K Number
K990663

Validate with FDA (Live)

Manufacturer
Date Cleared
1999-05-27

(87 days)

Product Code
Regulation Number
876.1500
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Intromit is inserted into the patient's abdomen to provide abdominal access for the surgeon's hand while preserving the peritoneum during laparoscopic surgery.

Device Description

Intromit™ will be sold as a kit that includes a flexible access tube, medical grade adhesive, sterile lubricant and incise template. The access tube consists primarily of an outer sleeve, an inner sleeve, and an adhesive coated flange, which is welded to the distal end of the outer sleeve. The inner and outer sleeves are welded together. The principal components of the access tube are made of polymer film. Intromit™ is packaged into sterile double tyvek pouches and is labeled as a single-patient use device.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and the study proving the device meets them, structured as requested:

Device Name: Intromit™ Hand Access Port for Hand Assisted Laparoscopic Surgery

1. Table of Acceptance Criteria and Reported Device Performance

The provided 510(k) summary does not explicitly state pre-defined acceptance criteria with specific numerical thresholds for the clinical study. Instead, it frames the study results in terms of "equivalence" or "no more likely/higher/equivalent" compared to a control group or existing procedures.

Acceptance Criteria (Implied)Reported Device Performance (Intromit™ HALC vs. LC with trocars)
Procedure time is not longerOverall procedure time for HALC with the Intromit™ is at least as short as the overall procedure time for LC with trocars.
Conversion to open surgery is not more likelyPatients in the Intromit™ device group are no more likely to convert to open surgery than patients in the LC group.
Instrument changes are not more frequentNo more instrument changes are required for HALC with the Intromit™ than LC with trocars.
Mortality is not related to the deviceThe one death that occurred during the clinical trial was not related to the Intromit™ device.
Morbidity is equivalentThe morbidity for HALC with the Intromit™ is equivalent to the morbidity for LC with trocars.
Hospital costs are not higherThe direct hospital costs associated with HALC using the Intromit™ are no higher, and possibly lower, than the direct hospital costs associated with LC using trocars.
Hospital stays are equivalentThe hospital stays for patients who receive HALC with the Intromit™ are equivalent to the hospital stays for patients who receive LC with trocars.
Shorter reconvalescence compared to open surgery is maintainedThe HALC with the Intromit™ has retained one of the major advantages of laparoscopic surgery, i.e., a shorter reconvalescence when compared to open surgery.
Overall Safety and EffectivenessThe clinical study shows that the HALC with the Intromit™ are as safe and effective as LC with trocars.

2. Sample Size and Data Provenance for the Test Set

  • Sample Size: Forty-four (44) patients were enrolled in the clinical study.
  • Data Provenance: The text does not specify the country of origin. It describes a "clinical study," which implies it was prospective in nature, as patients were "enrolled" and their outcomes compared between two surgical approaches.

3. Number of Experts and Qualifications for Ground Truth

  • The document does not explicitly mention the number of experts used to establish ground truth or their specific qualifications. For a surgical device study like this, the "ground truth" would be established by the performing surgeons and other medical staff involved in the patient's care (e.g., assessing procedure time, complications, hospital stay). However, the document does not elaborate on an independent ground truth review panel.

4. Adjudication Method for the Test Set

  • The document does not mention an explicit adjudication method (e.g., 2+1, 3+1). For a clinical trial comparing surgical procedures, patient outcomes (such as procedure time, conversion to open surgery, morbidity, mortality) are typically recorded and analyzed according to predefined study protocols, rather than being subject to an "adjudication" process akin to image review by multiple experts.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

  • No, an MRMC comparative effectiveness study was not explicitly described. The study compared surgical procedures (Hand Assisted Laparoscopic Colectomies with Intromit™ vs. Standard Laparoscopic Colectomies with trocars), not the diagnostic performance of a device with and without AI assistance involving human readers.
  • Therefore, there is no effect size reported on how much human readers improve with AI vs. without AI assistance.

6. Standalone (Algorithm Only) Performance

  • The Intromit™ is a physical surgical device, not an AI algorithm. Therefore, no standalone (algorithm only) performance study was done. The performance described is of the device as used by a surgeon.

7. Type of Ground Truth Used

  • The ground truth for this device's efficacy and safety was based on clinical outcomes data from a comparative study of surgical procedures. This includes metrics like procedure time, rates of conversion to open surgery, instrument changes, mortality, morbidity, direct hospital costs, and hospital stay duration.

8. Sample Size for the Training Set

  • The concept of a "training set" is typically associated with machine learning or AI models. Since the Intromit™ is a physical surgical device, there is no training set in this context. The clinical study of 44 patients serves as the primary performance evaluation.

9. How the Ground Truth for the Training Set Was Established

  • As there is no training set for a physical surgical device, this point is not applicable. The clinical outcomes mentioned above served as the basis for evaluating the device's performance in the clinical study.

{0}------------------------------------------------

MAY 27 toon

510(k) Summary

MEDTECH LIMITED'S INTROMITTM HAND ACCESS PORT FOR HAND ASSISTED LAPAROSCOPIC SURGERY

Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared

  • Martin Caldwell Medtech Limited Unit 3 70 Heather Road Sandyford Industrial Estate Dublin 18 Ireland
    Phone: 011-353-1-295-0744 Facsimile: 011-353-1-294-3319

Contact Person:

Jonathan S. Kahan Hogan & Hartson 555 13th Street, N.W. Washington, DC 20004

(202) 637-5794 Phone: Facsimile: (202) 637-5910

Regulatory Counsel to Medtech Limited

Date Prepared: May 4, 1999

Name of Device:

Intromit™ Hand Access Port for Hand Assisted Laparoscopic Surgery

Common or Usual Name:

Surgical Access Port

{1}------------------------------------------------

Classification Name:

Endoscopic Accessory - Extended Laparoscopy Device

Product Code: GCJ

Predicate Devices:

    1. Pilling Weck's Dexterity® Pneumo Sleeve and Protector® Retractor;
    1. Karl Storz Endoscopy America, Inc.'s trocars and cannulas for endoscopic and laparoscopic procedures;
    1. Solos Endoscopy Inc. GS-4300 Cannula and Trocar and GS-4500 Cannula and Trocar:
    1. Marlow Surgical Technologies Hasson SAC Cannula;
    1. Ethicon, Inc.'s Endopath Trocar; and
    1. Innerdyne Medical 'Step' Radical Expansion Obturator/Cannula.

Intended Use

Intromit™ is intended to be used as a port into surgical sites. The Intromit™ is inserted into the patient's abdomen to provide abdominal access for a surgeon's hand while preserving the pneumoperitoneum during laparoscopic surgery.

Principles of Operation

Intromit™ is a hand access port for use in Minimally Invasive Surgery (MIS). The aim of the designers of Intromit™ was to minimize the trauma to the patient while permitting ease of access to an inflated abdomen. By creating a device whereby the incision required is no larger than that necessary to allow passage of the surgeon's hand and forearm, the first criterion has been achieved. The single handed entry method, with no external clamping or sealing mechanism, has ideally achieved the second criterion. The flexible nature of Intromit™ makes it universal in relation to the size of a surgeon's arm and the depth of insertion into the abdomen.

Intromit™ is easy to use. The incise template is used to mark the location and size of the incision and the perimeter of the adhesive flange site. An incision is made through the patient's abdominal wall into the abdominal cavity. The distal end of the inner sleeve is inserted into the abdominal cavity through the incision. The flange is adhered to the patient's skin using the drawn perimeter line as a guide to its application, creating a pressure-tight seal. Insufflation gas is then

{2}------------------------------------------------

administered through the insufflation valve into the access tube, which makes it easier for the surgeon to insert his or her hand through it into the abdomen.

The insufflation pressure forces the walls of the inner sleeve together at two points, which forms pressure-tight seals. The distal end of the inner sleeve contains a taut valve that forces the walls of the sleeve together, creating a seal in the abdominal cavity. In addition, the proximal end of the access tube is tapered to produce a seal between its lip edge and the surgeon's forearm or wrist. The active seal maintains the abdominal pressure by preventing gas from escaping through the mouth of the access tube.

Technological Characteristics

Intromit™ will be sold as a kit that includes a flexible access tube, medical grade adhesive, sterile lubricant and incise template. The access tube consists primarily of an outer sleeve, an inner sleeve, and an adhesive coated flange, which is welded to the distal end of the outer sleeve. The inner and outer sleeves are welded together. The principal components of the access tube are made of polymer film. Intromit™ is packaged into sterile double tyvek pouches and is labeled as a single-patient use device.

Substantial Equivalence

Intromit™ has the same intended use as Pilling Weck's Dexterity® Pneumo Sleeve and Protector® Retractor. Both devices to be inserted into the patient's abdomen to provide abdominal access for a surgeon's hand while preserving the pneumoperitoneum during laparoscopic surgery. Both the predicate device and Intromit™ achieve this in similar fashion: both comprise a flexible polymer sleeve that when adhesively attached to the patients abdomen, form an extension to the abdomen; both have distal and proximal seals and means of affording protection at the incision site; and both utilize similar materials and manufacturing methods in their construction.

Both Intromit™ and Dexterity® are similar in the way in which they retain pneumoperitoneal pressure during laparoscopic surgery in that they are nominally cylindrical, gas impermeable, flexible polymer sleeves with an integral adhesive flange for sealing and attachment to the patients abdomen. Both incorporate two sealing means, (1) a proximal seal for the containment of pressurized gas with a surgeon's arm passing through the device and (2) a distal seal for the containment of pressurized gas in the absence of a surgeon's arm. In addition, both devices contain a mechanism for protection of the mini-laparotomy incision by way of a polymer film barrier that exists between the incision and the access lumen for the surgeon's "gloved" hand and forearm.

{3}------------------------------------------------

Intromit™ also has the same general intended use as its predicate device trocars and cannulas, listed above. As with the Intromit™, these devices are indicated for laparoscopic surgery using gas pressure to distend the abdominal cavity. These predicate devices have similar principles of operation and technological characteristics, namely they provide access into the surgical site and insufflation gas may be administered through several of the predicate devices, in order to maintain abdominal gas pressure.

Performance Data

Both nonclinical and clinical testing have been conducted on Intromit™. Bench testing has been successfully performed on each component of Intromit™ in order to ensure that the components function as intended during use (i.e., tensile strength testing was conducted on the Intromit™'s outer sleeve, inner sleeve, feather valve, flange, and taut valve). The Intromit™ material was also evaluated for its ability to cause skin sensitization in guinea pigs, and no positive skin responses were obtained for any of the guinea pigs. A clinical study was then conducted to compare hand assisted laparoscopic colectomies ("HALC") with the Intromit™ and standard laparoscopic colectomies ("LC") with trocars. Forty four patients were enrolled in the study. This study showed that:

  • The overall procedure time for HALC with the Intromit™ is at least as short as the overall procedure time for LC with trocars;
  • Patients in the Intromit™ device group are no more likely to convert to open . surgery than patients in the LC group;
  • No more instrument changes are required for HALC with the Intromit™ than . LC with trocars:
  • The one death that occurred during the clinical trial was not related to the . Intromit™ device;
  • The morbidity for HALC with the Intromit™ is equivalent to the morbidity for . LC with trocars;
  • The direct hospital costs associated with HALC using the Intromit™ are no . higher, and possibly lower, than the direct hospital costs associated with LC using trocars;
  • The hospital stays for patients who receive HALC with the Intromit™ are . equivalent to the hospital stays for patients who receive LC with trocars; and

{4}------------------------------------------------

  • The HALC with the Intromit™ has retained one of the major advantages of . laparoscopic surgery, i.e., a shorter reconvalescence when compared to open surgery.
    Thus, the clinical study shows that the HALC with the Intromit™ are as safe and effective as LC with trocars.

Conclusion

Based on a narrative comparison of the Intromit™, the Dexterity® device, and the trocars as summarized above and the clinical data which also is summarized above, Intromit™ is substantially equivalent to its predicate devices.

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three wavy lines that resemble a stylized human figure. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the caduceus symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 2 7 1000

Medtech, Limited c/o Mr. Jonathan S. Kahan, Esq. Hogan & Hartson, L.L.P. Columbia Square 555 Thirteenth Street, NW Washington, D.C. 20004

K990663 Re:

Trade Name: Intromit® Hand Access Port for Hand Assisted Laparoscopic Surgery Regulatory Class: GCJ Product Code: II Dated: March 1, 1999 Received: March 1, 1999

Dear Mr. Kahan:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{6}------------------------------------------------

Page 2 – Mr. Jonathan S. Kahan, Esq.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

510(k) Number (if known):__K990663

Device Name: Medtech Ltd.'s Intromit Hand Access Port

Indications for Use:

The Intromit is inserted into the patient's abdomen to provide abdominal access for the surgeon's hand while preserving the peritoneum during laparoscopic surgery.

(PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices K990663.
510(k) Number

Prescription Use
(Per 21 C.F.R. 801.109)
OR
Over-The-Counter Use

(Optional Format 1-2-96)

§ 876.1500 Endoscope and accessories.

(a)
Identification. An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification —(1)Class II (special controls). The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.