRP measurement method.
- Clinical correlation: Correlation with clinical diagnoses of inflammation, infection, or tissue injury, likely using a combination of clinical assessment, other laboratory markers, and patient outcomes.
However, for this specific submission, which is a modification of intended use, the "ground truth" for supporting the expanded indications ("associated diseases") is based on a "large, published literature base" demonstrating the clinical utility of CRP in these conditions. This is essentially literature-based evidence supporting the clinical relevance of the analyte output.
8. The Sample Size for the Training Set
Not applicable. This is not a machine learning device that requires a training set in the typical sense. The "training" for the assay's performance characteristics would be the analytical validation studies conducted for its initial clearance, but no specific sample sizes for such studies are provided in this summary.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as above. For the initial analytical validation, ground truth would have been established through methods like those described in point 7.
{0}------------------------------------------------
OCT 2 5 1999
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92
Dade Behring N High Sensitivity CRP
Summary of Safety and Effectiveness
N High Sensitivity CRP is a nephelometric assay for the quantitative measurement of C-reactive protein in human serum, or heparinized or EDTA plasma. FDA has previously cleared this assay via the 510(k) process for the intended use for "the detection and evaluation of inflammatory disorders, tissue injury and infection".
The proposed revision is limited to modifications in the intended use. The intended use will be "for the detection and evaluation of tissue injury, inflammatory disorders, and associated diseases."
There are no other changes to the assay reagents or test system.
This labeling change is supported by a large, published literature base that uses both the Dade Behring methodology as well as other high sensitivity CRP methods.
Rebecca S. Aycock
Reberra S Manager, Regulatory Affairs
9/28/99
Date
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a triple-stranded helix intertwined with a staff. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA". The text is arranged in a circular fashion around the caduceus.
OCT 2 5 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Rebecca S. Ayash Manager, Regulatory Affairs Dade Behring Inc. Glasgow Building 500, Mailbox 514 P.O. Box 6101 Newark, Delaware 19714
Re: K991385
Trade Name: Dade Behring N High Sensitivity CRP Assay Regulatory Class: II Product Code: DCK Dated: September 28, 1999 Received: September 29, 1999
Dear Ms. Ayash:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895, A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{2}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Indications Statement
Device Name: Dade Behring N High Sensitivity CRP assay
- Indications for Use: The Dade Behring N High Sensitivity CRP assay is an in vitro diagnostic test for the quantitative determination of C-reactive protein in human serum as well as heparin- and EDTA-plasma by means of particle-enhanced immunonephelometry using Dade Behring Nephelometer Systems. In acute phase response, increased levels of a number of plasma proteins, including Creactive protein, are observed. Measurement of CRP is useful for the detection and evaluation of infection, tissue injury, inflammatory disorders, and associated diseases.
ban looph
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K 991385
Rebecca S. Ayers
becca S. Avash Manager, Regulatory Affairs
9/20/99
Date
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
K991385
510(k) Number
Division Sign-Off Office of Device Evaluation
prescription use
§ 866.5270 C-reactive protein immunological test system.
(a)
Identification. A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.(b)
Classification. Class II (performance standards).