K Number
K990856
Date Cleared
2000-01-11

(302 days)

Product Code
Regulation Number
868.5905
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The nasal CPAP unit is intended for treatment of adult Obstructive Sleep Apnea (OSA). Obstructive sleep apnea is defined as the absence of air movement for ten seconds during sleep. Obstructive sleep apnea is usually diagnosed by a sleep study. The study obtains the optimum level of pressure required to maintain an unobstructed airway, the positive pressure allows the is ve or prosuropen. The nasal CPAP unit is intended only for spontaneously breathing adult patients.

Device Description

The SoftAire Nasal CPAP System is intended to provide continuous positive airway pressure for the care and treatment of individuals suffering from obstructive sleep apnea. The positive pressure is clinician adjustable within the designated operating range and a clinician adjustable timer will allow timed rise to pressure ("Ramp" mode). The user controls will be limited to an on/off switch and option of using the timed pressure rise. This product is designed to be used with a 22 millimeter (smooth bore) tube air delivery system and user selected nasal mask. This mask is not marketed by Bird. Power will be provided by AC to DC adapter s which accommodate a variety of input voltages and frequencies. This adapter will plug into a 12 VDC connector on the side of the unit. The SoftAire Nasal CPAP System does not utilize an actively controlled exhalation valve. The patient circuit consists of two essential parts: a large-bore tube from the ventilator to the patient and a required exhaust port. The exhaust is an orifice which may be part of the adapter between the tubing an the patient mask, or may be an orifice which is part of the patient mask.

AI/ML Overview

Acceptance Criteria and Device Performance for SoftAire Nasal CPAP System

1. Table of Acceptance Criteria and Reported Device Performance

SpecificationRequirementReported Device Performance
Adjustable Therapeutic Pressure Range4-18 cmH2O☑ (Verified to operate as designed and intended)
Adjustable Beginning Pressure Range4 and 18 cmH2O and below therapeutic range☑ (Verified to operate as designed and intended)
Pressure Accuracy± 10% to the nearest displayed digit (calibrated at 500ft elevation)☑ (Verified to operate as designed and intended)
Timer0, 5, 10, 15, 20 minutes, ± 2 minutes☑ (Verified to operate as designed and intended)
Input Voltage12 ± 10% VDC, 2A☑ (Verified to operate as designed and intended)
Noise Level< 38dB☑ (Verified to operate as designed and intended)
Dimensions11X8X4 inches☑ (Verified to operate as designed and intended)
Weight5.2 lb☑ (Verified to operate as designed and intended)
Tubing6 ft, 22mm female connectors☑ (Verified to operate as designed and intended)
Electrical SafetyEN 60601-1☑ (Verified to operate as designed and intended)
EMI/RFIEN 60601-1-2 & Draft Reviewers Guidance for Ventilators 1995☑ (Verified to operate as designed and intended)
EnvironmentalIEC 68-2☑ (Verified to operate as designed and intended)
Operating Temperature+10 to +40°C, 30 to 75% RH☑ (Verified to operate as designed and intended)
Storage Temperature-20 to +60°C, up to 95% RH☑ (Verified to operate as designed and intended)

2. Sample Size and Data Provenance (Test Set)

The document does not provide details on specific sample sizes used for the test set or the data provenance. It states, "Performance testing was conducted in the laboratory to confirm operation compliance to the device specifications." This implies that the testing was likely conducted in a controlled lab environment and not necessarily with patient data from a specific country, nor does it specify if it was retrospective or prospective.

3. Number of Experts and Qualifications (Ground Truth for Test Set)

The document does not mention the use of experts or their qualifications for establishing ground truth for the test set. The performance testing appears to be based on engineering and regulatory standards rather than clinical expert evaluation.

4. Adjudication Method (Test Set)

There is no mention of an adjudication method in the provided text. The performance was confirmed against predefined specifications.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

No MRMC comparative effectiveness study was done. The document describes a medical device clearance process, not a study comparing human readers with and without AI assistance.

6. Standalone Algorithm Performance

This section is not applicable as the device is a physical medical device (CPAP system), not an algorithm or AI system for diagnostic or interpretive purposes. The performance assessment is for the device's functional specifications.

7. Type of Ground Truth Used

The "ground truth" for the performance testing was based on pre-defined engineering and regulatory specifications and requirements (e.g., pressure accuracy, noise level, safety standards). It did not utilize expert consensus, pathology, or outcomes data in the context of clinical performance, but rather compliance with technical specifications.

8. Sample Size for the Training Set

This question is not applicable. The SoftAire Nasal CPAP System is a hardware device, not a machine learning model, and therefore does not have a "training set" in the context of AI/ML development.

9. How the Ground Truth for the Training Set Was Established

This question is not applicable for the same reason as point 8.

{0}------------------------------------------------

510(K) SUMMARY


SoftAire Nasal CPAP System

Bird Products Corporation

Tom Gutierrez P.E. Regulatory Compliance Engineer Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262-6267

760.778.7341 (voice) 760.778.7274 (fax)

June 18, 1999

{1}------------------------------------------------

510(k) Summary for the tem, page 2 of 4

General Information

. . .

Device Trade Name: SoftAire Nasal CPAP System

Device Common/Classification Name:

The SoftAire Nasal CPAP System is Bird Products Corporation ventilator classified under 73 BZD "Non-Continuous Ventilator"

  • · 868.5905 Non-Continuous Ventilator, 73 BZD
Predicate Device:ResMed Sullivan III Nasal CPAP SystemFDA 510(k) No: 930656
Year 2000 Compatibility:The SoftAire Nasal CPAP System does not perform any datecalculation or processing and is therefore year-2000 compliant.

Device Description

The SoftAire Nasal CPAP System is intended to provide continuous positive airway pressure for the care and treatment of individuals suffering from obstructive sleep apnea. The positive pressure is clinician adjustable within the designated operating range and a clinician adjustable timer will allow timed rise to pressure ("Ramp" mode). The user controls will be limited to an on/off switch and option of using the timed pressure rise. This product is

designed to be used with a 22 millimeter (smooth bore) tube air delivery system and user selected nasal mask. This mask is not marketed by Bird.

Power will be provided by AC to DC adapter s which accommodate a variety of input voltages and frequencies. This adapter will plug into a 12 VDC connector on the side of the unit.

Image /page/1/Picture/12 description: The image shows a close-up of a black electrical box with a label on it. The label contains the model number 15682 and serial number KJTC00313. It also includes electrical requirements of 12VDC 2 amps and several warnings and cautions. The label also has several certification logos.

Bottom View of SoftAire Nasal CPAP System

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a close-up of a gas mask. The mask is black and has a boxy shape. Two corrugated tubes extend from the mask, one on each side. The tubes are also black and appear to be made of a flexible material.

SoftAire Nasal CPAP System with tube and nasal mask.

The SoftAire Nasal CPAP System does not utilize an actively controlled exhalation valve. The patient circuit consists of two essential parts: a large-bore tube from the ventilator to the patient and a required exhaust port. The exhaust is an orifice which may be part of the adapter between the tubing an the patient mask, or may be an orifice which is part of the patient mask.

Electrical/Pneumatic System

Image /page/2/Figure/5 description: This image is a block diagram of a system. The system includes a 100 PPI filter, a muffler, and a 12 VDC blower with a 22 MM outlet. The system also includes a power supply that can accept either 110 VAC or 220 VAC, a connector that is 2.5MM x 5.5MM, an electronic module, a user input, and a temp fuse.

Flow to the patient is provided via 12 VDC Brushless Motor driving an impeller. Air intake passes through a filter and muffler into the motor, which operates at a preset speed established by the (clinical) operator. Patient tubing pressure is verified by an independent pressure measurement.

An external 12-Volt power source supplies a 5-Volt regulator, which provides power to all interval components. The CPU runs at 4 MHz. There are 5 switches underneath the unit for Mode, Plus, Minus, "Ramp Mode On" and "Ramp Mode Off",

{3}------------------------------------------------

Comparison to Predicate Device

The SoftAire Nasal CPAP System is not significantly different from the predicate device, the ResMed Sullivan III Nasal CPAP, first cleared for market under 510(k) K930656. Both devices utilize Blower type flow generators with the Bird SoftAire Nasal CPAP having a maximum working pressure of 22cmH,O.


Summary of Performance Testing

Performance testing was conducted in the laboratory to confirm operation compliance to the device specifications. Testing to Environmental, EMI/RFI and Electrical Safety Standards were performed by certified test facilities. The following table specifies all SoftAire Nasal CPAP functions. All functions were verified to operate as designed and intended, as denoted by the checkmark.

SpecificationRequirement
Adjustable TherapeuticPressure Range4-18 cmH2O.
Adjustable BeginningPressure Range4 and 18 cmH2O and below therapeutic range.
Pressure Accuracy$\pm$ 10% to the nearest displayed digit (claibrated at 500ft elevation)
Timer0, 5, 10, 15, 20 minutes, $\pm$ 2 minutes
Input Voltage12 $\pm$ 10% VDC, 2A
Noise Level< 38dB
Dimensions11X8X4 inches
Weight5.2 lb
Tubing6 ft, 22mm female connectors
Electrical SafetyEN 60601-1
EMI/RFIEN 60601-1-2 & Draft Reviewers Guidance for Ventilators 1995
EnvironmentalIEC 68-2
Operating Temperature+10 to +40°C, 30 to 75% RH
Storage Temperature-20 to +60°C, up to 95% RH

Performance testing verified that the SoftAire Nasal CPAP meets all of it's performance requirements and that this device is substantially equivalent to medical devices currently legally marketed in the United States.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a caduceus, a symbol often associated with medicine and healthcare, featuring a staff with a serpent entwined around it.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 1 1 2000

Mr. Tom Gutierrez Bird Products Corporation 1100 Bird Center Drive Palm Springs, CA 92262

Re: K990856 SoftAire Nasal CPAP System Regulatory Class: II (two) Product Code: 73 BZD Dated: October 13, 1999 Received: October 14, 1999

Dear Mr. Gutierrez:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{5}------------------------------------------------

Page 2 - Mr. Tom Gutierrez

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Jeanroth W. Westerhausen for,

Celia M. Witten, Ph.D., M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

Pageof
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Acres of Childrenl1
1

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

Device Name: _________________________________________________________________________________________________________________________________________________________________

Indications for Use:

The nasal CPAP unit is intended for treatment of adult Obstructive Sleep Apnea (OSA). Obstructive sleep apnea is defined as the absence of air movement for ten seconds during sleep. Obstructive sleep apnea is usually diagnosed by a sleep study. The study obtains the optimum level of pressure required to maintain an unobstructed airway, the positive pressure allows the is ve or prosuropen. The nasal CPAP unit is intended only for spontaneously breathing adult patients.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF REQUIRED) * ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ ^ -----------------------------------------------------------

Concurrence of CDRH, Office of Device Evaluation (ODE)

Wutcherson

(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K990856 510(k) Number ________________________________________________________________________________________________________________________________________________________________

Prescription Use

OR

Over-the-Counter Use

(Per 21 CFR 801.109)


(Optional format 1-2-96)

§ 868.5905 Noncontinuous ventilator (IPPB).

(a)
Identification. A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.(b)
Classification. Class II (performance standards).