K Number
K991014
Date Cleared
1999-10-20

(208 days)

Product Code
Regulation Number
N/A
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CASE 8000 exercise testing system is a diagnostic device intended to monitor and record electrocardiographic data from the patient during exercise, as well as perform resting ECG analysis.

The intended use of the CASE 8000 exercise testing system is to acquire, process, record, archive, analyze, and output data during a period of physiologic stress or during a resting ECG condition. The CASE 8000 exercise testing system has the ability to calculate prognostic scores. A user selectable option can provide printout of prognostic scores on select reports. A vector loops report is also available.

Ancillary functions include control of an external device (typically a treadmill or ergometer) and communications with centralized electronic/digital storage system.

This device uses a computerized analysis program, which can be used as a tool in ECG tracing interpretation.

The device records 12 lead ECG data taken in resting and exercise modes, as well as real-time rhythm and median morphology recordings. The device can perform a 12 lead interpretive analysis on the resting ECG, providing the user with statements of morphology, rhythm and conduction.

The CASE 8000 exercise testing system is intended to be used by trained operators under the direct supervision of a licensed health care practitioner when exercise or resting ECG records are required.

The CASE 8000 exercise testing system is intended to be used primarily in hospital based exercise testing laboratories, but can be used in clinics, physician offices, outreach centers or wherever exercise testing is performed. This device is not intended for use with high frequency surgical units.

The CASE 8000 system is not intended to be used as a vital signs physiological monitor.

This equipment will not cause abnormal operation of a patient's cardiac pacemaker or other electronic stimulator.

Device Description

The CASE 8000 exercise testing system is a PC-based device capable of being operated as a stand-alone system or in conjunction with a centralized electronic/digital storage system such as the MUSE CV Information System - K980495 (MUSE). Major components include a CPU, display, modem, mouse, printer, keyboard, power supply, acquisition module and a cart. It is intended to be used primarily in hospital based exercise testing laboratories. It can also be used in clinics and outreach centers or wherever exercise testing is performed.

The primary application of the device is to acquire, process, record, archive, analyze and output data during a period of physiologic stress or during rest. Ancillary functions include control of an external exercise device, typically a treadmill or ergometer, and communication with a centralized electronic/digital storage system such as the MUSE.

When used in conjunction with MUSE, the CASE 8000 exercise testing system is capable of bringing computerized information management to the exercise laboratory. Demographic and order information can be downloaded to the system to simplify order management and data entry operation. Results from previous tests can be reviewed on the system after network download from the MUSE. New results can be quickly sent to the MUSE for long term storage.

AI/ML Overview

The provided text describes the GE Marquette Medical Systems' CASE 8000 exercise testing system and its 510(k) submission (K991014) to the FDA. However, the document does not contain specific acceptance criteria or a detailed study report that quantitatively proves the device meets such criteria.

The "Test Summary" section {2} mentions general quality assurance measures applied during development but lacks concrete metrics, study designs, sample sizes, or performance results. The "Conclusion" states "The results of these measures demonstrate that the CASE 8000 exercise testing system is as safe, as effective, and performs as well as the predicate devices," but provides no data to support this.

Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor can I provide details about sample sizes, data provenance, expert ground truth establishment, adjudication methods, MRMC studies, standalone performance, or training set information.

Based on the provided text, the following information is either not available or only vaguely mentioned:

  1. Table of acceptance criteria and the reported device performance: Not provided. The document states the device "complies with the voluntary standards as detailed in Section 9 Specific Standards and Guidances of this submission" and that "the results of these measures demonstrate that the CASE 8000 exercise testing system is as safe, as effective, and performs as well as the predicate devices." However, no specific performance metrics or their comparison against acceptance criteria are presented.
  2. Sample sized used for the test set and the data provenance: Not provided.
  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not provided.
  4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not provided.
  5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs. without AI assistance: Not provided. The device performs "12SL Analysis" which is a computerized interpretive analysis on resting ECGs, but no MRMC study details are given. The document does not describe AI assistance for human readers.
  6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: The device performs a "12 lead interpretive analysis on the resting ECG" {2}. This implies standalone algorithmic performance, but no specific study or metrics for this standalone performance are provided.
  7. The type of ground truth used: Not explicitly stated. For the 12-lead interpretive analysis, it would typically be compared against expert cardiologist interpretations, but this is not confirmed or detailed in the document.
  8. The sample size for the training set: Not provided.
  9. How the ground truth for the training set was established: Not provided.

{0}------------------------------------------------

OCT 18 1999 10:03 FR GE MARQUETTE FINANCE 414 362 2420 TO 913014804204

OCT 20 1999

9910/4

Section 2 Summary and Certification

2.1 510(k) Summary of Safety and Effectiveness

Date:March 25, 1999
Submitter:GE Marquette Medical Systems8200 West Tower AvenueMilwaukee, WI 53223 USA
Contact Person:Laura L McComisCorporate Regulatory AffairsGE Marquette Medical SystemsPhone: (414) 362-2688Fax: (414) 355-3790
Device:Trade Name:CASE 8000 exercise testing system
Common/Usual Name:ECG analysis computer
Classification Names:
21 CFR 870.1425 Programmable diagnostic computerClass II74DQK
21 CFR 870.2920 Transmitters and Receivers, Electrocardiograph,TelephoneClass II74DXH
21 CFR 870.2340 ElectrocardiographClass II74DPS
21 CFR 870.2340 System, ECG AnalysisClass III74LOS
Predicate Devices:K951130 CardioSoft/CardioSys

CardioSys K973403 12SL Analysis System

{1}------------------------------------------------

Device Description:

The CASE 8000 exercise testing system is a PC-based device capable of being operated as a stand-alone system or in conjunction with a centralized electronic/digital storage system such as the MUSE CV Information System - K980495 (MUSE). Major components include a CPU, display, modem, mouse, printer, keyboard, power supply, acquisition module and a cart. It is intended to be used primarily in hospital based exercise testing laboratories. It can also be used in clinics and outreach centers or wherever exercise testing is performed.

The primary application of the device is to acquire, process, record, archive, analyze and output data during a period of physiologic stress or during rest. Ancillary functions include control of an external exercise device, typically a treadmill or ergometer, and communication with a centralized electronic/digital storage system such as the MUSE.

When used in conjunction with MUSE, the CASE 8000 exercise testing system is capable of bringing computerized information management to the exercise laboratory. Demographic and order information can be downloaded to the system to simplify order management and data entry operation. Results from previous tests can be reviewed on the system after network download from the MUSE. New results can be quickly sent to the MUSE for long term storage.

{2}------------------------------------------------

Intended Use:The intended use of the CASE 8000 exercise testing system is toacquire, process, record, archive, analyze, and output dataduring a period of physiologic stress or during a resting ECGcondition. The CASE 8000 exercise testing system has the abilityto calculate prognostic scores. These include DUKE treadmillscore and Risk Factor Prediction, which are Coronary HeartDisease Risk Factor and Stroke Risk Factor. A user selectableoption can provide printout of prognostic scores on selectreports. A vector loops report is also available.Ancillary functions include control of an external device (typicallya treadmill or ergometer) and communications with centralizedelectronic/digital storage system.The device records 12 lead ECG data taken in resting andexercise modes, as well as real-time rhythm and medianmorphology recordings. The device can perform a 12 leadinterpretive analysis on the resting ECG, providing the user withstatements of morphology, rhythm and conduction. No changeshave been made to this algorithm - 12SL Analysis System K973403.The CASE 8000 exercise testing system is intended to be used bytrained operators under the direct supervision of a licensedhealth care practitioner when exercise or resting ECG records arerequired.The CASE 8000 exercise testing system is intended to be usedprimarily in hospital based exercise testing laboratories, but canbe used in clinics, physician offices, outreach centers or whereverexercise testing is performed.
Technology:CASE 8000 exercise testing system employs the same technologyas the predicate devices.
Test Summary:The CASE 8000 exercise testing system complies with thevoluntary standards as detailed in Section 9 Specific Standardsand Guidances of this submission. The following quality assurancemeasures were applied to the development of CASE 8000exercise testing system:• Risk analysis• Requirements specification reviews• Code inspections• Software and Hardware Testing• Safety testing• Environmental testing• Final validation
Conclusion:The results of these measures demonstrate that the CASE 8000exercise testing system is as safe, as effective, and performs aswell as the predicate devices.

.

.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 20 1999

Ms. Laura L. McComis GE Marquette Medical Systems 8200 West Tower Avenue Milwaukee, WI 53223

Re: K991014 CASE 8000 Exercise Testing System Regulatory Class: III (three) Product Code: 74 LOS Dated: July 21, 1999 Received: July 22, 1999

Dear Ms. McComis:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{4}------------------------------------------------

Page 2 - Ms. Laura L. McComis

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

joannett Westerheim

Wolf Sapirstein, M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

K991014

Section 11 Indications for Use Statement

510/k) Number lif known):

Unknown - 510/k) filed March 25, 1999

Device Name:

CASE 8000 exercise testing system

Indications For Use:

The CASE 8000 exercise testing system is a diagnostic device intended to monitor and record electrocardiographic data from the patient during exercise, as well as perform resting ECG analysis.

The intended use of the CASE 8000 exercise testing system is to acquire, process, record, archive, analyze, and output data during a period of physiologic stress or during a resting ECG condition. The CASE 8000 exercise testing system has the ability to calculate prognostic scores. A user selectable option can provide printout of prognostic scores on select reports. A vector loops report is also available.

Ancillary functions include control of an external device (typically a treadmill or ergometer) and communications with centralized electronic/digital storage system.

This device uses a computerized analysis program, which can be used as a tool in ECG tracing interpretation.

The device records 12 lead ECG data taken in resting and exercise modes, as well as real-time rhythm and median morphology recordings. The device can perform a 12 lead interpretive analysis on the resting ECG, providing the user with statements of morphology, rhythm and conduction.

The CASE 8000 exercise testing system is intended to be used by trained operators under the direct supervision of a licensed health care practitioner when exercise or resting ECG records are required.

The CASE 8000 exercise testing system is intended to be used primarily in hospital based exercise testing laboratories, but can be used in clinics, physician offices, outreach centers or wherever exercise testing is performed. This device is not intended for use with high frequency surgical units.

The CASE 8000 system is not intended to be used as a vital signs physiological monitor.

This equipment will not cause abnormal operation of a patient's cardiac pacemaker or other electronic stimulator.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use_
(Per 21 CFR 801.109)

OR

(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) NumberK991014
------------------------
Over-The-Counter Use_
(Optional Format 1-2-96)

49

N/A