K Number
K991315
Manufacturer
Date Cleared
1999-11-12

(207 days)

Product Code
Regulation Number
862.1645
Reference & Predicate Devices
N/A
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

MICROALBUSTIX Reagent Strips (OTC) are intended for home use by testing persons at risk for developing kidney damage
MICROALBUSTIX Reagent Strips (OTC) are firm plastic strips that contain two reagent areas to test for small amounts of albumin in urine (microalbuminuria), creatinine in urine, and also determine the albumin-to-creatinine ratio in urine. The strips are read visually by comparing the color of the reacted strip to the color chart on the bottle label. MICROALBUSTIX Reagent Strips (OTC) provide semiquantitative results and are intended for home use by persons at risk of developing kidney disease.

Device Description

MICROALBUSTIX Reagent Strips (OTC) are firm plastic strips that contain two reagent areas to test for microalbumin (low concentration of albumin) and creatinine in urine. MICROALBUSTIX Reagent Strips (OTC) are dipped into a urine specimen and "read" visually by comparing the color of the strip to a color chart on the label. In addition to providing an albumin and a creatinine result, an albumin-to-creatinine ratio can also be determined. Semi-quantitative results are available within one minute.

AI/ML Overview

The provided text describes the MICROALBUSTIX™ Reagent Strips (OTC) and the study conducted to demonstrate their performance. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in a quantitative table. However, the performance assessment concludes and effectively acts as the acceptance criteria that were met.

Metric / Acceptance Criteria (Implied)Reported Device Performance
Albumin DetectionDetects urinary albumin at 20 to 40 mg/L
Albumin-to-Creatinine RatioDetects albumin-to-creatinine ratio of 30 to 300 mg/g
User PerformanceLay users can obtain clinical test results that are substantially equivalent to current methods.
Comparison to Current MethodsResults comparable to results obtained by health care professionals.

2. Sample Size Used for the Test Set and Data Provenance:

  • Sample Size: Not explicitly stated. The document mentions "clinical and home use settings" and that "studies demonstrated that lay users can obtain clinical test results," implying a test set, but the number of participants or urine samples is not provided.
  • Data Provenance: Not explicitly stated, though it's implied the studies were conducted in the US, as it's a US FDA submission. It refers to "clinical and home use settings," but doesn't specify if the data was retrospective or prospective.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Number of Experts: Not explicitly stated. The study compared results obtained by lay users to "currently used laboratory tests for microalbumin and creatinine in urine" and to "results obtained by health care professionals." This implies that the ground truth was established by laboratory methods and/or healthcare professionals, but the specific number or qualifications of these professionals beyond their general role are not provided.

4. Adjudication Method for the Test Set:

  • Not explicitly stated. The document indicates a comparison between lay user results and "current methods" or "health care professionals," but doesn't describe any specific adjudication process (e.g., 2+1 review, etc.).

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and the Effect Size:

  • MRMC Study: The description implies a type of comparative effectiveness study in that lay users' performance was compared to "current methods" or "health care professionals." However, it is not formally described as a "Multi-Reader Multi-Case (MRMC) comparative effectiveness study" in the context of radiology or similar fields where that term is often used. It's more of a comparison of a new device's home-user performance against established laboratory/professional methods.
  • Effect Size: Not quantified. The study concludes that lay users can obtain "substantially equivalent" and "comparable" results, but no specific effect size (e.g., in terms of sensitivity, specificity, accuracy improvement, or statistical significance of performance differences) is provided.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

  • Yes, in a way. The device itself is a "standalone" diagnostic device (reagent strip with visual read). The "human-in-the-loop" is the lay user who reads the strip. The study explicitly tests the performance of the device as read by lay users against established laboratory methods, which is the intended standalone performance from the user's perspective. It's not an "algorithm-only" study in the modern AI sense, but rather a study of the device's performance when used as intended without professional interpretation.

7. The Type of Ground Truth Used:

  • Ground Truth: The ground truth was established using "currently used laboratory tests for microalbumin and creatinine in urine." This suggests a laboratory reference method or clinical gold standard as the ground truth.

8. The Sample Size for the Training Set:

  • Not mentioned. The document describes performance assessment through clinical and home-use studies, but does not provide information about a separate training set. This is typical for traditional (non-AI/machine learning) diagnostic devices, which don't have a "training set" in the same way.

9. How the Ground Truth for the Training Set Was Established:

  • Not applicable, as a "training set" in the context of machine learning is not described for this device. The device's performance relies on the chemical reactions of the reagent strips.

{0}------------------------------------------------

.

510(k) SAFETY AND EFFECTIVENESS SUMMARY

Prepared:April 15, 1999
Submitter:Bayer Corporation, Business Group Diagnostics
Address:1884 Miles Avenue, P.O. Box 70Elkhart, IN 46515(219) 262-6928
Contact:George M. Tancos, R.A.C.Manager, Regulatory Compliance
Device:Trade/Proprietary Name: MICROALBUSTIX™ Reagent Strips (OTC)Common/Usual Name: Test for albumin in urineTest for creatinine in urineDocument Control Number: K99 1315
Classification:Division of Clinical Laboratory DevicesPanel - Clinical Chemistry and ToxicologyClassification Code: 75 JIR [Protein or Albumin (Urinary, Non-Quantitative)]75 JFY (Creatinine, enzymatic)
Predicate Devices:MICROALBUSTIX™ Reagent Strips (Professional)Manufactured by Bayer Corporation
Device Description:MICROALBUSTIX Reagent Strips (OTC) are firm plastic strips thatcontain two reagent areas to test for microalbumin (low concentration ofalbumin) and creatinine in urine. MICROALBUSTIX Reagent Strips(OTC) are dipped into a urine specimen and "read" visually bycomparing the color of the strip to a color chart on the label. In additionto providing an albumin and a creatinine result, an albumin-to-creatinineratio can also be determined. Semi-quantitative results are availablewithin one minute.
Intended Use:MICROALBUSTIX Reagent Strips (OTC) are intended for home use bytesting persons at risk for developing kidney damage

{1}------------------------------------------------

K991315

Bayer Corporation, Business Group Diagnostics MICROALBUSTIX™ Reagent Strips (OTC) S&E Summary Page 2 of 2

Technological Characteristics:

MICROALBUSTIX Reagent Strips (OTC) are dip-and-read reagent strips to detection urinary albumin at 20 to 40 mg/L, and the albumin-to-creatinine ratio of 30 to 300 mg/g. The albumin test is based on dye binding using a high affinity sulfonephthalein dye. At a constant pH, the development of a blue color is due to the presence of albumin. The creatinine test is based on the peroxidase-like activity of a copper creatinine complex that catalyzes the reaction of diisopropylbenzene dihydroperoxide and 3,3',5,5'-tetramethylbenzidine. The color of the reagent areas are read visually by comparing to a color chart on the label. The albumin-to-creatinine ratio is determined according to a table in the package insert.

Assessment of Performance:

The performance of MICROALBUSTIX Reagent Strips (OTC) was studied in clinical and home use settings by typical lay users and results compared to currently used laboratory tests for microalbumin and creatinine in urine. The studies demonstrated that lay users can obtain clinical test results that are substantially equivalent to current methods.

Conclusion:

MICROALBUSTIX Reagent Strips (OTC) provide a convenient method for testing for microalbumin and creatinine in urine. Studies show that the product can be used by lay persons to obtain results comparable to results obtained by health care professionals.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three swooping lines representing its wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Nov 1 2 1999

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

Mr. George M. Tancos Manager, Regulatory Compliance Bayer Corporation Business Group Diagnostics 1884 Miles Avenue P.O. Box 70 Elkhart, Indiana 46515-0070

Re: K991315

Trade Name: MICROALBUSTIX™ Reagent Strips (OTC) Regulatory Class: II Product Code: JFY Regulatory Class: I Product Code: JIR Dated: August 2, 1999 Received: August 3, 1999

Dear Mr. Tancos:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Butman

Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(k) Number (if known): K991315

Device Name:

MICROALBUSTIXTM Reagent Strips (OTC)

Indications for Use:

MICROALBUSTIX Reagent Strips (OTC) are firm plastic strips that contain two reagent areas to test for small amounts of albumin in urine (microalbuminuria), creatinine in urine, and also determine the albumin-to-creatinine ratio in urine. The strips are read visually by comparing the color of the reacted strip to the color chart on the bottle label. MICROALBUSTIX Reagent Strips (OTC) provide semiquantitative results and are intended for home use by persons at risk of developing kidney disease.

Lisa Cooper
(Division Sign-Off)

Division of Clinical Laboratory Devices 510(k) Number .

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use

(Optional Format 1-2-96)

§ 862.1645 Urinary protein or albumin (nonquantitative) test system.

(a)
Identification. A urinary protein or albumin (nonquantitative) test system is a device intended to identify proteins or albumin in urine. Identification of urinary protein or albumin (nonquantitative) is used in the diagnosis and treatment of disease conditions such as renal or heart diseases or thyroid disorders, which are characterized by proteinuria or albuminuria.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 862.9.